AJP - Heart and Circulatory Physiology, Vol 232, Issue 4 434-H436, Copyright © 1977 by American Physiological Society
ARTICLES |
M. M. Laks, D. Garner, J. Beazell and J. Piscatelli
Previous methods of internal calibration for cineangiography have made use of reference objects such as catheters or wires of known length or diameter. Such devices have either insufficient X-ray resolution, or require a specific orientation within the ventricle which is difficult to confirm. External calibration methods, while eliminating these errors, can only estimate actual position of the left ventricle. The use of a Swan-Ganz balloon catheter as a calibration object diminishes these problems. When positioned within the left ventrical and filled with a radiopaque medium, the balloon is of sufficient size and density to avoid errors induced by previously employed objects. The inflated balloon geometry is simple and reproducible. The use of this internal calibration catheter markedly diminishes major sources of errors in the determination of ventriculograms for both dogs and humans.