AJP - Heart and Circulatory Physiology, Vol 270, Issue 2 760-H768, Copyright © 1996 by American Physiological Society
ARTICLES |
R. F. Labadie, J. F. Antaki, J. L. Williams, S. Katyal, J. Ligush, S. C. Watkins, S. M. Pham and H. S. Borovetz
Department of Surgery, University of Pittsburgh, Pennsylvania 15261, USA.
We have constructed and performed initial validation of an innovative perfusion system that allows exposure of intact segments of vascular tissue to realistic physiological and hemodynamic environments ex vivo. Computer-controlled opening and closing of an in-line gate valve allows generation of arterial pressure waveforms. The control algorithm predicted resultant pressure waveforms with a high degree of accuracy (Pearson correlation coefficient > 0.97). To document vascular homeostasis ex vivo, vasomotor bioassays and morphological studies were performed. The bioassays consisted of injecting epinephrine (2 x 10(-3) mg/ml) into the perfusion system followed by acetylcholine (100 microM) while concurrently measuring vessel diameter with a laser micrometer, significant vasomotion was measured for canine carotid arteries (n = 4) bioassayed after 1, 24, and 48 h of perfusion (P < 0.03). Additionally, human saphenous vein segments were perfused for 24 h (n = 4) and viewed with laser confocal scanning microscopy and transmission electron microscopy; photomicrographs show typical vascular morphology. We conclude that the vascular perfusion system described herein is well suited for investigating the response of intact vascular tissue to hemodynamic variables.
This article has been cited by other articles:
![]() |
R. J. Jankowski, R. L. Prantil, M. O. Fraser, M. B. Chancellor, W. C. de Groat, J. Huard, and D. A. Vorp Development of an experimental system for the study of urethral biomechanical function Am J Physiol Renal Physiol, February 1, 2004; 286(2): F225 - F232. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Peng, F. A. Recchia, B. J. Byrne, I. S. Wittstein, R. C. Ziegelstein, and D. A. Kass In vitro system to study realistic pulsatile flow and stretch signaling in cultured vascular cells Am J Physiol Cell Physiol, September 1, 2000; 279(3): C797 - C805. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |