Vol. 273, Issue 4, H2044-H2061, October 1997
MODELING IN PHYSIOLOGY
Left ventricular pressure response to small-amplitude,
sinusoidal volume changes in isolated rabbit heart
Kenneth B.
Campbell1,2,
Yiming
Wu1,
Robert D.
Kirkpatrick1, and
Bryan K.
Slinker1
1 Departments of Veterinary and
Comparative Anatomy, Pharmacology, and Physiology and
2 Department of Biological Systems
Engineering, Washington State University, Pullman, Washington
99164-6520
 |
ABSTRACT |
The objective was to determine the dynamics of
contractile processes from pressure responses to small-amplitude,
sinusoidal volume changes in the left ventricle of the beating heart.
Hearts were isolated from 14 anesthetized rabbits and paced at 1 beats/s. Volume was perturbed sinusoidally at four
frequencies ( f ) (25, 50, 76.9, and 100 Hz) and five amplitudes (0.50, 0.75, 1.00, 1.25, and 1.50% of baseline
volume). A prominent component of the pressure response occurred at the
f of perturbation [in-frequency
response,
(t)].
A model, based on cross-bridge mechanisms and containing both pre- and
postpower stroke states, was constructed to interpret
(t).
Model predictions were that
(t)
consisted of two parts: a part with an amplitude rising and falling in
proportion to the pressure around that which
(t)
occurred
[Pr(t)],
and a part with an amplitude rising and falling in proportion to the
derivative of
Pr(t)
with time. Statistical analysis revealed that both parts
were significant. Additional model predictions concerning response
amplitude and phase were also confirmed statistically. The model was
further validated by fitting simultaneously to all
(t) over the full range of f and
V in a
given heart. Residual errors from fitting were small
(R2 = 0.978) and
were not systematically distributed. Elaborations of the model to
include noncontractile series elastance and distortion-dependent cross-bridge detachment did not improve the ability to represent the
data. We concluded that the model could be used to identify cross-bridge rate constants in the whole heart from responses to 25- to
100-Hz sinusoidal volume perturbations.
cross-bridge model; cross-bridge detachment; cross-bridge power
stroke; heart muscle
 |
INTRODUCTION |
PRESSURE RESPONSES to controlled volume perturbation of
the left ventricle (LV) of the beating heart have long been used to characterize ventricular mechanodynamics. These pressure responses have
been elicited using a variety of volume perturbation protocols including 1) sustained constant-flow
volume withdrawal over periods sufficient to achieve a given volume at
a specified time in the cardiac cycle (29-31, 35);
2) rapid small-volume withdrawal at each of several times during the cardiac cycle (1, 13, 25, 26, 38);
3) small-volume withdrawal of
varying amplitude and rate at the time of peak systolic pressure (4, 5,
9, 27); and 4) sinusoidal volume
change over the entire time course of the cardiac cycle (32, 33).
Analyses of the resultant data have related pressure to volume and
flow. Conclusions from all these analyses point to an organ with
complex mechanodynamics that varies with time over the course of the
cardiac cycle. These analyses have emphasized global features such as
chamber elastance, viscous resistance, and series elastance and have
made only indirect associations between these global features and the
underlying muscle properties responsible for them. Recently, there has
been an effort to interpret LV pressure responses directly in terms of
underlying muscle mechanisms (3, 5, 7, 27). The promise from these
early studies is that by using carefully controlled small-amplitude
perturbations and appropriate model-based analysis, detailed kinetic
behavior of cardiac muscle may be elucidated from observations of
pressure responses to volume changes in the whole heart.
In this study, we pursued that promise by using one of the original
volume perturbation protocols: a continuous high-frequency, small-amplitude, sinusoidal volume change delivered over the entire cardiac cycle (32, 33). We chose this protocol not only because of the
history of its previous use but also because it provides a means for
identifying underlying contractile behavior from observations in the
whole heart; these observations extend over the entire cycle period and
are not confined to a very brief 20-ms interval at the time of peak
isovolumic systole as in our previous studies (5, 27, 28). Furthermore,
we employed a cross-bridge model with pre- and postpower stroke
elastance states to describe, predict, and explain the observed
pressure responses. We concluded that these experimental and analytic
techniques could be used to extract information about underlying
cross-bridge mechanisms from observations made in the whole heart.
Glossary
| A1 |
Amplitude scaling factor for pressure response component that varies
proportionately with Pr(t)
|
| A2 |
Amplitude scaling factor for pressure response component that varies
proportionately with time drivative of
r(t)
|
| AIC |
Aikake Information Criterion
|
| Ap |
Amplitude of dynamic passive component
|
| b, d |
Rate constants governing formation and dissolution of prepower stroke
state
|
| Bn |
Amplitude of the nth harmonic of
Pd(t)
|
P(t) |
Pressure response to volume perturbation
|
Pd(t) |
Depressive component of pressure response
|
Pf(t) |
In-frequency component of pressure response
|
Pfa(t) |
Active part of in-frequency response
|
Pp(t) |
Dynamic passive component of pressure response
|
V |
Measured amplitude of volume perturbation
|
V(t) |
Time-varying volume perturbation
|
Vc |
Computer-commanded amplitude of volume perturbation
|
| Ee0(t) |
Elastance of pressure generators in the prepower stroke state
|
| Eep(t) |
Elastance of pressure generators in the postpower stroke state
|
| ESE |
Elastance of series-coupled noncontractile element
|
| ci |
regression coefficient
|
 |
Elastance of a single generator
|
| f |
Frequency of volume perturbation
|
| g |
Rate constant governing cross-bridge detachment
|
| h |
Rate constant governing power stroke
|
| K |
Number of parameters
|
lm/2 |
Change in half-mass wall circumference
|
| N |
Number of sampled data points
|
| Ne0(t) |
Number of pressure generators in the prepower stroke state
|
| Nep(t) |
Number of pressure generators in the postpower stroke state
|
| P(t) |
Pressure of perturbed beat
|
| Piso(t) |
Pressure of isovolumic (unperturbed) beat
|
| Pr(t) |
Pressure around which Pf (t)
occurred
|
| Q10 |
Relative rate of change with a 10°C increase in temperature
|
1 |
Phase of pressure response component that varies proportionately with
Pr(t)
|
2 |
Phase of pressure response component that varies proportionately with
r(t)
|
p |
Phase of dynamic passive component
|
| RSS |
Residual Sum of Squares
|
| SC |
Schwartz Criterion
|
| T |
Period of a heartbeat
|
n |
Phase of the nth harmonic of
Pd(t)
|
| VBL |
Baseline volume
|
| VW |
Wall volume
|
 |
Wall stress
|
 |
Angular frequency
|
| X0 |
Average isovolumic distortion of postpower stroke generators
|
| Xe0(t) |
Average distortion of prepower stroke generators
|
| Xep(t) |
Average distortion of postpower stroke generators
|
| Zep |
Total distortion among all ep generators
|
 |
EXPERIMENTAL METHODS AND PROCEDURES |
Experimental Preparation
Hearts were isolated from 14 adult male rabbits (avg wt = 3.1 kg).
Procedures for isolating the heart and attaching it to a volume-servo
device have been described in detail elsewhere (7, 19). Briefly, the
brachiocephalic artery was cannulated, and perfusion was begun with
oxygenated relaxing solution (in mM: 121.4 Na+, 35.0 K+, 137.4 Cl
, 0.1 Ca2+, 1.1 Mg2+, 21.0
, 0.36
, and 11.1 glucose and 2.5 U/l
insulin) to stop the heart before it was isolated from the animal. The
perfusate was oxygenated by vigorously bubbling with 95%
O2-5%
CO2.
The heart was transferred to a perfusion support system consisting of a
gas-exchange chamber, a roller pump, a constant-pressure chamber, and
an environmental chamber. The heart was placed within an environmental
chamber where the coronary arteries were perfused at 90 mmHg.
Temperature was kept constant at 30°C. The heart was submerged in
perfusate at all times by allowing the coronary effluent to accumulate
in the environmental chamber until it reached the chamber overflow at
the level of the base of the heart. The perfusate was not recirculated.
A thin latex balloon, secured to the piston cylinder of a volume-servo
system, was drawn into the LV chamber such that its tip was anchored
through a puncture in the apex, which also served as a vent for any
fluids between the balloon and chamber wall. A draw-string suture in
the mitral annulus was tightened around the obturator of a
piston-cylinder device, which secured the balloon in the LV
chamber. The balloon was filled with degassed distilled water until passive chamber pressure reached 10 mmHg. Balloons were
sized to fill the LV without excessive folding and without developing
pressure at the volumes encountered in these ventricles. Thus balloons
did not contribute to measured pressure.
The perfusing solution was changed from the relaxing solution to one
that allowed spontaneous beating (in mM: 148.4 Na+, 7.4 K+, 139.1 Cl
, 1.24 Ca2+, 1.1 Mg2+, 21.0
, 0.36
, and 11.1 glucose and 2.5 U/l
insulin). The heart, suspended in the spent perfusate, was paced at 1 Hz with field stimulation using 5-ms pulses of 15 mV and 250 mA from 5 cm × 5 cm copper plates placed 4.5-cm apart on either side of the
heart.
The volume-servo system consisted of a linear motor, a piston-cylinder
device, and a linear variable differential transformer (LVDT, model
0294-0000, Trans-Tek). The piston-cylinder device was a modified
5-ml glass syringe (East Rutherford Syringes) with two side ports. One
side port allowed calibrated infusion of fluid into the LV balloon to
establish a baseline volume
(VBL). The second port was used
to introduce a 5-Fr catheter-tip pressure transducer (Millar, Houston,
TX) into the balloon. The piston was driven by the armature shaft of
the linear motor. Motions of the piston produced LV volume changes
around VBL at a resolution of
0.001 ml. Both the pressure measurement system and the LVDT system had
frequency responses of 1 kHz.
Motion of the motor armature, and consequently piston motion, was
controlled to achieve specified changes in LV volume by feeding back
the position signal from the LVDT transducer, comparing it with a
reference position signal from a supervisory-control computer, and
passing the difference through an analog
proportional-integral-derivative compensator. Output from the
compensator was used to drive a high-current amplifier, which delivered
electrical current to the motor, causing piston position to match the
volume command.
The supervisory-control computer controlled experimental protocols
according to programmed instructions and also acquired data for later
analysis. Pressure and volume signals were amplified to make maximal
use of the 12-bit range of an analog-to-digital converter and were
acquired at a 2-kHz sampling rate.
Experimental use of animals was approved by the Animal Care and Use
Committee at Washington State University. The investigation conforms
with the Guide for the Care and Use of Laboratory
Animals published by the National Institutes of Health
(NIH publication No. 85-23, Revised 1985).
Protocols
A single-beat Frank-Starling protocol (7) was conducted to establish
VBL for each heart.
VBL was chosen as the volume equal to 80% of the volume at which maximum pressure was developed. This
protocol was also used to establish the passive pressure-volume relationship. A monoexponential equation was fit to points over the
range of end-diastolic pressure and volume values generated in this
protocol. Thus the contribution to pressure by parallel passive
structures at any volume was estimated and removed from all ensuing
data records in order to allow us to focus on just active contractile
properties.
After VBL was established, a
high-frequency volume perturbation protocol was conducted as follows.
Twenty pairs of data records consisting of pressure and volume signals
were taken. One record in a pair contained a single volume-perturbed
beat, and the other record contained an unperturbed beat that served as
a reference. Volume perturbation was administered only on a selected
single beat. On the perturbed beat, the linear motor was commanded to deliver a sinusoidal volume change at one of four frequencies (100, 76.9, 50, or 25 Hz, corresponding to periods of 10, 13, 20, or 40 ms)
and one of five amplitudes (0.5, 0.75, 1.0, 1.25, or 1.5% of
VBL). Repeated records of
perturbed beats were taken until all combinations of frequencies and
amplitudes (20 perturbed beats) were recorded. Pressure responses to
the volume perturbation were then analyzed.
Because the volume-servo system was underdamped, the actual volume
perturbation did not exactly equal the commanded sinusoid from the
supervisory-control computer. The frequency
( f ) of actual and commanded
signals was the same, but there were differences between actual and
commanded amplitudes (
V and
Vc, respectively), and there
was 1-2 ms delay in the actual signal relative to the commanded
signal. Consequently for some analyses (see below), each of the
measured volume perturbation signals was fitted with the analytic
function
|
(1)
|
where
was a phase relative to the recorded time window.
Equation 1 fitted all measured
V(t) with a correlation
coefficient (R2) > 0.99 and was thus judged to be an adequate representation of the actual
perturbation signal for specific analyses. The underdamped character of the volume-servo system produced an actual
V that, for
a given
Vc, increased with
frequency over the 25- to 100-Hz frequency range with the result that
the actual
V at 100 Hz was 145% of that at 25 Hz. Thus
V from
the fit to the measured signal was used rather than the commanded
Vc in all data analyses.
After the high-frequency volume perturbation protocol, a second
single-beat Frank-Starling protocol was conducted to generate a
Frank-Starling curve that could be compared with the one collected previously. This allowed detection of any deterioration of the preparation during the course of an experiment. No detectable deterioration occurred.
 |
PRESSURE RESPONSE |
Peak isovolumic pressure generated by these 14 hearts (averaged over
all 280 observations) was 120.2 ± 13.6 (SD) mmHg at an average VBL of 2.11 ± 0.09 ml.
The average LV weight, including the septum plus LV free wall, was 5.96 ± 0.54 g.
The pressure response
[
P(t)] to
V(t) was defined as the
difference between active pressure of the reference isovolumic beat [Piso(t)],
i.e., the pressure that would have developed had no volume perturbation
been administered, and active pressure of the perturbed beat,
P(t)
|
(2)
|
Representative
Piso(t),
P(t), and
P(t) are shown in Fig.
1 (f = 50 Hz,
Vc = 1%
VBL). All responses
[
P(t)] contained two components: a depressive response,
Pd(t)
[called "depressive" because it represented a sustained
decrease in pressure below Piso(t)
that was not at the perturbation frequency] and an in-frequency response
[
(t)]
(that part of the response at the perturbation frequency). Thus
|
(3)
|
The
Pd(t)
was extracted from
P(t) by
fitting a curve to
P(t) that did
not contain frequency content of the perturbation frequency.
Pd(t)
was taken as the sum of the first 10 harmonics in the Fourier series
|
(4)
|
where
n is the harmonic number, Bn and
n are harmonic amplitude and phase,
respectively, of n, and
T is the heart period. Because the
shortest heart period used in these studies was 1 s, the 10th harmonic
(10 Hz) was well below 25 Hz, the lowest frequency used for volume
perturbation. The amplitude and phase parameters for the
ith harmonic
(Bi and
) had no particular
significance other than to give a
Pd(t)
waveshape, identifiable within
P(t), that did not include
components of the in-frequency response. Once
Pd(t) was identified by fitting with Eq. 4,
it was subtracted from
P(t) to
yield
(t).
Subtraction of
Pd(t)
from
Piso(t) generated a signal representing the pressure around which
(t)
took place
[Pr(t)].
Pr(t),
with corresponding
P(t),
Pd(t), and
(t),
is shown in Fig. 1.

View larger version (18K):
[in this window]
[in a new window]

View larger version (14K):
[in this window]
[in a new window]
|
Fig. 1.
Method of determining pressure response [volume
perturbation = 50 Hz, 1% baseline volume
(VBL)].
A: 2 left panels:
Piso(t),
pressure of an isovolumic beat in which no volume perturbation was
applied; P(t), pressure of a beat
that received volume perturbation. Middle panel:
P(t), pressure response to volume
perturbation [= P(t) Piso(t)].
Right panel:
Pd(t),
depressive component of P(t)
obtained by low-pass filtering P(t). Bottom panel:
(t),
in-frequency component of P(t)
[= P(t) Pd(t)].
B: pressure around which in-frequency
response occurred
[Pr(t)]
was obtained by subtracting
Pd(t)
from
Piso(t).
Vertical scale is in mmHg, horizontal scale is in s.
|
|
This report concerns just
(t);
the
Pd(t)
is the subject of another report (unpublished observations). Twelve of
twenty
(t)
responses obtained in one heart are shown in Fig.
2.
(t)
rose and fell during the course of the heartbeat but also contained a
small contribution that was present during diastolic periods when there
was no active contraction. This small component was assumed to be due
to dynamic features of parallel passive properties that were not
included in the static passive pressure-volume relationship, which had already been subtracted from the response. Furthermore, these passive
dynamic properties were assumed to be expressed continuously throughout
the period of the heartbeat and, for a given
V(t) sinusoid, they were
represented as
|
(5)
|
This
passive dynamic response was subtracted from the response signal using
data-fitting procedures described below. The great majority of
(t)
waxed and waned as pressure rose and fell and was considered to be an
expression of active processes. Thus
|
(6)
|
where
(t)
was the contribution of active process to
(t).

View larger version (33K):
[in this window]
[in a new window]
|
Fig. 2.
(t)
over single heartbeat for 12 of 20 responses in one heart, showing
extremes and midrange of responses to various amplitudes and
frequencies of V(t). Note growth
in amplitude of response with both amplitude and frequency of
V(t).
A: 0.5%;
B: 1.0%;
C: 1.5%.
|
|
It is clear from Fig. 1 that, in accordance with results from several
studies (1, 13, 25, 29, 32, 38),
(t) waxed and waned during the heart period as
Pr(t)
rose and fell. One objective of the current work was to predict from
model considerations whether other time-varying components were
contained within
(t) and, then, to test these predictions.
 |
MODEL DESCRIPTION |
A model for describing and predicting the active part of the pressure
response
[
(t)]
was constructed on the assumption that elements responsible for force
generation in cardiac muscle (i.e., cross-bridges between thick and
thin myofilaments) were also responsible for pressure generation in the
LV chamber. Furthermore, it was assumed that there was a
straightforward linear transformation between force-length
relationships in the wall of the heart and pressure-volume
relationships in the LV chamber. Acceptability of the linear
transformation assumption requires small-amplitude perturbations and
homogenous myocardium. Criteria for satisfying the small-amplitude
requirement are detailed in the
APPENDIX. Evidence that the homogenous
myocardium requirement is satisfied is given in the findings relative
to the unimportance of noncontractile series elasticity in these hearts
(see MODEL VALIDATION). However, a strong
reason for employing the linear transformation assumption is the
success that has been achieved with its application in earlier studies
(3, 5, 6, 27).
It was further assumed that during a heartbeat mechanodynamics were
from two sources: 1) dynamics of
activation as activator Ca2+ comes
and goes and numbers of force-bearing cross-bridges rise and fall, and
2) dynamics of cross-bridge cycling
as myosin heads cyclically attach to and detach from the actin binding
site. In accordance with an earlier hypothesis (6), we argue that the only dynamics expressed within the brief cycle period of frequencies
25 Hz were those associated with steps in the cross-bridge cycle and
that the dynamics of activation were too slow to contribute to changes
within these brief time periods. Such separation of time scales in the
study of muscle dynamics is in accordance with analyses conducted by
Kawai and co-workers (17, 24, 39) and in accordance with our recent
demonstration that cooperativity between force-bearing cross-bridges
and activation can cause activation to be slow relative to cross-bridge
dynamics (2).
We refer to cross-bridges as force generators. Generators contributing
to the pressure response were assumed to be in two states:
1) a state that possessed elastance
but did not, under isometric (isovolumic) conditions, generate pressure
(state
e0), and
2) a state that both possessed
elastance and also generated isometric (isovolumic) pressure
(state
ep). When we assume linear, independent, and parallel generators, the elastance associated with
each state is the number of parallel generators in that state (N) times the elastance of a single
generator (
). Assuming that all generators in
states
e0 and
ep possess the same
, we show the
net elastance of all parallel generators in each of the two states as
|
(7)
|
|
(8)
|
where,
because of our linear transformation assumption,
Ee0(t)
and
Eep(t)
may be taken as volumetric elastances (with units of mmHg/ml).
Generators are in continual transition as they progress from one state
to another in the cross-bridge cycle (Fig.
3). We assumed that
state
e0 preceded
state
ep. Furthermore, transitions into, out
of, and between states
e0 and
ep were assumed to be governed by rate
constants b,
d, g,
and h.
State
0 in Fig. 3 is without elastance and
is a precursor to state
e0;
state
00 is also without elastance and
follows state
ep. These nonelastance states
represent all other states needed to complete a cross-bridge cycle.
Given these relationships between states and assuming that there is no
noncontractile series elastance (4), it is shown in the APPENDIX that
Ee0(t)
and
Eep(t) may be calculated from
Pr(t)
and its first time derivative
[
r(t)] according
to
|
(9)
|
and
|
(10)
|
where
X0 is a parameter
representing average volumetric distortion among generators in
state
ep during isovolumic conditions. The
transitional step between states
e0 and
ep, which is governed by
h, is the cross-bridge power stroke,
and this step is responsible for inducing
X0 distortion in
generators as they go through the power stroke to enter the
ep state. The dissolution of the
postpower stroke (ep) state, which
is governed by g, is the cross-bridge detachment step.

View larger version (17K):
[in this window]
[in a new window]
|
Fig. 3.
Schematic drawing of states in cross-bridge cycle.
Ni is the
number of generators in ith state.
Generators in states e0 and
ep possess elastance, whereas
generators in states 0 and
00 do not. Generators possessing
elastance may be distorted during a volume perturbation such that both
contribute to pressure response. Under isovolumic conditions,
generators enter state e0 without distortion and do not
generate pressure. Transitions between states are governed by rate
constants b,
d, h,
and g. Transition between
state e0 and
state ep is the power stroke and induces a
baseline distortion in postpower stroke
(ep) generators, which, as a result
of their elastance, causes development of isovolumic pressure.
Isovolumic pressure is modified during a volume perturbation by induced
distortion in postpower stroke and prepower stroke states,
ep and
e0, respectively, and by whatever
influence volume perturbation has on recruitment of generators into or
out of cross-bridge cycle.
|
|
Because of their elastic nature, generators within each of these states
are distorted during volume perturbation. The net volume-induced
distortion is determined by the rate of volume change relative to the
rates of formation and dissolution of the respective states.
Differential equations for these volume-induced distortions
[
Xe0(t)
and
Xep(t) in states
e0 and
ep, respectively] are derived
in the APPENDIX as
|
(11)
|
|
(12)
|
A
dot over a variable indicates its first time derivative. In words,
Eqs. 11 and 12 state that the time rate of change
of generator distortion is negatively related to distortion itself and
is positively related to the first time derivative of volume. Thus
volume-induced distortion is driven directly by flow (velocity) and
dies away at a rate proportional to the distortion.
Given these relationships, the model equation predicting the active
part of the in-frequency pressure response
[
fa(t)] is
|
(13)
|
where
the "hat" in

fa(t)
indicates that the quantity is model predicted.
Equations 9-13 constitute a
2-state, 4-parameter model. In this model,
V(t) is the input [although
distortion is driven directly by

(t)];
Ee0(t),
Eep(t), and their time derivatives combine to form time-varying parameters;
Xe0(t)
and
Xep(t)
are state variables; and

fa(t)is
the output variable. Equations 11 and 12 are a set of linear, uncoupled,
first-order, time-varying differential equations.
 |
MODEL PREDICTIONS |
Two Dynamic Components of In-Frequency Response
In-frequency response consists of two dynamic components:
1) a component with an amplitude
varying with
Pr(t)
and 2) a component with an amplitude
varying with the time derivative of
Pr(t).
The model output equation Eq. 13 can
be rearranged by substituting elastance Eqs.
9 and 10 into
Eq. 13 to create an alternative
formulation.
|
(14)
|
The first term on the right-hand side of Eq. 14 is a response component with an amplitude varying
proportionately with
Pr(t), and the second term on the right-hand side is a component with an
amplitude varying proportionately with
r(t).
This development clearly identifies the contribution of the prepower
stroke, e0 state as the sole source of
the dynamic response component with an amplitude rising and falling in
proportion to the derivative of the pressure around which the response
is occurring.
To determine the relative roles of these two components, an approximate
sinusoidal solution of the model equations Eqs.
11 and 12 was
developed as follows. When we ignore the influence of the time-varying
part of the coefficients in Eqs. 11 and 12, a steady-state solution of
these equations for
Xe0(t)
and
Xep(t)
when
V(t) is a volume sinusoid of frequency f results in an expression for the
first and second terms on the right-hand side of Eq. 14 as
|
(15)
|
Such that for a volume sinusoid of frequency
f
|
(16)
|
The relative role of the
r(t)
component in the observed responses was determined by an incremental
approach. Equations 6 and 16 were combined and fit to the
observed
(t) by first excluding the
r(t)
component
|
(17)
|
and then including the
r(t)
component
|
(18)
|
Fitting of Eqs. 17 and 18 to
(t)
was by an heuristic search algorithm (Levenberg-Marquardt algorithm,
Argonne National Laboratory) to minimize the residual sum of squares
(RSS).
To test for degradation or improvement in the representation of
(t)
signal with the addition of the two additional parameters in
Eq. 18 that relate to the
r(t)
component, the Aikake Information Criterion (AIC) and the Schwartz
Criterion (SC) were calculated from fits with both
Eqs. 17 and 18 according to Landaw and DiStefano
(20) as
|
(19)
|
where N is the number of
sampled data points, and K is the
number of parameters. Note that the first term on the right-hand side
of Eq. 19 is a measure of how well the
model fit the data, whereas the second term is a penalty function based
on the number of model parameters. Therefore, increasing the number of
parameters increases AIC and SC unless there is a more than
compensating reduction in RSS. In considering two competing equations
such as Eqs. 17 and 18, the better representation is the
one with the smallest AIC and SC. To further determine whether
significant reduction in the RSS occurred with Eq. 18 compared with Eq. 17, an incremental
F-test was used (11).
When fit to
(t)
of each of the 280 data records obtained in these hearts (14 hearts
times 20 records/heart), both Eq. 17,
which did not include the
r(t)
component, and Eq. 18, which did
include this component, fit
(t) very well with median
R2 of 0.980 and
0.981, respectively. The contribution of the
r(t) component was quite small as judged by the fact that
A2 was always two-orders of magnitude less than
A1. However,
despite this small contribution, the inclusion of the
r(t)
component in Eq. 18 consistently reduced the AIC (only one exception in 280 instances; median reduction:
0.78%; range: 0.02 to
10.1%) and SC (14 exceptions in
280 instances; median reduction:
0.63%; range: 0.09 to
9.93%), suggesting that the addition of the two parameters
associated with the
r(t) component improved the representation of the information in the
(t)
signals. Furthermore, of the 280-response records analyzed, the
incremental F-test generated an
F-statistic that was significant at
the P < 0.01 level in all 280 instances. Thus, although making only a small contribution in
accounting for the total variability in
(t),
the
r(t) component contributed significantly to representing its information content and in reducing the RSS.
To summarize these results, the model predicted that there would be an
in-frequency response component with an amplitude rising and falling in
proportion to the pressure around which the response occurred
[Pr(t)]
and another component with an amplitude rising and falling in
proportion to the derivative of the pressure around which the response
occurred
[
r(t)].
Analysis of all the response data revealed that the response was
dominated by the
Pr(t)
component, although a small but significant component existed with an
amplitude of which was proportional to
r(t).
Given the very small contribution by the
r(t)
component, the model was further used to test the importance of
including this term in validation procedures described below.
Amplitude Ratio and Phase of Response
Additional model predictions resulted from considering the nature of
the response just around the time of peak
Pr(t),
when
r(t)
approximated zero. When transients are ignored and it is assumed that
steady state had been achieved at this time, an argument can be made
that during a short interval around the time of peak pressure, the
e0 state does not contribute to the
response and the model reduces to
|
(20)
|
Input-output relationships between
Xep(t)
and
V(t) for this reduced model
may be derived as
|
(21)
|
where
is angular frequency in radians and equals
2
f,
|
(
)/
V(
)|
is the magnitude of an input-output amplitude ratio equivalent to
A1/
V from the
previous sinusoidal fits, and
(
) is the phase difference between
output pressure response and input volume sinusoid.
Two predictions result from Eq. 21:
1) the amplitude ratio will increase
with frequency up to some plateau, provided the frequencies examined
are in the vicinity of the characteristic frequency, g. At frequencies either far below or
far above g, the amplitude ratio will
change only weakly with frequency.
2) The phase of
(t)
will lead the phase of
V(t) by as
much as 90° at low frequencies, but this phase lead will decline
and approach zero as frequencies increase above
g.
To test these model predictions, the amplitude ratio,
A1/
V, was evaluated for
its dependence on f and
V. As noted above, A1/
V will change sharply with
f over a frequency range around the
characteristic frequency of the underlying process. Additionally, A1/
V is not expected to be dependent on
the amplitude of the
V input. Any dependence of
A1/
V on the amplitude of
the input is an indication of nonlinear processes that are not part of
the current model. Nonlinearities may also show up as dependence of A1/
V on product
combinations of f and
V. Stepwise
regression analysis was used for these determinations. Regression
equations were formulated as
|
(22)
|
where
the ci
values are regression coefficients and
(f,
V) represents one or more of
four candidate interaction terms:
f · (
V),
root-mean-squared (rms) flow;
f 2 · (
V),
rms acceleration;
V2, squared
rms volume amplitude; and
(f ·
V)2,
squared rms flow amplitude. The regression procedure used dummy variables and effects coding to account for between-subjects
differences, and the subject dummy variables were forced into the
stepwise regression (11). A candidate predictor variable was considered significant only when the P value for
its inclusion was <0.05.
The dependence of
A1/
V on
f at the various commanded
Vc for one heart is shown in
Fig. 4. At all
Vc,
A1/
V increased
with f. Furthermore, at
f equal to 25, 50, and 76.9 Hz,
A1/
V had
virtually no dependence on
Vc;
there was an apparent small dependence on
Vc at 100 Hz. Regression
analysis of pooled data from all hearts revealed that, of all potential
predictor variables, there was a significant dependence on
f and
V2; the simplest best
regression equation (leaving out terms for between subjects
variability) was
|
(23)
|
No interaction terms were found to be significant in this
regression. Of the two significant predictor variables,
f was the more important variable. For
example, at f = 50 Hz and
V = 1%, the term in Eq. 23 due to
f adds 0.55 units to
A1/
V, whereas
the term due to
V2 subtracts
only 0.03 units, an 18-fold difference in sensitivity of
A1/
V on these
variables. The three important outcomes of this regression analysis are
1)
A1/
V increased
with f in accordance with

(t) acting as
the effective driving function; 2)
because of the strong dependence on f,
the characteristic frequency of the dynamic processes responsible for
the
Pr(t)
response component lies either within or not far distant from the 25- to 100-Hz frequency range; and 3)
the very small dependence of the
Pr(t)
response component on
V indicated that nonlinearities were not a
large part of the underlying processes.

View larger version (7K):
[in this window]
[in a new window]
|
Fig. 4.
Amplitude ratio
(A1/ V) of
(t)
component proportional to
Pr(t)
as a function of frequency (f) and
commanded amplitude ( Vc) of
perturbation. , Vc = 0.50%
VBL;
+,
Vc = 0.75%
VBL; ×,
Vc = 1.00%
VBL;
* Vc = 1.25%
VBL; ,
Vc = 1.50%
VBL.
A1/ V increases
with frequency and has very little dependence on
Vc, as predicted by model.
|
|
Predictions with regard to the phase lead were not analyzed
exhaustively. Rather, single cycles, spanning the time of peak Pr(t)
of response to 1%
Vc at each
frequency, were examined. In these, the phase of
(t)
led that of
V(t) at 25 Hz by
~30-40°, and this phase lead decreased progressively to
approach zero at 76.9 and 100 Hz.
To summarize these results, model predictions with regard to frequency
dependence of input-output amplitude ratio and phase relations were
confirmed. These indirect confirmations of the model suggested a more
rigorous model validation test.
 |
MODEL VALIDATION |
Ability to Fit the Data
Model validation was, in part, by evaluating how well the model fit the
full time course of the pressure response over an entire cardiac cycle.
Model fitting was by the following procedure. Initial values of the
four model parameters (g,
h, d,
and X0) and the
two dynamic passive pressure parameters
(Ap and
p) were assigned. Derivatives
of measured
V(t) and
Pr(t)
were calculated using a five-point Lagrangian polynomial method.
Measured
Pr(t), the calculated derivatives, and the then-current parameter values were
fed into the differential equations Eqs.
11 and 12, allowing these equations to be solved numerically by integrating with a fourth-order Runge-Kutta algorithm (integration step size = 0.0005 s)
to obtain predictions of
Xe0(t)
and
Xep(t). These were then used with measured
Pr(t)
and its derivative to compute a
(t)
according to Eq. 13 and a
Pp(t) according to Eq. 5, which were then
added to obtain a
(t).
The RSS between predicted and observed
(t) was calculated. Values of model parameters (g,
h, d,
and X0) and dynamic passive pressure parameters
(Ap and
p) were then adjusted according to the rules of a Levenberg-Marquardt heuristic search algorithm, and the processes were repeated iteratively until RSS was
minimized. Median model parameter values obtained with this procedure
in the 14 hearts are reported in Table 1.
Unlike the sinusoidal approximation of Eq. 18, which could be fit only to individual responses to
a single perturbation, model Eqs.
9-13 are more general and could be fit to groups
of responses to perturbations of multiple frequencies and amplitudes.
By fitting simultaneously to responses to the 20 perturbations imposed
in any one heart, we required this single model with a single set of
parameters to account for a wide range of behaviors resulting from many
different perturbations as in Fig. 2. Successful reproduction of this
broad range of dynamic responses was taken as compelling evidence for
validity of the basic model.
By all measures, the basic 2-state, 4-parameter model fit the response
data very well. An example of this good fit in one heart is shown for a
single heartbeat ( f = 50 Hz;
Vc = 1%
VBL) in Fig.
5. In evaluating Fig. 5, it must be kept in
mind that the fit that generated the predicted response was to
responses obtained from the complete set of four frequencies and five
amplitudes of inputs delivered to each of 20 beats and not just to the
single beat shown in Fig. 5. Because rapid cycling at 50 Hz generated a
dense pattern on display from which model-predicted and measured waveforms could not be discriminated, individual cycles in the response
were identified that 1) spanned the
point on the ascending limb of
Pr(t)
at which
Pr(t) = 1/2 its peak value, 2) spanned the
peak value of
Pr(t),
and 3) spanned the point on the
descending limb at which
Pr(t) = 1/2 its peak value. These three cycles were then expanded in
row B of Fig. 5 such that predicted
and observed waveforms could be compared. The comparatively small
values of the differences between predicted and observed waveforms
(residuals) are given in row C of Fig.
5. In this particular example, but not true in all cases, the residuals
appeared to be random at all times during the heart period and
exhibited no transient systematic character. Systematic patterns in the
residuals will exhibit as a periodicity at the frequency of
perturbation.

View larger version (28K):
[in this window]
[in a new window]
|
Fig. 5.
A: response to a single perturbation
f = 50 Hz;
Vc = 1%
VBL. Model was fit to complete set
of 4 frequencies and 5 amplitudes of inputs delivered to 20 beats, not
just to single beat shown. B:
model-predicted vs. measured waveforms. These overlain waveforms cannot
be discriminated in left panel. Waveforms of single cycles (identified
by period between pairs of vertical lines in left panel) are displayed
in 3 panels to right. These cycles spanned:
1) the point on ascending limb of
Pr(t)
at which
Pr(t) = 1/2 its peak value, 2) the peak
value of
Pr(t),
and 3) the point on descending limb
at which
Pr(t) = 1/2 its peak value. C: residuals
between model-predicted and observed waveforms. Residuals are small
valued and do not have periodicity of perturbation. Thus, in this
example, they are apparently random.
|
|
To demonstrate the goodness of the fit over the full set of 20 perturbed beats collected in the single heart featured in Fig. 5,
predicted and measured values were plotted against one another to
generate Fig. 6. In keeping with an
R2 value of 0.98 in this heart, all the points cluster tightly around a line that is not
clearly distinguishable from the line of identity. It can be seen that
deviations of predicted from observed
(t) were never large, and an excellent fit was achieved at all frequencies and amplitudes of perturbation.

View larger version (17K):
[in this window]
[in a new window]
|
Fig. 6.
Model predicted vs. measured
(t)
for all 20 perturbed beats in one heart; 40,000 points shown (20 beats,
each of 1-s duration, sampled at 2 kHz). Points in densest part of
cluster around origin represent data generated at smallest
Vc, at lowest
f, at lowest values of
Pr(t),
and during zero crossings at all perturbations. Points in less dense
vertices of cluster represent data generated at highest
Vc, at highest
f, and at peak of
Pr(t).
All points cluster tightly around line of identity. Deviations of
predicted from observed
(t)
were never large, and an excellent fit was achieved at all frequencies
and amplitudes of perturbation.
|
|
An additional demonstration of model goodness of fit comes from
comparing the actual responses displayed in Fig. 2 with the corresponding model-predicted responses displayed in the same format in
Fig. 7. Visual comparison of these two
figures reveals no important differences. Yet another demonstration of
correspondence between model prediction and measured responses is seen
in Fig. 8, where it is shown that there was
good agreement between
A1/
V for all
20 responses from the sinusoidal analysis and the 20 model-derived equivalents. This agreement was with respect to both absolute magnitudes and to systematic variations with
f and
V, strong dependence on
f, and little or no dependence on
V.

View larger version (30K):
[in this window]
[in a new window]
|
Fig. 7.
Alternative view of some of data plotted in Fig. 6. Twelve of 20 model-predicted responses are shown, corresponding to 12 measured
responses shown in Fig. 2. A: 0.5%;
B: 1.0%;
C: 1.5%.
|
|

View larger version (7K):
[in this window]
[in a new window]
|
Fig. 8.
, Model-predicted equivalent of
A1/ V for each
of 20 volume perturbations. There is virtually no dependence of
model-predicted
A1/ V on
perturbation amplitude but a strong dependence on frequency. ,
A1/ V from
sinusoidal analysis, i.e., same data shown in Fig. 4.
|
|
That good fits were obtained in all 14 hearts is supported by a median
R2 value of 0.978 with a narrow range between 0.968 and 0.987. These R2 values were
slightly smaller than the
R2 value obtained
when the sinusoidal approximation (Eq. 18) was fit to a single response; these
R2 values often
exceeded 0.98. However, it must be kept in mind that the sinusoidal
equation was fit individually to each of the 20 responses in each
heart, and the resultant
R2 values
reflected the ability of the sinusoidal equation to reproduce only the
variability seen in a single response with no consideration of the
causal relation between
V(t) and
the response. This contrasts with the
R2 obtained with
the model in which causal relationships between
V(t)and
(t)
were stipulated and in which the model was asked to reproduce not just
the variability in a single response but rather the variability due to
changes in
V and f in the input
V(t) over the whole family of 20 responses in each heart. Given the resulting range of response
variation to these broad ranges in input
V and
f (Fig. 2), the median
R2 value of 0.978 obtained with the basic model is remarkable.
This good fit by the basic model left only 2.2% of the total variation
in the data unaccounted. Of this 2.2% it was very difficult to
ascertain how much was simply random noise that could not be reduced
with model improvement and how much was systematic and subject to
reduction with model improvement. Systematic error would appear as
systematic variation in the residuals at the frequency of perturbation,
whereas random error would appear as residual variation unrelated to
the frequency of perturbation. Unfortunately, the residuals did not
lend themselves to analysis because systematic error, when it could be
observed, did not persist throughout the period of the cardiac cycle.
Rather, detectable systematic error would appear briefly during, say,
just early systole or just during late systole and then disappear into
apparently random error for the rest of the cycle. Such comings and
goings of systematic error made it impossible to quantify the degree to
which the residuals were random. Qualitatively, our impression was that
systematic variation in the residuals was only a small part of the
total residual variation. Because there was only a very small
systematic error, our impression was that there was very little room
for improvement with model modifications. However, results of this analysis do not eliminate the possibility that other features are
present or even that a simpler model may be as good.
Basic Model vs. Competing Alternatives
Further model validation was confirmed by comparing the basic model
against both simpler and more complex alternatives.
Simpler model, without the prepower stroke state, does
not work as well. Because of the very small
contribution of the
r(t) component, it was important to test whether the 2-state, 4-parameter model, which predicted that there would be a
r(t)
response component, was actually needed, or whether a simpler model
would suffice. A model that discounts the contribution by prepower
stroke state e0 and considers only contributions
from the postpower stroke state
ep is more parsimonious. Such a model
predicts a response that is composed of a single component proportional
only to
Pr(t) and possesses just 1 state and 2 parameters as given by
|
(24)
|
|
(25)
|
The simple 1-state, 2-parameter (Eqs.
10 and 24-25)
and the basic 2-state, 4-parameter (Eqs.
9-13) models were each fit to the data and
compared to determine which one was the best based on reduction in the
AIC and SC and on the incremental
F-test as described above.
The basic 2-state, 4-parameter model was superior to the simple
1-state, 2-parameter model in fitting the data. The improvement in
median R2 for
data from the 14 hearts was on the order of 3% (from 0.951 to 0.978).
Smaller AIC and SC were associated with the 2-state, 4-parameter model
in all 14 hearts with a median reduction of 8.5% (range =
1.22
to
13.6%) and 8.5% (range =
1.22 to
13.6%), respectively. Furthermore, the incremental
F-test generated an F-statistic that was significant at
P < 0.01 in all 14 hearts. These are
strong results for accepting the 2-state, 4-parameter model over the
simpler 1-state, 2-parameter model. Another way of stating the outcome
of the incremental F-test is that the
precision of the estimates of parameters remain within an acceptable
range in the 2-state, 4-parameter model. Precision of parameter
estimates were quantified by using the standard error of parameter
estimate (11) using the RSS and covariance information returned from the optimization algorithm. Over all 14 hearts, the median and range of
the standard error of the estimates of
X0,
g, h,
and d were (expressed as percentage of
the respective parameter estimate) 0.19 (0.08-0.44), 0.38 (0.00-0.65), 1.88 (0.78-31.32), and 0.60 (0.20-26.93).
The heart that generated the upper extremes in the ranges for
h and
d was an outlier among these 14 hearts
because the heart with the next highest standard error of the estimate for these two parameters exhibited values of 8.48 and 8.60%,
respectively. Except for estimates of
h and
d in one heart, the respective
parameters were generally estimated with more than adequate precision.
Despite these results, certain caveats in representing the prepower
stroke e0 state in the model must be
made clear. In contrast to the good correspondence between
A1/
V and its
model-derived equivalent (Fig. 8), which is a response primarily
attributable to the postpower stroke state, only a rough correspondence
was obtained between
A2/
V and its
model-derived equivalent (Fig. 9), which is
a response attributable to the prepower stroke state. The order of
magnitude of values of
A2/
V and its
model-derived equivalent were the same, but dependencies of these
quantities on f and
V were not the
same: there was no dependence on f and strong dependence on
V for
A2/
V, whereas
there was weak dependence on f and no
dependence on
V for the model-derived equivalent. Employing the
procedures used in the derivation of Eq. 25, we show that an approximate model-derived
equivalent of
A2/
V is
|
(26)
|
where
|
(27)
|
Because
median h + d was 2,555 s
1 (see Table 1), at 100 Hz
|
(28)
|
This
model approximate value, which is on the order of values found for
A2/
V, is more
than two orders of magnitude less than that for
A1/
V,
implicating virtually no role for the prepower stroke state. However,
the contribution of the prepower stroke state is not only determined by
the relative values of
A1/
V and A2/
V, but also
because the amplitudes of components associated with pre- and postpower
stroke states depend on the relative values of
Pr(t)
and
r(t),
which change with time over the cycle. A more valid assessment of
contributions by the prepower stroke state is in the ratio of
A2
r(t)/A1Pr(t).
This ratio varied with frequency, but representative values ranged
between 1 and 5%. Even with these considerations, the contribution by
the prepower stroke state is relatively small. Thus despite the
statistical evidence suggesting good precision of the estimates in most
cases and significant reduction in RSS, AIC, and SC, the prepower
stroke state is contributing very little to the response and caution must be exercised in placing unqualified confidence in the estimates of
associated parameters h and
d.

View larger version (10K):
[in this window]
[in a new window]
|
Fig. 9.
, Model-predicted equivalent of
A2/ V for each
of 20 volume perturbations. There is virtually no dependence of
model-predicted
A2/ V on
perturbation amplitude and only a weak dependence on frequency. ,
A2/ V from
sinusoidal analysis. There is only a rough correspondence between
model-predicted and sinusoidal
A2/ V.
|
|
The reason for the relatively small dynamic contribution of the
prepower stroke state in the 25- to 100-Hz range is the very high value
of the characteristic frequency of responsible underlying processes
h + d = 2,555 s
1. As a
consequence, the prepower stroke state is very labile and, once a
generator enters this state, it does not last long enough to undergo
appreciable distortion when the volume is perturbed at frequencies in
the 25- to 100-Hz frequency range. It would take frequencies in the
400-Hz (
2,555/2
) range to generate appreciable distortion in the
e0 state and make a sizable
contribution to the response. Consequently, the
e0 state makes only a small
contribution to the response over the frequencies used in these
studies, and the associated parameters are estimated with less
precision than those associated with the postpower stroke state.
There were additional consequences to incorporating the prepower stroke
state in the model. Including this state doubled the estimated
g from a median of 69.0 s
1 in the 1-state,
2-parameter model to 115.6 s
1 in the 2-state,
4-parameter model. Also, including this state increased the robustness
of the parameter estimates, making their values less sensitive to the
manner in which the data were prepared for analysis, i.e., values of
filter settings, complete removal of trends, etc. (results not shown).
We concluded that although it was important to include contributions
from the prepower stroke state to account for variability in the data
and for achieving robustness in the estimates of the parameters, we
could identify only the rough magnitude of the underlying dynamic
processes, because it made only small contributions to the overall
response at even the highest frequency used in these studies (100 Hz).
Addition of noncontractile series elastance did not
improve model. An important assumption in the model was
that only cross-bridges were contributing to active chamber properties,
i.e., there was no contribution by noncontractile structures in
parallel or in series with the cross-bridge force generators.
Contributions by noncontractile parallel structures were identified
from passive behavior and removed by previously described methods. Our
previous findings (4) indicated that there was no important
noncontractile series elasticity participating in LV mechanical
behavior to rapid ramp changes in volume. However, there was a need to
confirm that this was so for sinusoidal volume changes.
To test for the contributions by noncontractile series elastic
elements, the basic 2-state, 4-parameter model was compared with
an elaborated model containing series elastance and consisting of
3 states and 5 parameters. The third-state variable was
required to account for internal shortening during
Pr(t),
and the fifth parameter was a noncontractile series elastance.
Equations for the model with series elastance are given in the
APPENDIX. In addition to the extra
state and parameter, these equations differed from the basic model
Eqs. 9-13 in that the two
distortional variables
Xe0(t)
and
Xep(t)
were no longer uncoupled, i.e., the differential equations describing the derivatives of each of these variables depended on both variables. The model with series elastance was fit to the data and then compared with the basic model without series elastance using the AIC, SC, and
the incremental F-test.
The improvement in fit when series elastance was added to the basic
model was scarcely observable, the median
R2 value was
0.978 both with and without the series elastance. Furthermore, the AIC
was reduced in only 9 of 14 hearts, and the SC was reduced in only 8 of
14 hearts. The incremental F-statistic
was significant in 8 of 14 hearts (P < 0.05). In the six hearts where the series elastic element was found
not to be significant, the estimated value was
>1010 mmHg/ml. In these six
hearts, the model was equally well fit with any large value of series
elastance, leading to considerable imprecision in parameter estimates
and correspondingly insignificant reduction in RSS. In the eight hearts
where series elastance was found to be significant, the average
estimated value was 17,892 mmHg/ml. This compares with an average peak
generator elastance [i.e., the maximal value of
Ee0(t) + Eep(t)] of 411 mmHg/ml. Thus when it was found to be significant (in roughly half of all hearts), series elastance was 43 times greater than maximal
generator elastance and, consequently, played only a minor role in
chamber mechanical properties. Furthermore, the addition of the series
elastance element did not appreciably change the estimated value of the
other model parameters (Table 1). In accordance with previous findings
(4), we concluded that there was no important role for noncontractile
series elastance in these hearts and no compelling reason for
complicating the basic model by incorporating noncontractile series
elastance.
Incorporation of distortion-dependent cross-bridge
detachment did not improve model. Another fundamental
model assumption was that the kinetic rate constants were truly
constants and did not depend on behavior such as distortion of the
generators. There are several reasons why this assumption required
testing. First, distortion-dependent, cross-bridge rate constants have
been a central tenet in much of the theoretical work in muscle
mechanics (14) and have been a major feature of several models to
explain aspects of mechanical behavior (10, 15, 22, 23, 34). Second,
enhanced cross-bridge detachment (i.e., elevated detachment rate
constant) secondary to vibration-induced, cross-bridge distortion in
cardiac muscle has been hypothesized to be the mechanism by which force
was depressed (12, 18, 21, 36, 37) and relaxation period shortened (16)
in cardiac muscle, and this depressive response was a prominent part of
the response observed in these experiments (as in Fig. 1). Third, our
results indicated a small but significant effect of
V on
A2/
V through
the
V2 term in
Eq. 23, which suggested
amplitude-related nonlinearity such as distortion-dependent detachment.
To test whether distortion-dependent kinetic factors accounted for any
of the observed behavior, we incorporated these effects into an
expanded form of the basic model and compared the results with those
from the basic 2-state, 4-parameter model. To introduce distortional
effects, we considered that both distortional variables,
Xe0(t)
and
Xep(t),
vary with
V(t) according to
Eqs. 11 and 12. Because the rate constant governing detachment of the prepower stroke state
(h + d) was very large compared with both
the frequency of perturbation and the detachment rate constant for the
postpower stroke state (g), a
V(t) would induce a much greater
Xep(t) than
Xe0(t)
and, because of this, it is reasonable to consider only the effects of
V-induced changes in
Xep(t) on g, the rate constant of
cross-bridge detachment. We used two functional forms of
distortion-dependent g, where
g varied around its isometric
(isovolumic) value (g0)
according to either
|
(29)
|
or
|
(30)
|
These
two formulations differ importantly as follows. In Eq. 29, when
Xep(t)
is negative, g takes on values less
than g0, i.e.,
reduced generator distortion below baseline would slow rate of
detachment, whereas increased distortion would increase rate of
detachment. This contrasts with the formulation of Eq. 30, where g is greater
than g0 for both
positive and negative values of
Xep(t),
i.e., both reduced and increased distortion increase
g and rate of detachment.
A fundamental difference between equations of the model with
distortion-dependent g compared with
those of the basic model (Eqs.
9-13) was that, with distortion-dependent
g, it could no longer be assumed that
there were no changes in elastance at the perturbation frequency.
Accordingly, it was necessary to add a third-state variable to account
for in-frequency variation in
Eep(t). Equations for this formulation are given in the
APPENDIX.
Comparison of competing model formulations was by fitting the model
with each g formulation, then
comparing results from Eq. 29 with
those from the basic model using AIC, SC, and incremental F-test, and then comparing results
from Eq. 30 with those from the basic
model using the same criteria.
There was no systematic improvement in accounting for variation in the
data (R2) in
reducing the AIC and SC or in yielding a significant incremental F-statistic when either form of
distortion-dependent detachment (Eqs.
29 and 30) was
incorporated into the model (Table 1). Furthermore, the standard error
of the estimate of the parameter associated with distortion dependence,
in those cases where the parameter was found to be significant by the
incremental F-test, was, on the
average 5-20 times larger than the standard error of the other parameter estimates. Such lack of precision in the parameter estimate indicated that the parameter played a lesser role in accounting for
variability in the data. Finally, the values of the other four
parameters were virtually unaffected by whether distortion-dependent detachment was part of the model or not (Table 1). Thus the effect of
including distortion dependence in the model was negligible even when
statistically significant.
We concluded that the added complexity as a result of incorporating
distortion-dependent g into the basic
model, in either of the forms tested, was not worth the negligible
increase in ability to explain the in-frequency component of the
response to high-frequency volume perturbations.
From the ability of the basic 2-state, 4-parameter model to account for
a wide variety of observed features and the inability of variants to
bring about any significant improvement, we adopted it as the best of
several options and as a proper representation of mechanisms
responsible for the active part of the in-frequency response. This
representation not only satisfies statistical requirements, but it is
consistent with all of the major phenomenology seen in this analysis.
 |
OVERALL SUMMARY AND CONCLUSIONS |
Dynamic Processes Between 25 and 100 Hz Are Driven by Derivative of
Volume Rather Than Volume Itself
Evidence indicating that LV dynamic processes in the 25- to 100-Hz
range are responding to the derivative of volume change comes from the
model-independent observations that the phase of
(t)
pressure leads
V(t) in
frequency-dependent manner in this frequency range and that the
A1/
V amplitude
ratio increases with frequency rather than decreases. This is strong evidence that the underlying dynamic processes are obeying behavioral patterns similar to those represented by the distortional
Eqs. 11 and 12 of the model. Thus when we employ
the model equations to predict the observations, they reproduce these
observations with a high degree of fidelity. We argue that this
distortion actually resides in the cross-bridges because we can find no
evidence in this work and in previous work (4) that there is any
effective series elastance in noncontractile structures in these
isolated hearts.
Nonlinearities Are Not an Important Part of Response to
Small-Amplitude Volume Sinusoids
We found no convincing evidence for important nonlinearity in these
responses. First, when we plotted the
A1/
V amplitude ratio in Fig. 4 and conducted the regression analysis, there was only a
slight indication that any nonlinearities were operative. Furthermore,
when we fit the data with the model and examined the correspondence
between model-fitted and measured pressure as in Fig. 6, there was no
indication for unrepresented nonlinearity; unrepresented nonlinearity
would have exhibited itself by imposing some crescent shape to the
oblong cluster of points in Fig. 6. Absolutely no evidence of crescent
shape to this cluster was detectable. Finally, when we introduced the
most likely candidate for nonlinearity into the model in the form of
distortion-dependent, cross-bridge detachment, we were unable to detect
any improvement.
Parameters From Current Sinusoidal Technique Agree With Those From
Previous Studies and With Those Obtained From Isolated Muscle
It is instructive to compare results for
g and
X0 obtained in
this sinusoidal analysis with estimates of these parameters from
earlier studies done in this laboratory. Using ferret hearts at
30°C, ramp-volume withdrawals at the time of peak isovolumic pressure, and a model equivalent to the 1-state, 2-parameter model of
the current study, we (5, 30) found values of
g of ~70-80 s
1 and of
X0 of 0.31 ml.
These compare with values of 115 s
1 for
g and 0.29 ml for
X0 in the current
study. Thus g, as estimated here, was
1.5 times greater than our previous estimates with other techniques,
whereas X0 was
not different. [Interestingly, the value of
g estimated from the 1-state,
2-parameter model in the current sinusoidal study (69 s
1) is the same as the
value we previously obtained using essentially this model in the
ramp-withdrawal studies.] In results reported elsewhere
(unpublished observations) that were obtained using the sinusoidal
technique in beating rabbit hearts at a lower temperature of 25°C,
we found average g from the 2-state,
4-parameter model to be 49 s
1, which is the same as we
previously obtained at 25°C from ferret hearts using ramp-volume
withdrawals and the 1-state, 2-parameter model (28). Furthermore, the
49 s
1 value at 25°C in
beating rabbit hearts from the current sinusoidal method is the same as
a value of 52 s
1 estimated
in constantly activated (Ba2+),
nonbeating rabbit hearts at 25°C, using responses over a dense spectrum of sinusoidal frequencies between 0.1 and 30 Hz (6). In
summary, values of g estimated in the
current study at 30°C are higher than values obtained previously
with the ramp-withdrawal technique at 30°C, but results with the
current sinusoidal technique at 25°C in beating hearts are
identical with previous results with the ramp-withdrawal technique in
beating hearts and sinusoidal techniques in constantly activated
hearts.
In earlier studies with the ramp-withdrawal technique (28), a
Q10 for
g of 2.15 was obtained. When the
25°C value of g reported elsewhere
(unpublished observations) is used with the 30°C value reported
here to calculate a Q10 for
g from the sinusoidal technique, a
value close to 5 is obtained. Although this value is far above our
previous estimate, it compares favorably with a
Q10 of 4.6 for maximal velocity of
shortening (which is proportional to
g) in membrane-intact rat trabecular
muscle (9). At this time we do not know whether the 2-state,
4-parameter model detects features that respond more strongly to
temperature than do features of the 1-state, 2-parameter model.
Resolution of these contrasting results and the effects of temperature
will have to await the conclusion of on-going studies.
Results obtained here in isolated hearts may also be compared with
results from isolated muscle studies by Kawai and co-workers (17, 24,
39). These workers used skinned, maximally activated, papillary and trabecular muscles from ferret (17, 24) and porcine (39)
hearts and conducted sinusoidal perturbation experiments at 20°C.
Our g and
h correspond with their rate constants
for the detachment step, 2
c, and
the energy transduction step, 2
b, respectively. Because 2
c and
2
b change dramatically with ATP, ADP, and Pi, which were varied in
their experiments, the proper comparison is with values of these
constants obtained when fiber-bathing solution concentrations were
comparable to a membrane-intact condition. Reported values of
comparable 2
c and
2
b were approximately 30 and 8.6 s
1 (24) or 37 and 15 s
1 (17), respectively, in
ferret heart muscle; and 23 and 15 s
1, respectively, in
porcine heart muscle (39). Despite differences in preparations,
species, interpretive models, and analysis, there are remarkable
similarities between our findings in the beating, isolated heart and
those of Kawai's group in skinned, constantly activated, isolated
muscle. First, the rate constant of the energy transduction step was on
the same order of that of the detachment step in Kawai's studies
(2
b/2
c = 1/3 to 2/3) as it was in our study
(h/g = 2/3). Second, given the potential high
Q10 for these processes, it can be
argued that the isolated muscle values found at 20°C are in
good agreement with those we found in the isolated heart at 30°C.
Again, future studies at equivalent temperatures and with the
same species are needed to establish how closely the
estimates from the isolated heart agree with those from isolated muscle.
Cross-Bridge Mechanisms Can be Observed in Whole Heart Behavior
The most important outcome of this study is that aspects of
cross-bridge kinetic behavior may be observed from pressure responses to volume perturbations in the isolated heart. This is based on evidence for the validity of the proposed cross-bridge model. Model
validity was demonstrated from concordance between the sinusoidal analysis and model predictions, from how well the model fit the data,
and from inability of elaborated models to improve representation of
the data.
No model is perfect, including this one. Particularly, it is troubling
that one of the model components, the component associated with the
prepower stroke state, contributes so very little to the magnitude of
the response. Thus a fair amount of caution must be exercised in
relying on the accuracy of estimated parameter values associated with
this state when estimation is made from responses to sinusoidal
frequencies no higher than 100 Hz. We predict that frequencies of 400 Hz or more would be required to precisely estimate parameters
associated with the prepower stroke state. More work is needed to
secure greater confidence in this parameter estimate. However, given
the overall good performance of the model and agreement of our
estimated parameters with equivalent parameters found by others in
isolated muscle, there is every reason to accept its general form and
basic premises and to proceed with additional work for its verification
and application.
In summary, pressure responses to small-amplitude, high-frequency
volume perturbations in the isolated heart provide dynamic information
relative to cross-bridge kinetics. The use of a model to analyze these
responses gives estimates of specific kinetic parameters relative to
the power stroke and cross-bridge detachment. The application of these
methods to isolated heart studies promises to be a powerful
experimental tool for elucidating mechanisms involved in changes in
LV contractile function in response to a variety of
interventions, including inotropic agents, ischemia, and chronic
pathophysiological states such as cardiomyopathy.
 |
APPENDIX |
Linear Transformation Between Incremental Wall Stress and Strain
and Incremental Chamber Pressure and Volume
To justify the assumption that incremental linear pressure-volume
elastance of the LV chamber can coexist with linear stress-strain muscle elasticity in the LV wall, we need to consider the relationship between representative lineal dimensions in the wall and chamber volume
and between representative wall stress and chamber pressure. If changes
in chamber volume linearly transform into changes in wall lineal
dimensions and if changes in chamber pressure linearly transform into
changes in wall stress, then incremental volumetric elastance can be
taken as linearly related to incremental wall elasticity.
When it is assumed that the LV is a thick-walled sphere with a
reference volume of VBL, then it
can be shown that a change in the half-mass wall circumference
(
lm/2) is
related to a change in chamber volume (
V) according to
|
(A1)
|
For
a representative ventricle of VBL = 2 ml and VW = 6 ml, a
V of
±0.015 VBL will produce a
lm/2 of
+0.1996% and
0.2004% of the reference
lm/2. With the
equivalent positive and negative changes in
lm/2 for equal
positive and negative changes in
V, there is, for all intents and
purposes, a linear relationship between
lm/2 and
V
over this range of
V. Similarly, for this same ventricle the
relationship between average wall stress and chamber pressure is given
by
|
(A2)
|
In
this relationship, pressure transforms approximately linearly into
stress if the value of the coefficient on P is not changed appreciably
with
V. For
V equalling ±0.015
VBL in these representative
hearts, the coefficient on P varies between ±1.22% its baseline
value of 0.658. This small change in the coefficient justifies the
approximation of a constant coefficient and a linear transformation
between P and
over the range in volume given by
V.
Relationship Between Elastance and Measured Pressure
By definition, chamber pressure equals elastance times volumetric
distortion. Invoking the parallel generator assumption and the model of
Fig. 3
|
(A3)
|
where
Xe0(t)
and
Xep(t)
are the respective average volumetric distortions of
Ne0(t) and
Nep(t)
generators.
Consider that elastances of the perturbed beat consist of isovolumic
elastance plus changes in elastance induced by the perturbation. Furthermore, distortion consists of isovolumic distortion plus perturbational-induced in distortion. Thus
|
(A4)
|
where isovolumic conditions are indicated with the iso superscript
and changes induced by the sinusoidal perturbation are indicated with
the
. Because
Xisoe0 = 0, Eq. A4 can be rewritten
as
|
(A5)
|
where the first term in brackets on the right-hand side is
the pressure around which the distortional response is occurring
|
(A6)
|
and
the second term in brackets is the distortional or in-frequency
response
|
(A7)
|
Thus
Comparing
Eqs. A5 and A3, we make the following
assignments
|
(A8)
|
Then from Eq. A6, it can be
written that
|
(A9)
|
An
expression for
Ee0(t)
can be developed by considering the differential equation for the
ep state. From Fig. 3
|
(A10)
|
which,
after substituting Eqs. 16 and 17 in the text, can be rewritten in
terms elastances as
|
(A11)
|
Solving
Eq. A11 for
Ee0(t)
|
(A12)
|
Substituting
for
Eep(t)
from Eq. A9 gives the following
expression for
Ee0(t)
|
(A13)
|
Differential Equations for Distortional Variables
The differential equations for
Xep(t)
may be derived by considering the summed distortion among all
generators within a given state. The total distortion among all
ep generators is given by
|
(A14)
|
where
Xep(t)
is the average distortion among the
Nep(t)
generators.
At some t +
t,
Zep(t +
t) can be written as
Zep(t +
t) = Zep(t) + (added distortion due to change in volume over
t) + (added distortion due to
formation of new units with baseline distortion over
t)
(lost distortion due
to detachment of distorted units over
t), where (added distortion due
to change in length over
t) = (externally imposed
V) · (no. of
ep generators existent at
t) =
V
Nep(t),
and where (added distortion due to formation of new units with
X0 distortion
over
t) = [h · Ne0(t)]
t(X0 +
V/2), and (lost distortion due to detachment of distorted units
over
t) = [g · Nep(t)]
t[Xep(t)] Therefore
|
(A15)
|
and
|
(A16)
|
Taking the limit as
t
0
|
(A17)
|
Note
that differentiation of Eq. A10 yields
|
(A18)
|
Equating
Eqs. A17 and A18, making appropriate substitutions
for
ep(t)
from Eq. A10, and
solving for
ep(t) gives
|
(A19)
|
To
remove the dependence on
Ne0(t),
we rearrange Eq. A10 to get
|
(A20)
|
Substituting
Eq. A20 into Eq. A19 and writing in terms of the incremental variable,
Xep(t) = Xep(t)
X0
|
(A21)
|
Finally,
substituting for
Nep(t)
its equivalent in terms of
Eep(t),
as given by Eq. 8 in the text, gives
the final form of the differential equation as given by
Eqs. 12 and 11 in the text.
Equations When There is Noncontractile Series Elastance
When there is noncontractile series elastance, there is internal
shortening during isovolumic or reference beats. Under assumptions made
in the text
|
(A22)
|
where
Xepr(t)
is no longer X0
as was true when the series elastance was considered to be infinitely
stiff. Given a measured
Pr(t),
it can be shown that when there is internal shortening during a
reference beat, generator distortion as a result of internal shortening
is given by
|
(A23)
|
where
ESE is the
elastance of series-coupled, noncontractile elements. Solving
Eq. A23 for
Xepr(t)
allows calculation of
Eep(t)
using Eq. A22. Furthermore, the
presence of a compliant series elastance changes the state equations as
follows
|
(A24)
|
|
(A25)
|
After terms were collected and expressed explicitly in terms
of the state variables, it is seen that these are no longer independent
but are now coupled equations.
Equations When There is Distortion-Dependent "g"
In addition to incorporating the functional form of
g into the differential equations of
the basic model, it was necessary to expand these equations because, as
g changed its value during the period
of a volume sinusoid according to the distortions that were imposed,
Nep(t)
and, consequently,
Eep(t) also changed their values subsequent to these variations in
g. These in-frequency variations in
Eep(t)
were treated as follows. Let
|
(A26)
|
where g0 and
epr(t)
are the values around which variations occurred and
g(t)
and
Eep(t)
are the variations themselves. From Eq. A11
|
(A27)
|
where
|
(A28)
|
and
|
(A29)
|
We
do not need Eq. A28 because
Eepr(t)
can be calculated from
Eepr(t)
Pr(t)/X0.
Substituting Eepr(t)
into Eq. A29 and then solving that
equation numerically allows the calculation of
(t). Thus all elements of
Eep(t)
are determined and the complete model solution with the
distortion-dependent g may be
conducted.
 |
ACKNOWLEDGEMENTS |
This work was support by National Heart, Lung, and Blood Institute
Grant HL-21462.
 |
FOOTNOTES |
Address for reprint requests: K. Campbell, Depts. of VCAPP & Biological
Systems Engineering, Washington State Univ., Pullman, WA 99164.
Received 14 January 1997; accepted in final form 2 July 1997.
 |
REFERENCES |
1.
Boom, H. B.,
and
H. Wijkstra.
The step response of the left ventricular pressure to ejection flow: a system oriented approach.
Ann. Biomed. Eng.
20:
99-126,
1992[Medline].
2.
Campbell, K. B.
Rate constant of muscle force re-development reflects cooperative activation as well as crossbridge kinetics.
Biophys. J.
72:
254-262,
1997[Medline].
3.
Campbell, K. B.,
L. W. Campbell,
J. E. Pinto,
and
T. D. Burton.
Contractile-based model interpretation of pressure-volume dynamics in the constantly activated (Ba2+) isolated heart.
Ann. Biomed. Eng.
22:
550-567,
1994[Medline].
4.
Campbell, K. B.,
R. D. Kirkpatrick,
A. H. Tobias,
H. Taheri,
and
S. G. Shroff.
Series-coupled passive elastance is functionally unimportant in the isolated heart.
Cardiovasc. Res.
28:
242-251,
1994[Abstract/Free Full Text].
5.
Campbell, K. B.,
S. G. Shroff,
and
R. D. Kirkpatrick.
Short time-scale LV systolic dynamics: evidence for a common mechanism in both LV chamber and heart-muscle mechanics.
Circ. Res.
68:
1532-1548,
1991[Abstract/Free Full Text].
6.
Campbell, K., B.,
H. Taheri,
R. D. Kirkpatrick,
and
B. K. Slinker.
Single perturbed beat vs. steady-state beats for assessing systolic mechanical function in the isolated heart.
Am. J. Physiol.
262 (Heart Circ. Physiol. 31):
H1631-H1639,
1992[Abstract/Free Full Text].
7.
Campbell, K. B.,
H. Taheri,
R. D. Kirkpatrick,
T. Burton,
and
W. C. Hunter.
Similarities between dynamic elastance of left ventricular chamber and papillary muscle of rabbit heart.
Am. J. Physiol.
264 (Heart Circ. Physiol. 33):
H1926-H1941,
1993[Abstract/Free Full Text].
9.
De Tombe, P. P.,
and
H. E. D. J. ter Keurs.
Force and velocity of sarcomere shortening in trabeculae from rat heart. Effects of temperature.
Circ. Res.
66:
1239-1254,
1990[Abstract/Free Full Text].
10.
Eisenberg, E.,
T. L. Hill,
and
Y. Chen.
Cross-bridge model of muscle contraction.
Biophys. J.
29:
195-227,
1980[Medline].
11.
Glantz, S. A.,
and
B. K. Slinker.
Primer of Applied Regression and Analysis of Variance. McGraw-Hill, 1990.
12.
Honda, H.,
Y. Koiwa,
and
T. Takashima.
Mathematical model of the effects of mechanical vibration on crossbridge kinetics in cardiac muscle.
Jpn. Circ. J.
58:
416-425,
1994[Medline].
13.
Hunter, W. C.,
J. S. Janicki,
K. T. Weber,
and
A. Noordergraaf.
Systolic mechanical properties of the left ventricle: effects of volume and contractile state.
Circ. Res.
52:
319-327,
1989[Abstract/Free Full Text].
14.
Huxley, A. F.
Muscle structure and theories of contraction.
Prog. Biophys. Mol. Biol.
7:
257-318,
1957.
15.
Huxley, A. F.,
and
R. M. Simmons.
Proposed mechanism of force generation in striated muscle.
Nature
233:
533-538,
1971[Medline].
16.
Janssen, P. M. L.,
H. Honda,
Y. Koiwa,
and
K. Shirato.
The effect of of diastolic vibration on the relaxation of rat papillary muscle.
Cardiovasc. Res.
32:
334-350,
1996.
17.
Kawai, M.,
Y. Saeki,
and
Y. Zhao.
Crossbridge scheme and the kinetic constants of elementary steps deduced from chemically skinned papillary and trabecular muscles of the ferret.
Circ. Res.
73:
35-50,
1993[Abstract].
18.
Koiwa, Y.,
N. Hoshi,
T. Ohyama,
T. Takagi,
J. Kikuchi,
H. Honda,
and
T. Takashima.
Evaluation of left ventricular function by application of external minute vibration.
Front. Med. Biol. Eng.
2:
207-210,
1990[Medline].
19.
Kirkpatrick, R. D.,
K. B. Campbell,
D. L. Bell,
and
H. Taheri.
Method for studying arterial wave-transmission effects on left ventricular function.
Am. J. Physiol.
260 (Heart Circ. Physiol. 29):
H1003-H1012,
1991[Abstract/Free Full Text].
20.
Landaw, E. M.,
and
J. J. DiStefano.
Multiexponential, multicompartmental, and noncompartmental modeling. I. Data analysis and statistical considerations.
Am. J. Physiol.
246 (Regulatory Integrative Comp. Physiol. 15):
R665-R677,
1984.
21.
Nishioka, T.,
Y. Goto,
K. Hata,
T. Takasago,
A. Saeki,
T. W. Taylor,
and
H. Suga.
Mechanoenergetics of negative inotropism of ventricular wall vibration in dog heart.
Am. J. Physiol.
270 (Heart Circ. Physiol. 39):
H583-H593,
1996[Abstract/Free Full Text].
22.
Pate, E.,
and
R. Cooke.
A model of crossbridge action: the effects of ATP, ADP and Pi.
J. Muscle Res. Cell Motil.
10:
181-196,
1989[Medline].
23.
Podolski, R. J.,
and
A. C. Nolan.
Muscle contraction transients, cross-bridge kinetics, and the Fenn effect.
Cold Spring Harb. Symp. Quant. Biol.
37:
661-668,
1974.
24.
Saeki, Y.,
M. Kawai,
and
Y. Zhao.
Comparison of cross-bridge dynamics between intact and skinned myocardium from ferret right ventricles.
Circ. Res.
68:
772-781,
1991[Abstract/Free Full Text].
25.
Schiereck, P.,
and
H. B. K. Boom.
Left ventricular active stiffness: dependency on time and inotropic state.
Pflügers Arch.
374:
135-143,
1978[Medline].
26.
Schiereck, P.,
and
H. B. K. Boom.
Left ventricular force-velocity relations measured from quick volume changes.
Pflügers Arch.
379:
251-258,
1979[Medline].
27.
Shroff, S. G.,
K. B. Campbell,
and
R. D. Kirkpatrick.
Short time-scale LV systolic dynamics: pressure vs. flow clamps and effects of activation.
Am. J. Physiol.
264 (Heart Circ. Physiol. 33):
H946-H959,
1993[Abstract/Free Full Text].
28.
Shroff, S. G.,
K. B. Campbell,
D. E. Miller,
R. D. Kirkpatrick,
and
H. Taheri.
Effect of temperature on short time-scale left ventricular contractile dynamics (Abstract).
Circulation
86:
I553,
1992.
29.
Shroff, S. G.,
J. S. Janicki,
and
K. T. Weber.
Left ventricular systolic dynamics in terms of its chamber mechanical properties.
Am. J. Physiol.
245 (Heart Circ. Physiol. 14):
H110-H124,
1983.
30.
Shroff, S. G.,
J. S. Janicki,
and
K. T. Weber.
Evidence of quantitation of left ventricle resistance.
Am. J. Physiol.
249 (Heart Circ. Physiol. 18):
H358-H370,
1985.
31.
Suga, H.,
L. Demer,
and
K. Sagawa.
Determinants of instantaneous pressure in canine left ventricle.
Circ. Res.
46:
314-322,
1980.
32.
Templeton, G. H.,
and
L. R. Nardizzi.
Elastic and viscous stiffness of the canine left ventricle.
J. Appl. Physiol.
36:
123-127,
1974[Free Full Text].
33.
Templeton, G. H.,
K. Wildenthal,
J. T. Willerson,
and
J. H. Mitchell.
Influence of acute myocardial depression on left ventricular stiffness and its elastic and viscous components.
J. Clin. Invest.
56:
278-285,
1975.
34.
Thorson, J.,
and
D. C. S. White.
Role of cross-bridge distortion in the small-signal mechanical cynamics of insect and rabbit striated muscle.
J. Physiol. (Lond.)
343:
59-84,
1983[Abstract/Free Full Text].
35.
Vaartjes, S. R.,
and
H. B. K. Boom.
Left ventricular internal resistance and unloaded ejection flow assessed from pressure-flow relations: a flow-clamp study on isolated rabbit hearts.
Circ. Res.
60:
727-737,
1987[Abstract/Free Full Text].
36.
Vukas, M.,
I. Malek,
and
A. Hjalmarson.
Myocardial depressant effect of vibration in isolated rabbit heart.
Scand. J. Clin. Lab. Invest.
38:
421-424,
1978[Medline].
37.
Vukas, M.,
R. Silvertsson,
and
B. Ljung.
Inhibitory effects of vibration on contractiltiy of isolated rabbit papillary muscle.
Scand. J. Clin. Lab. Invest.
38:
415-419,
1978[Medline].
38.
Wijkstra, H.,
and
H. B. Boom.
Deactivation in the rabbit left ventricle induced by constant ejection flow.
IEEE Trans. Biomed. Eng.
36:
1113-1123,
1989[Medline].
39.
Zhao, Y.,
and
M. Kawai.
Inotropic agent EMD-53998 weakens nucleotide and phosphate binding to cross bridges in porcine myocardium.
Am. J. Physiol.
271 (Heart Circ. Physiol. 40):
H1394-H1406,
1996[Abstract/Free Full Text].
AJP Heart Circ Physiol 273(4):H2044-H2061
0363-6135/97 $5.00
Copyright © 1997 the American Physiological Society