Laboratoire d'Anesthésie, Faculté de Médecine du
Kremlin-Bicêtre, Université de Paris Sud, F-94276 Le
Kremlin-Bicêtre Cedex, France
Colored microspheres have become popular
compared with radioactive microspheres because they do not use
radioactivity. However, they suffer from a much greater variability in
their determination. We have developed a new method for assaying the
dye using high-performance liquid chromatography (HPLC) with internal
standard. This technique permits accurate determination of
400
spheres in rat blood, heart, kidney, liver, and brain with a relative
error [coefficient of variation (CV)] <10%. To date,
only three colors (white, yellow, and red) may be used because, of the
five colors tested, one (violet) served as internal standard and
another (blue) exhibited marked degradation during extraction. Compared
with the classical spectrophotometric technique, HPLC allows a three to
five times improvement in reproducibility with a relative error
significantly lower (P < 0.01) than
with direct spectrophotometry. Although this new technique appears to
be more time consuming than the classical method, its use seems to be
preferable because of the improvement in measurement sensitivity.
 |
INTRODUCTION |
THE MICROSPHERE TECHNIQUE is still considered the
"gold standard" for measuring regional blood flow, either as an
absolute value or relative to other organ flow (4, 13). Until recently, only radioactive spheres were used, but the emergence of colored microspheres provides the opportunity for a simple and safe technique, particularly by avoiding the use of radioactivity (9, 12). Briefly, the
technique is as follows. A known number of microspheres is injected in
a mixing chamber such as the left atria or the left ventricle, and a
reference arterial sample is drawn at a known withdrawal rate. After
animals are euthanized, the number of spheres is measured both in the
reference sample and in the sample organ of interest. Regional blood
flow is then calculated relative to the reference withdrawal rate. Thus
errors in both measurements will participate in the error in regional
blood flow calculation. In fact, measurements of regional blood flow by
microspheres suffer from lack of precision due to distributional
variation governed by the Poisson distribution and due to assay
imprecision (2, 5, 6). The assay error is most often neglected in case
of radioactive microspheres because increasing counting time meets the
required precision (2). However, in the case of colored microspheres,
the assay error may be the primary cause of imprecision (see
RESULTS). One may thus be tempted to increase
the number of microspheres injected. However, increasing
the number of spheres in both tissue and reference samples will not
decrease the error because the assay generally has a constant
coefficient of variation (CV). Moreover, if one increases the number of
spheres injected per measurement in the rat, serious hemodynamic
disturbances are expected to occur (11). It is, then, of utmost
importance to reduce the assay error when one uses colored
microspheres. In that respect, the usual direct spectrophotometric
method derived from radioactivity measurement lacks precision (see
RESULTS). If some studies have addressed the question of
recovery of dye (8), none (in our knowledge) has studied the problem of
reproducibility of the assay.
We now propose 1) the use of an
internal standard technique and 2) a
new assay using high-performance liquid chromatography (HPLC), which we
compare with the direct spectrophotometric assay in terms of
reproducibility, i.e., relative error.
 |
METHODS |
Materials and reagents.
Colored (white, yellow, violet, red, and blue) microspheres (DyeTrak
Triton 15.5 µm) were purchased from Bioseb (Paris, France). Inasmuch
as the whole study is based on the principle of the reference sample,
we used the particle concentration recorded on the manufacturer's chart (3,000,000 spheres/ml except for the violet spheres, whose concentration was recorded to be 3,300,000 spheres/ml) with no further
verification. N,N,-dimethylformamide
(DMF) (HPLC grade) and polyoxyethylenesorbitan monooleate (Tween 80)
were purchased from Sigma (St. Quentin Fallavier, France). Acetonitrile
(HPLC grade), ethanol (low grade with 4% methanol; a preliminary
experiment has shown that the quality of extraction and background
noise were not modified whether HPLC-grade or low-grade ethanol was used), NaCl,
NaH2PO4,
and
H3PO4
were purchased from Prolabo (Paris, France).
Chromatography.
Chromatographic analysis was performed under isocratic conditions at
room temperature. The HPLC system consisted of a Shimadzu LC9A pump
(Touzart et Matignon, Les Ulis, France), a variable-wavelength ultraviolet (UV)-visible detector (model 1050, Hewlett-Packard, Les
Ulis, France), and a Linseis L 6512B pen recorder (Bioblock, Illkirch,
France). The separation was performed on a Waters Resolve 5-µm
spherical C18, 3.9 × 300 mm
column (Waters, St. Quentin en Yvelines, France). The mobile phase
consisted of 35% 0.01 M
NaH2PO4 buffer containing 0.001 M sodium dodecyl sulfate and 65% acetonitrile (vol/vol) and was set at a flow rate of 1.2 ml/min. Twenty-five microliters of the final extract were injected in the system. Detection
was made at wavelengths of 370, 448, 268, and 530 nm, respectively, for
the white, yellow, violet, and red dye. The blue spheres were not used
because a preliminary experiment showed that a marked degradation of
the dye occurred during extraction with the appearance of multiple
peaks after chromatographic separation.

View larger version (20K):
[in this window]
[in a new window]
|
Fig. 1.
Top: chromatograms obtained after
extraction of 0.5 ml blood containing 25 yellow (2), 50 white (1), and
50 red spheres (4) with 1,500 violet spheres (3) as internal standard
(IS) (A); 125 yellow (2), 250 white
(1), and 250 red spheres (4) with 1,500 violet spheres (3) as IS
(B); or 1,000 yellow (2), 2,000 white (1), and 2,000 red spheres (4) with 1,500 violet spheres (3) as
IS (C).
Bottom: chromatograms obtained after
extraction of rat brain samples with the following measured content: 27 white (1), 11 yellow (2), and 20 red spheres (4) with 1,500 violet
spheres (3) as IS (D) and 451 white
(1), 445 yellow (2), and 335 red (4) spheres with 1,500 violet spheres
(3) as IS (E). In
D, U is unknown peak.
|
|
Spectrophotometric procedure.
Sample spectra were recorded using the same variable-wavelength
UV-visible detector as used for HPLC, set in the scanning mode. One
hundred to one hundred twenty microliters of the extract were injected
using a direct injection port with minimal dead volume (Rheodyne,
Touzart et Matignon, Les Ulis, France). Absorption was measured at 335, 375, 405, 435, 475, 505, 535, 565, and 595 nm. Dye concentration was
determined using a modification of the procedure described by Schosser
et al. (15). This modification allows the use of a number of wavelength
measurements greater than the number of unknown dye concentrations by
using linear least-squares regression using the Faddeev-Leverrier
algorithm (10) rather than a simple matrix inversion procedure. Because background noise in biological samples is greater in the lower wavelength spectrum (typically at wavelengths between 190 and 400 nm),
we compared the results obtained using the set of nine wavelengths to
the results obtained using a reduced set of seven wavelengths (from 405 to 595 nm) and to the results obtained with the classical
four-point technique [number of measurements equal to number of
colors (405, 475, 535, and 595 nm)].
Extraction.
The extraction procedure was always performed using 1% Tween 80. The
spheres were gently vortex-mixed for 30-60 s (at low speed to be
sure that no foam was formed), then put in an ultrasonic bath and
hand-mixed just before sampling with a micropipette. Direct extraction
of dye from the spheres was performed using DMF and was considered
complete. Extraction from biological material was performed as follows.
Heparinized blood (500 µl) or 200- to 300-mg tissue samples (heart,
kidney, liver, or brain) obtained from Sprague-Dawley rats were spiked
with known amounts of spheres in 5-ml polypropylene tubes. The spheres
were introduced into the tubes by pipetting known numbers of spheres
(in 50 µl saline with 1% Tween 80). In spiked samples to which a
known number of violet spheres were added, this number served as an
internal standard reference. For the lower concentrations tested (50 yellow, 100 white, and 100 red spheres) and for extraction after in
vivo injection, 1,500 violet spheres were added to the sample as
internal standard. Two milliliters of distilled water with 1% Tween 80 were added, and the tubes were hand-shaken. One milliliter of KOH (12 M) was added, and the tubes were put in a water bath shaker at 70°C
for 2 h 30 min. After being cooled, the tubes were centrifuged at 2,500-3,000 revolutions/min for 20 min. The lower part was
carefully aspired with a thin needle and discarded, leaving the spheres in the upper layer (0.5-0.75 ml). Four milliliters of ethanol containing 1% Tween 80 and 100 µl
H3PO4
(6 M) were added, and the tubes were vortex-mixed for 30-40 s and
ultrasonicated for 4-5 min. After centrifugation at
2,500-3,000 revolutions/min for 10 min, the supernatant was
discarded and the procedure was repeated. The residue was then
evaporated to dryness under vacuum at 35°C for 40 min. The dry
residue was extracted with 200 µl DMF, vortex-mixed for 30-40 s,
and ultrasonicated for 4-5 min. After centrifugation for 10 min at
4,000 revolutions/min, 25 µl of the supernatant were injected into
the HPLC system or 100-120 µl were directly injected in the
spectroscopic cell for measurement by the traditional technique.
Protocol design.
Experiments were conducted using five replicates. First, we assessed
the pipetting variability 1) by
pipetting 50 µl of solutions containing 4,000, 400, and 40 white
microspheres (these concentrations were obtained by serial dilutions)
into test tubes and 2) by pipetting 150,000 white microspheres (50 µl) in a polyethylene tube (PE-50, Guerbet Louvres, France) using a model 44 Harvard pump (Harvard, Les
Ulis, France) and rinsing the tube with 2 ml water containing 1% Tween
80. Because each dye did not display the same absorbance intensity, we
normalized the mixtures of spheres according to their respective
absorbance, i.e., to the intensity of their respective signal. Recovery
and reproducibility of extraction and detection in biological samples
were tested using three different mixtures: 1) 50 yellow, 100 white, 100 red,
and 200 violet spheres (violet spheres were not added at this time when
an internal standard procedure was used);
2) 400 yellow, 800 white, 800 red,
and 1,600 violet spheres; and 3)
1,500 yellow, 3,000 white, 3,000 red, and 6,000 violet
spheres.
In vivo experiments.
Two Sprague-Dawley rats weighing 380-400 g were anesthetized with
pentobarbital (50 mg/kg ip). The left femoral artery was cannulated
using a 22-gauge Teflon short catheter (Insyte,
Beckon-Dickinson, Dun Laoghaire, UK), and a polyethylene catheter
(PE-50, Guerbet Louvres, France) was advanced in the left ventricle via
the right carotid artery for dye injection (8). Rat
1 received a mixture of 300,000 white, 300,000 yellow,
and 300,000 red spheres over a 40-s period using a model 22 Harvard
pump (Harvard, Les Ulis, France), and rat
2 received a mixture of 100,000 white, 100,000 yellow,
and 100,000 red spheres. The animals were euthanized with an overdose
of pentobarbital, and the organs (heart, brain, and right and left
kidneys) were dissected and cut into five samples weighing 150-300
mg each (200-400 mg for brain tissue). After extraction as
described above and elution with 200 µl DMF, 25 µl were injected
into the HPLC system and then 100-120 µl were injected directly
into the spectrophotometer.
Data are presented as means of replicate experiments, and CVs (relative
standard deviation or relative error) are expressed as percentages.
Comparisons between CVs used the Fisher-Snedecor F test (1).

View larger version (16K):
[in this window]
[in a new window]
|
Fig. 2.
Measured vs. spiked dye concentration obtained after fitting an 8-point
standard curve (7 points for violet)
(A), and the corresponding weighted
residuals (B). Weighting as the
inverse of concentration squared led to small bias.
|
|
 |
RESULTS |
HPLC.
The separation was optimal with mobile phase containing 65%
acetonitrile (vol/vol) (Fig. 1). Under the
above-described conditions, retention times were as follows: white
spheres, 5.1 min (CV 1.3%); yellow spheres, 10.6 min (CV 1.0%);
violet spheres, 12.6 min (CV 0.7%); and red spheres, 13.2 min (CV
0.7%). The lower limits of detection of the dye (directly extracted
from the spheres) at four times the size of the basal noise were 2.5 spheres (white), <1 sphere (yellow), 12.5 spheres (violet), and 2.5 spheres (red). The CV within the same extracted sample at four times
the lower limit of detection were 0.6% (white spheres), 1.0% (yellow
spheres), 1.7% (violet spheres), and 3% (red spheres). The errors in
pipetting were, respectively, 4, 7, and 15% for the solutions
containing 4,000, 400, and 40 spheres. The error in pipetting 50 µl
of spheres in PE tubing was 4%. An eight-point calibration curve was
constructed in blood (7 points for violet spheres). The
curves were found to be linear from 20 to 5,000 spheres (yellow,
r = 0.998), from 40 to 10,000 spheres
(white, r = 0.996; red,
r = 0.995), and from 80 to 10,000 spheres (violet, r = 0.993) (Fig.
2). However, recovery of dye was incomplete even at
sphere concentrations
400 per sample (Table
1). Reproducibility was poor, but when
violet spheres were used as internal standard, reproducibility markedly
increased with a CV between 1.8 and 5.1% at the higher concentration
and between 4.1 and 18% at the lower concentration, depending on the type of spheres and type of tissue (Table 1).
Direct spectrophotometry.
In contrast to the above results concerning HPLC, the classical direct
spectrophotometric method showed very poor reproducibility, and the
difference between classical spectrophotometric and HPLC methods using internal standards was highly significant (Table 2). After scanning the samples,
the calculations used linear least squares with a
nine-point method (from 335 to 595 nm), a seven-point
method (obtained by deleting the first 2 points, at 335 and 375 nm,
because the high background noise observed at these wavelengths was
supposed to decrease reproducibility), or the classical four-point
method (number of measurements equal to number of unknowns). We
observed an improvement in the quality of results when the reduced
seven-point method was used compared with the nine-point technique. In
contrast, the four-point technique always showed an impairment in the
quality of results (except in the case of white spheres, for which the
3 techniques gave similar results). It was not possible to measure dye
content at sphere concentrations less than 200 (yellow), 400 (white and
red), or 800 spheres per milliliter (violet): CVs and bias both always exceeded 60% at these concentrations.
View this table:
[in this window]
[in a new window]
|
Table 2.
Intraday reproducibility of traditional spectrophotometric method with
and without internal standard compared with new HPLC technique
|
|
Because the HPLC technique showed that recovery of dye after extraction
was incomplete, we estimated the bias obtained with the direct
spectrophotometric technique by comparing the results with those
obtained by direct extraction of spheres in saline. With HPLC, mean
bias (n = 5) was in the range of
4.7 ± 12,
4.1 ± 5.6, and
3.3 ± 2.1%
for the low, medium, and high concentrations, respectively, whereas
with direct spectrophotometry (7 points with internal standard), mean
bias was in the range of
17 ± 19 and
6.8 ± 12%
for the medium and high concentrations, respectively.
In vivo comparison of the two techniques.
The results of the in vivo experiment are summarized in Table
3. With the exception of brain measurements
in rat 2 (which received only 100,000 spheres), the two methods gave similar mean results, but HPLC always
showed lower CVs than the traditional spectrophotometric technique
using internal standard. However, the variability within an organ was
still greater than the in vitro variability.
 |
DISCUSSION |
The present study shows that assaying dye content of colored
microspheres by HPLC with internal standard is a major improvement compared with the direct spectrophotometric assay. The extraction step
needs to be carefully performed. In fact, use of surfactant is
critical. Some authors use filtration in the washing-extraction process
(12). This may improve the sample washing procedure but with the risk
of losing spheres. We made the choice of working in a unique tube. This
procedure is assumed to lose fewer spheres but may result in incomplete
clean-up procedure. In fact, recovery was incomplete at all particle
concentrations, especially in blood and in the liver, as already
described by Hakkinen et al. (8). However, these authors used a much
larger number of spheres in their samples (5,000 spheres) compared with
the range of spheres used in our experiment.
The lower limit of detection of the HPLC technique was far below the
number of spheres usually encountered in tissue or blood samples. When
there are
400 microspheres in the sample, the relative error of the
assay is then between 2 and 10% with the yellow, white, and red
microspheres. To date, only three different measurements are made
possible because, of the five available colors, one served as internal
standard (violet) and another (blue) exhibited marked degradation
during the extraction or separation procedure, leading to the
unexpected observation of three different peaks. We chose violet as
standard because its absorbance is much lower than the absorbance
observed with the other colors, thus leading to a greater CV at the
critical value of 400 spheres/sample: ~1,500 spheres appears then to
be the adequate number to be added as internal standard (see Fig. 1).
The use of the internal standard markedly improved the precision of the
technique (Table 1). This is not surprising because recovery appears
incomplete even at high particle concentrations. It should also be
noted that we only measured recovery relative to dye directly extracted
from spheres in DMF and not relative to the pure dye molecule. The
technique of isocratic HPLC with UV-visible spectrophotometric
detection is simple and easy to implement. However, it is certainly
more time consuming than the direct spectrophotometric method; the use
of an autosampler is highly recommended.
At the critical value of 400 spheres/sample, the direct
spectrophotometric method cannot be recommended because of unacceptable error in flow measurement due to lack of precision of the assay. The
high background noise associated with the direct spectroscopic measurement method is responsible for this imprecision. In contrast, chromatography adds a separation step that markedly increases the
signal-to-noise ratio and permits a single determination of the
compound of interest.
In vivo injection of spheres in rats confirms these findings. The mean
number of spheres measured in samples by the two methods is not
markedly different. This confirms the results of the in vitro
experiments showing that bias was (on average) not very important.
Nevertheless, when the number of spheres in the samples was low (<300
in brain of rat 2), a systematic
difference in the number of spheres measured by the two techniques
became visible. The average intraorgan CV was three times greater with
traditional spectrophotometry than with HPLC (Table 3). In fact,
physiological variability in sphere deposition in organs is well
documented. For example, Bassingthwaighte et al. (3) showed that
relative dispersion of flows measured in baboon, sheep, and rabbits
using spheres or soluble markers was 23-30% (mean of 6-11
animals). Our results of single-organ variability in sphere content
(measured with HPLC) is consistent with these findings. Nevertheless,
the combined error made in measuring reference blood and tissue
contents is expected to increase the resulting CV of regional blood
flow.
Cost and time requirements.
The extraction procedure is the same for the two techniques. However,
these two techniques differ in equipment, in time needed for
spectrophotometry (traditional or after chromatographic separation), and in data processing requirements. The classical spectrophotometric technique requires the use of a spectrophotometer with an automatic scanning mode, leading either to a total spectrum or to the
measurement of absorbance at predetermined wavelengths, usually coupled
with a computerized recording system, whereas HPLC needs a
simpler spectrophotometer (with the same precision and sensitivity).
After the spectroscopic procedure, a computer program is needed for inverting the matrix of absorbance. On the other hand, HPLC requires the use of other equipments (pump, injection port, and column) and also
the use of solvents. Moreover, HPLC is time consuming, and the use of
an automatic sampler is highly recommended. The counting unit and
computer software used for traditional spectrophotometry cost
approximately $18,000, whereas an HPLC system costs approximately $17,000-$20,500 (isocratic pump $6,000, injection port and tubing $1,500, UV-visible detector $8,000, and pen recorder $1,500 or integrator $5,000). The replacement of the column ($500 for ~1,000 samples, i.e., $0.50 per sample assayed) and the extra cost of solvents
($30 for ~100 samples, considering mobile-phase recycling) have also
to be taken into account. Also, there is an extra cost due to the time
required for chromatographic separation. This cost is about $6-$8
per sample, considering $25 per hour for personnel costs. The addition
of an automatic sampler to the HPLC system is highly recommended. Its
cost ($6,000-$8,000) is then paid off after ~1,000 samples are
assayed. Finally, the use of HPLC adds an extra cost of about
$5,000-$8,000 for equipment and $0.80 per sample assayed compared
with the traditional spectrophotometric technique. However, many
laboratories already possess an HPLC system that can be used for
different purposes.
In conclusion, HPLC with internal standard appears to give much less
variability due to assay errors in the measurement of regional blood
flow with colored microspheres. Moreover, this new technique allows the
accurate detection of as low as 50-100 spheres, whereas the
traditional technique, even when corrected with the use of an internal
standard, cannot permit the detection of fewer than 200-400
spheres. Compared with other nonradioactive microsphere methods that
have recently emerged [e.g., fluorescence (7) and X-ray
fluorescence (14)], the colored microsphere technique has the
advantage of simplicity of use associated with moderate cost. The lower
limit of detection of colored spheres using HPLC favorably compares
with the limit of detection obtained with fluorescent microsphere
techniques (7), but the number of available different spheres is much
lower when using colored rather than fluorescent spheres. The benefit
of using HPLC with internal standard, then, is to increase the
precision of the method and to allow the injection of a limited number
of spheres, thereby decreasing the hemodynamic consequences of sphere
embolization (11).
Address for reprint requests: J. X. Mazoit, Laboratoire
d'Anesthésie, Faculté de Médecine du
Kremlin-Bicêtre, Université de Paris Sud, F-94276 Le
Kremlin-Bicêtre Cedex, France.
Received 25 November 1996; accepted in final form 13 November
1997.