Vol. 275, Issue 2, H653-H661, August 1998
Simultaneous LV and RV volumes by conductance catheter:
effects of lung insufflation on parallel conductance
Richard S.
Szwarc and
Howard A.
Ball
Department of Clinical Pharmacology, Novartis, CH-4002 Basel,
Switzerland
 |
ABSTRACT |
One aspect in the
measurement of ventricular volume using the conductance catheter
technique is the assessment of parallel electrical conductivity of
structures extrinsic to the ventricular blood pool. Because it is
sometimes necessary to make volume measurements during ventilation or
spontaneous respiration, the extent to which parallel conductance may
vary with lung insufflation was investigated. Anesthetized pigs
(11-15 kg) were ventilated and instrumented with both left (LV)
and right ventricular (RV) conductance and pressure-tip catheters and
end-hole catheters for injection of hypertonic saline into the inferior
vena cava and pulmonary artery. Data were recorded during ventilation
with tidal volumes of 10 and 20 ml/kg, and the associated fluctuations
to LV and RV end-diastolic (EDV) and stroke (SV) volumes were measured.
With the use of a saline dilution technique, parallel conductance
(Vc) was determined for each
ventricle with the ventilator off and lungs insufflated to 0, 10, and
20 ml/kg. Whereas ventilation caused marked oscillations in LV and RV
EDV and SV, these variations could not be attributed to
Vc, which remained statistically
unchanged from their baseline values of 34.1 ± 3.1 in the LV and
31.1 ± 4.4 in the RV. These results indicate that the fluctuations
that occur in conductance catheter-derived LV and RV volume signals
with ventilation are not caused by any significant changes to parallel
conductance.
conductance catheter; ventricular volume; ventricular interaction; parallel conductance
 |
INTRODUCTION |
THE CONDUCTANCE CATHETER method of measuring
ventricular volume has become an important advance in the assessment of
left ventricular (LV) (3, 4, 10, 20, 21) and, more recently, right
ventricular (RV) (7, 11, 18, 23) function. The multielectrode catheter
emits a low-current electrical field within the ventricle, allowing the
measurement of a time-varying, intracavity electrical conductance
assumed to be proportional to changes in ventricular blood volume. In
practice, however, in order to obtain absolute volume measurements,
both a gain and an offset must first be applied to the conductance
signal.
The conductance offset term is required because the electrical field is
not contained entirely within the ventricular cavity but extends
through the myocardium, contralateral ventricle, and surrounding
structures. Thus whereas the conductance catheter measures conductivity
of the ventricular blood pool, it also detects that of extrinsic
structures. This parallel conductance is manifested as a positive
offset of the conductance-derived volume signal and is usually assumed
to be a constant in any given subject.
Normally, conductance catheter measurements are made in the
anesthetized patient with ventilation temporarily suspended at end
expiration or in the awake patient during shallow or held respiration.
However, because the conductance catheter is used in a wider variety of
situations, one component of parallel conductance that must be
addressed is the lungs, especially if the method is employed in
spontaneously breathing subjects, during Muller and Valsalva maneuvers
or to assess the actual effects of ventilation on ventricular volume
and function. It is well known that both RV and LV volumes change in
response to breathing and ventilation (1, 6, 8, 12-14, 22), but
the effects of changing lung volume on parallel conductance are
unknown.
The present study was undertaken to establish whether LV and RV
parallel conductances are affected by lung insufflation volume and, if
so, to what extent such changes are manifested in the volume signal
fluctuations observed during ventilation in the anesthetized minipig.
Fluctuations in conductance-derived LV and RV end-diastolic and stroke
volumes associated with ventilation with various tidal volumes were
recorded. Multiple determinations of parallel conductance were made,
using a saline dilution technique, at end expiration and during
steady-state lung insufflation at different lung volumes.
This paper also introduces simultaneous biventricular conductance
volumetry, a method that may prove to be a valuable technique for the
assessment of ventricular interaction.
 |
METHODS |
Instrumentation. The study was
conducted on eight anesthetized Göttinger minipigs (11-15
kg). Anesthesia was induced with pentobarbital sodium (75 mg/kg ip),
and pigs were intubated and ventilated with 40%
O2 in air. The left femoral artery
was cannulated (5-Fr single lumen) for taking blood samples and
monitoring arterial blood pressure, and the right femoral vein was
cannulated (7-Fr triple lumen) for administration of anesthetic and
hypertonic saline. Anesthesia was maintained with a
continuous intravenous pentobarbital infusion (0.2 mg · kg
1 · min
1)
and adjusted to a level sufficient to suppress spontaneous breathing. Ventilation was adjusted to achieve normal blood gases. A Fleisch tube
pneumotachogram was connected in series with the endotracheal tube,
which was also connected via a sidearm to a pressure transducer. An
eight-electrode, 5-Fr conductance catheter (Webster Laboratories, Irvine, CA) was inserted through a left carotid artery cutdown and
advanced to the LV apex, and a 5-Fr micromanometer-tip catheter (Millar, Houston, TX) was advanced from the right carotid artery into
the LV. A second 5-Fr micromanometer tip catheter was advanced into the
RV through a right jugular vein cutdown. A 5-Fr, flow-directed, end-hole catheter was advanced from the left jugular vein into the
pulmonary artery, through which an 0.018-in. exchange wire (Cordis,
Miami, FL) was passed into a distal pulmonary artery. The flow-directed
catheter was exchanged for a 6-Fr eight-electrode conductance catheter
(Webster), which was advanced into the pulmonary artery. Interplay
between the conductance catheter and exchange wire was used to form a
curve in the catheter such that the proximal electrode was positioned
at the RV apex with the distal electrode remaining in the pulmonary
artery just above the pulmonary valve. A 5-Fr, flow-directed
thermodilution catheter (Baxter, Irvine, CA) was advanced to the
pulmonary artery from the left femoral vein. All catheters were
positioned under fluoroscopic guidance. The final positions of the
catheters within the heart are shown in Fig.
1.

View larger version (149K):
[in this window]
[in a new window]
|
Fig. 1.
Anteroposterior view showing position of catheters within the heart.
RV, right ventricular; LV, left ventricular.
|
|
Conductance catheter method. The
conductance catheters with outer electrode separations ranging from 4.9 to 5.6 cm were connected to separate volume signal-conditioning units
(BioMetrics, Las Vegas, NV and Cardiodynamics, Oegstgeest, The
Netherlands) and computer data acquisition systems. The
signal-conditioning units were set to different excitation frequencies
(20.0 and 23.5 kHz) allowing simultaneous operation without interfering
with one another. For both left and right ventricles, the five
time-varying, segmental conductance
(G) signals were combined to yield
total volume signals (V) according to
where
L is the conductance catheter
interelectrode separation,
is the measured specific electrical
resistance of blood,
is the conductance gain factor, and
Vc is the offset term required due
to parallel conductance. Correct conductance catheter positions and
electrode separations were ascertained both fluoroscopically and by
monitoring the pressure-volume loops of the segments in the LV and RV
outflow tracts. If, with one excitation electrode at the ventricular
apex, advancing the catheter slightly into the pulmonary artery (in the
case of the RV catheter) or withdrawing it slightly into the aorta (in
the case of the LV catheter) caused an outflow tract segmental
volume-phase shift to occur at end systole (Fig.
2), then electrode separation and position
were deemed to be correct. Synchronization of the two systems was
achieved by issuing an electronic pulse transmitted simultaneously to
both ventricular pressure channels at the beginning of each data
acquisition episode. Data were digitally acquired at 250 Hz and 12-bit
resolution with one system recording RV segmental volumes, RV pressure,
right atrial pressure, pulmonary artery pressure, and
electrocardiogram. The other system recorded LV segmental volumes, LV
pressure, arterial blood pressure, electrocardiogram, and tracheal flow
and pressure.

View larger version (10K):
[in this window]
[in a new window]
|
Fig. 2.
A: left ventricular pressure
segment 5 volume loops from one animal
with segment 5 (proximal segment)
completely in LV outflow tract. B:
temporary, very slight withdrawal of catheter (on the order of 1-2
mm), results in a figure 8-shaped loop. This is most likely the result
of the proximal segment being forced into aorta near end of systole by
ventricular long-axis shortening. With catheter properly positioned,
its tip at ventricular apex, formation of such a
segment 5 pressure-volume loop
suggests that the most proximal sensing electrode resides just below
the aortic valve. A similar loop is formed by RV
segment 1 when conductance catheter is
advanced slightly into the pulmonary artery.
|
|
Protocol. After a 1-h stabilization
period, specific electrical resistance of blood was measured, and
baseline LV and RV pressure-volume data were acquired with the
ventilator off at end expiration. Cardiac output was measured (Baxter
COM1) by thermodilution (3 determinations using 1 ml of 0-5°C
normal saline injections) and used to compute independent LV and RV
conductance gain factors (
). These values for
were assumed to be
constant throughout the protocol.
Parallel conductance was assessed using a saline dilution technique
whereby the conductivity of blood was transiently increased by
injection of a saturated saline solution (0.02-0.04 ml/kg). The
injection was administered into the femoral vein for assessment of RV
parallel conductance and into the pulmonary artery for LV assessment.
If ectopic beats occurred during saline washin or if heart rate or
systolic ventricular pressure changed noticeably, the injection was
repeated with a smaller volume. Analysis of the volume signal transient
used to determine Vc was by
nonlinear regression of apparent stroke volume against maximum and
minimum volumes, a method previously described for both the LV (21) and
RV (18). Briefly, the method combines both maximum and minimum volume
values in a single regression equation. Regressing maximum and
subsequent minimum volumes of each beat during the washin period
against the corresponding stroke volume yields two lines, the
intersection of which occurs at the
y-axis (stroke volume = 0 ml) at some
intercept Vc (Fig.
3). A nonlinear regression technique is
applied to solve for the parameters in a single regression equation
combining minimum, maximum, and stroke volumes
|
|
|
|
where
VV is the ventricular volume (uncorrected for gain or
offset),
mVmax
and
mVmin
are the slopes of the two regression lines, SV is stroke volume defined
as maximum volume less minimum volume, and
C is a coding variable set to 0 for
minimum volumes and 1 for maximum volumes. The regression yields
estimates for the parameters
mVmin
and
mVmax as well as SV0 and
Vc, the coordinate pair defining
the intersection point of the two regression lines. Saline injection
and Vc calculation were repeated
three times, and an average value was computed for each ventricle at
each step of the protocol.

View larger version (20K):
[in this window]
[in a new window]
|
Fig. 3.
Saline dilution method of determining parallel conductance volume
(Vc). Injection of hypertonic
saline into pulmonary artery results in a transient increase to the LV
volume signal (A). A nonlinear
regression method is then applied yielding 2 lines relating maximum and
minimum volumes recorded during washin phase to their corresponding
stroke volumes (B). Lines with
slopes
mVmax
and
mVmin
intersect at y-axis (0 stroke volume)
at a point Vc.
|
|
Synchronized biventricular pressure-volume data were recorded during
ventilation with tidal volumes of 10 and 20 ml/kg at 16 and 12 breaths/min, respectively. Data were also acquired under steady-state
conditions and during hypertonic saline washin, with the ventilator
off, at three lung insufflation volumes: first at ambient pressure, the
endotracheal tube vented to the atmosphere; next with lungs insufflated
to 10 ml/kg; and finally with an insufflation volume of 20 ml/kg. These
recordings were ~20 s long, with onset of steady state occurring
several cardiac cycles after the lungs were inflated and persisting for
another 7-10 cardiac cycles (Fig. 4).
A 5- to 10-min pause, during which normal ventilation was applied, was
allowed between each lung volume maneuver.

View larger version (63K):
[in this window]
[in a new window]
|
Fig. 4.
Simultaneous RV volume (A), RV
pressure (B), LV volume
(C), and LV pressure
(D) recorded during lung
insufflation maneuver. Vertical shaded bar represents initial lung
inflation, which was sustained for ~20 s. Horizontal bar indicates
that period of relative steady state during which hypertonic saline
injections were given.
|
|
Data analysis. To investigate the
effects of ventilation tidal volume on changes to conductance-derived
ventricular volume signals, end-diastolic and stroke volumes of each
cardiac cycle over three contiguous ventilation cycles were calculated.
Direct analysis of the relationship between these volumes and the
position of the cardiac cycle within the ventilation cycle was not
possible because of interanimal heart rate differences and relative
timing of cardiac and ventilation cycles. Thus the three ventilation cycles were divided into 24 sections, each
/4 radians wide (1/8 of a
ventilation cycle). End-diastolic and stroke volumes that would occur
at each
/4 radian division were extrapolated based on a linear
relationship between values falling immediately on either side of the
division. A 2 × 8 repeated-measures analysis of variance was then
applied to test for significant effects of tidal volume (2 levels) and
phase of the ventilation cycle (8 levels) on both LV and RV
end-diastolic and stroke volumes. The changes in end-diastolic and
stroke volumes that occurred during ventilation at both tidal volumes
were also expressed as a percentage of the mean values. The maximum
positive and negative deviations from the mean of both parameters were
computed. The mean LV and RV stroke volumes during ventilation were
compared using a paired sample t-test.
During steady-state lung insufflation, ventricular volume data are
based on an average of 7-10 consecutive cardiac cycles. An
analysis of variance for repeated measures (3 levels of lung inflation)
was used to test for a significant effect on LV and RV
Vc, end-diastolic volume, and
stroke volume. A paired sample t-test
was then used to establish whether the mean LV stroke volume was
different from the mean RV stroke volume at the three steady-state lung
insufflation volumes.
Hemodynamic stability during the measurement of parallel conductance at
each insufflation volume was established by computing the coefficients
of variation of heart rate and maximum rate of change of ventricular
pressure
(dP/dtmax) over
the cardiac cycles starting three beats before the onset of saline
washin and ending with the beat at the peak volume transient. If either
coefficient of variation for any given saline injection episode was
>5%, that run was deemed hemodynamically unstable and was excluded
from subsequent analysis. The precision of
Vc measurement was assessed by
computing the coefficient of variation of the three measurements of
parallel conductance used to determine
Vc for each ventricle at each lung
insufflation volume. Analysis of variance for repeated measures was
used to establish if Vc was
dependent on lung insufflation volume. A paired sample
t-test was used to determine whether
LV Vc was significantly different
from RV Vc.
 |
RESULTS |
Unless otherwise stated, all data are expressed as means ± SD. The
coefficient of variation of thermodilution measurements was 4.72 ± 2.13%. At baseline, with the ventilator off at end expiration, the LV
was 0.64 ± 0.12 and RV
was 0.61 ± 0.18.
Ventilation with tidal volumes of 10 and 20 ml/kg caused cyclic changes
to both LV and RV volumes and pressures (Fig.
5). Analysis of variance revealed that both
LV and RV end-diastolic and stroke volumes were dependent on the
position of the cardiac cycle within the ventilation cycle
(P < 0.0001 for all) with minimum values occurring near end inspiration (Fig.
6), the minimum LV stroke volume lagging
that of the RV by one beat. Furthermore, the magnitude of the cyclic
variation of both LV and RV end-diastolic and stroke volumes was
related to ventilation tidal volume (Table 1). The mean LV stroke volume of 10.31 ± 2.05 ml during ventilation with a 10 ml/kg tidal volume was not
significantly different from the RV stroke volume 10.09 ± 1.73. With a tidal volume of 20 ml/kg, the LV stroke volume was 9.80 ± 2.17 ml, also not different from the RV value of 9.86 ± 2.07 ml.

View larger version (36K):
[in this window]
[in a new window]
|
Fig. 5.
Simultaneous tracheal, LV and RV pressure and volume signals recorded
in one animal during 2 complete ventilation cycles, with a tidal volume
of 20 ml/kg, illustrating decreasing LV and RV volumes concomittant
with increasing tracheal pressure and lung volume.
|
|

View larger version (37K):
[in this window]
[in a new window]
|
Fig. 6.
Results from all 8 animals showing cyclic variation of end-diastolic
volume (A and
C) and stroke volume
(B and
D) with ventilation using tidal
volumes of 10 and 20 ml/kg as percent change from mean values. Circles,
LV values; triangles, RV values; horizontal dashed lines, mean. Shaded
areas depict inspiration portion of ventilation cycle.
|
|
View this table:
[in this window]
[in a new window]
|
Table 1.
Maximum, minimum, and maximum fluctuation of LV and RV end-diastolic
and stroke volumes observed during ventilation with 10 and 20 ml/kg
tidal volumes
|
|
Steady-state lung insufflation at 10 and 20 ml/kg caused shifts in both
LV and RV pressure and volume (Fig. 7).
Lung insufflation volume-dependent decreases in both LV stroke volume
(P < 0.002) and RV stroke volume
(P < 0.0001) were observed. However,
only the LV end-diastolic volume decreased
(P < 0.0001) with steady-state lung
insufflation, with RV end-diastolic volume showing no significant change (Table 2).

View larger version (14K):
[in this window]
[in a new window]
|
Fig. 7.
Simultaneous RV (A) and LV
(B) pressure-volume loops from one
animal recorded during steady-state lung insufflation of 0 (solid
line), 10 (dotted line), and 20 ml/kg (dashed line).
|
|
View this table:
[in this window]
[in a new window]
|
Table 2.
LV and RV stroke and end-diastolic volumes measured during steady-state
lung insufflation with volumes of 0, 10, and 20 mg/kg
|
|
Of the combined total of 144 saline runs recorded for this study, 1 was
omitted from LV analysis and 2 from RV analysis due to not meeting the
hemodynamic stability test. Analysis of variance did not reveal any
relationship between LV Vc or RV
Vc and the degree of lung
insufflation. Values of Vc are
presented in Table 3 as are the mean
coefficients of variation of the repeated
Vc determinations made at each
state of lung insufflation, which also were not related to lung
insufflation volume. Parallel conductance for the RV was significantly,
albeit slightly, lower than that measured for the LV
(P < 0.05).
 |
DISCUSSION |
The conductance catheter technique is being increasingly applied to
measure LV and RV volumes in both animals and humans. Whereas most
studies may benefit from the absence of hemodynamic perturbations
associated with breathing or ventilation, suspending respiration during
conductance measurements may not always be possible or even desirable
(16). In those instances, it will be important to know the extent to
which volume signal artifact is present. One possible component of that
artifact would be parallel conductance associated with the degree of
insufflation of the lungs. This study documented the changes to both LV
and RV conductance-derived volumes during ventilation and investigated
the effects of lung insufflation volume on LV and RV parallel
conductances in an anesthetized minipig model.
Theoretically, the volume of air, an electrical insulator, within the
lungs could affect conductance catheter parallel conductance. Changes
in lung volume associated with respiration and ventilation could
possibly be manifested as changes to conductance-derived volume
signals. Specifically, if the electrical conductivity of the lungs were
decreased during inspiration, a decrease in parallel conductance may
follow, resulting in underestimation of absolute ventricular volumes.
Another mechanism by which parallel conductance could be affected by
lung insufflation is the degree to which the lungs are engorged with
blood during the respiratory cycle. Whereas significant changes to both
LV and RV conductance-derived stroke and end-diastolic volumes were
observed with ventilation, assessment of parallel conductance during
steady-state lung insufflation did not reveal any relationship between
insufflation volume and parallel conductance, which remained
essentially unchanged throughout the protocol. This can be explained by
the relatively low conductance of air. Even a small amount of air, the
residual capacity of the lungs after expiration, can provide sufficient
electrical separation that increasing the lung volume further has
little effect on Vc. It may
require complete collapse of the lungs for this effect to have an
impact on parallel conductance.
An indirect mechanism of affecting parallel conductance by lung
insufflation would be that of changing volume of the contralateral ventricle, the blood pool of the opposite ventricle offering a source
of parallel conductance. However, the effects of RV volume on LV
parallel conductance are reportedly minimal (9, 21). Furthermore, some
studies have indicated Vc is
volume dependent, LV Vc being
somewhat dependent on LV volume itself (2, 5). However, it has also
been suggested that such volume dependency is more related to the
method of achieving volume reduction, specifically, inferior vena cava
occlusion (21). When LV parallel conductance was assessed at different
ventricular volumes achieved by volume loading and depletion (which
would necessarily have also affected RV volume), it was found to be
relatively unchanged by the procedures (21). In the present study,
parallel conductance was assessed under relatively steady-state
conditions. It is possible that, under the dynamic conditions of
respiration, Vc may respond
differently. However, because the effect of ventilation on
conductance-derived RV volume was double that of the LV, one would
expect that if Vc were being
affected to any extent by associated changes in actual LV or RV
volumes, the mean conductance-derived ventricular outputs would have
been different. Because this was not the case, we can assume that
either there is little effect or the result is masked because the
effects on Vc are similar in both
ventricles. In any case, if some ventricular volume-dependent effect on
Vc were present, at a normal tidal
volume of 10 ml/kg, end-diastolic volumes only changed by ~5%, thus
offering little cause for concern regarding this possible mechanism.
Ventilation caused significant fluctuations in LV and RV stroke
volumes, the magnitude of which were related to tidal volume. Such
cyclic changes in both LV and RV output during the respiratory cycle
are well known (1, 6, 8, 12-14, 22). RV stroke volume decreases
with inspiration as intrathoracic and intrapulmonary pressures
increase, while decreases in LV stroke volume are smaller and lag those
of the RV by a few beats. On expiration, RV stroke volume recovers
almost immediately while that of the LV increases gradually. Given that
ventilation causes larger changes to RV stroke volume, those changes
must be more abrupt with higher maximum and lower minimum values than
the LV in order to maintain equal mean RV and LV outputs. Such events
have led the RV to be described as a buffer of changes in systemic
venous return, effectively dampening respiratory-induced variations of
LV stroke volume (17). Observations made in the present study are in
keeping with the sited work as can be seen in Fig. 5; fluctuations in
RV stroke volume almost double those of the LV with maximum decreases
to both occurring near end inspiration, the LV effect lagging the RV by
at least a beat.
Some of the observed changes to stroke volume could have been a result
of changes to
; however, because of the design of the protocol,
possible effects on
can only be speculated on, which is a
limitation of this study. Whereas it is well established that both LV
(15, 16) and RV (7, 18)
values are somewhat inversely related to
ventricular volume, there is no reason to suspect that changes in lung
volume would directly affect this parameter. Rather, an indirect effect
could again be postulated in that lung insufflation does alter absolute
ventricular volumes and thereby could also change
. As ventricular
volumes decrease during inspiration, a concomitant increase in
may
be expected. In the present study, for each ventricle
was
determined at the beginning of the protocol and thereafter was assumed
to be constant. Thus both stroke and end-diastolic volumes may be
increasingly overestimated with increasing lung insufflation. Whereas
such overestimation would not affect the conclusions drawn by this study regarding Vc, changes to LV
and RV stroke and end-diastolic volumes with ventilation may have been
underestimated. If
were affected by ventilation, the change must
have been similar in both LV and RV because there was no significant
difference between the mean LV and RV stroke volumes during ventilation
with 10 and 20 ml/kg tidal volumes. Steady-state lung insufflation at
0, 10, and 20 ml/kg did result in insufflation volume-related decreases to both LV and RV stroke volume with that of the RV decreasing more
than the LV, although not significantly. Because during steady state
one would expect both ventricles to have the same stroke volume, this
could suggest that the LV
was affected more than that of the RV.
However, because the excursion of end-diastolic volume with ventilation
with a normal tidal volume of 10 ml/kg was relatively small, on the
order of 5%, substantial changes to
do not seem likely. The
changes in absolute ventricular volumes associated with the normal
ejection (stroke volume) of both ventricles, which are greater than
those due to normal lung inflation, have previously been shown to have
a minimal effect on
(18, 20).
Ventilation also resulted in fluctuations in both LV and RV
end-diastolic volume. Again, the effect on the RV was double that of
the LV, with minimum end-diastolic volumes occurring at end inspiration. As with stroke volume, recovery of RV end-diastolic volume
occurred almost immediately on expiration, whereas that of the LV was
more gradual, not reaching a maximum value plateau before onset of the
next inspiratory phase. Because this may be the first study to record
absolute ventricular volumes during respiration, there is little in the
literature with which to compare these findings. In the context of this
study, though, the most important observation regarding end-diastolic
volume is that absolute ventricular volumes only changed minimally
(relative to stroke volume for example) with insufflation, thus
limiting the impact that ventricular volume changes may have had on
either
or Vc.
When continuous lung insufflation was applied, following a transient of
a few cardiac cycles (Fig. 6), RV end-diastolic volume returned to its
preinsufflation value, whereas stroke volume remained decreased. RV
end-diastolic volume assumed a value falling between the maximum and
minimum observed during ventilation, whereas LV end-diastolic volume
reached a plateau at a value even less than the minimum observed during
ventilation. During this sustained insufflation, hemodynamic signals
remained stable sufficiently long enough to assess parallel
conductance, with only three saline runs being rejected from analysis
due to hemodynamic instability. Inflation of the lungs was also stable
during this period as confirmed by both elevated, yet constant,
end-diastolic pressures and stable intratracheal pressure. We are,
therefore, confident that parallel conductance measurements were
reliable and were, in fact, made at different lung volumes.
This paper introduces the use of the conductance catheter technique to
measure pressure-volume data of both ventricles, simultaneously offering an exciting prospect for a novel method of studying
ventricular interaction. No technical obstacles were apparent in using
the technique once it was ascertained that the two volume
signal-conditioning units did not interfere electrically with one
another. The method simply combines LV and RV techniques already
described with the addition of an electrical synchronizing pulse
recorded by both systems. The synchronizing signal would not have been
necessary if data were recorded on a single computer. The use of double bolus injections of hypertonic saline for assessment of LV and RV
parallel conductances also may not be required because the femoral vein
injection used for RV Vc also
resulted in a transient in the LV signal, which could be used for
computation of LV Vc.
In conclusion, in this study, there was no evidence that the degree of
lung insufflation affects either LV or RV conductance catheter parallel
conductance. The possibility that
may have changed as a result of
the effects of lung insufflation on absolute ventricular volumes should
not preclude use of the technique during normal ventilation or
respiration, since the absolute volume of neither ventricle changes a
great deal under those circumstances. It should be noted that these
results may not be valid under conditions of spontaneous respiration,
during which the diaphragm may move in and out of the conductance
catheter excitation field, thus altering parallel conductance.
Furthermore, pericardial or pleural effusions may increase both
mechanical and electrical coupling of both ventricles with the lungs
and other extracardiac structures, thus allowing the possibility of
changes to both
and Vc with ventilation.
 |
FOOTNOTES |
Address for reprint requests: R. S. Szwarc, 9708 Heirloom Ct., Las
Vegas, NV 89134.
Received 10 December 1997; accepted in final form 4 May 1998.
 |
REFERENCES |
1.
Abel, F. L.,
and
J. A. Walhhausen.
Respiratory and cardiac effects on venous return.
Am. Heart J.
78:
266-275,
1969[Medline].
2.
Applegate, R. J.,
C. P. Cheng,
and
W. C. Little.
Simultaneous conductance catheter and dimension assessment of left ventricular volume in the intact animal.
Circulation
81:
638-648,
1990[Abstract/Free Full Text].
3.
Baan, J.,
T. T. Aouw Jong,
P. L. M. Kerkhof,
R. J. Moene,
A. D. Van Dijk,
E. T. Van der Velde,
and
J. Koops.
Continuous stroke volume and cardiac output from intraventricular dimensions obtained with impedance catheter.
Cardiovasc. Res.
15:
328-334,
1981[Medline].
4.
Baan, J.,
E. T. Van der Velde,
H. G. De Bruin,
G. J. Smeenk,
J. Koops,
A. D. Van Dijk,
D. Temmerman,
J. Senden,
and
B. Buis.
Continuous measurement of left ventricular volume in animals and humans by conductance catheter.
Circulation
70:
812-823,
1984[Abstract/Free Full Text].
5.
Boltwood, C. M.,
R. F. Appleyard,
and
S. A. Glantz.
Left ventricular volume measurement by conductance catheter in intact dogs.
Circulation
80:
1360-1377,
1989[Abstract/Free Full Text].
6.
Charlier, A. A.,
P. M. Jaumin,
and
H. Pouleur.
Circulatory effects of deep inspirations, blocked expirations and positive pressure inflations at equal transmural pressures in conscious dogs.
J. Physiol. (Lond.)
241:
589-605,
1974[Abstract/Free Full Text].
7.
Dickstein, M. L.,
O. Yano,
H. M. Spotnitz,
and
D. Burkhoff.
Assessment of right ventricular contractile state with the conductance catheter technique in the pig.
Cardiovasc. Res.
29:
820-826,
1995[Medline].
8.
Hoffman, J. I. E.,
A. Guz,
A. A. Charlier,
and
D. E. L. Wilcken.
Stroke volume in conscious dogs: effect of respiration, posture, and vascular occlusion.
J. Appl. Physiol.
20:
865-877,
1965[Abstract/Free Full Text].
9.
Kass, D. A.,
M. Midei,
W. Graves,
J. A. Brinker,
and
W. L. Maughan.
Use of a conductance (volume) catheter and transient inferior caval occlusion for rapid determination of pressure-volume relationships in man.
Cathet. Cardiovasc. Diagn.
15:
192-202,
1988[Medline].
10.
Kass, D. A.,
T. Yamazaki,
D. Burkhoff,
W. L. Maughan,
and
K. Sagawa.
Determination of left ventricular end-systolic pressure-volume relationships by the conductance (volume) catheter technique.
Circulation
73:
586-595,
1986[Abstract/Free Full Text].
11.
Khy, D.,
H. McAlister,
B. Wilkoff,
T. Simmons,
Y. Rudy,
R. McCowan,
V. Morant,
L. Castle,
and
J. Maloney.
Continuous right ventricular volume assessment by catheter measurement for Antitachycardia System Control.
PACE
12:
1918-1926,
1989.
12.
Moreno, A. H.,
A. R. Burchell,
R. Woude,
and
J. H. Burke.
Respiratory regulation of splanchnic and systemic venous return.
Am. J. Physiol.
213:
455-465,
1967.
13.
Morgan, B. C.,
W. E. Martin,
T. F. Hornbein,
E. W. Crawford,
and
W. G. Guntherorth.
Hemodynamic effects of intermittent positive pressure respiration.
Anesthesiology
27:
584-590,
1966[Medline].
14.
Peters, J.,
C. Fraser,
R. S. Stuart,
W. Baumgartner,
and
J. L. Robotham.
Negative intrathoracic pressure decreases independently left ventricular filling and emptying.
Am. J. Physiol.
257 (Heart Circ. Physiol. 26):
H120-H131,
1989[Abstract/Free Full Text].
15.
Salo, R. W.
The theoretical basis of a computational model for the determination of volume by impedence.
Automedica
11:
299-310,
1989.
16.
Salo, R. W. Improvement in intracardiac impedance
volumes by field extrapolation. Eur. Heart
J. 13, Suppl. E:
35-39, 1992.
17.
Santamore, W. P.,
and
J. N. Amoore.
Buffering of respiratory variations in venous return by right ventricle: a theoretical analysis.
Am. J. Physiol.
267 (Heart Circ. Physiol. 36):
H2163-H2170,
1994[Abstract/Free Full Text].
18.
Stamato, T. M.,
R. S. Szwarc,
and
L. N. Benson.
Measurement of right ventricular volume by conductance catheter in closed-chest pigs.
Am. J. Physiol.
269 (Heart Circ. Physiol. 38):
H869-H876,
1995[Abstract/Free Full Text].
19.
Szwarc, R. S., and H. A. Ball.
Determination of left ventricular end-systolic elastance
(Ees) and preload recruitable
stroke work (PRSW) using lung inflation to reduce preload.
J. Mol. Cell Cardiol. 25, Suppl. I: II-P27, 1993.
20.
Szwarc, R. S.,
D. Laurent,
P. R. Allegrini,
and
H. A. Ball.
Conductance catheter measurement of left ventricular volume: evidence for nonlinearity within the cardiac cycle.
Am. J. Physiol.
268 (Heart Circ. Physiol. 37):
H1490-H1498,
1995[Abstract/Free Full Text].
21.
Szwarc, R. S.,
L. L. Mickleborough,
S. I. Mizuno,
G. J. Wilson,
P. Liu,
and
S. Mohamed.
Conductance catheter measurement of left ventricular volume in the intact dog: parallel conductance is independent of left ventricular size.
Cardiovasc. Res.
28:
252-258,
1994[Abstract/Free Full Text].
22.
Versprille, A.,
J. R. C. Jansen,
R. C Frietman,
A. R. Hulsman,
and
M. M. Klauw.
Negative effect of insufflation on cardiac output and pulmonary blood volume.
Acta Anaesthesiol. Scand.
34:
607-615,
1990[Medline].
23.
Woodard, J. C.,
C. D. Bertram,
and
B. S. Gow.
Detecting right ventricular volume changes using the conductance catheter.
PACE
15:
2283-2294,
1992.
Am J Physiol Heart Circ Physiol 275(2):H653-H661
0002-9513/98 $5.00
Copyright © 1998 the American Physiological Society