|
|
||||||||
1 Division of Pulmonary and Critical Care Medicine, 2 Department of Statistics, and 3 Department of Physiology and Biophysics, University of Washington, Seattle, 98195; and 4 The Mountain-Whisper-Light Statistical Consulting, Seattle, Washington 98112
| |
ABSTRACT |
|---|
|
|
|---|
Microsphere experiments are useful in measuring regional organ perfusion as well as heterogeneity of blood flow within organs and correlation of perfusion between organ pieces at different time points. A 400 microspheres/piece "rule" is often used in planning experiments or to determine whether experiments are valid. This rule is based on the statement that 400 microspheres must lodge in a region for 95% confidence that the observed flow in the region is within 10% of the true flow. The 400 microspheres precision rule, however, only applies to measurements of perfusion to a single region or organ piece. Examples, simulations, and an animal experiment were carried out to show that good precision for measurements of heterogeneity and correlation can be obtained from many experiments with <400 microspheres/piece. Furthermore, methods were developed and tested for correcting the observed heterogeneity and correlation to remove the Poisson "noise" due to discrete microsphere measurements. The animal experiment shows adjusted values of heterogeneity and correlation that are in close agreement for measurements made with many or few microspheres/piece. Simulations demonstrate that the adjusted values are accurate for a variety of experiments with far fewer than 400 microspheres/piece. Thus the 400 microspheres rule does not apply to many experiments. A "rule of thumb" is that experiments with a total of at least 15,000 microspheres, for all pieces combined, are very likely to yield accurate estimates of heterogeneity. Experiments with a total of at least 25,000 microspheres are very likely to yield accurate estimates of correlation coefficients.
organ perfusion; heterogeneity; correlation; modeling; statistics; Poisson noise
| |
INTRODUCTION |
|---|
|
|
|---|
THE MICROSPHERE METHOD introduced by Rudolf and Heymann (10) has become the gold standard for measuring regional organ blood flow. When microspheres are used to measure regional organ perfusion, a commonly held tenet is that at least 400 microspheres must lodge in each region of interest. This stipulation is derived from theoretical and experimental work of Buckberg and coworkers (2), in which they concluded that for 95% confidence that observed flow to a region is within 10% of the true flow, at least 384 microspheres must lodge within the organ sample. A subsequent study (9) reported that fewer microspheres are needed in each piece if the reference blood sample contains adequate numbers of microspheres, but this study is rarely cited. Regardless of the experimental goals, many investigators have interpreted Buckberg's guideline to mean that if an organ sample has <400 microspheres, the observed blood flow is so unreliable that the sample should be excluded from the analysis. Others consider results to be unreliable for an entire experiment if some or all of the pieces have <400 microspheres. If the experimental objectives are to examine blood flow rates to individual pieces of an organ, then the 400-microsphere rule is helpful. However, if investigators are interested in general descriptors of blood flow distribution, such as means, slopes (flow gradient over a spatial distance), coefficients of variation, and correlation coefficients, many fewer microspheres per region are required.
A second incentive for clarifying the 400 microsphere per piece rule is the recent introduction of histological methods that permit blood flow distributions to be determined from direct counting of colored and fluorescent microspheres at the microscopic level (3, 8). Higher resolution and smaller regions of interest result in many fewer microspheres being counted in each region of interest. Methods for smaller regions of interest are well suited to experiments with smaller animals, such as transgenic mice. Because the number of microspheres in an organ sample can be modeled by the Poisson distribution, the random variations, or "noise" component, can be estimated and equations for heterogeneity or other statistics can be mathematically corrected for noise. In this paper, we present experimental evidence in support of the Poisson assumption. The methodological or Poisson noise component of measurements becomes increasingly important in experiments in which the number of microspheres distributed to organ regions is small. We now present the theory and derivations of the formulas needed to correct measures of flow heterogeneity and correlations between flows for Poisson noise. We also present computer simulations to demonstrate the utility of these formulas. Finally, we present the results of an animal experiment that uses <400 microspheres per sample piece yet yields reliable estimates of heterogeneity and correlation.
The hypothetical example presented in Table
1 motivates the methods developed later and
shows that some firm conclusions can be made even when all sample
pieces have <400 microspheres. The first column shows
the expected number of microspheres for each of 10 pieces based on the
true flow to each piece, and the next two columns show
"observations" on these pieces, generated by random sampling from
the Poisson distribution with the expected number of microspheres for
each piece. As predicted from Buckberg's work, the percent error for
some pieces is unacceptably high. However, it should be noted that when
the mean and heterogeneity (coefficient of variation) are calculated
for these three columns, they are very similar, differing by <2% out
of 58%. Note also that the correlation between the two observations,
r = 0.985, is close to 1.0, which is
the correlation that would be observed in the absence of any random
variation (the correlation of column 1 with itself). Thus the experimental observations are very
similar to the "truth" (column
1), even though not one of the pieces has an expected
value or an observed value >400 microspheres. As a further example,
it seems clear from the data in experiment
1 that the piece with 6 microspheres and the piece with
101 microspheres represent very different flows. This is formally
confirmed by a
2 test (1 degree
of freedom), which yields P < 0.0001. Thus one can conclude with a great deal of confidence that there is a different flow to these two pieces, even though both pieces have many fewer than
400 observed microspheres. A number of other differences in observed
microsphere counts between pairs of pieces in Table 1 are also
statistically significant. The table also includes adjusted values for
the coefficient of variation (CV) and correlation, using the methods
developed later in this paper. In each case, the adjusted values are
closer to the true values than the values based on observed results.
The adjustments are quite minor.
|
We show that quite good estimates of heterogeneity and correlations can be calculated even when many regions (pieces) have relatively low microsphere counts. Furthermore, heterogeneity and correlation can also be adjusted to take account of the Poisson noise.
| |
METHODS |
|---|
|
|
|---|
Theory
For any given piece or region, there is a certain probability that a microsphere will lodge in the region at the time of injection. If microspheres act independently, every microsphere has the same probability of reaching a specified region. The expected number of microspheres reaching a region will be N × p, where N is the number of microspheres injected and p is the proportion of total flow to the organ, which perfuses the region. Thus p is also the expected proportion of microspheres captured by the specified region. Because N is large and p is small, the distribution of n, the number of microspheres actually reaching the region, can be well approximated by the Poisson distribution (12). Previous studies have attempted to correct for Poisson noise using the total microsphere count for the entire organ (5, 7), but the methodology presented here is more realistic and precise in its use of microsphere data for individual pieces of an organ.Notation.
We use the subscript "true" to indicate the unobservable true
value of a quantity, e.g., CVtrue
represents the coefficient of variation of true flow, measured without
error. We use the subscript "obs" to denote a quantity calculated
by conventional formulas from the observed data from an experiment or
from randomly simulated data that are intended to mimic observed data
from a real experiment. We use the subscripts or notation "adj,"
"adjusted," or "corrected" to indicate an estimate, based
on the equations developed in this paper, that attempts to remove the
effect of random error due to Poisson noise. We also use the hat symbol "
" to indicate
these adjusted estimates. Thus
true = CVadj is an estimate of the true
CV. Similarly, rtrue is the true unobservable
correlation, robs is the observed correlation,
and both radj and
true are estimates of the true
correlation incorporating the adjustments described in this paper.
Adjusting estimates of heterogeneity for noise.
When heterogeneity is expressed as the CV, then, with the use of the
observed number of microspheres per piece
(Xi for
piece i), the observed CV is calculated as
|
(1) |
is the
mean of Xi.
Random errors will tend to cancel in the denominator when the mean is
taken across all pieces, but the numerator, the standard deviation,
will be inflated when Poisson variation is added to the true variation
in flow.
The Poisson noise can be adjusted "out" of the CV estimate by
using a relatively simple equation that is derived in
APPENDIX 1.
Equation 2 provides an estimate of the
true CV
|
(2) |
are large,
this formula simplifies to
|
(3) |
.
Equation 3 is then equivalent to that
proposed by Iversen et al. (7), which is
CV2true = CV2obs
CV2noise.
Adjusting estimates of correlation for noise.
Pearson's correlation (r) measures
the tendency of flows to vary together (e.g., before and after an
experimental change, or time
1 vs.
time
2, etc.). The Pearson correlation
robs is
calculated from the observed data, pairs of values
(Xi,
Yi) for
N pieces, as
|
(4) |
|
(5) |
true and
true,
however, estimate heterogeneity and correlation of true absolute flow
or true relative flow. It is important in using Eqs.
2 and 5 that
microsphere counts be used in calculation of all means and standard
deviations. Absolute blood flow or relative blood flow should not be
used in the calculations.
Simulation Tests
The applicability and robustness of the derived formulas, Eqs. 2 and 5, were explored through computer simulations of organ blood flow and microsphere injections. Organ blood flow distributions were generated with the fractal model of van Beek et al. (11). This model produces heterogeneous flows that are skewed to the right (toward higher flows), similar to experimental observations of perfusion distributions to myocardium (11), lung (4), and skeletal muscle (6). The details of the construction of simulated organs can be found in APPENDIX 3.Single and paired flows for true organs.
"True" organs were generated by specifying the perfusion
heterogeneity and the number of pieces in the manner described in APPENDIX 3, and paired true
sets of piece flows were generated with specified correlations,
representing paired observations on pieces. Every generated true organ,
then, had a paired value of
(Xtrue,
Ytrue) for each
piece. A variety of true organs were generated by varying the nominal
perfusion heterogeneity (CV = 25, 50, 100%), the nominal correlation
(r =
0.7,
0.3, 0.0, 0.3, 0.7, 0.98), and the number of pieces
(N = 8, 16, 32, 64, 128, 256). Because
of the numerical procedures used in generating true organs, the CVs and
correlations can differ slightly from the specified target values. The
true CV values for generated organs ranged from 24.8 to 100.2%, and
the true correlations ranged from
0.742 to 0.995.
Simulation of observed organs.
The microsphere data for simulated observed organs were generated by
randomly drawing from the Poisson distribution to yield an observed
value or pair of values for each piece. The Poisson distribution used
for each piece had an expected value of
Xtrue (or
Ytrue). The
overall noise level of an experiment was specified by the expected
number of microspheres per piece, which varied across 14 levels: 3, 6, 12, 25, 50, 100, 200, 400, 800, 1,600, 3,200, 6,400, 12,800, and 25,600 microspheres. For each combination of perfusion heterogeneity, number
of pieces, number of microspheres/piece, and correlation, we generated
observed organs and paired organs with the same mean number of
microspheres/piece. Altogether, 1,512 combinations of these parameters
were used to test the CV correction (Eq. 2) and the correlation coefficient
(Eq. 5). We simulated 500 observed
organs per combination. For each of the 500 simulations, the value of
CVobs was corrected to
true using
Eq. 2 and the value of
robs was
corrected to
true using
Eq. 5. For each set of 500, we
calculated the mean, variance, and standard deviation of
CVobs,
true,
robs, and
true.
Experimental Protocol
We carried out an animal experiment to illustrate the use of our adjusted estimates with biological data. We also wanted to demonstrate that relatively small numbers of microspheres per sample piece could still yield quite accurate estimates of the CV and correlations.The study was approved by the University of Washington Animal Care Committee. A single 3-kg New Zealand White rabbit was chemically restrained with intramuscular ketamine, anesthetized with thiopental, and allowed to breathe spontaneously. An internal jugular catheter was placed. Fifteen-micrometer-diameter microspheres with four different radiolabels (153Gd, 113Sn, 103Ru, and 95Nb; NEN) were used to determine regional pulmonary blood flow. Radioactivity per microsphere for each radiolabel was determined by visually counting microspheres on filters and then determining the radioactive counts from each filter. The numbers of microspheres injected were intentionally chosen to provide large numbers of gadolinium and tin with small numbers of ruthenium and niobium. The microspheres were sonicated, vortexed, mixed in a single syringe, and then injected via the internal jugular catheter over 30 s. The animal was killed by anesthetic overdose, and the lungs were removed, air dried inflated, and diced into 100 pieces. Each piece was placed in an individual scintillation vial, and the radioactivity of each radionuclide was determined in a scintillation counter (Packard Minaxi gamma counting system, model 5550). Each sample was read for 20 min or until sufficient counts were obtained to provide a counting error <0.5% for each radionuclide. The radioactive counts for each radionuclide in each piece were corrected for spillover using a matrix inversion method and were also corrected for background and decay. The numbers of each microsphere in each piece were calculated from radioactive counts and the measured value of radioactive counts/microsphere.
| |
RESULTS |
|---|
|
|
|---|
Experimental Results
Our data consist of 100 pieces of lung, each with 4 radioactive microsphere measurements (153Gd, 113Sn, 103Ru, and 95Nb). On average, there are 1,599 153Gd, 880 113Sn, 131 103Ru, and 80 95Nb microspheres per organ piece. We examine the CVs and correlations in these data, as well as our assumption that the noise can be modeled by a Poisson distribution.Table 2 shows the conventional observed CV
for each microsphere measurement computed from Eq. 1, as well as the adjusted CV computed from
Eq. 2. With these data, our adjustment
changes the CV only slightly. The adjusted CV values for all four
measurements show extremely good agreement, differing by
0.008. The
observed values of CV are all also quite similar (differing by
0.01)
even though two of the injections have quite low mean counts of
microspheres/piece (103Ru and
95Nb), well below 400 microspheres/piece.
|
Table 3 shows the Pearson correlation
between each pair of microsphere measurements; for each correlation, we
show both the conventional observed value computed from
Eq. 4 and the adjusted value computed
from Eq. 5. The true correlation in
the absence of all noise should be 1, because the measurements are made
at the same time on the same pieces. The correction brings each
correlation closer to the true value of 1.0; however, all correlations
are very similar and very close to 1.0, even though some are based on
low microsphere counts (103Ru and
95Nb). Figure
1 shows the strong correlation between
103Ru- and
95Nb-based microsphere counts,
with little scatter around the least-squares line. The observed
correlations differ from 1.0 by
0.03, and the adjusted correlations
differ from 1.0 by
0.006. Clearly, Poisson noise has had little
influence on these correlations.
|
|
A further examination of the microsphere counts provides evidence in
favor of the Poisson noise assumption. True Poisson noise has a piece
variance equal to the piece mean. To check this assumption, we
estimated the variance and mean for each piece. We regressed each of
the low signal measurements (103Ru
and 95Nb) on the sum of the other
three measurements with a zero-intercept model. For each piece the
squared residual from the regression model is an estimate of the piece
variance; this can be compared with the regression-fitted expected
microsphere count for the same piece, which is an estimate of the true
piece mean. (A single outlier in the entire data set was excluded from
the analysis of 103Ru because its residual was >4
standard deviations from the regression line.) In a second analysis, we
regressed the estimated piece variance on the estimated mean. We found
that the regression slope of variance versus mean was close to 1.0, the
value expected if the microsphere counts follow the Poisson
distribution (Table 4). This supports (but
does not prove) the assumption of Poisson noise. The increment in slope
above 1.0 is most likely caused by variation not related to the
distribution of microspheres. Such variation would include random
variation in radioactive counting and laboratory error.
|
Definitively showing that the Poisson distribution fits microsphere count data would require an experiment that is currently impossible. One would have to carry out a large number of simultaneous microsphere injections, each with a unique radioisotope or other tag, and then compare the resulting microsphere count distribution to the Poisson distribution. Constraints on tagging microspheres and on body burden render this experiment impossible with current technology.
Simulation Tests
When the number of microspheres in an experiment is large, the unadjusted CV and correlation differ little from the true CV and correlation and either can be used. However, as the number of microspheres in an experiment is reduced, the adjusted CV and correlation are approximately unbiased and are accurate for much smaller numbers of microspheres than the unadjusted values, as shown by the simulation results below.We initially illustrate the simulation results with some examples and
then present the overall results. The relative performance of the
observed and adjusted CV are displayed for four true organs in Fig.
2, which shows simulation results for small
(CV = 25%) and large (CV = 100%) heterogeneity and for small
(n = 16) and large
(n = 256) numbers of pieces. Note in
all plots that the mean
true (noted as
"corrected") is closer to the true CV for considerably smaller
numbers of microspheres than the mean
CVobs (noted as "observed").
Also note that both CVtrue and
CVobs are quite close to the true
CV, with rather small standard deviations, for a number of the points
(observed organs) <400 microspheres/piece. Also, in the top panels of
Fig. 2, where the true CV is larger than in the bottom panels,
CVobs diverges less rapidly from
the true CV as microspheres/piece decreases. The larger true CV (the "signal") is much more prominent than the Poisson error (the
"noise"), and the noise has relatively less influence.
|
Although the observed CV of organ perfusion is dependent on the number
of pieces into which the organ is dissected (1), these simulations show
that unbiased estimates of the observed perfusion heterogeneity can be
obtained from an organ with as few as 16 pieces and an average of only
4 microspheres per piece by using Eq. 2 for
true. The primary
consequence of using such small numbers of organ pieces and
microspheres is a large variance in the estimate of heterogeneity and
hence a loss of confidence in the observation from a single experiment.
Because the corrected estimates are unbiased, observed CV values
averaged over a number of experiments will produce quite accurate
estimates of the true CV.
The simulation results for correlations show features similar to those
for the CV. Figure 3 illustrates
performance for low (true r = 0.27-0.34) and strong (true r = 0.98-0.99) correlations and for many
(n = 256) and few
(n = 16) pieces. The CV for all of the
simulations in these figures is ~50%. Again, both
robs (noted as
"observed") and
true
(noted as "corrected") are close to
rtrue, with small
standard deviations for a number of the points below 400 microspheres/piece. The mean
true stays
closer to the true r than
robs for much
smaller numbers of microspheres per piece. Also,
robs diverges
from the true correlation much more rapidly with decreasing
microspheres/piece when the correlation is large.
|
These examples illustrate that the observed CVs and correlations can be quite accurate for many experiments with <400 microspheres/piece and that use of corrected CVs and correlations extends accurate results into experiments with considerably smaller numbers of microspheres/piece.
We can summarize the accuracy of the corrected and uncorrected CV and
correlation estimates using two key measures: bias (systematic error,
e.g., expected value of CVobs
CVtrue) and root mean
square error [RMSE = (bias2 + variance)1/2]. The RMSE
incorporates both bias and variability. A small value of RMSE ensures
that both bias and variability are small and that the estimated
quantity is accurate. For completeness, we also present results on the
standard deviation of the CV and of the correlation. The standard
deviation is a concept of precision (not accuracy), but good precision
is not helpful if bias is large.
Table 5, based on the simulations, shows
the minimum size of experiments, expressed as the total number of
microspheres in the organ, that yielded an estimated bias, standard
deviation, and RMSE at or less than the amount shown, for all sets of
simulation parameters (number of pieces,
CVtrue, and
rtrue). The
1,512 simulated experiments used to construct the table were described
in Simulation Tests and include a wide
range of numbers of pieces, true CV, microspheres per piece, and
correlation. The corrected estimates have a clear
advantage in bias and RMSE; the uncorrected estimates require many more
microspheres to obtain levels of bias comparable to those of the
uncorrected estimates. For example, to have a quite accurate estimate
of the CV, as indicated by an RMSE of
2%, at least 25,260 microspheres are needed (all pieces combined) when the conventional
CVobs is used. However, if
CVtrue is used, only 12,803 microspheres are needed, about one-half as many. In an experiment
involving correlations, the conventional
robs would require >400,000 microspheres and
true would
require ~100,000 microspheres to yield very good accuracy with a
small RMSE of 0.02. The standard deviation results are about the same
for both corrected and uncorrected estimates. Thus the smaller values
of RMSE for the corrected estimates compared with the conventional observed estimates are primarily caused by the smaller values of bias
for the corrected estimates.
|
The blank cells in Table 5 indicate that there were some parameter sets for which the standard deviation or RMSE never reached the minimum level indicated in the table. The maximum number of microspheres (organ total) in the simulations was 6,553,600; using more microspheres would provide lower levels of bias, standard deviation, and RMSE, even in the extreme cases.
Effect on Experimental Design
We developed approximate equations for the variance of
true and
true that
can be used in planning the number of microspheres to be used in
experiments to achieve a specified level of accuracy. The exact
variance of
true and
true
cannot be written in closed form, and the asymptotic variances are
extremely complex. Thus we fit approximate models for these variances
using linear regression with the observed variances as dependent
variables and combinations of powers, products, and other functions of
simulation parameters (number of pieces, heterogeneity, microspheres
per piece, correlation) as independent variables. All of the
simulations were pooled for this modeling.
The fitted models for the variance of
true, expressed as a
percentage (Eq. 6), and the variance
of
true
(Eq. 7) can be used to estimate the
precision of an experiment. The model for the variance of
true fit well for
experiments with at least 1,500 total microspheres, yielding
R2 = 0.89. We
obtained R2 = 0.61 from the model for the variance of
rtrue for
experiments with at least 3,300 total microspheres
|
(6) |
. The
variance of CVtrue will,
therefore, be approximately the same for two experiments, one with 10 pieces and an average of 800 microspheres/piece and a second with 100 pieces and an average of 80 microspheres/piece
|
(7) |
true.
To illustrate the use of the variance equations, if we consider an
experiment with N = 200 pieces, an
average of 200 microspheres/piece for each of the two separate
microsphere injections, and CVtrue = 50% for both of our paired measurements, then the estimated standard
error (SE) of
true
(square root of the variance) would be 0.62%, and the estimated SE of
true would
be 0.017. This would lead to approximate 95% confidence intervals of
±1.2% for the adjusted CV estimate and ±0.033 for the adjusted
correlation estimate. This clearly indicates quite small uncertainty in
the estimates and, overall, a very accurate experiment.
Now consider an example with N = 100 pieces, an average of only 50 microspheres/piece for each of the two
separate microsphere injections, and true heterogeneity
(CVtrue = 50%) for both paired measurements. In this case, the estimated SE of
true is 1.75%, and
the estimated SE of
true is
0.048. This leads to approximate 95% confidence intervals of
±3.4% for the corrected CV estimate and ±0.094 for the
corrected correlation estimate. Although the variability here is larger
than the previous example, it is still small enough to be useful if
there is a large effect or if a rough estimate is sufficient. Note that
Eqs. 6 and 7 should be used only for planning
experiments and not for estimation of actual standard deviations for
completed experiments.
We can use Eqs. 6 and 7 to come up with a "rule of thumb" for the number of microspheres needed to provide estimates of the CV and correlation that are relatively free of Poisson noise and are reasonably precise. Although it may be somewhat arbitrary, let us consider that a relatively small amount of Poisson noise is acceptable if we are 95% confident that an estimated CV is within ±2% of the true value. These 95% confidence bounds imply a standard error of the CV of ~1% and a variance of (1%)2 = 1.0. Using this value for the variance in Eq. 6 and solving for Total, we get 15,330 or, rounded, 15,000 microspheres needed to provide 95% confidence of good precision of the CV. Note that this estimate is independent of the number of pieces.
We can provide a similar rule of thumb for the correlation coefficient.
Correlations range from
1.0 to +1.0. Let us consider that
Poisson noise is acceptable if we are 95% confident that the estimated
correlation coefficient is within ±0.05 of the true value. These
95% confidence bounds imply a standard error of 0.025 and a variance
of (0.025)2 = 0.000625. We can use
Eq. 7 for a "worst case"
scenario by setting each CV = 10%. This CV would imply an
extraordinarily small level of variability for a biological system, a
level that would rarely be encountered in practice, and thus, a highly
conservative numerical choice. Solving Eq. 7 for Total then yields Total = 24,481, which is
rounded up to an easily remembered value of 25,000. Again, this
estimate is independent of the number of pieces.
| |
DISCUSSION |
|---|
|
|
|---|
The 400 microsphere rule for ±10% precision of flow to a single
region or piece is certainly a useful and important guideline for
investigators for whom that is the single goal. We have shown both
experimentally and by simulation that experiments that use <400
microspheres/piece can yield valid and accurate estimates of
heterogeneity and of correlation after adjustment using
Eqs. 2,
3, and
5. Often, adjustments are so small
that no correction will be needed for heterogeneity and correlation
estimates. This methodology can be extended to include other types of
microsphere analyses as well, such as regression slopes expressing the
gradient in flow versus distance within an organ. When measurements of slope, heterogeneity, or correlation are made for several animals, hypotheses about this sample of statistics can be made using a single
sample t-test comparing the mean
heterogeneity, mean correlation, or mean slope to a specified value,
such as zero. If a statistic, such as
true, is unbiased for
each animal, then the mean for the collection of animals is unbiased.
Naturally, the use of more microspheres per animal will increase the
precision of the statistic within each animal and allow a greater power
to detect effects in the collection of animals. However, if a
statistically significant result is obtained for a mean (e.g., mean CV)
from experiments involving several animals with small numbers of
microspheres, then the hypothesis test, the
P-value, and the confidence intervals are all valid, regardless of the number of microspheres per piece.
The driving force behind precision and bias is the magnitude of Poisson noise in relation to the size of the effect being considered. Thus, if true heterogeneity is large, Poisson noise may be relatively unimportant. If correlations are very strong, they are likely to be quite evident, even without correction. Correction will usually give a more precise estimate of heterogeneity and correlation.
A rule of thumb for the number of microspheres to use in an experiment is 15,000 if an accurate coefficient of variation is of interest and 25,000 if the correlation coefficient is of interest. An accurate correlation coefficient requires more microspheres (almost twice as many) than the number required for the CV, because two sets of microspheres are involved.
Our methodology does not need to take account of reference flows. Reference flow microspheres play no role when heterogeneity and correlation are being estimated using the methods developed in this article or when microspheres are individually counted with histological methods. Reference flows also play no role when the mean flow or mean normalized flow to pieces is adjusted to a fixed value, such as 1.0, that is used for all animals in an experiment. Reference flows will be important, however, when the absolute flow (e.g., ml/min) to a region is of interest. In this case, the methods developed by Buckberg et al. (2) and Nose et al. (9) can be used to achieve a satisfactory level of precision.
In conclusion, measurements of heterogeneity and correlation can be corrected for Poisson noise. The 400 microspheres/piece rule does not apply to many analyses. At least 15,000 microspheres (total) should be used for accurate estimates of heterogeneity and at least 25,000 microspheres for accurate estimates of correlation coefficients.
| |
APPENDIX 1 |
|---|
|
|
|---|
Derivation of Adjusted CV
Let Xi represent the number of microspheres counted in the ith piece for i = 1, ... , N pieces.The observed coefficient of variation is
|
is the
mean of Xi.
|
|
|
|
|
2,
, and
N, solve for
2, and then express
CV2true in terms of
E(CV2obs),
,
2, and
N.
|
|
|
|
|
|
|
|
|
2
and combine them into a new estimator for
CV2. Define
|
|
|
|
|
|
| |
APPENDIX 2 |
|---|
|
|
|---|
Derivation of Adjusted Correlation Coefficient
We have N pieces with two different measurements of flow, based on the number of microspheres in each piece. We are interested in the correlation between flows in the absence of Poisson noise. An estimate of this true correlation coefficient can be derived.Let X and Y be two distributions of microspheres per piece. Xi and Yi represent the number of microspheres in the ith piece for i = 1, ... , N pieces.
|
|
The deviation of piece
i from µ can be broken into three
components,
,
, and
, defined as follows.
xi is the expected common
deviation of piece
i from
µx, in common with
yi, except for a scale factor depending on
the number of microspheres injected.
yi is
the expected common deviation of piece i from
µy, in common with
xi, except for a scale factor depending on
the number of microspheres injected.
xi
is the unique deviation of piece
i from µx, not in common with the
deviation from the Y injection.
yi is the unique deviation of
piece
i from
µy, not in common with the
deviation from the X injection.
xi is the Poisson noise for
piece
i, X
injection.
yi is Poisson noise
for piece
i, Y
injection. Xi is
~Poisson(µx
xi
xi).
Yi is
~Poisson(µy
yi
yi)
|
is the ratio of total microspheres injected
(X
spheres/Y spheres), then
|
|
|
2cx = Var(
xi), the variance of the
common part;
2µx = Var(
xi), the variance of the
unique part; and
2
x = Var(
xi), the variance of the
Poisson noise.
A similar set of definitions holds for
2cy,
2µy, and
2
y. Note that, by
definition of the Poisson distribution,
2
x = µx and
2
y = µy.
The population correlation coefficient is calculated as
|
|
|
|
|
|
|
|
|
|
yi +
yi). The derivation of the
adjusted correlation coefficient leads to Eq. 5 with or without this change, so our adjusted
correlation coefficient can be used regardless of whether the
correlation is positive or negative.
| |
APPENDIX 3 |
|---|
|
|
|---|
Method for Generating Simulated Organs
The fractal model of van Beek (11) was used to generate organ blood flow distributions. In this model, regional blood flows are determined by repetitively subdividing an organ into pieces. After k subdivisions there are 2k organ pieces. At each subdivision, blood flow to a piece is asymmetrically portioned into its two daughter pieces. A fraction of blood (
) is portioned to one
piece and the remaining flow (1
) is portioned to the other
piece. At each subdivision,
is randomly determined from a normal
distribution with mean of 0.5 and a standard deviation of
. The
value of
is constrained to lie between zero and unity. If
= 0, blood flow will be uniformly distributed to all organ pieces. As
deviates from 0, blood flow becomes more heterogeneous. At the smallest
piece size, the fractional amount of blood flow, fi, to each
piece i is determined.
The expected numbers of microspheres depositing in these simulated organs are determined by multiplying the number of microspheres injected into the organ, M, by the fractional flow to each organ piece, fi. Each quantity of the distribution Xtrue,i = M · fi, is rounded to the nearest integer and designated as the "true" blood flow to each organ piece. The true coefficient of variation, CVtrue, is determined for each organ. These true CV values serve as a comparison for the noise-corrected CV values generated with simulated data.
We also simulated observed and corrected correlations. Starting with
Xtrue,i,
the expected number of microspheres deposited in the same piece at
another time (e.g., under another experimental condition) is simulated
by creating a second distribution, Ytrue,i,
with a specified correlation to
Xtrue,i. We consider nonnegative (zero or positive) correlations and negative correlations separately. For nonnegative values of the specified correlation r, the second distribution
is generated with the following equation
|
(8) |
i is a random normal number
from a distribution with a mean of zero and a variance of 1.0. The
constants K1,
K2, and
K3 are chosen by
the user to yield the desired correlation between
Xtrue and
Ytrue while
keeping the total flow of Y equal to
the total flow of X. We also attempt
to keep the true CV of Y close to the
true CV of X. By chance, negative
values of Yi may
occur; these numbers are discarded and a new
i is redrawn. Because of the
random generation of
Ytrue,i, the actual correlation between
Xtrue and
Ytrue may differ
slightly from the desired correlation
r. The true correlation,
rtrue, is determined from the generated
Xtrue,i
and
Ytrue,i, rather than from the target value, r.
If needed, the process is repeated until
rtrue is
sufficiently close to the desired r,
and the final set of values
Xtrue,i
and
Ytrue,i is designated as the true organ observed under two conditions.
When the desired r is negative, we use
the method described above to generate
Ytrue using the
absolute value of r as the target correlation. The maximum and minimum values of
Yi are then
determined. Each
Yi is replaced by
the value of (maximum + minimum
Yi), and then this new
Ytrue is
normalized so that the total flow of Y
is equal to the total flow of
X. As with the
nonnegative correlations, the process is repeated until
rtrue is
sufficiently close to the desired negative
r, yielding a final set of
Xtrue,i
and Ytrue,i,
which comprise a true organ observed under two conditions.
Observed microsphere distributions,
Xobs,i,
are generated by adding Poisson noise to the true number of
microspheres in each organ piece. Poisson noise is estimated by
choosing a random number (m) from a
uniform distribution between 0 and 1 and then determining the largest
value of ki such
that
|
(9) |
is used to
estimate
Xobs,i
for
Xtrue,i > 100. For the paired simulated true flow distribution,
Ytrue,i, the observed microsphere distribution,
Yobs,i,
is generated in the same manner as the
Xobs,i,
using Eq. 9 or the normal distribution.
| |
FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. §1734 solely to indicate this fact.
Address for reprint requests and other correspondence: N. L. Polissar, 1827 23rd Ave. East, Seattle, WA 98112 (E-mail: polissar{at}biostat.washington.edu).
Received 30 November 1998; accepted in final form 25 June 1999.
| |
REFERENCES |
|---|
|
|
|---|
1.
Bassingthwaighte, J. B.,
R. B. King,
and
S. A. Roger.
Fractal nature of regional myocardial blood flow heterogeneity.
Circ. Res.
65:
578-590,
1989
2.
Buckberg, G. D.,
J. C. Luck,
D. B. Payne,
J. I. E. Hoffman,
J. P. Archie,
and
D. E. Fixler.
Some sources of error in measuring regional blood flow with radioactive microspheres.
J. Appl. Physiol.
31:
598-604,
1971
3.
Glenny, R. W.,
S. Bernard,
C. Barlow,
J. Kelly,
and
H. T. Robertson.
Spatial distribution of pulmonary blood flow at a microscopic scale of resolution (Abstract).
Am. J. Respir. Crit. Care Med.
151:
A518,
1995.
4.
Glenny, R. W.,
and
H. T. Robertson.
Fractal modeling of pulmonary blood flow heterogeneity.
J. Appl. Physiol.
70:
1024-1030,
1991
5.
Iversen, P. O.,
and
G. Nicolaysen.
Heterogeneous blood flow distribution within single skeletal muscles in the rabbit: role of vasomotion, sympathetic nerve activity, and effect of vasodilation.
Acta Physiol. Scand.
137:
125-133,
1989[Web of Science][Medline].
6.
Iversen, P. O.,
and
G. Nicolaysen.
Fractals describe blood flow heterogeneity within skeletal muscle and within myocardium.
Am. J. Physiol. Heart Circ. Physiol.
268:
H112-H116,
1995
7.
Iversen, P. O.,
M. Standa,
and
G. Nicolaysen.
Marked regional heterogeneity in blood flow within a single skeletal muscle at rest and during exercise hyperaemia in rabbit.
Acta Physiol. Scand.
136:
17-28,
1989[Web of Science][Medline].
8.
Luchtel, D. L.,
J. C. Boykin,
S. L. Bernard,
and
R. W. Glenny.
Histologic methods to determine blood flow distribution with fluorescent microspheres.
Biotech. Histochem.
73:
291-309,
1998[Web of Science][Medline].
9.
Nose, Y.,
T. Nakamura,
and
M. Nakamura.
The microsphere method facilitates statistical assessment of regional blood flow.
Basic Res. Cardiol.
80:
417-429,
1985[Web of Science][Medline].
10.
Rudolph, A. M.,
and
M. A. Heymann.
The circulation of the fetus in utero. Methods for studying distribution of blood flow, cardiac output and organ blood flow.
Circ. Res.
21:
163-184,
1967
11.
Van Beek, J. H. G. M.,
S. A. Roger,
and
J. B. Bassingthwaighte.
Regional myocardial flow heterogeneity explained with fractal networks.
Am. J. Physiol. Heart Circ. Physiol.
257:
H1670-H1680,
1989
12.
Zar, J. H.
Biostatistical Analysis. Englewood Cliffs, NJ: Prentice-Hall, 1984.
This article has been cited by other articles:
![]() |
M. Marxen, J. G. Sled, L. X. Yu, C. Paget, and R. M. Henkelman Comparing microsphere deposition and flow modeling in 3D vascular trees Am J Physiol Heart Circ Physiol, November 1, 2006; 291(5): H2136 - H2141. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. J. E. Lamm, S. L. Bernard, W. W. Wagner Jr., and R. W. Glenny Intravital microscopic observations of 15-{micro}m microspheres lodging in the pulmonary microcirculation J Appl Physiol, June 1, 2005; 98(6): 2242 - 2248. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. K. M. Decking, V. M. Pai, E. Bennett, J. L. Taylor, C. D. Fingas, K. Zanger, H. Wen, and R. S. Balaban High-resolution imaging reveals a limit in spatial resolution of blood flow measurements by microspheres Am J Physiol Heart Circ Physiol, September 1, 2004; 287(3): H1132 - H1140. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. C. Marshall, P. Powers-Risius, B. W. Reutter, A. M. Schustz, C. Kuo, M. K. Huesman, and R. H. Huesman Flow heterogeneity following global no-flow ischemia in isolated rabbit heart Am J Physiol Heart Circ Physiol, February 1, 2003; 284(2): H654 - H667. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Matsumoto, H. Tachibana, Y. Ogasawara, and F. Kajiya New double-tracer digital radiography for analysis of spatial and temporal myocardial flow heterogeneity Am J Physiol Heart Circ Physiol, January 1, 2001; 280(1): H465 - H474. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. L. Bernard, J. R. Ewen, C. H. Barlow, J. J. Kelly, S. McKinney, D. A. Frazer, and R. W. Glenny High spatial resolution measurements of organ blood flow in small laboratory animals Am J Physiol Heart Circ Physiol, November 1, 2000; 279(5): H2043 - H2052. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. W. Glenny, S. L. Bernard, and W. J. Lamm Hemodynamic effects of 15-{micro}m-diameter microspheres on the rat pulmonary circulation J Appl Physiol, August 1, 2000; 89(2): 499 - 504. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. W. Glenny, S. L. Bernard, and H. T. Robertson Pulmonary blood flow remains fractal down to the level of gas exchange J Appl Physiol, August 1, 2000; 89(2): 742 - 748. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Gerbino, S. McKinney, and R. W. Glenny Correlation between ventilation and perfusion determines VA/Q heterogeneity in endotoxemia J Appl Physiol, June 1, 2000; 88(6): 1933 - 1942. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Sinclair, S. McKinney, R. W. Glenny, S. L. Bernard, and M. P. Hlastala Exercise alters fractal dimension and spatial correlation of pulmonary blood flow in the horse J Appl Physiol, June 1, 2000; 88(6): 2269 - 2278. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. A. Altemeier, S. McKinney, and R. W. Glenny Fractal nature of regional ventilation distribution J Appl Physiol, May 1, 2000; 88(5): 1551 - 1557. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |