Vol. 279, Issue 3, H1291-H1298, September 2000
New optical method for analyzing cortical blood
flow heterogeneity in small animals: validation of the method
Istvan
Schiszler,
Minoru
Tomita,
Yasuo
Fukuuchi,
Norio
Tanahashi, and
Koji
Inoue
Department of Neurology, School of Medicine, Keio University,
Tokyo-160, Japan
 |
ABSTRACT |
In
pentobarbital-anesthetized male Sprague-Dawley rats, a small
cranial window was trephined, and the cortex was transilluminated with
a fine glass fiber inserted into the brain parenchyma. The light
intensity at the surface area of 2 × 2 mm was recorded during intracarotid injection of 25 µl of carbon black (CB) solution. The
region of interest (ROI) was divided into a 50 × 50 matrix, and the
mean transit time of CB transport was calculated in each matrix
element. We found rapid transits of CB along the microvasculature, with
considerable heterogeneity in the avascular area, and heterogeneous efficiency in autoregulatory capacity in the ROI during hypotension. The method was validated by comparison with laser-Doppler flowmetry. The average mean difference was 0.03 ± 0.05%. Five percent
CO2 inhalation increased the flow by 85%, but
heterogeneously. We concluded that the technique is exclusively
sensitive to indicator transits in a very small area on the brain
surface with potential usefulness in detecting regional heterogeneity
in blood flow.
cerebral blood flow autoregulation; mean transit time; carbon
dioxide reactivity
 |
INTRODUCTION |
THE ORIGINAL PHOTOELECTRIC
METHOD developed by Tomita et al. (15) permitted
repeated measurements of the cortical mean transit time (MTT) by use of
a nondiffusible intravascular tracer, which necessitated minor
assumptions. The observed region was, however, relatively large
(3 × 5 mm); flow values in arteries, capillaries, and veins were
averaged all together. For a modification of the original technique, we
attempted to combine it with the two-dimensional (2-D) method described
by Eke (3). The main difference between the method of Eke
and the present one used by us is the method of illumination; namely,
Eke used reflected light, whereas our method employed transmitted
light. This difference is important in that it improves spatial
resolution and quantification of dye concentration and is free from
contamination by artifactual aberrant surface-light reflection.
The most significant problem in any system using optical signals for
monitoring cerebral blood flow or neural function is light scattering
by the brain parenchyma. Light scattering causes the optical path
length between the light source and the detector to become longer than
the physical distance. This difference was estimated by use of the
Monte Carlo simulation and was determined in vivo by time-of-flight
measurements by Delpy et al. (2). Their results indicated
that the optical path length was ~4.5 times longer than the
interoptode spacing. This ratio is known as the differential
path-length factor (DPF), which is a function of the tissue absorption
coefficient, the tissue scattering coefficient, and the tissue
geometry. The demarcation of the cortical MTT requires one to measure
the relative dye concentrations in the brain parenchyma. This is
possible only if the DPF is constant during dye transport. As mentioned
above, the DPF changes, albeit slightly, simultaneously with
changes in the absorption coefficient. In this paper, we will examine
whether the relative dye concentration can be measured with the use of
the Lambert-Beer approach. If the answer is in the affirmative, the
microregional MTT values can be calculated, and the 2-D map of the
reciprocal MTT values will represent the microregional flow
distribution, because the blood volume in infinitesimally small
cortical areas is relatively constant. The new evidence emphasizes the
importance of 2-D cortical blood flow (CBF) measurements: studies using
intravital microscopy have shown spontaneous fluctuations in red blood
cell (RBC) velocity in capillaries (7), which cannot be
explained by a simple "stopcock" theory (10).
Functional activation was found to decrease the heterogeneity of
capillary plasma perfusion in the rat barrel cortex (17).
Hudetz et al. (6), also using intravital
microscopy, reported the maintenance of capillary RBC flow velocity
during hypotension. The significance of 2-D flow mapping was pointed
out by Heimann et al. (5), who employed the scanning
laser-Doppler flow (LDF) method. Intravital microscopy and confocal
laser-scanning microscopy are very versatile tools for analyzing
individual capillaries but are inadequate if one chooses to observe the
interrelation of microflows or flow heterogeneity in a wider area of
the cerebral cortex. The technique described herein may solve this problem.
 |
MATERIALS AND METHODS |
Preparations.
Experiments were carried out on male Sprague-Dawley rats
(n = 20) weighing 300-350 g. The animals were
anesthetized by intraperitoneal administration of 50 mg/kg of
pentobarbital. Additional doses of the anesthetic were used
intravenously if needed, according to tail pinch and cornea reflex.
During the experiment, the rectal temperature was kept constant by use
of a thermostat-controlled heated holder. Both femoral arteries and one
femoral vein were catheterized [polyethylene (PE)-50 catheter] to
measure blood pressure, blood withdrawal, and drug administration. The
most critical procedure was the insertion of a catheter into the
carotid artery for the dye injection without disturbing the internal
carotid flow. Under additional local anesthesia (lidocaine), a midline skin incision was made at the neck. A tracheal cannula was inserted first to secure respiration. Taking care to avoid excessive bleeding and nerve injuries, we moved the muscle layers to the sides and carefully explored the branches of the external carotid artery. Approximately 1.5 cm from the bifurcation, an incision to one-third of
the diameter of the external carotid artery was made, and a PE-10
catheter filled with heparinized saline was led proximally into the
carotid artery until its tip reached the bifurcation. The external
carotid artery was tied to the catheter just above the bifurcation to
avoid dye escaping to the external carotid system. The skin was sutured
to avoid fluid evaporation. The animals were placed on a head holder
with ear bars. A 5-mm-diameter cranial window was trephined above the
right temporoparietal cortex, leaving the dura intact. To
transilluminate the region of interest (ROI), we inserted a
200-µm-outer diameter (OD) glass-fiber light source from a
xenon lamp into the brain parenchyma through a hole drilled 5 mm to the
back of the cranial window. An ultraviolet (UV) filter was employed to
reduce UV light. The glass fiber was fixed to the edge of the skull
with dental cement, so that the light-emitting tip was located beneath
the gray matter of the ROI. A laser-Doppler probe (Advance laser
flowmeter, 21R, 1-mm probe) was applied above the thinned bone on the
contralateral side of the skull. During the drilling process, the skull
was cooled with a continuous superfusion of saline. We employed a
silicon intensifier target (SIT) camera with a Nikon lens to
monitor the transmitted light intensity focused on the pertinent brain
surface. When we used saline to measure the MTT, we employed a green
filter between the lens and the detector (
= 548 ± 15 nm,
one of the isosbestic points of hemoglobin). The camera was linked to
an image processor (Argus 10, Hamamatsu Photonics), a personal computer
(NEC), a video cassette recorder (ED Beta, Sony), and a Macintosh
computer through an 8-bit frame-grabber card (Scion LG-3).
Experiments.
In the first group (n = 6), rats were used for basic
experiments. In three rats, concentrated carbon black (CB) dye was
injected intravenously, and the light intensity was measured
continuously in the cranial window by a silicon photodiode (SPD). Blood
samples (0.2 ml) were repeatedly taken from the arterial catheter and were anticoagulated with 0.05 ml of 3.1% citrate. The RBCs were hemolyzed to reduce light scattering by addition of 0.5 ml of distilled
water. The light extinction of the samples was measured in a
transilluminated glass chamber with the use of the same SPD. The in
vivo and in vitro CB concentrations were calculated according to the
method described in Data analysis. These were not absolute concentrations, however; the maximum was defined as 1, and the relative
values were calculated. The corresponding in vivo and in vitro data
were compared by use of linear regression analysis. The neurological
consequences of the surgical operation and brain damage were estimated
in three other rats. Short-cut glass fibers were implanted into the
brains of the animals, and, after they regained consciousness, their
behaviors were monitored for 8 h.
In the second group (n = 8), the experimental protocol
started 60 min after the completion of the preparations. During the experiments, the LDF and the arterial blood pressure (BP) were recorded
continuously with the use of a laser flowmeter (Advance, 21R) with a
1-mm-diameter flow probe, a pressure transducer (Nihon Koden, TP,
400T), an analog-to-digital (A/D) board (Biopac, MP-100), and a
Power Macintosh. To determine the regional MTT, 25 µl of diluted
(1:30) CB dye or saline were injected into the internal carotid artery
through the external carotid artery. To validate the method, the CBF
was subjected to change, by hemorrhagic hypotension below the
autoregulatory range, and reinfusion (n = 8). The CB injections were performed during the control experiment during exsanguination and immediately after reinfusion. Arterial blood gases
were monitored during the experiment.
In the third group (n = 6), the effect of 5%
CO2-95% O2 gas inhalation was analyzed by use
of only the new technique. In this group, we replaced the CB with
saline, a negative indicator. CB represents plasma transit, whereas
saline indicated the transit of the RBC (a negative indicator). The
measurements were performed before CO2 inhalation and at 2 min after the commencement of CO2 initiation.
Data analysis.
To analyze dye-dilution curves, we examined whether the relationship
between the dye concentration and the output amplitude of the SIT
camera coupled to the A/D board adhered to the principle of
superposition. We filled a glass chamber with saline containing various
concentrations of CB and measured the light intensity on the surface,
transilluminating it with the same light source. When the recorded
outputs were plotted against the corresponding concentrations in a
semilogarithmic chart, the relationship was found to be linear. The
linearity in blood as a scattering medium has been previously tested in
cats by Tomita et al. (15). To determine the regional MTT,
200 consecutive frames of CB dye transport through a 2 × 2-mm ROI
were captured by the Power Macintosh at a rate of 15 frames/s. The
captured images were stored in a magnetooptical disk. Sequential frame
analysis was performed by MATLAB by employing a Gateway 2000 desktop
computer. The ROI was initially divided into 250 × 250 matrices.
However, we found that the unit pixels exhibited a low signal-to-noise
ratio and inevitably required averaging. Therefore, we averaged each
element of the matrix, which represented either a 5 × 5-pixel
area (20 × 20 µm) of the ROI (ROIH) or a 10 × 10-pixel area (~40 × 40 µm; Fig.
1) of the ROI (ROIL). For the
shorter calculation time, 2-D flow maps were calculated on the basis of
the latter (ROIL) system, yielding 2,500 dilution curves.
One of the merits of this method was that the flow maps could be
recalculated for finer spatial resolution from the stored data. The
pixel resolution, however, is not the same as the actual resolution
because of the effects of sideward light scatter.

View larger version (190K):
[in this window]
[in a new window]
|
Fig. 1.
The first step of the analysis. The region of interest
(ROI) is divided into a matrix, with each of its elements representing
a 10 × 10-pixel area of the video image or the equivalent of a
40 × 40-µm area of the ROI. The video pixels were averaged to
reduce noise level.
|
|
According to the modified Lambert-Beer equation (Ref.
2), the attenuation (
) can be expressed as
|
(1)
|
where I is the intensity of the transmitted light,
I0 is the intensity of the input light, DP is the
differential path length (
/
µ
),µ
is the
tissue absorption coefficient, and G is an unknown
geometry-dependent factor. With the assumption that DP is constant
during dye transport through the ROI (validated by Fig. 3), the MTT can
be calculated from the individual dilution curves by use of numerical
integration, according to Eq. 2
|
(2)
|
where ct is the dye concentration,
µ
t is the change in tissue absorption coefficient,
and It is the transmitted light intensity at
time t; Icontrol is the transmitted light
intensity at time 0. The baseline of the dilution curve was
set as (initial c) = 0, and any baseline shift in the dilution curves was corrected by use of linear interpolation. The reciprocal MTT
values were displayed on a 2-D map.
To validate the method, comparisons of CBF values were made with those
measured by the LDF method. Because neither our method nor the LDF
yielded absolute flow values, we correlated the relative values to CBF
(CBFrel), which were calculated as
|
(3)
|
where CBFmeasured and CBFcontrol are
measured and control values, respectively.
Evaluation of eight experiments yielded 47 relative LDF value
(LDFrel) versus 1/relative MTT value (MTTrel)
pairs. These pairs were analyzed statistically by use of linear
regression and correlation analysis and by calculation of the limits of
agreement (1), which are defined as follows: limits of
agreement = mean difference
2 standard deviations (SD).
 |
RESULTS |
General observations.
Small arterioles and veins with diameters ranging from 300-20 µm
were well focused with clear contours, suggesting that the light
transmitted through the complicated deep-tissue structures illuminated
homogeneously only the surface layer, where the vessels were located.
Figure 2 shows typical 2-D flow maps. It
should be noted that the picture was quite similar to the photograph shown in Fig. 1, although the 2-D flow map was constructed from an
entirely different element. The rats that had received implantation of
short-cut glass fibers, as described in MATERIALS AND
METHODS, awakened within 1.5 h, ultimately becoming alert
and attentive and responsive to noise. Two hours later, they recovered
completely, showing no signs of neural disturbance. The insertion of
the glass fiber seemed to have no affect on their behavior.

View larger version (102K):
[in this window]
[in a new window]
|
Fig. 2.
A and B: typical series of
two-dimensional (2-D) maps of the reciprocal mean transit time (MTT)
values in the ROI at different perfusion pressure levels and after
reinfusion. The maps were generated with the use of interpolated colors
between the neighboring pixels. The units on both axes represent 40 µm. The arrows indicate pial arterioles. The cerebral blood volume
(CBV) did not change during the experiments, so that the reciprocal MTT
represented the relative cortical blood flow (CBF). A: flow
was maintained at a 75-mmHg perfusion pressure, and arteriolar dilation
was noticeable. B: flow was already deteriorated in some
parts of the ROI but was maintained in other parts. In both cases, the
flow decreased at 50 mmHg and showed a marked increase after
reinfusion. Note the heterogeneous flow during the reinfusion phase
during the experiment. C: 2-D maps during control and at
10 s and 2 min of 5% CO2 inhalation. The map
generated at 10 s reveals a heterogeneous relative flow increase
along the vessels. The relative flow further increased and became
homogeneous after 2 min of inhalation. Note the intensified mixing in
the vein indicated by the disappearance of the dark central area.
|
|
In vivo tissue and in vitro blood dye concentration.
We observed a linear relationship between the in vitro and in vivo
measured concentrations. Figure 3 shows
the scatter plot of the corresponding values and the best-fitting
line (y = 0.93x + 0.12, r2 = 0.92, P < 001).

View larger version (24K):
[in this window]
[in a new window]
|
Fig. 3.
In three rats, concentrated carbon black (CB) dye was
injected in a stepwise manner intravenously. Tissue and blood CB
concentrations were calculated according to the routine described in
MATERIALS AND METHODS. A linear relationship was found
between the in vitro and in vivo measured CB concentrations. The
scatter plot of the corresponding values and the best-fitting line are
shown (y = 0.93x + 0.12, r2 = 0.92, P < 0.001).
Both axes represent relative dye concentrations.
|
|
Exsanguination and reinfusion.
The arterial blood gas values were within the normal range during the
experiments: arterial partial pressure of O2
(PaO2), 97 ± 7 mmHg (means ± SD); arterial
partial pressure of CO2 (PaCO2), 39.7 ± 1.7; and pH, 7.37 ± 0.04. In the control group, the average mean arterial blood pressure was 109 ± 11 mmHg. The average MTT of CB in the total ROI was 2.11 ± 0.67 s in the control;
increased to 2.61 ± 0.79 and 3.26 ± 1.18 s at mean
blood pressures of 75 and 50 mmHg, respectively; and decreased to
1.67 ± 0.28 s after reinfusion. The respective values of LDF
were 21.4, 17.4, 15.9, and 27.9. Neither the transmitted-light
intensity nor the mass signal of the laser-Doppler flowmetry changed
throughout the experiment. Figure 4 shows
the relative changes in the measured parameters. Figure 2, A
and B, shows the obtained 2-D flow maps in the ROI in two
control experiments at BPs of 75 and 50 mmHg and after reinfusion. In
Fig. 2A, two pial arterioles are visible in the flow maps,
descending from the top. They appear to be dilated at a BP
of 75 mmHg, as the two yellow bands are wider. During the reperfusion
phase, the arterioles were still dilated, with the bright color
indicating reactive hyperemia. Between the two arterioles, a pial vein
was clearly discernible, especially during hypotension (BP 50 mmHg),
when the flow decreased in the vessel, and during the reinfusion phase,
when the reactive hyperemia was visible in the vein as well. The tissue
perfusion remained constant when the BP decreased to 75 mmHg but
decreased markedly in the tissue flow at a BP of 50 mmHg. During the
reinfusion phase, an increase in the tissue perfusion was visible, but
its degree was quite heterogeneous throughout the ROI. In Fig.
2B, two tortuous arterioles are seen emerging from the
left (arrows) as a pial vein exits the ROI on the
right. In this case, the tissue perfusion became quite
heterogeneous at a BP of 75 mmHg: some regions showed good
autoregulation, maintaining the regional blood flow, but the perfusion
decreased in other areas of the ROI. The tissue perfusion remained
heterogeneous even during the reinfusion phase. It should be noted that
the "laminar" flow was visualized in the vein by use of this
technique during reinfusion as two distinct layers of
different flow velocities. In Fig.
5, the regional differences in
the CBF autoregulation are more visible. Figure 5 was created by
calculation of the relative changes in the local CBF values. The light
tones represent those areas of the ROI where the CBFrel remained unchanged at a perfusion pressure of 75 mmHg, during which time the flow decreased in the dark regions. Figure
6 shows representative arterial,
venous, and tissue dilution curves obtained during the experiment. It
should be noted that the distance between the peaks of the arterial and
venous curves correlated well with the measured CBF. The analysis of
MTT values in avascular areas, arteries, and veins from four
experiments is presented in Fig. 7.

View larger version (33K):
[in this window]
[in a new window]
|
Fig. 4.
Relative changes (means ± SE, calculated as
measured value/control) in the laser-Doppler flow (LDF), 1/MTT, and
"red blood cell mass" values during exsanguination at 75 and 50 mmHg and after reinfusion. The CBV was derived from the light
intensity, which did not change during the exsanguination-reinfusion
procedure.
|
|

View larger version (103K):
[in this window]
[in a new window]
|
Fig. 5.
Relative changes in the CBF during hypotension (75 mmHg)
compared with the control state. The light tones represent
well-autoregulating areas, and the dark grades indicate
nonautoregulating areas. The relative change was calculated in each
pixel as control MTT value/MTT value at 75 mmHg.
|
|

View larger version (29K):
[in this window]
[in a new window]
|
Fig. 6.
Representative dilution curves from arteries, avascular
areas, and veins. The curves represent the light intensity in the given
pixels as a function of time. The curves were "well behaved," with
a very low noise level even when they were obtained from a small area
of 40 × 40 µm2. The y-axis indicates the
measured light intensity in arbitrary units, and the x-axis
is the time in seconds. A: control. B:
hypotension, 75 mmHg. C: hypotension, 50 mmHg. D:
reinfusion.
|
|

View larger version (40K):
[in this window]
[in a new window]
|
Fig. 7.
MTT values measured in arteries, avascular areas
(tissue), and veins. The error bars indicate SE (n = 4).
|
|
CO2 inhalation.
In this group (n = 6), the control MTT ranged from 0.98 to 1.49 s, with an average of 1.41 ± 0.14 s (mean ± SD). At 2 min of CO2 inhalation, the MTT decreased to
0.76 + 0.22 s. This indicates an 85% increase in the CBF.
Figure 2C shows flow maps from three consecutive states of
CO2 inhalation. The first map represents the control state.
In the central area of the map, a big cortical vein is visible, and the
flow in the surrounding tissue is relatively homogeneous. The second
map reflects the flow alteration 10 s after the initiation of the
CO2 inhalation. The tissue flow became very heterogeneous
during this phase, increasing initially in the area surrounding the
feeding arteries. The map reveals different flow layers in the pial
vein. The third map shows the highly increased CBF after 2 min of
CO2 inhalation. During this phase, the pial arteriole,
which was not visible in the previous maps, is clearly discernible on
the right side of the map. The mixing in the vein intensified, as
indicated by the disappearance of the dark central area.
Comparison between laser-Doppler flowmetry and the new optical
method.
During the exsanguination-reinfusion procedure, we performed 55 MTT
measurements in eight rats. In each case, except for the controls, we calculated the relative change as
CBFmeasured/CBFcontrol. The results
were paired with the corresponding
LDFmeasured/LDFcontrol values. The
statistical analysis yielded a significant correlation between the LDF
and the optical flow measurement (Pearson coefficient = 0.847, P < 0.001). Figure 8
shows the scatter plot of the data set and the best-fitting curve,
obtained by linear regression analysis (y = 0.93x + 0.05). The distribution of the calculated differences (CBFrel
LDFrel) was
virtually normal, and the average mean difference was 0.03 ± 0.05% (means ± 95% confidence interval), indicating a 3%
overestimation of the CBF alteration as measured by the optical
technique compared with the LDF. An analysis of the discrepancy in the
function of relative CBF change revealed a significant divergence
between the two methods when the measured CBF was below 80% of the
control (Fig. 9). In this case, the mean difference was 14% (P < 0.001); our method resulted
in lower CBF values. When the CBF decrease was <20% and when it
increased during the reinfusion phase, the differences were not
significant. The limits of agreement were
27 and 21%.

View larger version (16K):
[in this window]
[in a new window]
|
Fig. 8.
Linear regression analysis between the corresponding
relative (r) flow values, calculated from the reciprocal MTT and the
LDF. The best curve was described by y = 0.91x + 0.12 (r2 = 0.84, P < 0.001).
|
|

View larger version (22K):
[in this window]
[in a new window]
|
Fig. 9.
Mean difference ± SE between the two techniques at
various relative flow values. The discrepancy was significant when the
flow was <80% of the control (P < 0.001, n = 8), with our method measuring lower CBF values.
|
|
 |
DISCUSSION |
Tissue damage.
One possible criticism of our method might be that it is invasive. The
present method requires insertion of a glass fiber of 200-µm OD into
brain tissue, which inevitably causes brain tissue damage. The diameter
is less than that of a hydrogen electrode used for CBF measurements or
as a probe during microdialysis and is almost equivalent to that in
electrodes used for measuring deep action potentials, a technique also
requiring insertion of the instrument into brain tissue. However, it
should be noted that the latter three methods measure the flow at the
sensitive surface of the probes, i.e., the most damaged portion facing
the sensor. On the other hand, our method measures the undamaged layer of the cortex sandwiched between the light source and the detector. One
other factor that must be taken into consideration is the spreading
depression (SDE). We confirmed that SDE after insertion of a microglass
pipette spread in a wavering manner at a speed of 1.6 mm/min, and the
cortex was stabilized within 30 min (unpublished observation).
Therefore, measurements performed 60 min after the insertion were
considered free from the artifacts generated by SDE waves. It is also
known that the SDE never spread to the contralateral side where the
laser-Doppler probe was employed. As confirmed by the results, paired
values of CBF obtained by the laser-Doppler technique and the present
optical method correlated well, ruling out the influence of SDEs on the
measurements. Autoregulation was partially maintained, and vascular
responses to CO2 were comparable with those reported in the
literature. Despite the damage, it is concluded that the present method
can measure CBF within the limits of these experimental conditions.
Light scattering.
Because the tissue consists of numerous particles and structures in the
deep layer, transmitted light from the glass fiber is subjected to
light-pass change by reflection and refraction and, therefore,
scattering. Tomita et al. (16) noticed a complete mixing
of transmitted light beams when they characterized the three-dimensional light scattering in blood in a thin transparent tube.
Even though a parallel beam was projected through a specific point on
the tube, the light emerged from the surface of the blood at all angles
from the injected plane (detected by a small photodetector sliding on
the surface). The blood in the tube was as luminous as a glow lamp,
resulting in scattered light throughout the cerebral cortex on the
brain surface.
In vivo tissue and in vitro dye concentration.
We investigated whether the intravascular dye concentration could be
determined by use of the Lambert-Beer approach. Because we believed
that the brain CB filling was a function of the arterial CB
concentration, we compared the in vivo data, which were measured in the
highly scattering medium, with those obtained in vitro from the
arterial blood samples. To reduce light scattering, we decided to
hemolyze the RBCs. Because plasma has an even lower scattering
coefficient, however, the repeated blood samplings might have altered
the systemic hematocrit, and thus we chose to measure the total light
absorption of the blood. The linear relationship shown in Fig. 3
confirms the validity of the calculated MTT values in the brain tissue.
However, the in vivo measured concentrations were higher than the
corresponding in vitro values, with the best-fitting curve intercepting
the y-axis at 0.12. This might be the result of higher
scattering losses in tissue and CB pooling in the veins or CB adhesion
on the endothelial surface. Cerebral vasodilatation due to slight
exsanguination could have contributed to the discrepancy.
Autoregulation.
We found that the mean CBF was rather maintained when compared with
decreases in the mean BP. However, we found heterogeneous distribution
of areas well autoregulated and nonautoregulated, as shown in Fig. 5.
This visualization of autoregulation in such a small area would be the
first occurrence in the literature, although Leninger-Follert and
Lübbers (9) reported heterogeneous perfusion in
capillaries, according to the metabolic requirement. Our observation
that the cerebral blood volume (CBV) did not change during the
exsanguination-reinfusion procedure is in accordance with the data
reported in the literature (12, 14). Hudetz et al. (6) reported that the flow autoregulation in the
cerebral microcirculation was effectuated through the preservation of
RBC capillary transit time instead of the recruitment of new
capillaries. It must be kept in mind that CB is a plasma indicator;
thus the maintenance of the RBC velocity could remain unrecognized if
the plasma flow decreased selectively. This could explain the slightly higher discrepancy between the two methods when the CBF was <80% of
the control value (Fig. 9). This point should be studied in the future.
The present method may provide a versatile tool for that purpose.
CO2 effects.
For the analysis of the CO2 inhalation, we used saline as a
nondiffusible tracer. The blood-brain barrier is relatively impermeable to sodium under physiological conditions. The difference between saline
and CB, as we emphasized in MATERIALS AND METHODS,
is that saline is an RBC indicator and CB is a plasma indicator. Thus we can expect lower MTT values measured with saline as a tracer, because RBC moves at a more rapid rate in the microcirculation (13). Our results are consistent with this, as the average
MTT was 1.41 ± 0.06 s with saline as a tracer and 2.11 ± 0.24 s with CB as a tracer, indicating that the RBC rate was
~50% faster than that of plasma. The systemic inhalation of
CO2 caused an 85% increase in the CBF. The flow maps
presented in Fig. 2C show that the flow alteration was
diffuse throughout the ROI but was heterogeneous in the early phase.
The average rate of increase of microflow was again in accordance with
previous reports (8, 14).
Comparison with an accepted method.
A comparison of our technique with laser-Doppler flowmetry reveals a
very low mean difference, but with relatively wide limits of agreement.
This may be due to practical and theoretical dissimilarities between
the two techniques: the CB is a plasma dye, and its transport refers to
plasma flow; meanwhile, the LDF measures RBC velocity. Any shift in the
plasma flow-to-RBC flow ratio can cause a bias in the comparison. The
surface areas over which the measurements were performed were
different: at the optical method, the ROI was 2 × 2 mm; on the
other hand, we employed a 1-mm-diameter LDF probe. MTT values from
larger veins and arteries contributed to the optical CBF measurement;
however, the LDF probe was positioned above the "avascular" area. A
more important difference is the contribution of deeper cortical layers
to the measurement. Because the light source was positioned below the
cortex in the case of the optical technique, the signal provided
information about flow in layers IV-VI. The LDF,
nevertheless, focuses on the top layers. Taking into account these
factors, we think that results of the comparison are convincing enough
to validate our method.
In addition to the valid flow values, we were able to detect very
important information about the flow heterogeneity, which has been
smeared out by averaging with a wide sensitive area. Thus the biggest
advantage of our method is in its ability to produce 2-D flow maps in a
small area (2 × 2 mm2). Because the instant flow map
consists of a 50 × 50 square matrix, its spatial resolution is 40 µm/pixel. As stated in MATERIALS AND METHODS, we averaged
100 video points during the data analysis to increase the
signal-to-noise ratio; however, reduction of the number to 25 would
still have yielded a satisfactory signal level. Thus, if necessary, the
spatial resolution can be increased up to 20 µm/pixel although
the actual resolution becomes lower than the pixel resolution
because of sideward light scatter. We were able to observe acinous
spots in the 2-D flow maps (Fig. 2), which could constitute functional
microcirculatory units. This correlates well with the functional
architecture of the cerebral cortex, because the cortical barrels in
rats reportedly have a diameter of ~160 µm (4).
However, the diameters of the units appear to be much smaller with our method.
In conclusion, our new technique enables one to visualize microflow in
the cerebral cortex of small animals with high spatial resolution. The
measurement is not continuous, but it can be repeated frequently, with
~15-s intervals. The measured changes during hypotension or
CO2 inhalation were comparable with those in the reported
data, and the relative flow values correlated well with those measured
by use of laser-Doppler flowmetry. Our new method may provide a
powerful tool for studies of regulatory changes in cerebral
microcirculation, for example, functional or SDE-induced hyperemia.
 |
ACKNOWLEDGEMENTS |
I. Schiszler is a Japan Society for the Promotion of Science fellow
from the Department of Physiology, Semmelweis University, Budapest.
 |
FOOTNOTES |
Address for reprint requests and other correspondence: I. Schiszler, Dept. of Neurology, School of Medicine, Keio Univ.,
35-Shinanomachi, Shinjuku-ku, Tokyo-160, Japan.
The costs of publication of this
article were defrayed in part by the
payment of page charges. The article
must therefore be hereby marked
"advertisement"
in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 12 October 1999; accepted in final form 13 March 2000.
 |
REFERENCES |
1.
Bland, J,
and
Altman DG.
Statistical methods for assessing agreement between two methods of clinical measurement.
Lancet
1:
307-310,
1986[Web of Science][Medline].
2.
Delpy, DT,
Cope M,
Zee P,
Arridge S,
Wray S,
and
Wyatt J.
Estimation of optical pathlength through tissue from direct time of flight measurement.
Phys Med Biol
33:
1433-1442,
1988[Web of Science][Medline].
3.
Eke, A.
Reflectometric mapping of microregional blood flow and blood volume in the brain cortex.
J Cereb Blood Flow Metab
2:
41-53,
1982[Web of Science][Medline].
4.
Elston, GN,
Pow DV,
and
Calford MB.
Neuronal composition and morphology in layer IV of two cancer vibrissal barrel subfields of rat cortex.
Cereb Cortex
7:
422-431,
1997[Abstract/Free Full Text].
5.
Heimann, A,
Kroppenstedt S,
Ulrich P,
and
Kempski OS.
Cerebral blood flow autoregulation during hypobaric hypotension assessed by laser Doppler scanning.
J Cereb Blood Flow Metab
14:
1100-1105,
1994[Web of Science][Medline].
6.
Hudetz, AG,
Feher G,
Weigle CGM,
Knuesse DE,
and
Kampine JP.
Video microscopy of cerebrocortical capillary flow: response to hypotension and intracranial hypertension.
Am J Physiol Heart Circ Physiol
268:
H2202-H2210,
1995[Abstract/Free Full Text].
7.
Hudetz, AG,
Roman RJ,
and
Harder DR.
Spontaneous flow oscillations in the cerebral cortex during acute changes in mean arterial pressure.
J Cereb Blood Flow Metab
12:
491-499,
1992[Web of Science][Medline].
8.
Lauritzen, M.
Long-lasting reduction of cortical blood flow of the brain after spreading depression with preserved autoregulation and impaired CO2 response.
J Cereb Blood Flow Metab
4:
546-554,
1984[Web of Science][Medline].
9.
Leninger-Follert, E,
and
Lübbers DW.
Interdependence of capillary flow and regional blood flow in the brain.
In: Cerebral Circulation and Metabolism, edited by Landfitt TW,
McHenry J,
Reivich LCM,
and Wollman H.. New York: Springer, 1975, p. 46-48.
10.
Mchedlishvili, G.
Vascular Mechanism of the Brain. New York: Plenum, 1986.
11.
Naora, H.
Schwarzschild-Villiger effect in microspectrophotometry.
Science
115:
248-249,
1952[Free Full Text].
12.
Sándor, PJ,
Cox van Put J,
de Jong W,
and
de Wied D.
Continuous measurement of cerebral blood volume in rats with the photoelectric technique: effect of morphine and naloxone.
Life Sci
39:
1657-1665,
1986[Web of Science][Medline].
13.
Tomita, M,
and
Gotoh F.
Which circulates faster through the cerebral microcirculatory system, red cells or plasma?
Stroke
13:
722,
1982[Free Full Text].
14.
Tomita, M,
Gotoh F,
Amano T,
Tanahashi N,
Kobari M,
Shinohara T,
and
Mihara B.
Transfer function through regional cerebral cortex evaluated by a photoelectric method.
Am J Physiol Heart Circ Physiol
245:
H385-H398,
1983.
15.
Tomita, M,
Gotoh F,
Sato T,
Amano T,
Tanahashi N,
Tanaka K,
and
Yamamoto M.
Photoelectric method for estimating hemodynamic changes in regional cerebral tissue.
Am J Physiol Heart Circ Physiol
235:
H56-H63,
1978.
16.
Tomita, M,
Gotoh F,
Yamamoto M,
Tanahashi N,
and
Kobari M.
Effects of hemolysis, hematocrit, RBC swelling, and flow rate on light scattering by blood in a 0.26 cm ID transparent tube.
Biorheology
20:
485-494,
1983[Web of Science][Medline].
17.
Vogel, J,
and
Kuschinsky W.
Decreased heterogeneity of capillary plasma flow in the rat whisker-barrel cortex during functional hyperaemia.
J Cereb Blood Flow Metab
16:
1300-1306,
1996[Web of Science][Medline].
Am J Physiol Heart Circ Physiol 279(3):H1291-H1298
0363-6135/00 $5.00
Copyright © 2000 the American Physiological Society