|
|
||||||||
1 University of Texas Health Science Center at San Antonio, San Antonio 78284-7872; and 2 Biomedical Engineering Program, University of Texas at Austin, Austin, Texas 78712-1084
| |
ABSTRACT |
|---|
|
|
|---|
Transgenic mice offer a valuable way to relate gene products to phenotype, but the ability to assess the cardiovascular phenotype with pressure-volume analysis has lagged. Conductance measurement offers a method to generate an instantaneous left ventricular (LV) volume signal in the mouse but has been limited by the volume signal being a combination of blood and LV muscle. We hypothesized that by developing a mouse conductance system that operates at several simultaneous frequencies, we could identify and correct for the myocardial contribution to the instantaneous volume signal. This hypothesis is based on the assumption that mouse myocardial conductivity will vary with frequency, whereas mouse blood conductivity will not. Consistent with this hypothesis, we demonstrated that at higher excitation frequency, greater end-diastolic and end-systolic conductance are detected, as well as a smaller difference between the two. We then empirically solved for LV blood volume using two frequencies. We combined measured resistivity of mouse myocardium with an analytic approach and extracted an estimate of LV blood volume from the raw conductance signal. Development of a multifrequency catheter-based system to determine LV function could be a tool to assess cardiovascular phenotype in transgenic mice.
transgenic mouse; pressure-volume analysis; myocardial resistivity; left ventricular function
| |
INTRODUCTION |
|---|
|
|
|---|
CONDUCTANCE MEASUREMENT offers a method to generate an instantaneous left ventricular volume signal in the mouse (13). It uses an electric field generated from electrodes at the apex and immediately above the left ventricle to sense the instantaneous conductance change as the left ventricle fills and ejects blood. A signal proportional to the left ventricular blood volume is required for use in physiological studies. Unfortunately, the presently available instantaneous conductance output is a combination of blood and left ventricular muscle (2-4, 17, 27). We hypothesized that by developing a mouse conductance system that operates at several simultaneous frequencies, we could identify and possibly correct for the myocardial contribution to the instantaneous volume signal.
This hypothesis is based on the assumption that mouse myocardial
conductivity will vary with frequency, whereas mouse blood conductivity
will not. Prior work has shown that blood has constant electrical
resistivity over a wide range of frequencies (2-100 kHz)
(22). In contrast, the resistivity of myocardium is known to change with frequency: specifically, the resistivity of myocardium is lower at increased excitation frequency (8,
23-25, 33). We propose (see Fig. 1) that
at lower frequencies, there is a maximal gradient between the
resistivity of blood and myocardium such that the electric field
generated will be primarily confined to the left ventricular cavity
and, to a lesser degree, the myocardium. At higher frequencies, there
will be a minimal gradient between the resistivity of blood and
myocardium, and the electric field generated will not be confined to
the left ventricular cavity but will extend into the myocardium.
Accordingly, we hypothesized that the higher the excitation frequency,
the greater the apparent end-diastolic and end-systolic conductance
detected by the miniaturized conductance catheter. In addition, if this
construct is correct, there should be a slight reduction in the
difference between end-diastolic and end-systolic conductance at higher
frequencies because the relative proportion of the signal changing from
systole to diastole is smaller.
|
This approach could have an important advantage over the traditional conductance method for determining measures of ventricular function such as end-systolic elastance. Because elastance is generated during beat-by-beat changes in loading conditions, a method to determine and correct for instantaneous parallel conductance is critical and does not exist. The use of multiple simultaneous frequencies has the potential to solve this problem. The application of this approach would be in transgenic mice. There is a need to relate specific gene products to phenotype. Unfortunately, the ability to rigorously assess the cardiovascular phenotype in very small animals has lagged (6, 15). Such analysis has been available in larger animals by measurement of simultaneous left ventricular pressure and volume to examine cardiac performance in the pressure-volume plane (1, 16). The application of this approach to mice has been difficult because of the small size of the mouse heart and the rapid heart rate. Creating the technology to generate an accurate instantaneous volume signal in the transgenic mouse to generate pressure-volume relationships during occlusion of the inferior vena cava was a goal of this study.
In an effort to exploit this differential frequency response, we sought to determine whether we could empirically solve for left ventricular blood volume using a multifrequency input signal. By combining experimental data with an analytic approach consisting of a series of equations, we were able to extract an accurate estimate of left ventricular blood and myocardial components.
| |
METHODS |
|---|
|
|
|---|
Conductance catheter. A 1.4-Fr miniaturized pressure-volume catheter (SPR-719; Millar Instruments, Houston, Texas) was used in these studies. The catheter has four platinum electrodes, each 0.25 mm in length, with an interelectrode spacing of 0.5, 4.5, and 0.5 mm between electrodes 1 and 2, 2 and 3, and 3 and 4, respectively. A constant excitation current (17 µA root mean square) was applied to the two outermost electrodes by using a custom signal generator/processor and bridge amplifiers developed by us and subsequently modified (MCS-100; Millar Instruments), and the two intermediate electrodes were used to measure the instantaneous voltage signal. A high-fidelity pressure transducer (frequency response flat to 10 kHz) was mounted between electrodes 2 and 3 to measure ventricular pressure.
The theory behind the determination of volume with the use of the conductance catheter in larger animals has been described in detail elsewhere (19). Briefly, the tetrapolar catheter generates an intraventricular electric field between the outer electrodes. The potential differences between the inner electrode pair are measured continuously and yield an instantaneous voltage output. The method of conversion of instantaneous voltage to instantaneous conductance for both single and dual frequency is given below. The time-varying ventricular volume, Vi(t), was estimated from
|
(1) |
is the ratio of flow probe stroke volume (SV) to
conductance SV,
is the blood resistivity, L is the
distance between the sensing electrodes,
Gi(t) is the instantaneous
conductance, and Gpi is the
instantaneous conductance of the surrounding structures, particularly
the left ventricular myocardium, or parallel conductance. Assuming that
the ratio of flow probe SV to conductance SV is 1, and assuming no
parallel conductance, then Eq. 1 simplifies to
|
(2) |
Calculation of mouse blood volume from single-frequency data.
The signal processor provides an analog voltage output. A calibration
procedure was developed to convert this voltage output into conductance
(µS). Known resistors were used to calibrate our instrumentation.
Because conductance is the inverse of resistance, these resistors were
connected to our system as the input, and a corresponding voltage was
derived. The relationship of the input conductance to the output
voltage was then known. This relationship was added to our data
acquisition software and allowed us to obtain instantaneous conductance
from the beating mouse heart. This generated Gi,
which was converted to volume using Baan's simplified Eq. 2
and not Eq. 1, because we did not know
Gpi. We then subtracted steady-state
parallel conductance (Vp), derived by the
hypertonic saline method (described in detail below) from the raw
volume signal, and then multiplied by
to generate a corrected
volume signal.
is a correction factor. The electrical field generated by the
conductance catheter is known to be inhomogeneous, and as a result, the
raw conductance signal underestimates SV. The raw conductance signal is
therefore multiplied by
, defined as the ratio of flow probe
conductance SV to raw conductance SV.
Vp was determined via the hypertonic saline
technique described by Baan et al. (1), which was modified
for the mouse. A 20-µl bolus of hypertonic saline (1.5%) was
directly injected into the right ventricle with a 30-gauge needle
attached to PE-10 tubing. As the saline entered the left ventricle
following transit through the lungs, it transiently increased blood
conductivity. This resulted in an increase in the left ventricular
conductance signal and, theoretically, no change in left ventricular
hemodynamics. Conductance at end-systole (Ges)
and end-diastole (Ged) are related during saline
wash-in as
|
(3) |
|
(4) |
Calculation of mouse blood volume from multifrequency data.
The conductance signal output generated from the mouse varies with time
and is a combination of signals arising from both the blood and
myocardium. Derivation of left ventricular chamber volume requires the
assumption that, whereas resistivity of myocardium varies as a
definable function of frequency, resistivity of blood is independent of
frequency. As such
|
(5) |
|
(6) |
|
(7) |
|
(8) |
|
(9) |
|
(10) |
f1 and
f2 in other mice of the same strain, we could then solve for the shape constant k in
Eq. 10. Once k was known, we could derive the
instantaneous muscle conductance for either frequency according to
Eq. 9, substitute muscle conductance into Eq. 5
or 6, and then solve for blood conductance.
Equation 1 was used to convert conductance to volume.
Assuming that
= 1, maximal and minimal volume were determined
as well as their difference (SV).
was then calculated as the ratio
of flow probe SV to dual-frequency-derived SV. Finally, the raw
dual-frequency volume was multiplied by
to derive the corrected
dual-frequency volume.
System calibration.
We wanted to demonstrate that the miniaturized mouse conductance
catheter would generate a constant conductance output at different
frequencies. Saline with resistivities of 62, 78, 93, 109, and 125
· cm, which span the reported resistivities of blood
(28), were used. The saline was placed in 6-mm-diameter test tubes in a H2O bath at 37°C, and the conductance
catheter was centered in the saline. Voltage output was determined at
frequencies ranging from 500 Hz to 100 kHz.
Studies in mice.
The protocol was approved by the Institutional Animal Care and Use
Committee at the University of Texas Health Science Center at San
Antonio and conformed with Guidelines for the Care and Use of
Laboratory Animals (NIH Publication No. 86-23, Revised 1985) and
Principles of Laboratory Animal Care (published by the National Society for Medical Research). CD-1 mice (n = 19) weighing 20-30 g were anesthetized by administration of
methoxyflurane (Metofane; Pitman-Moore, Mundelein, IL) in a closed
chamber, followed by urethan (1,000 mg/kg ip) and etomidate (25 mg/kg
ip). Respiration was controlled through a tracheostomy cannula, and the
animals were mechanically ventilated with room air at 60 breaths/min
using a rodent ventilator (model 680; Harvard Apparatus, South Natick, MA). The chest was entered via an anterior thoracotomy. A small animal
blood flow meter (T 106; Transonic Systems, Ithaca, NY) was placed
around the aorta. The flowmeter was placed on the descending thoracic
aorta immediately above the level of the diaphragm. For technical
reasons, it could not be placed on the ascending thoracic aorta. In
five additional CD-1 mice examined by echocardiography (10), 72.5 ± 4.8% of the SV ejected into the
ascending thoracic aorta was determined to be present in the descending
thoracic aorta. Because the flow probe was placed around the descending thoracic aorta, flow probe SV (as presented in Table
1) was normalized to correct for SV lost
to the vessels of the aortic arch.
|
Determination of mouse muscle and blood resistivity. To determine the conductance of in vivo myocardium, a customized suction tetrapolar electrode device was developed using silver-coated copper wire. The four electrodes ended as blunt probes separated by 0.25 mm that were contained within a 1-mm plastic tube that could be attached to a vacuum, similar to the device developed for the canine by Steendijk et al. (24). This enabled the device to be applied directly to the epicardium of the beating mouse myocardium to determine frequency-resistivity relationships at the same time that the miniaturized mouse conductance catheter was inserted into the left ventricle to determine frequency-conductance relationships (n = 8).
We also wanted to demonstrate that the resistivity of mouse blood was constant at different frequencies. Six additional CD-1 mice were bled. The blood from each animal was placed in 6-mm-diameter test tubes in a H2O bath at 37°C, and the conductance catheter was centered in the blood. Resistivity was determined at frequencies ranging from 500 Hz to 100 kHz.Evidence that the electric field is extending into the mouse myocardium. To demonstrate the limitations of the hypertonic saline technique and the importance of an alternative technique such as multifrequency measurement, we sought to demonstrate the load dependence of Vp. Six additional CD-1 mice underwent the in vivo protocol outlined above. After the baseline conductance and flow probe data were acquired at 10 kHz, sustained aortic occlusion was performed by placing tension on a suture around the descending thoracic aorta until a new steady-state systolic pressure was achieved with increased intraventricular pressure. Hypertonic saline determination of Vp was performed at baseline and at increased load.
Mouse heart morphological measurements. To determine the wall thickness-to-chamber ratio (11, 20), three mouse hearts were cut longitudinally in 4-µm sections with a microtome and mounted on microscope slides. Staining was performed with hematoxylin and eosin. Measurements of anterior and posterior wall thickness and chamber diameter were made at the midpapillary muscle level with planimetry.
Calculations.
Conductance-derived pressure-volume data were analyzed with software
developed by us (PVAN) and modified by Millar Instruments. Heart rate
was determined as 1/R-R interval, end-systolic pressure was pressure at
the point of maximal pressure-to-volume ratio, end-diastolic pressure
was the pressure at the R wave, end-systolic volume (ESV) was the
minimal left ventricular volume, end-diastolic volume (EDV) was the
maximal left ventricular volume, and SV was the difference (EDV
ESV).
Statistics. Within each mouse, relationships between frequency (logarithmic scale) and resistivity and between frequency and conductance were examined by scatter plots and by computing Spearman's nonparametric rank correlation coefficients. Nonparametric analyses were used because most relationships were monotone but not necessarily linear. Cochran-Mantel-Haenszel statistics based on ranks were computed to obtain global measures of correlation controlling for individual mouse effects. The average Spearman's correlation coefficients, averaged over all pairs of mice, were then computed from the resulting Friedman's test statistics (7). All computations were performed using SAS (version 6.11; SAS Institute, Cary, NC).
| |
RESULTS |
|---|
|
|
|---|
Figure 2 shows an example of data
from an individual mouse during calculation of single-frequency
Vp with bolus injection of hypertonic saline.
Figure 2A shows the conductance-derived left ventricular
pressure-conductance relationships during hypertonic saline washin that
were used to calculate Ges and
Ged on a beat-to-beat basis, shown in Fig.
2B. The beat-to-beat changes were greater in
Ged than in Ges. Figure
2B demonstrates the calculation of Vp
using Eqs. 3 and 4 in METHODS.
Vp was 554.24 µS.
|
Figure 3 shows the regression
lines of frequency (kHz) versus conductance (µS) of the signal
processor (MCS, Houston, TX). These studies demonstrate that the
conductance output of the signal processor is constant over the
frequencies examined. This is critical because voltage was converted to
conductance, and a varying source current would confound interpretation
(because G = I/V, where G is
conductance, V is voltage, and I is current;
I must be constant over the frequencies examined for
V to be converted to G). Moreover, the constant
current output persisted despite a doubling of resistivity of saline
from 62 to 125
· cm.
|
Figure 4 is a plot of the resistivity
versus frequency of mouse blood, determined by placing the miniaturized
mouse conductance catheter in a tube containing blood, and of mouse
left ventricular myocardium, derived from the epicardial suction
tetrapolar electrodes. The mean and standard deviations from eight mice
are shown. The resistivity of mouse blood is constant despite
frequencies ranging from 500 Hz to 100 kHz
(r2 = 0.183, where
r2 is the correlation coefficient;
P = 0.015). In contrast, the myocardial resistivity
falls as frequency increases over the same frequency range
(r2 =
0.441, P = 0.001).
|
Figure 5 is a plot of measured
conductance, as a function of frequency, measured in vivo by the
conductance catheter placed in the same eight mice shown in Fig. 4. The
conductance is not corrected for parallel conductance
(Vp) or electrical field inhomogeneity (
). As
frequency increases, both end-diastolic conductance
(r2 = 0.987, P = 0.001) and
end-systolic conductance (r2 = 0.985, P = 0.001) increase. In contrast, as frequency
increases, there is a decrease in the difference between end-diastolic
and end-systolic conductance (
G;
r2 =
0.200, P = 0.008),
although the correlation coefficient is low.
|
Table 1 is a comparison of the EDV, ESV, and SV values calculated by
using the two methods. The first method uses multiple frequencies as
proposed in this manuscript, according to Eqs. 5-10.
The second method uses that proposed by Baan and co-workers (1) where the raw conductance signal is corrected for
Vp and
. For this analysis, two frequencies
were used (10 and 100 kHz), and these frequencies were delivered
simultaneously. These frequencies correspond to maximal differences in
myocardial resistivity (Fig. 4) and a flat current output of the signal
processor (Fig. 3).
The baseline hemodynamics of these six mice included an end-systolic
pressure of 57 ± 6 mmHg, heart rate of 439 ± 19 beats/min, and flow probe SV of 16.19 ± 3.29 µl/beat. The use of two
simultaneous frequencies (10 and 100 kHz) to calculate volumes
according to the new method proposed in this manuscript yielded a raw
SV of 6.0 ± 1.3 µl/beat (not corrected by
Vp or
), EDV of 40 ± 8 µl, and ESV of
24 ± 7 µl.
was 2.75. The shape constants
ked and kes were 6.1 ± 0.6 and 5.3 ± 0.4 mm, respectively (P < 0.001). The standard method (single frequency
saline) yielded a
raw SV of 4.3 ± 0.8 µl/beat (not corrected by
Vp or
), EDV of 23 ± 4 µl, and ESV of
8 ± 2 µl.
was 3.63 ± 1.04. As the
multifrequency-derived SV, ESV, and EDV increased, so did the same
parameters derived with the standard method of calculation (SV,
r2 = 0.916; ESV,
r2 = 0.713; and EDV,
r2 = 0.933). The raw dual-frequency SV was
larger than the raw single-frequency SV (P < 0.001, not corrected by Vp or
).
Table 2 is a comparison of
Vp at steady-state conditions and when afterload
was increased by sustained aortic occlusion. Studies were performed on
six additional mice at a single frequency (10 kHz). The baseline
Vp of 14.6 ± 7.0 µl increased to
19.2 ± 7.9 µl with sustained afterload (P < 0.01). The heart rate did not change (446 ± 74 to 439 ± 43 beats/min, P = not significant), but end-systolic
pressure (52 ± 8 to 80 ± 11 mmHg, P < 0.01) and end-diastolic pressure both increased (3 ± 1 to 4 ± 2 mmHg, P < 0.01).
|
A plot of left ventricular pressure-volume relationships during the
identical caval occlusion are shown in Fig.
6. The dual-frequency data were
determined at 10 and 100 kHz and the single-frequency data at 10 kHz,
and the latter were corrected with the hypertonic saline technique. It
was anticipated that the slope of end-systolic elastance would differ
if determined with single- versus dual-frequency conductance. Figure 6
demonstrates that a 20.9% difference in elastance was found (23.68 vs.
19.59 mmHg/µl, single vs. dual frequency, respectively) in this
single example.
|
| |
DISCUSSION |
|---|
|
|
|---|
This study has demonstrated a new concept concerning the impact of
excitation frequency from a miniaturized mouse conductance catheter on
electric field generation. We proposed that at lower frequencies, the
field would be confined to the left ventricular cavity and, to a lesser
degree, the myocardium. Likewise, at higher frequencies, the electric
field generated would not be confined to the left ventricular cavity
but extend into the myocardium. The basis for this proposal, which we
confirmed in the in vivo mouse heart, is that myocardial resistivity
decreases with increasing frequency (8,
23-25, 33), whereas blood resistivity is
constant at different frequencies (22). The resulting
hypothesis, stating that at higher excitation frequency greater
end-diastolic and end-systolic conductance and smaller
G
would be generated, has been confirmed in this study. Finally, we were
able to empirically estimate left ventricular blood volume using two
frequencies simultaneously. We combined measured resistivity of mouse
myocardium with an analytic approach and were able to extract an
estimate of left ventricular blood volume from the raw conductance signal.
One advantage of dual-frequency conductance documented in the present
study was the detection of a larger amount of SV. The flow probe
measured 16.19 ± 3.29 µl/beat compared with 4.3 ± 0.8 µl/beat for the single-field conductance technique. The dual-field technology detected 6.0 ± 1.3 µl/beat. As a result, the
correction factor for electrical field inhomogeneity (
) as the ratio
of flow probe SV to raw conductance SV would be smaller with dual frequency. Because the final step in converting the instantaneous voltage output generated by the conductance catheter to volume is
multiplication by
, there would be less magnification of any error.
A second advantage is that the dual-field technique eliminates the need
to administer hypertonic saline to determine steady-state parallel
conductance. The conductance technique and Eq. 1 require knowledge of the resistivity of blood, which is always changing in a
given mouse as more hypertonic saline is injected. Given the very small
blood volume of mice, it is not practical to measure this changing
resistivity for every condition during each experiment.
The mean ejection fraction (EF) for the dual-frequency method calculated for the six mice presented in Table 1 is 42 ± 7%. The mean EF for these same mice calculated with the single-frequency method is 68 ± 9%. Given the open-chest preparation, the heart rate, and the instrumentation used in the present study, 42% is likely a more realistic value. Why does single-frequency conductance overestimate EF in the mouse? The SV determined by the dual- and single-frequency methods will be equal because both are corrected by the flow probe SV. The single-frequency EDV is smaller than the dual-frequency EDV. Therefore, the calculated EF will always be larger with the single-frequency method.
The smaller EDV determined with single frequency compared with dual frequency is due to overestimation of parallel conductance (Vp). The hypertonic saline method as originally described by Baan et al. (1) assumes no significant effect on left ventricular volume or performance. That finding has been confirmed in larger animals where ~2% of the left ventricular blood volume is injected. In the mouse, the injection of 20 µl of hypertonic saline in the right ventricle represents a large portion of left ventricular blood volume. This causes significantly greater beat-by-beat changes in end-diastolic than in end-systolic conductance, as demonstrated in Fig. 2A. This is anticipated because the bolus of hypertonic saline produces a sudden increase in preload without a change in contractility. To determine Vp, Ges is plotted on the ordinate and Ged on the abscissa as shown in Fig. 2B, and Vp is determined as the intercept of the extrapolated data with the line of identity. Because Ged is changing faster than Ges, violating the assumptions of Baan et al. (1), the intersection with the line of identity will be artificially increased. The result will be an overestimation of Vp, a smaller EDV, and an overestimated EF with the single-frequency technique.
The derivation of left ventricular blood volume from multifrequency data requires two assumptions. The first is that mouse myocardial resistivity determined in vivo on the epicardial surface by the suction tetrapolar electrode is similar to the myocardial resistivity component of the miniaturized mouse conductance signal generated from within the left ventricular chamber. The traditional method used to determine myocardial resistivity is to place in vitro epicardial pins or electrode holders onto the myocardium (8, 18, 21, 30). This method is limited by injury from the electrodes and ischemia to the myocardium as an in vitro preparation. Ischemia has previously been shown to significantly increase myocardial resistivity (9, 29). The use of the blunt tetrapolar suction electrodes in the current study solves both of these problems and is similar in concept to that employed by Steendijk and co-workers (24).
The second assumption is that mouse myocardial resistivity varies with frequency, whereas mouse blood resistivity does not vary with excitation frequency. The observation that muscle resistivity decreases as the frequency of the input signal increases was established in the 19th century (14) and has been reconfirmed by several groups (8, 23-25, 33). This construct holds that at high frequency, the cell membranes should be effectively short-circuited, with tissue conductance equal to the conductivity of cytoplasm. As a result, resistivity falls. In contrast, at low frequency, the field also travels through the cell membrane that has higher resistivity. We specifically designed the tetrapolar suction electrodes with electrode spacing of 0.25 mm to be significantly less than the 1.2 ± 0.1-mm (n = 3) average anterior myocardium thickness determined morphologically. A ratio of myocardial thickness to electrode separation >1 ensures that the electrical field is confined to the myocardium and does not extend into the left ventricular blood (24). Moreover, because the intramyocardial conductance signal will pass through a substantial depth of myocardium, the fiber orientation through which it passes will vary. Because myocardial blood volume varies with the phase of the cardiac cycle and resistivity of cardiac muscle differs with fiber direction (8, 21, 24, 25), these could all be sources of variation in determining in vivo myocardial resistivity. We therefore choose to present a lumped value for resistivity, independent of fiber orientation or phase of the cardiac cycle.
There was a previous study attempting to estimate left ventricular offset volume using two stimulation frequencies. Gawne and co-workers (12) also took advantage of the fact that myocardial resistivity is dependent on input frequency, whereas blood resistivity is independent of input frequency. They hypothesized that SV would be independent of input frequency, whereas minimal volume (ESV) would vary with input frequency. Therefore, the change in ESV with frequency would be due to quantifiable Vp. However, their approach could only be applied during steady-state conditions and therefore could not be applied with changing loading conditions to generate end-systolic elastance. Furthermore, as shown in Fig. 1, our hypothesis predicts that apparent SV should not be constant but should fall at increasing frequency due to a less dense electric field in the left ventricular chamber. We confirmed that apparent SV fell with increasing input frequency, whereas Gawne et al. (12) did not. The frequencies examined by Gawne et al. did not exceed 33 kHz, a frequency maximum in the mouse that did not change myocardial resistivity. Finally, Gawne et al. wrote that an analytic solution to this problem of using multifrequency signals to solve for instantaneous parallel conductance was not possible due to the complex, changing geometry of the heart. We have solved this problem with the introduction of a shape constant, k. By determining the instantaneously changing shape constant during the cardiac cycle and during changing loading conditions, we have a more robust approach that can now be applied not just during steady state but also during the generation of load-independent measures of contractility.
The use of dual-frequency conductance is relevant to crossing the major hurdle of applying conductance measurements to evaluate left ventricular function in the pressure-volume plane. A traditional limitation of conductance measurements has been changing parallel conductance during occlusion of the inferior vena cava to generate end-systolic elastance, effective arterial elastance, and additional measures of ventricular function generated from pressure-volume analysis (2, 5). The currently accepted technique of using small injections of hypertonic saline to correct for parallel conductance is only accurate during steady-state conditions (1, 17). Small and physically insignificant changes in parallel conductance occur throughout the cardiac cycle (17, 26, 31). However, during a change in loading conditions, parallel conductance is also changing (2, 5). This was confirmed in the current study where sustained aortic occlusion, which allowed a significant increase in aortic pressure, increased Vp from 14.6 ± 7.0 to 19.2 ± 7.9 µl (P < 0.01, Table 2) in six mice. Occlusion of the inferior vena cava results in the left ventricle shrinking around a fixed electric field generated from a conductance catheter. Therefore, the amount of field leakage into the left ventricular myocardium and surrounding structures changes instantaneously. For these reasons it has not been possible to determine the absolute slope of end-systolic elastance with conductance measurements.
There are two limitations of this study. The first is whether dual-frequency conductance will only be successfully applied to the mouse or whether it can be generalized to larger species. The mouse has a greater ratio of left ventricular wall thickness to chamber radius than larger species [mouse 1.0 ± 0.3 (n = 3) vs. human 0.45] (11, 20). Therefore, more of the mouse conductance signal will derive from the myocardium, making it an ideal species for multifrequency studies. This may not be true in larger dilated hearts. For instance, in patients with dilated cardiomyopathy, little of the electrical field generated from the conductance catheter reaches the myocardium (32). Dual-frequency measurement may not be useful to determine an accurate estimate of left ventricular volumes in this setting.
Second, it is not currently known how the shape constant k varies during inferior vena cava occlusion. The ability to generate instantaneous Gp i with multifrequency measurements is dependent on determining k not only at end diastole and end systole but also throughout the cardiac cycle and with different loading conditions. The conductance of myocardium at a given frequency is equal to the resistivity of myocardium multiplied by the shape constant. Because the shape varies with the cardiac cycle and load, and because myocardial resistivity varies with frequency and not the cardiac cycle or load, once the relationship between k and cardiac cycle/loading condition is defined, correction for parallel conductance at any time during the cardiac cycle or loading condition will be feasible. In the present study, k decreased from 6.1 ± 0.6 to 5.3 ± 0.4 mm between end diastole and end systole. An assessment of how k varies with changing loading conditions will require further studies.
In conclusion, we have been able to estimate left ventricular blood volume in the mouse with a miniaturized multifrequency mouse conductance system. Such an approach has the capability to determine instantaneous parallel conductance and determine an accurate estimate of ventricular function with left ventricular pressure-volume analysis, which has not been previously possible. The value of this approach would be increased if the conductance catheter could be placed and calibrated from the carotid artery. Attempts to add a velocity transducer to the shaft of the conductance catheter are in progress. This would eliminate the need for the flow probe. Dual frequency would allow calibration for parallel conductance, eliminating the need for bolus hypertonic saline. Resistivity of mouse myocardium may vary in different strains, especially transgenic strains. However, once these resistivities are determined and published, investigators may be able to use them without measuring them in every mouse. With such refinements, the development of the multifrequency catheter-based system to determine left ventricular function could be a tool to relate specific gene products to phenotype in the transgenic mouse.
| |
ACKNOWLEDGEMENTS |
|---|
We thank Danny Escobedo for excellent surgical assistance. Gary M. Clark, in the Biostatistics and Data Processing Unit, performed statistics at the University of Texas Health Science Center in San Antonio, with support from Cancer Center Support Grant PO CA54174.
| |
FOOTNOTES |
|---|
This study was supported by the University of Texas Health Sciences Center in San Antonio, the Department of Veterans Affairs, and a research grant from Millar Instruments, Houston, TX.
Address for reprint requests and other correspondence: M. D. Feldman, Medicine/Cardiology, 7703 Floyd Curl Dr., San Antonio, TX 78284-7872 (E-mail: feldmanm{at}uthscsa.edu).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 17 August 1999; accepted in final form 11 February 2000.
| |
REFERENCES |
|---|
|
|
|---|
1.
Baan, J,
Van Der Velde ET,
De Bruin HG,
Smeenk GJ,
Van Dijk AD,
Temmerman D,
Senden J,
and
Buis B.
Continuous measurement of left ventricular volume in animals and humans by conductance catheter.
Circulation
70:
812-823,
1984
2.
Boltwood, CM,
Appleyard RF,
and
Glantz SA.
Left ventricular volume measurement by conductance catheter in intact dogs: parallel conductance volume depends on left ventricular size.
Circulation
80:
1360-1377,
1989
3.
Burkhoff, D,
Van Der Velde E,
Kass D,
Baan J,
Maughan WL,
and
Sagawa K.
Accuracy of volume measurement by conductance catheter in isolated, ejecting canine hearts.
Circulation
72:
440-447,
1985
4.
Cabreriza, SE,
Dean DA,
Jia CX,
Dickstein ML,
and
Spotnitz HM.
Electrical isolation of the heart: stabilizing parallel conductance of left ventricular volume measurement.
ASAIO J
43:
M509-M514,
1997[Web of Science][Medline].
5.
Cassidy, SC,
and
Teitel DF.
The conductance catheter technique for measurement of left ventricular volume in young piglets.
Pediatr Res
31:
85-90,
1992[Web of Science][Medline].
6.
Christensen, G,
Wang Y,
and
Chien K.
Physiologic assessment of complex cardiac phenotypes in genetically engineered mice.
Am J Physiol Heart Circ Physiol
272:
H2513-H2524,
1997
7.
Conover, WJ.
Practical Nonparametric Statistics. New York: Wiley, 1971, p. 271-272.
8.
Epstein, BR,
and
Foster KR.
Anisotropy in the dielectric properties of skeletal muscle.
Med Biol Eng Comput
21:
51-55,
1983[Web of Science][Medline].
9.
Fallert, MA,
Mirotznik MS,
Downing SW,
Savage EB,
Foster KR,
Josephson ME,
and
Bogen DK.
Myocardial electrical impedance mapping of ischemic sheep hearts and healing aneurysms.
Circulation
87:
199-207,
1993
10.
Feldman, MD,
Erikson JM,
Mao Y,
and
Freeman GL.
Comparison of three techniques to assess mouse ventricular volume (Abstract).
Circulation
100:
I-759,
1999.
11.
Ganau, A,
Devereux RB,
Roman MJ,
De Simone G,
Pickering TG,
Sergio P,
Vargiu P,
Simongini I,
and
Laragh JH.
Patterns of left ventricular hypertrophy and geometric remodeling in essential hypertension.
J Am Coll Cardiol
19:
1550-1558,
1992[Abstract].
12.
Gawne, TJ,
Kristen SG,
and
Goldstein RE.
Estimating left ventricular offset volume using dual-frequency conductance catheters.
J Appl Physiol
63:
872-876,
1987
13.
Georgakopoulos, D,
Mitzner WA,
Chen CH,
Byrne BJ,
Millar HD,
Hare JM,
and
Kass DA.
In vivo murine left ventricular pressure-volume relations by miniaturized conductance micromanometry.
Am J Physiol Heart Circ Physiol
274:
H1416-H1422,
1998
14.
Hermann, L.
Ueber eine wirkung galvanischer Strome auf Muskeln und Nervern.
Pflügers Arch Ges Physiol
5:
223-275,
1871.
15.
James, JF,
Hewett TE,
and
Robbins J.
Cardiac physiology in transgenic mice.
Circ Res
82:
407-415,
1998
16.
Kass, DA,
Yamazaki T,
Burkhoff D,
Maughan WL,
and
Sagawa K.
Determination of left ventricular end-systolic pressure-volume relationships by the conductance (volume) catheter technique.
Circulation
73:
586-595,
1986
17.
Lankford, EB,
Kass DA,
Maughan WL,
and
Shoukas AA.
Does parallel conductance vary during a cardiac cycle?
Am J Physiol Heart Circ Physiol
258:
H1933-H1942,
1990
18.
Lofgren, B.
The electrical impedance of a complex tissue and its relation to changes in volume and fluid distribution.
Acta Physiol Scand
23:
3-51,
1951.
19.
Mur, G,
and
Baan J.
Computation of the input impedances of a catheter for cardiac volumetry.
IEEE Trans Biomed Eng
31:
448-453,
1984[Web of Science][Medline].
20.
Reichek, N,
and
Devereux RB.
Reliable estimation of peak left ventricular systolic pressure by M-mode echographic-determined end-diastolic relative wall thickness: identification of severe valvular aortic stenosis in adult patients.
Am Heart J
103:
202-209,
1982[Web of Science][Medline].
21.
Rush, S,
Abildskov JA,
and
McFee R.
Resistivity of body tissues at low frequencies.
Circ Res
12:
40-50,
1963
22.
Schwan, HP.
Electrical properties of blood and its constituents: alternating current spectroscopy.
Blut
46:
185-197,
1983[Web of Science][Medline].
23.
Schwan, HP,
and
Kay CF.
Specific resistance of body tissues.
Circ Res
4:
664-670,
1956
24.
Steendijk, P,
Mur G,
Van Der Velde E,
and
Baan J.
The four-electrode resistivity technique in anisotropic media: theoretical analysis and application on myocardial tissue in vivo.
IEEE Trans Biomed Eng
40:
1138-1148,
1993[Web of Science][Medline].
25.
Steendijk, P,
Mur G,
Van Der Velde E,
and
Baan J.
Dependence of anisotropic myocardium electrical resistivity on cardiac phase and excitation frequency.
Basic Res Cardiol
89:
411-426,
1994[Web of Science][Medline].
26.
Szwarc, RS,
Laurent D,
Allegrini PR,
and
Ball HA.
Conductance catheter measurement of left ventricular volume: evidence for nonlinearity within cardiac cycle.
Am J Physiol Heart Circ Physiol
268:
H1490-H1498,
1995
27.
Szwarc, RS,
Mickleborough LL,
Mizuno SI,
Wilson GJ,
Liu P,
and
Mohamed S.
Conductance catheter measurements of left ventricular volume in the intact dog: parallel conductance is independent of left ventricular size.
Cardiovasc Res
28:
252-258,
1994
28.
Trautman, ED,
and
Newbower RS.
A practical analysis of the electrical conductivity of blood.
IEEE Trans Biomed Eng
30:
141-153,
1983[Web of Science][Medline].
29.
Van Oosterom, A,
De Boer RW,
and
Van Dam RT.
Intramural resistivity of cardiac tissue.
Med Biol Eng Comput
17:
337-343,
1979[Web of Science][Medline].
30.
Weidman, S.
Electrical constants of trabecular muscle form mammalian heart.
J Physiol
210:
1041-1054,
1970
31.
White, PA,
Chaturvedi RR,
Shore D,
Lincoln C,
Szwarc RS,
Bishop AJ,
Oldershaw PJ,
and
Redington AN.
Left ventricular parallel conductance during cardiac cycle in children with congenital heart disease.
Am J Physiol Heart Circ Physiol
273:
H295-H302,
1997
32.
Wu, CC,
Skalak TC,
Schwenk TR,
Mahler CM,
Anne A,
Finnerty PW,
Haber HH,
Weikle RM,
and
Feldman MD.
Accuracy of the conductance catheter for measurement of ventricular volumes seen clinically: effects of electrical field homogeneity and parallel conductance.
IEEE Trans Biomed Eng
44:
266-277,
1997[Web of Science][Medline].
33.
Zheng, E,
Shao S,
and
Webster JG.
Impedance of skeletal muscle from 1 Hz to 1 MHz.
IEEE Trans Biomed Eng
31:
477-483,
1984[Web of Science][Medline].
This article has been cited by other articles:
![]() |
R J Hassink, H Nakajima, H O Nakajima, P A Doevendans, and L J Field Expression of a transgene encoding mutant p193/CUL7 preserves cardiac function and limits infarct expansion after myocardial infarction Heart, July 15, 2009; 95(14): 1159 - 1164. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Hassink, K. B. Pasumarthi, H. Nakajima, M. Rubart, M. H. Soonpaa, A. B. de la Riviere, P. A. Doevendans, and L. J. Field Cardiomyocyte cell cycle activation improves cardiac function after myocardial infarction Cardiovasc Res, April 1, 2008; 78(1): 18 - 25. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Nielsen, S. B. Kristiansen, S. Ringgaard, T. T. Nielsen, A. Flyvbjerg, A. N. Redington, and H. E. Botker Left ventricular volume measurement in mice by conductance catheter: evaluation and optimization of calibration Am J Physiol Heart Circ Physiol, July 1, 2007; 293(1): H534 - H540. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Jegger, A. S. Mallik, M. Nasratullah, X. Jeanrenaud, R. d. Silva, H. Tevaearai, L. K. von Segesser, and N. Stergiopulos The effect of a myocardial infarction on the normalized time-varying elastance curve J Appl Physiol, March 1, 2007; 102(3): 1123 - 1129. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Reyes, M. E. Steinhelper, J. A. Alvarez, D. Escobedo, J. Pearce, J. W. Valvano, B. H. Pollock, C.-L. Wei, A. Kottam, D. Altman, et al. Impact of physiological variables and genetic background on myocardial frequency-resistivity relations in the intact beating murine heart Am J Physiol Heart Circ Physiol, October 1, 2006; 291(4): H1659 - H1669. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. L. Lindsey, G. P. Escobar, L. W. Dobrucki, D. K. Goshorn, S. Bouges, J. T. Mingoia, D. M. McClister Jr., H. Su, J. Gannon, C. MacGillivray, et al. Matrix metalloproteinase-9 gene deletion facilitates angiogenesis after myocardial infarction Am J Physiol Heart Circ Physiol, January 1, 2006; 290(1): H232 - H239. [Abstract] [Full Text] [PDF] |
||||
![]() |
O.-J. How, E. Aasum, S. Kunnathu, D. L. Severson, E. S. P. Myhre, and T. S. Larsen Influence of substrate supply on cardiac efficiency, as measured by pressure-volume analysis in ex vivo mouse hearts Am J Physiol Heart Circ Physiol, June 1, 2005; 288(6): H2979 - H2985. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. S. Ikonomidis, J. W. Hendrick, A. M. Parkhurst, A. R. Herron, P. G. Escobar, K. B. Dowdy, R. E. Stroud, E. Hapke, M. R. Zile, and F. G. Spinale Accelerated LV remodeling after myocardial infarction in TIMP-1-deficient mice: effects of exogenous MMP inhibition Am J Physiol Heart Circ Physiol, January 1, 2005; 288(1): H149 - H158. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Uemura, T. Kawada, M. Sugimachi, C. Zheng, K. Kashihara, T. Sato, and K. Sunagawa A self-calibrating telemetry system for measurement of ventricular pressure-volume relations in conscious, freely moving rats Am J Physiol Heart Circ Physiol, December 1, 2004; 287(6): H2906 - H2913. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A Alvarez, M. Reyes, D. Escobedo, G. L Freeman, M. E Steinhelper, and M. D Feldman Enhanced left ventricular systolic function early in type 2 diabetic mice: clinical implications Diabetes and Vascular Disease Research, October 1, 2004; 1(2): 89 - 94. [Abstract] [PDF] |
||||
![]() |
C. J. Hartley, A. K. Reddy, S. Madala, M. L. Entman, L. H. Michael, and G. E. Taffet Noninvasive ultrasonic measurement of arterial wall motion in mice Am J Physiol Heart Circ Physiol, September 1, 2004; 287(3): H1426 - H1432. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. E. J. M. Creemers, J. N. Davis, A. M. Parkhurst, P. Leenders, K. B. Dowdy, E. Hapke, A. M. Hauet, P. G. Escobar, J. P. M. Cleutjens, J. F. M. Smits, et al. Deficiency of TIMP-1 exacerbates LV remodeling after myocardial infarction in mice Am J Physiol Heart Circ Physiol, January 1, 2003; 284(1): H364 - H371. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. N. Lorenz A practical guide to evaluating cardiovascular, renal, and pulmonary function in mice Am J Physiol Regulatory Integrative Comp Physiol, June 1, 2002; 282(6): R1565 - R1582. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Janssen, J. Debets, P. Leenders, and J. Smits Chronic measurement of cardiac output in conscious mice Am J Physiol Regulatory Integrative Comp Physiol, March 1, 2002; 282(3): R928 - R935. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. D. Feldman, J. M. Erikson, Y. Mao, C. E. Korcarz, R. M. Lang, and G. L. Freeman Validation of a mouse conductance system to determine LV volume: comparison to echocardiography and crystals Am J Physiol Heart Circ Physiol, October 1, 2000; 279(4): H1698 - H1707. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |