|
|
||||||||
Department of Medical Engineering and Systems Cardiology, Kawasaki Medical School, Kurashiki, Okayama 701-0192 Japan
| |
ABSTRACT |
|---|
|
|
|---|
A new high-resolution digital radiographic technique based on the deposition of 125I- and 3H-labeled desmethylimipramine (IDMI and HDMI, respectively) was developed for the assessment of spatial and temporal myocardial flow heterogeneity at a microvascular level. The density distributions of two tracers, or relative flow distributions, were determined by subtraction digital radiography using two imaging plates of different sensitivity. The regions resolved are comparable in size to vascular regulatory units (400 × 400 µm2). This method was applied to the measurement of within-layer myocardial flow distributions in Langendorff-perfused rabbit hearts. The validity of this method was confirmed by the strong correlation between regional densities of two tracers injected simultaneously (r = 0.89 ± 0.03, n = 8). The temporal flow stability was evaluated by a 90-s continuous IDMI injection and subsequent bolus HDMI injection (n = 8). Regional densities of the two tracers were fairly correlated (r = 0.86 ± 0.03), indicating that the spatial pattern of flow distribution was stable even at a microvascular level over a 90-s period. The effect of microsphere embolization on the flow distribution was also investigated by the sequential injections of IDMI, 15-µm microspheres, and HDMI at 20-s intervals (n = 8). Microembolization increased the coefficient of variation of tracer density from 19 to 25% (P < 0.05), whereas the regional densities of two tracers were still correlated substantially, as in the case of no embolization (r = 0.84 ± 0.06). Thus the microsphere embolization enhanced flow heterogeneity with increasing flow differences between control high-flow and control low-flow regions but rather maintained the pattern of flow distribution. In conclusion, double-tracer digital radiography will be a promising method for the spatial and temporal myocardial flow analysis at microvascular levels.
molecular flow tracer; high-resolution flow imaging; spatial and temporal flow variation; coronary microembolization
| |
INTRODUCTION |
|---|
|
|
|---|
SPATIAL HETEROGENEITY OF MYOCARDIAL FLOW (2, 9, 15) and its marked stability over periods of several hours (10, 23, 33) have been well documented. Recently, it has been demonstrated that regional heterogeneity of O2 consumption is matched by flow heterogeneity (13, 26, 34, 36). Such regional flow adjustments are made through arteriolar vasomotor tone, which is the essential determinant of flow distribution (1). Therefore, the pattern of flow heterogeneity will be regulated down to arteriolar-to-capillary levels; in other words, responding to disturbances in the balance between O2 supply and consumption, the spatial pattern of flow distribution is possibly changeable at these microvessel levels. When the resultant flow distribution no longer fulfills the local O2 supply-demand matching, it possibly leads to focal necrosis (38) or arrhythmia through the coexistence of metabolically disturbed cardiac cells and normal ones (5). Hence, the evaluation of flow pattern changes at a microvascular level after the various interventions is of great physiological and clinical importance with respect to the understanding of flow and metabolism matching, the nature of local ischemia, and the development of arrhythmias.
Desmethylimipramine (DMI), an
2-adrenoceptor antagonist,
is an ideal molecular tracer for microregional myocardial flows (24, 25). DMI is distributed to tissue in proportion to
local flow, nearly completely extracted during a single pass, and
stably deposited mainly at capillary-tissue units; therefore, local DMI depositions are proportional to the local flow. It has been shown that
the radionuclide-labeled DMI deposition technique combined with
quantitative autoradiography (27, 28, 35) provides a
high-resolution imaging method for regional myocardial flow distributions. The regions resolved in this method are smaller than the
regions supplied by single arterioles (20, 21) or vascular
regulatory units (27). Before now, however, this method has not been extended to a two-time measurement in the same heart. In
this study, double-tracer digital radiography with 125I-
and 3H-labeled DMI (IDMI and HDMI, respectively) was
developed for a high-resolution, two-time measurement of myocardial
flow distribution, e.g., before and after intervention. This method was
applied to the study of microregional flow distribution in isolated
rabbit hearts with the aim of evaluating 1) time-to-time
flow variability in the same region or temporal flow stability and
2) the effect of microsphere embolization on the spatial
pattern of flow distribution.
| |
MATERIALS AND METHODS |
|---|
|
|
|---|
Double-Tracer Digitalradiography
In the presently developed double-tracer digital radiography, two tracers, IDMI and HDMI, from NEN Life Science Products (Boston, MA), and two imaging plates of different sensitivity, BAS-TR2040 (TR) and BAS-MS2040 (MS), from Fujix (Tokyo, Japan), were used. The TR plate detects both 125I and 3H radioactivity, and the MS plate detects only 125I radioactivity; accordingly, quantitative digital radiography using the TR and MS plates can visualize the double distribution of IDMI and HDMI and the sole IDMI distribution, respectively. The HDMI distribution can then be separated through the image subtraction processing in which the following measurements are required in advance.Quantification of digital radiography for tissue tracer content.
To confirm the linearity between tracer content in the myocardium and
digital radiographic (DRG) density, we quantitated the tissue tracer
content (in cpm/µg) by gamma-ray counting (IDMI) or liquid
scintillation counting (HDMI) together with DRG measurements. Dog
myocardial walls were homogenized together with IDMI, and several
tissue pastes of different radioactivity were prepared. Fractions of
these pastes were frozen in a
80°C freezer. The frozen pastes were
divided into 10-µm-thick slices using Tissue-Tek tissue mount in a
cold environment (
25°C) of a cryostat microtome. The slices were
put onto slide glasses and allowed to air-dry. The TR plate was then
exposed to them for 2 days. The radioactive energy stored on the TR
plate was scanned by the imaging plate reader (BAS2000, Fujix) and
converted into DRG density in arbitrary units (AU) and with a
resolution of 100 pixels/mm2 through our image processing
system (Macintosh G3 with an aid of MATLAB 5.2 software, MathWorks,
Natick, MA). The density was extremely uniform in every slice. The mean
density within each slice was then obtained after correcting for
background density. During the DRG processing, other fractions of the
remaining pastes were weighed and subjected to liquid scintillation
counting. The first calibration line relating tissue IDMI content to
DRG density, determined with the TR plate, was then obtained by linear
fitting. The MS plate was also exposed to the same slices for 1 day,
and, in like manner, the second calibration line relating tissue IDMI content to DRG density, determined with the MS plate, was obtained. Likewise, the third calibration line relating tissue HDMI content to
DRG density determined with the TR plate was obtained. The MS plate
could not detect HDMI at all because the sensitivity of the MS plate is
too low to detect the extremely low radiant energy of 3H.
The slices of tracer-containing myocardial paste were utilized later as
the 3H- and 125I-graded scales for the image
subtraction processing.
Point-spread functions in IDMI digital radiography. The distribution of DRG density inevitably included blurring, characterized by point-spread functions (PSF), the one-half band width of which increases with radiant energy of the tracers. The blurring is negligible in DRG measurements of HDMI because of the tiny radiant energy of 3H; however, the effect of 125I-radiation spread on density distribution of IDMI is measurable and dependent on the sensitivity of imaging plates. This IDMI blurring effect must be taken into account when conducting the image subtraction and comparing the IDMI and the HDMI distributions. To this end, four cylindrical IDMI point sources of <30-µm diameter and 10-µm height were made of silk threads soaked in a dilute solution of IDMI in advance. The PSF of the two plates (TR-PSF and MS-PSF) were determined by imaging the 125I point sources using the TR and MS plates, respectively, with the assumption that the PSF is circular symmetric.
Image subtraction processing.
In the quantitative double-tracer digital radiography, the IDMI-
and HDMI-containing myocardial slices of 10-µm thickness were placed
in contact with the TR plate for 2 days along with the above
3H- and 125I-graded scales and then placed in
contact with the MS plate along with the same graded scales for 1 day.
The former resulted in the composite DRG density distribution of HDMI
and IDMI, and the latter resulted in the DRG density distribution of
IDMI only. In parallel with DRG measurements, the tissue tracer
contents of the graded scales were measured for the preparation of
calibration lines. Corrected for background density, the two DRG
density distributions derived from each sample were so superimposed
that their correlation coefficient (r) of regional DRG
density reached the maximal value using the image processing system.
The sample outlines identified by the two DRG density images coincided
with each other nearly perfectly because DRG density in the region free
from the tracer radiant energy was <5% of the mean DRG density of the
tracers. The central 128 × 128-mm2 portion, belonging
to the same region of the myocardial tissue sample, was then easily
drawn out from each digitalradiogram. The sectioned DRG density
distributions of IDMI and IDMI plus HDMI were advanced to the following
image subtraction processing (Fig. 1).
|
Animal Studies
This study was conducted in accordance with the Guiding Principles of the American Physiological Society and with prior approval of the Committee on Animal Research of the Kawasaki Medical School.General experimental procedures.
Japanese white male rabbits weighing 2.5-3.5 kg were
anesthetized with intravenous administration of pentobarbital sodium (30 mg/kg). Heparin was administered intravenously (500 U · ml
1 · kg
1) to prevent
coagulation. The hearts were excised quickly, and the aortas cannulated
according to the Langendorff technique with Tyrode solution at 37°C.
The perfusion of hearts was discontinued only for a few tens of
seconds, and, meanwhile, the hearts did not stop beating. The perfusion
medium, containing (in mM) 130 NaCl, 10 NaHCO3, 5 CaCl2 · 2H2O, 4 KCl, 1 MgCl2 · 6H2O, 0.435 NaH2PO4 · H2O, and 5.56 dextrose, was bubbled continuously with a mixture of 95%
O2-5% CO2 and filtered through a 5-µm
Millipore filter. The coronary sinus flow and thebesian flow into the
right ventricle were collected via a cannula placed through the
pulmonary artery, and an effluent flow was measured with an inline
electromagnetic flowmeter. Aortic root pressure was measured with a
pressure transducer connected to the aortic cannula.
Group I: study of myocardial tracer extraction and retention (n = 6). To ensure that HDMI and IDMI were highly extracted during a single passage through the myocardium, the indicator dilution study was done. A mixed bolus of HDMI (10 µCi) and IDMI (1 µCi) was injected with a glass syringe over a period of 2-3 s into the perfusion line at time (t) = 0 s. The dead space volume between the aortic root and the injection site was 3 ml, and the perfusion rate was 60 ml/min at the lowest, leading to the dead time of <3 s. The effluent was collected at 2-s intervals from t = 0 to 40 s and, subsequently, at 5-s intervals until 180 s. The hearts were then arrested by perfusing a saturated KCl solution and weighed.
The radioactivity in the effluent was assessed by gamma-ray counting (IDMI) or liquid scintillation counting (HDMI). The fraction of the tracer retained in myocardium at a given time t, i.e., the residue function [R(t); in percent], was given by
|
(1) |
) is the
concentration (in cpm/ml) of tracer at t =
, and
C0 is the injected dose of tracer (in cpm).
The tracer amounts for the injection were determined so that the IDMI
and HDMI digital radiograms have comparable DRG densities, which are
much higher than background density. These conditions were required to
lower the effect of possible deviation from the linear relation between
DRG densities of IDMI determined with the MS and TR plates (the first
and second calibration lines) on the separation of HDMI density
distribution. The same amounts of tracers were used in the regional
flow assessments described below.
Group II: study of methodological error (n = 8).
To estimate the methodological error in the double-tracer digital
radiography, the myocardial deposition patterns of simultaneously injected HDMI and IDMI were examined. After a mixed bolus injection of
HDMI and IDMI over 2-3 s, we allowed the hearts to beat for 20 s, arrested them by perfusing a saturated KCl solution, and weighed them. The full wall thickness sample was excised from the
equatorial portions of the left ventricle. After the papillary muscles
were removed, the sample was sandwiched in aluminum sheets without
compression and immediately put in a
80°C freezer. A frozen sample
in a platelike shape was then divided into 10-µm-thick slices from
subendocardium to subepicardium using a cryostat microtome. The
subepicardial layer, including the large coronary vessels, and the
subendocardial layer that were subjected to a warping effect due to an
uneven endocardial surface were omitted from the study. The slices were
put onto slide glasses and allowed to air-dry. Twelve slices from each
heart were then subjected to the DRG processing.
Group III: study of temporal stability of spatial flow heterogeneity (n = 8). To evaluate the temporal stability of spatial flow distribution, IDMI was injected continuously over 90 s, and, subsequently, HDMI was injected over 2-3 s. IDMI was used for the continuous injection because it was reported that IDMI retention was much larger than HDMI retention (25). This matter was also confirmed in this study (group I). After the hearts were allowed to beat for 20 s, we applied the same procedure as described for group II.
Group IV: study of flow distribution with microembolism (n = 8). To evaluate the effect of multifocal embolism on the myocardial perfusion pattern, flow distributions before and after the microsphere injection were determined. IDMI, microspheres, and HDMI were injected sequentially; each injection was performed over 2-3 s at intervals of 20 s. The microsphere-injection solution contained 8 × 104 nonradioactive microspheres (16.5 ± 0.1 µm in diameter, NEN Life Science Products) suspended in 0.2 ml of 0.01% Tween 80 in isotonic saline. Before the microsphere injections, we throughly dispersed the microspheres by vortex mixing. Microscopic observation of the sliced samples obtained later showed almost no aggregation of microspheres. The hearts were allowed to beat for 20 s after the HDMI injection, and the same procedure was then applied as described for group II except that 27 slices were studied from each heart.
Flow distribution analysis. The resultant DRG density distributions (128 × 128 mm2 resolved into 100 × 100-µm2 pixels) were reconstructed with a pixel size of 400 × 400 µm2 (i.e., the order of magnitude of vascular regulatory units) to reduce random error incident to DRG measurements; thus the quantification was enhanced at the cost of spatial resolution. The reconstructed distributions were then normalized with their respective mean DRG densities. The similarity of those normalized distributions, i.e., relative flow distributions, was evaluated with a linear scatter plot of relative DRG density of IDMI versus HDMI in every myocardial region. The spatial flow heterogeneity was quantitated by the coefficients of variation (CV) of the tracer DRG density [CV (in percent) = SD/mean × 100]. The regional flow dispersion due to the methodological error and the stability of flow distribution were evaluated with the local density difference (in percent) given by the average difference of relative DRG density between IDMI and HDMI in one myocardial region (400 × 400 µm2) divided by their mean × 100. It should be noted that the focus of this study was the spatial pattern of flow distribution; that is, region-to-region flow variability was evaluated on a relative basis. Evaluation of absolute flow changes was beyond this study.
Data are given as means ± SD. Regression analyses were based on individual measurements using Spearman's rank correlation coefficient. Differences between CV values determined with IDMI and HDMI deposition in each group (groups II-IV) were assessed by Wilcoxon matched-pairs signed rank test. Differences among groups II, III, and IV were tested with Mann-Whitney U-test. A value of P < 0.05 was assessed to indicate a significant difference.| |
RESULTS |
|---|
|
|
|---|
Calibration for Quantitative Digital Radiography
An example of the linear relations between tissue tracer content [x-axis (in cpm/µg)] and its DRG density [y-axis (in AU)] is shown in Fig. 2. There are highly linear relations between them. These calibration lines were obtained at every DRG processing of flow distribution measurements for groups II-IV, and the calibration equation averaged y = (202.73 ± 41.83)x + (
12.66 ± 8.15) (r2 = 0.996 ± 0.003, 0.988-0.999) for IDMI with the MS plate, y = (168.00 ± 35.88)x + (
9.14 ± 6.27)
(r2 = 0.997 ± 0.003, 0.990-0.999) for IDMI with the TR plate, and y = (10.31 ± 1.89)x + (
2.56 ± 2.57)
(r2 = 0.995 ± 0.005, 0.986-0.999) for HDMI with the TR plate. Furthermore, these
results show that the DRG density of IDMI is more than 10 times as high
as that of HDMI for the same tissue tracer content. Accordingly, 1 µCi IDMI and 10 µCi HDMI were used in the following animal
experiments.
|
Figure 3 shows the PSF for IDMI digital
radiography. The DRG densities were normalized with the density at the
location of an IDMI point source. The lines were obtained by fitting
spline functions to the mean data from all digitalradiograms of point sources (n = 4). The shape of the PSF was assumed to be
axial symmetry. The one-half band widths determined with the TR and MS
plates were estimated to be 160 and 260 µm, respectively. The inset of Fig. 3 is the microscopic image of a point source
embedded in Tissue-Tek tissue mount of 10 µm thickness.
|
Animal Studies
In group I, the heart rate, mean perfusion pressure, and mean perfusion rate were 183 ± 23 beats/min, 81 ± 3 mmHg, and 8.1 ± 1.2 ml · min
1 · g
1,
respectively. The mixed bolus injection of two tracers had no effect on
these values. Figure 4 shows plots of
typical residue functions [R(t)] for IDMI and
HDMI versus time and their means ± SD at t = 20, 40, 60, ..., 120 s. The R(20) for IDMI
and HDMI are 95.3 ± 1.8 and 96.4 ± 0.8%, respectively. The
high myocardial retention was maintained for IDMI over 120 s
[R(120) = 95.0 ± 0.9%] but not
for HDMI [R(60) = 90.7 ± 3.1% and
R(120) = 86.2 ± 3.9%]. Thus IDMI
was efficiently extracted and well retained, and HDMI was not retained
as much as IDMI; nevertheless, for up to about 30 s, enough HDMI
was well retained for it to be a satisfactory tracer.
|
In group II, heart rate, mean perfusion pressure, and mean
perfusion rate were 182 ± 20 beats/min, 80 ± 6 mmHg, and
8.8 ± 2.1 ml · min
1 · g
1,
respectively. The tracer injection had no effect on these values. The
representative results for one myocardial slice are shown in Fig.
5 (left), showing relative
density distributions of IDMI and HDMI deposition and the scatter plot
of relative densities of IDMI versus HDMI. The relative density of the
tracers, which normalized with the mean density of tracer distribution,
is hereafter referred to as density for the sake of convenience. The
density was visualized in 256 levels of black and white gradations with a spatial resolution of 400 × 400-µm2 pixels.
Intensities are proportional to local flow. Apparently, the two tracer
distributions are very similar to each other. The correlation
coefficient, the slope of the regression line, CV values of IDMI and
HDMI deposition densities, and the local density difference are
summarized for each heart (means ± SD of 12 slices) in Table
1. As a whole (12 slices each from 8 hearts), these averaged 0.89 ± 0.03, 0.93 ± 0.08, 21.2 ± 4.0, 21.9 ± 3.8, and 7.9 ± 1.0%,
respectively. Differences between CV values of IDMI and HDMI deposition
were not significant.
|
|
In group III, heart rate, mean perfusion pressure, and mean
perfusion rate were 173 ± 22 beats/min, 81 ± 3 mmHg, and
8.8 ± 1.3 ml · min
1 · g
1,
respectively. The representative results for one myocardial slice are
shown in Fig. 5 (middle). The distributions of two tracers are also very similar to each other. The scatter of density plots is
not discernibly different from that in group II.
The correlation coefficient, the slope of the regression line, CV
values of IDMI and HDMI deposition densities, and the local percent
difference of density summarized for each heart (means ± SD of 12 slices) are shown in Table 1. As a whole (12 slices each from 8 hearts), these averaged 0.86 ± 0.03, 0.93 ± 0.10, 19.7 ± 3.3, 21.0 ± 3.7, and 8.7 ± 1.4%, respectively.
Differences between CV values of IDMI and HDMI deposition were also not
significant. Although correlation coefficient values were significantly
lower than those of group II, the difference was slight, and
distributions of two tracers correlated linearly with statistical
significance. Moreover, differences of the local density difference
between groups II and III were not significant;
assuming the independent methodological and temporal contributions to
the local variation of flow, the temporal component is estimated at
only 3.6% [= (8.72
7.92)1/2]. These indicate that regional flow
is temporally stable over 90 s at least.
In group IV, mean perfusion pressure increased from 84 ± 4 to 86 ± 5 mmHg and mean perfusion rate decreased from
8.2 ± 1.4 to 7.7 ± 1.3 ml · min
1 · g
1 due to the
microsphere embolization of 9 ± 2 spheres/mg (wet weight),
resulting in the increase in coronary resistance by 8.6 ± 3.3%.
Heart rate (195 ± 22 beats/min) was not affected by the embolization. The representative results for one myocardial slice are
shown in Fig. 5 (right). The tracer distributions before and after the embolization have a remarkable pattern similarity; control high- or low-flow regions still received high or low flows even after
the embolization. The scatter of density plots is slightly larger than
those in groups II and III; however,
the distributions of two tracers correlated significantly. The
correlation coefficient, the slope of the regression line, CV values of
IDMI and HDMI deposition densities, and the local density difference
are summarized for each heart (means ± SD of 27 slices) in Table
1. As a whole (27 slices each from 8 hearts), these averaged 0.84 ± 0.06, 1.09 ± 0.13, 19.1 ± 4.5, 24.6 ± 5.4, and
10.8 ± 2.4%, respectively. Differences between CV values of IDMI
and HDMI deposition were significant; CV, i.e., flow heterogeneity,
increased after the embolization. Although correlation coefficient
values were significantly lower than those of group II,
distributions of two tracers still correlated linearly to the similar
degree in group III. The slope of the regression
line was significantly higher in group IV than in
groups II and III; accordingly,
regional high and low flows were more sharply contrasted due to the
embolization. The local density difference was significantly higher in
group IV than in groups II and III.
Assuming the independent methodological, temporal, and
microsphere-derivative contributions to the local variation of flow,
the component ascribable to the microembolization is estimated at 6.4%
[= (10.82
7.92
3.62)1/2].
| |
DISCUSSION |
|---|
|
|
|---|
The present study demonstrated that double-tracer digital radiography using IDMI and HDMI made it possible to measure the region-to-region myocardial flow variability on a relative basis at two different points of time, i.e., the spatial and temporal myocardial flow distribution with resolution of 400 × 400 µm2. The regions resolved are comparable in size to vascular regulatory units. The application of this method to isolated rabbit hearts showed that the spatial pattern of flow distribution was fairly stable even at microvascular levels and that the microsphere embolization enhanced flow differences between control high- and control low-flow regions but rather kept the pattern of flow distribution almost as it was.
The microsphere method for intraorgan flow measurements has conflicting problems on statistical and physiological sides. A recent study (31) has demonstrated that even less than 400 spheres per piece are enough to quantitate the flow heterogeneity; nevertheless, a higher-resolution measurement requires a larger number of microspheres per piece. However, the larger the number of microspheres per piece is, the more regional flows are disturbed due to the embolization. In other words, the effect of embolization on flow is augmented as the piece of myocardium decreases. Accordingly, a fairly common practice of using 1-g-order pieces of the myocardium is a compromise, underestimating the true heterogeneity of flows because such pieces are not internally uniform. Although the microsphere method can provide the acceptable accuracy for flow distribution measurements when the regions resolved are diminished down to a 0.1-g piece, even so, the methodological error is two to four times that for the two differently labeled IDMI deposition technique (3, 4).
The delivery of microspheres to small myocardial regions is not governed by flow alone. There are systemic biases in the deposition of microspheres (6, 12). The larger bias, in accordance with the flow proportionality at equivalent binary branch points, is toward preferential or excessive deposition in region of higher flow; the small bias, which caused by preferential mobility of microspheres along straight vessels penetrating the epicardium toward the endocardium, is toward excessive deposition in the subendocardial versus subepicardial regions. This bias, though, will be greater in larger animals because the increased lengths of penetrating vessels offer a greater number of branchings where skimming or separation of spheres and fluid might occur. DMI deposition technique is free from artifacts due to these rheological or geometric effects on delivery of particles into the microvasculture.
The methodological error in the present digital radiography is
attributable mainly to the spread nature of IDMI radiation. In fact,
IDMI radiation is not perfectly equal in all directions, and the decay
of IDMI density deviates somewhat from the assumed circular-symmetrical
spread functions. Furthermore, IDMI digital radiography is under the
influence of IDMI deposition over sample thickness because the IDMI
radiation ranges over several hundred micrometers. In contrast, HDMI is
detected exclusively in a sample surface attached to the imaging plate
(i.e., the TR plate) because of the low radiant energy of
3H (its
-particle path length is ~1 µm on average).
However, the error arising from the radiant energy difference between
two tracers would be small because the ventricular free wall was sliced
most thinly to a 10-µm thickness, and, moreover, the thickness was quite reduced after drying. Decreases in tracer retention also lowered
the accuracy of this method; in this study, however, the periods for
IDMI and HDMI to be exposed to perfusate were 120 and 30 s at
most, respectively. Washout effect over such short periods is
negligible because the retention of the tracers is still nearly 95%.
Despite the evaluation with the considerably fine resolution, the temporal flow fluctuations observed over a 90-s period were almost negligible, as inferred from the temporal flow variation of 3.6% (group III). This indicates that the spatial flow distribution in the myocardium is quite stable over this time period and that microvascular vasomotion or arteriolar twinkling, which is considered to cause temporal flow fluctuations, barely contributes to the spatial flow heterogeneity, as reported previously (37). In this study, however, temporal flow fluctuations may be somewhat underestimated because in crystalloid-perfused Langendorff hearts, microvascular vasomotion might be attenuated (7, 30). The absence of blood cells may also reduce the temporal flow fluctuations. The cell distributions to vessels at branching with a low-flow velocity were reported to be highly time dependent (11), and, actually, the relative temporal fluctuations were greatest in low-flow regions, although diminished in high-flow regions (22).
The microsphere injection increased flow heterogeneity significantly (group IV). The CV value increased by >5% with the local flow variation of 6.4%. However, the regional correlation between normalized local flows before and after the microsphere injection was still significantly high, indicating that the patterns of flow distribution were well maintained and stable even under the microsphere microembolization. Furthermore, that the slope of the regression lines was greater in group IV than groups II and III shows a tendency of flows in control high- and low-flow regions to be more sharply contrasted. The mechanism responsible for this phenomenon remains unproven. However, we postulate a potential mechanism as follows. In control higher-flow regions, the microsphere embolization occurred with higher frequency, inducing greater release of vasodilator agents and, accordingly, greater hyperemic response in those regions. Thus, in high-flow regions, the local hyperemic response would overcome the loss of functioning microvessels; in low-flow regions, however, the latter would prevail against the former. Consequently, the relative flow disparity between high- and low-flow regions would be broadened after the embolization with a slight decrease in perfusion rate. A reduced systolic function in the ischemic regions and an enhanced systolic function in the nonischemic regions (18) may also contribute to the increased relative flow difference by attenuating and enhancing the flow impairment due to the microvessel blockades during systole, respectively.
The microsphere embolization consistently increased the heterogeneity of regional flow distribution; however, this embolizing effect on flow distribution weakens when the flow heterogeneity is evaluated with resolution much lower than the present one. That is, the local flow disturbance caused by the embolization is more dispersed and, accordingly, less discerned in larger sample regions. Actually, when the flow heterogeneity was evaluated with a pixel size of 1,600 × 1,600 µm2, CV values before and after the embolization were 14.2 ± 4.0 and 17.7 ± 4.6, respectively, and the percent differences between CV values slightly but significantly decreased from 25 ± 12%, evaluated with a pixel size of 400 × 400 µm2 to 22 ± 15%. These results affirm the previous studies (3, 4) leading to the conclusion that 16.5-µm microsphere deposition densities, if evaluated with 0.1- to 1-g-order myocardial pieces, provide a fairly good measure of regional myocardial flows.
In the present study, the microsphere embolization decreased perfusion rate, slightly increased perfusion pressure, and, accordingly, increased flow resistance (8.6%). However, in studies with in vivo hearts (14, 16, 17) or blood-perfused isolated hearts (29), such amounts of microspheres as used in this study increased coronary flow through massive release of adenosine or other endogenous vasodilators in the adjacent area of ischemic foci. Such hyperemic response of coronary flow is likely to occur only when coronary flow reserve is sufficiently remained. Actually, repetitive injections of microspheres increased coronary flow as long as the flow reserve was maintained at the control level with no embolization, but, decreased coronary flow after the flow reserve began to decrease progressively (17). That is, whether coronary flow increases or decreases during microsphere injections will depend on the balance between a coronary resistance decrease due to the remaining local hyperemic response and a coronary resistance increase due to the loss of functioning microvessels. In crystalloid-perfused rabbit hearts, however, coronary flow reserve was nearly exhausted (30) or almost a quarter of that in blood-perfused hearts (8). In the present Tyrode-perfused heart experiments, coronary flow reserve would also be reduced considerably, so that hyperemic response due to the embolization could not overcome the loss of functioning microvessels. Consequently, the microsphere embolization did not increase but rather decreased the total perfusion rate. With thoroughly preserved coronary reserve, the microsphere injection would increase the perfusion rate and enhance contrasts between flows in control high- and low-flow regions.
The method developed in this study can be applicable to hearts in situ;
as a matter of fact, this method a promising method for in vivo
measurements of myocardial flow distribution because of the lower
washout effect on tracer escape. Myocardial perfusion rate is mostly
lower in blood-perfused hearts than in crystalloid-perfused hearts,
and, accordingly, the retention of both tracers will be larger in
blood-perfused hearts. Blood perfusion is likely to be more effective
in prolonging the HDMI retention. Indeed, HDMI is better extracted than
IDMI in blood-perfused hearts, although IDMI was highly extracted in
crystalloid-perfused hearts (25). DMI can be expected to
be delivered to the tissue in proportion to local whole blood flow
because it is carried by both plasma and erythrocytes
(25). Errors due to hematocrit changes should be small, as
shown by the same authors. DMI might have its own effect on coronary
flow by diminishing
2-adrenergic activity because
cardiac receptor affinity is 104 higher for DMI than for
catecholamines (32). In the present study, however, there
has been no change in cardiac performance and coronary flow during HDMI
and IDMI injections either continuously over 90 s or through a
bolus. Furthermore, in the previous in vivo studies (27,
28), no changes in both hemodynamics and cardiac performance
were observed during HDMI injections even in the hypoxic state. In
addition, it has been reported that
2-adrenergic vasoconstriction was negligible in the normal heart both at rest and
during exercise (7, 19). Thus
2-antagonistic effects of DMI on regional flow will be
minimal. Therefore, this double-tracer method using DMI will be
suitable to the measurement of the spatial pattern of flow distribution
in the heart or other organs where the extraction and retention of DMI
is complete.
In conclusion, double-tracer digital radiography with HDMI and IDMI will be of great use in myocardial flow distribution analysis at microvascular bed levels under different pathophysiological conditions in the same heart. The application study of this method showed that the spatial pattern of myocardial flow distribution was temporally stable at a microvascular bed level even when subjected to the microsphere embolization. However, the embolization enhanced the relative flow differences between high- and low-flow regions and, accordingly, increased flow heterogeneity.
| |
ACKNOWLEDGEMENTS |
|---|
We thank Chikako Tokuda and Kazue Hirose for expert secretarial assistance.
| |
FOOTNOTES |
|---|
This study was supported by a Japan Heart Foundation and IBM Japan Research Grant (1998), Okayama Science and Technology Association Research Grant (1999-2000), and by Grants-In-Aid for Encouragement of Young Scientists (11770388) from the Ministry of Education, Science, Sports, and Culture, Japan.
Address for reprint requests and other correspondence: T. Matsumoto, Dept. of Medical Engineering and Systems Cardiology, Kawasaki Medical School, 577 Matsushima, Kurashiki, Okayama 701-0192 Japan (E-mail: matsumoto{at}me.kawasaki-m.ac.jp).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 27 June 2000; accepted in final form 15 August 2000.
| |
REFERENCES |
|---|
|
|
|---|
1.
Austin, RE, Jr,
Smedira NG,
Squiers TM,
and
Hoffman JI.
Influence of cardiac contraction and coronary vasomotor tone on regional myocardial blood flow.
Am J Physiol Heart Circ Physiol
266:
H2542-H2553,
1994
2.
Bassingthwaighte, JB,
Liebovitch LS,
and
West BJ.
Intraorgan flow heterogeneities.
In: Fractal Physiology. New York: Oxford University Press, 1994, p. 236-262.
3.
Bassingthwaighte, JB,
Malone MA,
Moffett TC,
King RB,
Chan IS,
Link JM,
and
Krohn KA.
Molecular and particulate depositions for regional myocardial flows in sheep.
Circ Res
66:
1328-1344,
1990
4.
Bassingthwaighte, JB,
Malone MA,
Moffett TC,
King RB,
Little SE,
Link JM,
and
Krohn KA.
Validity of microsphere depositions for regional myocardial flows.
Am J Physiol Heart Circ Physiol
253:
H184-H193,
1987
5.
Brasch, F,
Neckel M,
Volkmann R,
Schmidt G,
Hellige G,
and
Vetterlein F.
Mapping of capillary flow, cellular redox state, and resting membrane potential in hypoperfused rat myocardium.
Am J Physiol Heart Circ Physiol
277:
H2050-H2064,
1999
6.
Chien, S,
Tvetenstrand CD,
Epstein MA,
and
Schmid-Schonbein GW.
Model studies on distributions of blood cells at microvascular bifurcations.
Am J Physiol Heart Circ Physiol
248:
H568-H576,
1985.
7.
Cousineau, D,
Goresky CA,
Bach GG,
and
Rose CP.
Effect of
-adrenergic blockade on in vivo norepinephrine release in canine heart.
Am J Physiol Heart Circ Physiol
246:
H283-H292,
1984
8.
Deng, Q,
Scicli AG,
Lawton C,
and
Silverman NA.
Coronary flow reserve after ischemia and reperfusion of the isolated heart. Divergent results with crystalloid versus blood perfusion.
J Thorac Cardiovasc Surg
109:
466-472,
1995
9.
Deussen, A.
Blood flow heterogeneity in the heart.
Basic Res Cardiol
93:
430-438,
1998[ISI][Medline].
10.
Deussen, A,
Flesche CW,
Lauer T,
Sonntag M,
and
Schrader J.
Spatial heterogeneity of blood flow in the dog heart. II. Temporal stability in response to adrenergic stimulation.
Pflügers Arch
432:
451-461,
1996[ISI][Medline].
11.
Ellis, CG,
Wrigley SM,
and
Groom AC.
Heterogeneity of red blood cell perfusion in capillary networks supplied by a single arteriole in resting skeletal muscle.
Circ Res
75:
357-368,
1994
12.
Fung, YC.
Stochastic flow in capillary blood vessels.
Microvasc Res
5:
34-48,
1973[ISI][Medline].
13.
Groeneveld, AB,
and
Visser FC.
Correlation of heterogeneous blood flow and fatty acid uptake in the normal dog heart.
Basic Res Cardiol
88:
223-232,
1993[ISI][Medline].
14.
Grund, F,
Sommerschild HT,
Lyberg T,
Kirkeboen KA,
and
Ilebekk A.
Microembolization in pigs: effects on coronary blood flow and myocardial ischemic tolerance.
Am J Physiol Heart Circ Physiol
277:
H533-H542,
1999
15.
Hoffman, JI.
Heterogeneity of myocardial blood flow.
Basic Res Cardiol
90:
103-111,
1995[ISI][Medline].
16.
Hori, M,
Inoue M,
Kitakaze M,
Koretsune Y,
Iwai K,
Tamai J,
Ito H,
Kitabatake A,
Sato T,
and
Kamada T.
Role of adenosine in hyperemic response of coronary blood flow in microembolization.
Am J Physiol Heart Circ Physiol
250:
H509-H518,
1986
17.
Hori, M,
Tamai J,
Kitakaze M,
Iwakura K,
Gotoh K,
Iwai K,
Koretsune Y,
Kagiya T,
Kitabatake A,
and
Kamada T.
Adenosine-induced hyperemia attenuates myocardial ischemia in coronary microembolization in dogs.
Am J Physiol Heart Circ Physiol
257:
H244-H251,
1989
18.
Ihara, T,
Komamura K,
Shen YT,
Patrick TA,
Mirsky I,
Shannon RP,
and
Vatner SF.
Left ventricular systolic dysfunction precedes diastolic dysfunction during myocardial ischemia in conscious dogs.
Am J Physiol Heart Circ Physiol
267:
H333-H343,
1994
19.
Ishibashi, Y,
Duncker DJ,
and
Bache RJ.
Endogenous nitric oxide masks
2-adrenergic coronary vasoconstriction during exercise in the ischemic heart.
Circ Res
80:
196-207,
1997
20.
Kassab, GS,
and
Fung YC.
Topology and dimensions of pig coronary capillary network.
Am J Physiol Heart Circ Physiol
267:
H319-H325,
1994
21.
Kassab, GS,
Rider CA,
Tang NJ,
and
Fung YC.
Morphometry of pig coronary arterial trees.
Am J Physiol Heart Circ Physiol
265:
H350-H365,
1993
22.
King, RB,
and
Bassingthwaighte JB.
Temporal fluctuations in regional myocardial flows.
Pflügers Arch
413:
336-342,
1989[ISI][Medline].
23.
King, RB,
Bassingthwaighte JB,
Hales JR,
and
Rowell LB.
Stability of heterogeneity of myocardial blood flow in normal awake baboons.
Circ Res
57:
285-295,
1985
24.
Little, SE,
and
Bassingthwaighte JB.
Plasma-soluble marker for intraorgan regional flows.
Am J Physiol Heart Circ Physiol
245:
H707-H712,
1983.
25.
Little, SE,
Link JM,
Krohn KA,
and
Bassingthwaighte JB.
Myocardial extraction and retention of 2-iododesmethylimipramine: a novel flow marker.
Am J Physiol Heart Circ Physiol
250:
H1060-H1070,
1986.
26.
Loncar, R,
Flesche CW,
and
Deussen A.
Coronary reserve of high- and low-flow regions in the dog heart left ventricle.
Circulation
98:
262-270,
1998
27.
Matsumoto, T,
Ebata J,
Tachibana H,
Goto M,
and
Kajiya F.
Transmural microcirculatory blood flow distribution in right and left ventricular free walls of rabbits.
Am J Physiol Heart Circ Physiol
277:
H183-H191,
1999
28.
Matsumoto, T,
Goto M,
Tachibana H,
Ogasawara Y,
Tsujioka K,
and
Kajiya F.
Microheterogeneity of myocardial blood flow in rabbit hearts during normoxic and hypoxic states.
Am J Physiol Heart Circ Physiol
270:
H435-H441,
1996
29.
Monroe, RG,
LaFarge CG,
Gamble WJ,
Kumar AE,
and
Manasek FJ.
Left ventricular performance and coronary flow after coronary embolization with plastic microspheres.
J Clin Invest
50:
1656-1665,
1971.
30.
Podesser, BK,
Hallstrom S,
Schima H,
Huber L,
Weisser J,
Kroner A,
Furst W,
and
Wolner E.
The erythrocyte-perfused "working heart" model: hemodynamic and metabolic performance compared with crystalloid perfused hearts.
J Pharmacol Toxicol Methods
41:
9-15,
1999[ISI][Medline].
31.
Polissar, NL,
Stanford DC,
and
Glenny RW.
The 400 microsphere per piece "rule" does not apply to all blood flow studies.
Am J Physiol Heart Circ Physiol
278:
H16-H25,
2000
32.
Raisman, R,
Sette M,
Pimoule C,
Briley M,
and
Langer SZ.
High-affinity [3H]desipramine binding in the peripheral and central nervous system: a specific site associated with the neuronal uptake of noradrenaline.
Eur J Pharmacol
78:
345-351,
1982[ISI][Medline].
33.
Ritman, EL.
Temporospatial heterogeneity of myocardial perfusion and blood volume in the porcine heart wall.
Ann Biomed Eng
26:
519-525,
1998[ISI][Medline].
34.
Sonntag, M,
Deussen A,
Schultz J,
Loncar R,
Hort W,
and
Schrader J.
Spatial heterogeneity of blood flow in the dog heart. I. Glucose uptake, free adenosine and oxidative/glycolytic enzyme activity.
Pflügers Arch
432:
439-450,
1996[ISI][Medline].
35.
Stapleton, DD,
Moffett TC,
Baskin DG,
and
Bassingthwaighte JB.
Autoradiographic assessment of blood flow heterogeneity in the hamster heart.
Microcirculation
2:
277-282,
1995[Medline].
36.
VanBeek, JH,
vanMil HG,
King RB,
deKanter FJ,
Alders DJ,
and
Bussemaker J.
A 13C NMR double-labeling method to quantitate local myocardial O2 consumption using frozen tissue samples.
Am J Physiol Heart Circ Physiol
277:
H1630-H1640,
1999
37.
Yipintsoi, T,
Dobbs WA, Jr,
Scanlon PD,
Knopp TJ,
and
Bassingthwaighte JB.
Regional distribution of diffusible tracers and carbonized microspheres in the left ventricle of isolated dog hearts.
Circ Res
33:
573-587,
1973
38.
Zuurbier, CJ,
van Iterson M,
and
Ince C.
Functional heterogeneity of oxygen supply-consumption ratio in the heart.
Cardiovasc Res
44:
488-497,
1999
This article has been cited by other articles:
![]() |
T. Matsumoto, H. Tachibana, T. Asano, M. Takemoto, Y. Ogasawara, K. Umetani, and F. Kajiya Pattern differences between distributions of microregional myocardial flows in crystalloid- and blood-perfused rat hearts Am J Physiol Heart Circ Physiol, April 1, 2004; 286(4): H1331 - H1338. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |