Vol. 280, Issue 6, H2936-H2943, June 2001
Assessment of the time constant of relaxation: insights from
simulations and hemodynamic measurements
S.
De Mey1,
J. D.
Thomas2,
N. L.
Greenberg2,
P. M.
Vandervoort3, and
P. R.
Verdonck1
1 Institute Biomedical Technology, Ghent University, 9000 Gent, Belgium; 2 Cardiovascular Imaging Center, Cleveland
Clinic Foundation, Cleveland, Ohio 44195; and 3 Heart Center
Limburg, 3600 Genk, Belgium
 |
ABSTRACT |
The objective of this
study was to use high-fidelity animal data and numerical simulations to
gain more insight into the reliability of the estimated relaxation
constant derived from left ventricular pressure decays, assuming a
monoexponential model with either a fixed zero or free moving pressure
asymptote. Comparison of the experimental data with the results of the
simulations demonstrated a trade off between the fixed zero and the
free moving asymptote approach. The latter method more closely fits the
pressure curves and has the advantage of producing an extra coefficient
with potential diagnostic information. On the other hand, this method
suffers from larger standard errors on the estimated coefficients. The method with fixed zero asymptote produces values of the time constant of isovolumetric relaxation (
) within a narrow confidence interval. However, if the pressure curve is actually decaying to a nonzero pressure asymptote, this method results in an inferior fit of the
pressure curve and a biased estimation of
.
hemodynamics; left ventricular relaxation constant; simulation
 |
INTRODUCTION |
THE ABILITY TO
QUANTIFY the left ventricular (LV) relaxation rate in normal and
pathological conditions is important in investigating myocardial pump
function. Despite advances in noninvasive assessment of relaxation
(7), invasive measurement of the ventricular relaxation
rate during isovolumic relaxation remains the "gold standard." Such
invasive parameters include the first derivative of LV pressure with
respect to time during isovolumic relaxation (dP/dt) and the
time constant of isovolumic relaxation (
).
The time course of the fall in LV pressure during isovolumic relaxation
has been modeled using a monoexponential function with three
parameters, described in Eq. 1 as follows
|
(1)
|
where P(t) is LV pressure as a function of time (in
mmHg), t is time (in ms), P
is the asymptote
to which LV pressure declines (in mmHg), and P0 is LV
pressure (in mmHg) at peak negative dP/dt (where
t = 0 ms). Measured pressure data during isovolumetric pressure decay is fitted to this model to obtain an estimation of
(in ms).
Initially, Eq. 1 was linearized to avoid difficult
calculation. Weiss et al. (1, 13) further simplified the
situation by assuming a zero asymptote, yielding the reduction of a
three-parameter monoexponential model to a two-parameter model
|
(2)
|
The assumption of a zero asymptote allows linearization by taking
the natural logarithm of both sides of Eq. 2, from which linear regression analysis can be used to determine the least mean
squared error (MSE) solution for
and P0. Although
useful physiological insight has been gained from this approach, a
disadvantage of the logarithmic transformation is to give undue
weighting to data points (and noise) at low pressures.
A refinement of Weiss' log transformation approach is to substitute
the differentiated monoexponential function back into Eq. 1,
which allows linearization without the assumption of a zero asymptote
(9). Again, linear regression analysis can be used to
obtain an estimate of
, but the differentiation process is very
sensitive to noise in the signal.
The improved performance of contemporary computer hardware and software
allows direct solution of Eq. 1 using nonlinear least squares parameter estimation techniques, most commonly the
Levenberg-Marquardt method. This nonlinear technique allows for an
accurate estimation of P0 and
both with and without the
assumption of a zero asymptote. The use of this nonlinear technique for
calculation of
was initially validated by Bernardi et al.
(1).
With nonlinear techniques widely available, they have largely
superceded both the log transform and differentiation methods for
solving Eq. 1. Nevertheless, the issue of choosing a two- or
three-parameter model is still an open question because conflicting results have been reported when using the three-parameter model (free
moving asymptote) versus the two-parameter model (zero pressure asymptote assumed) for calculation of
. A key issue is the trade off
between accuracy of fit to the observed data (which should be better
with three parameters) and the confidence intervals of the derived
parameters (which may worsen with three parameters if there is
significant collinearity between them). The objective of this study,
therefore, was to use high-fidelity animal data and numerical
simulations to gain more insight into the reliability of the estimated
relaxation constant when assuming either a zero or free moving pressure asymptote.
 |
METHODS |
LV pressure decay data obtained from both an animal experiment
and Monte Carlo simulations were analyzed.
was determined using the
nonlinear Levenberg-Marquardt technique both with (two-parameter exponential model; LM2) and without (three-parameter
exponential model; LM3) the assumption of a zero pressure
asymptote. The differences between the two approaches were then
compared for goodness of fit to the pressure curves and the confidence
intervals of the estimated
.
Animal experiment.
The investigation conformed to the Guide for the Care and Use of
Laboratory Animals (8) published by the National
Institutes of Health and was approved by the Animal Research Committee
of the Cleveland Clinic Foundation. Eight healthy adult mongrel dogs of
either sex weighing 29.7 ± 7.4 kg were studied. The dogs were anesthetized with 25 mg/kg intravenous pentobarbital sodium, and anesthesia was maintained throughout the experiments with additional aliquots of pentobarbital sodium. After the dogs underwent tracheal intubation, positive pressure mechanical ventilation was instituted using room air. A micromanometer catheter (Millar; Houston, TX) was
introduced into the left atrium (LA) through the LA appendage and
positioned across the mitral valve with the pressure sensor in the LV.
LA pressure was recorded by an additional single sensor catheter.
Pressure and electrocardiogram signals were digitally acquired with
1-ms/12-bit resolution using a multifunction input-output board
(AT-MIO-16, National Instruments; Austin, TX) interfaced with a
computer workstation (Intel 80486 PC) using customized software
developed using LabView version 5.0 (National Instruments). Data acquisition was performed at baseline (for each dog experiment), during isoproterenol infusion (for 6 dog experiments), and during esmolol infusion (for 3 dog experiments). Baseline runs were initiated after allowing sufficient time for hemodynamics to stabilize before starting the experiment. Esmolol or isoproterenol medication runs were
initiated after completion of a satisfactory number of baseline acquisition runs. Isoproterenol was infused at 0.025-0.4
µg · kg
1 · min
1
intravenously, and data acquisition runs were initiated after sufficient washin time for an appropriate heart rate response and
hemodynamics to stabilize. Similarly, esmolol was infused at
0.2-0.3
mg · kg
1 · min
1
intravenously, with data acquisition after hemodynamic stabilization. For the eight dogs, 45 recordings during baseline, 12 recordings during
isoproterenol infusion, and 8 recordings during esmolol infusion were
registered, with each recording containing ~7 consecutive heartbeats.
We thus captured 340, 94, and 56 pressure decays at baseline and during
isoproterenol and esmolol infusion, respectively. Post acquisition
numerical analysis of raw pressure data was performed using another
custom numerical analysis program developed in LabView. In this study,
dP/dt was calculated to define the period of isovolumic relaxation as the time period between maximum negative pressure change
and the first LA to LV pressure crossover.
was determined from the
pressure curves with the use of the nonlinear Levenburg-Marquardt technique both with and without the assumption of a zero pressure asymptote. With each derivation of the coefficients, a MSE value was
calculated as a measure of "goodness-of-fit" of the specific model
to the pressure data (Eq. 3) as follows
|
(3)
|
In this equation, (xi,
yi) are the input data points, f
(xi;
a1... aM) = f (X, A) is the nonlinear function
(where a1... aM are coefficients), N the number of input data points, and
i the variance. To analyze the efficiency of the
different models in estimating
, the accompanying standard error for
each derivation of
was calculated. Method-dependent differences
were analyzed.
Monte Carlo simulation.
One hundred instances of 125 different diastolic pressure curves were
created with the use of Monte Carlo simulation in the following manner.
First, an exact monoexponential curve was constructed using Eq. 1 with the coefficients P0 = 70 mmHg,
= 60 ms, and P
= 0 mmHg. By adding Gaussian noise
(mean value 0 mmHg and SD 0.4 mmHg) randomly, 100 "data curves"
were created from this exact monoexponential pressure decay. The
simulation of one pressure curve is illustrated in Fig.
1, showing the exact monoexponential curve (A), the Gaussian noise (B), and the
simulated curve (C). From each of the 100 data sets,
and
P0 were estimated using LM2, and
,
P0, and P
were estimated using
LM3. With each derivation of the coefficients, the MSE
value (Eq. 3) was calculated as a measure of
goodness-of-fit of the specific model to the simulated pressure data.
The estimated coefficients from the 100 data curves with the
accompanying standard error were compared with the actual coefficients
that produced the original pressure curve. To analyze a range of
parameter values, the simulation was repeated with P0
varying from 70 to 110 mmHg in steps of 10 mmHg,
varying from 40 to
120 ms in steps of 20 ms, and P
varying from
5 to +5
mmHg in steps of 2.5 mmHg. Thus a total of 5 × 5 × 5 = 125 combinations of the coefficients are simulated, yielding the
analysis of 12,500 pressure curves. The results were compared with the
findings of the dog experiment.

View larger version (16K):
[in this window]
[in a new window]
|
Fig. 1.
Monte Carlo simulation of a pressure curve: exact
monoexponential curve (A), Gaussian noise (B),
and simulated monoexponential curve containing noise (C).
|
|
Statistics.
All statistics were performed using SPSS version 9.0 (Chicago, IL).
Values are means ± SD. Normally distributed variables, calculated
using the different models, were compared using repeated-measures ANOVA. Post hoc testing was performed using either a Bonferroni t-test (equal variances assumed) or a Dunnett's
t-test (equal variances not assumed). Nonnormally
distributed variables were compared using the nonparametric Friedman
test for related variables. The level of significance was set at a
P value of 0.05.
 |
RESULTS |
Animal experiment.
Table 1 summarizes the results of the
determinations of
in the dog experiment using the nonlinear
Levenburg-Marquardt method with a zero pressure asymptote
(LM2) and a nonzero moving asymptote (LM3) for
the baseline measurements and during infusion of isoproterenol and
esmolol. The mean values for
are the result of averaging the
calculated
values obtained from LV pressure recordings using LM2 and LM3. Method-dependent differences were
observed. At baseline,
values obtained with LM2 were
consistently shorter than the values obtained with LM3
(P < 0.001), whereas during either isoproterenol or
esmolol infusion,
values obtained with LM2 were
consistently higher than the values obtained with LM3.
LM3 most closely fits the pressure decays, as reflected by
the significant difference in MSE between original and fitted pressure
decays at baseline and during isoproterenol and esmolol infusion
(P < 0.001). In contrast, however, the standard error
of the estimate was significantly higher when
was calculated using
LM3 compared with LM2 at baseline as well as
during isoproterenol and esmolol infusion (P < 0.001). LM3 showed that P
significantly increases
during isoproterenol and esmolol infusion compared with baseline
values. For both LM2 and LM3,
decreased
with isoproterenol and increased with esmolol infusion
(P < 0.001). However, the relative change compared
with baseline values for both isoproterenol and esmolol infusion was different when using LM2 compared with LM3. For
LM2, the infusion of isoproterenol resulted in a decrease
of
of 26%, whereas for LM3 the decrease of
was
42%. With the use of LM2, the infusion of esmolol resulted
in an increase of
of 67%, whereas LM3 resulted in an
increase of
of 30%.
View this table:
[in this window]
[in a new window]
|
Table 1.
Estimation of using a fixed zero asymptote monoexponential model
(LM2) and a free moving monoexponential model
(LM3) under baseline conditions, during isoproterenol
infusion, and during esmolol infusion
|
|
Monte Carlo simulation.
Table 2 shows the results from a
representative 2 of 125 pressure simulations (
= 60 ms,
P0 = 70 mmHg, and P
= 0 mmHg, and
= 60 ms, P0 = 70 mmHg, and
P
=
2.5 mmHg). In these and all other
simulations, the standard errors of the
estimates were
significantly smaller for the zero asymptote model (LM2)
compared with the moving asymptote model (LM3)
(P < 0.001).
View this table:
[in this window]
[in a new window]
|
Table 2.
Monte Carlo simulation of the estimation of from pressure curves
with coefficients of = 60 ms, P0 = 70 mmHg, and P = 0 mmHg and = 60 ms, P0 = 70 mmHg, and P = 2.5 mmHg
|
|
First,
, P0 and P
were determined from
100 pressure decays created with Monte-Carlo simulation starting from a
monoexponential curve using the Levenburg-Marquardt technique with a
fixed zero asymptote (LM2) and a free-moving asymptote
(LM3). The results are shown in Table 2. Both methods had a
comparable MSE. For each method, the mean values of the calculated
coefficients were a good approximation of the exact coefficients.
However, the estimate of
calculated using LM3 had a
larger SE (P < 0.001). Figure 2A shows the regression
between the calculated
and the calculated P0 for both
LM2 and LM3. From this graph, it is obvious
that, for this particular simulation, the nonlinear method with fixed zero asymptote (LM2) approach is the better one because
this approach provides the smallest confidence interval on the
estimated coefficients.

View larger version (21K):
[in this window]
[in a new window]
|
Fig. 2.
Regression of left ventricular (LV) pressure at peak
negative first derivative of LV pressure with respect to time during
isovolumetric relaxation (P0) and the time constant of
isolumetric relaxation ( ) obtained from the analysis of 100 Monte
Carlo simulations of an exponential pressure decay with coefficients of
= 60 ms, P0 = 70 mmHg, and the asymptote at
which LV pressure declines (P ) = 0 mmHg
(A) and = 60 ms, P0 = 70 mmHg, and
P = 2.5 mmHg (B). The analysis was
performed using a Levenberg-Marquardt model with two parameters
(LM2; solid lines, regression line and 99% prediction
interval on the estimates) and three parameters (LM3;
dashed line, regression line and 99% prediction interval).
|
|
A second Monte-Carlo simulation was done starting from a
monoexponential curve with a negative pressure asymptote
(P
=
2.5 mmHg). Again
, P0, and
P
were estimated using LM2 and
LM3 (cf. Table 2). In this second simulation,
LM3 had the smallest MSE (P < 0.001).
Moreover, the mean values of the calculated coefficients, estimated
using LM3, were good approximations of the exact
coefficients. The standard error on the estimated values was comparable
to the standard error accompanying this method in the first simulation.
By analogy with the first simulation, using LM2 resulted in
a significantly lower standard error for
. However, LM2
significantly underestimated the exact values of
, as illustrated in
Fig. 2B, which shows the regression between the calculated
and the calculated P0 for both LM2 and
LM3. Simulation of pressure curves with a positive instead
of a negative pressure asymptote revealed an overestimation instead of
an underestimation of
when using the fixed zero asymptote approach.
The trade off between the magnitude of the variance on the estimated
coefficients versus under/overestimation of the exact values is
demonstrated in Fig. 3 at
P0 = 70 mmHg and reference values of
= 40 ms
(A), 80 ms (B), and 120 ms (C). The
plots are illustrating the method-dependent sensitivity of the
estimates (
± SD) for variations in P
between 0 and
5 mmHg. In the case of a zero asymptote, both methods estimate
well because
do not significantly differ from the reference
values (P > 0.05). LM2 had the smallest
standard deviation compared with LM3 (P < 0.001). However, with increasing absolute values of the pressure asymptote, the values obtained using LM2 were moving away
from the exact values. Independent of the magnitude of the pressure asymptote, LM2 kept the smallest confidence interval
(P < 0.001).

View larger version (29K):
[in this window]
[in a new window]
|
Fig. 3.
Monte Carlo simulation of the influence of the magnitude
of the actual pressure asymptote on the estimation of using the
fixed zero asymptote approach (LM2) and the free moving
asymptote method (LM3). The estimated values (±SD) of using LM2 (open bars) and LM3 (hatched bars)
for values of the pressure asymptote = 0, 2.5, and 5 mmHg are
plotted for reference values of = 40 ms (A), 80 ms
(B), and 120 ms (C).
|
|
To evaluate the efficiency of the two different estimators of
, on
the basis of LM2 and LM3, respectively, a MSE
value was calculated as MSE = (Variance + Bias2)
for each estimator (14). This MSE was similar to the
variance on an estimated coefficient except that it was measured around the true target rather than around the (possibly biased) mean of the
estimator. Formally, we can compare two estimators by calculating the
relative efficiency (RE) as the proportion of the two MSE values. The
RE values comparing LM2 and LM3 for estimating
are plotted in Fig. 4 for pressure
asymptotes varying from
5 to +5 mmHg and
values ranging from 40 to 120 ms. Values for RE > 1 indicated a superior estimate of
when using LM3. Thus, in case of a zero pressure asymptote,
LM2 is always closer at estimating
. However, with an
increasing positive or negative pressure asymptote and decreasing
,
LM3 becomes superior.

View larger version (23K):
[in this window]
[in a new window]
|
Fig. 4.
Monte Carlo simulation of the relative efficiency (RE;
y-axis) of the estimators of using LM2 and
LM3 for P0 = 70 mmHg, P
varying from 5 to +5 mmHg (x-axis), and varying from
40 to 120 ms. Values for RE > 1 are an indication for a superior
estimate of when using LM3.
|
|
 |
DISCUSSION |
In this study, we exclusively used nonlinear techniques for
estimation of
from pressure decays using a monoexponential model: 1) the nonlinear Levenberg-Marquardt method with a fixed
zero pressure asymptote (LM2), and 2) the
Levenberg-Marquardt method with a variable pressure asymptote
(LM3). The two methods were chosen to observe for any
divergence in the resultant
.
Overall, both methods determined comparable values of
for the data
collected. The average values of the MSE when using LM3 were smaller compared with the values obtained when using
LM2. The smaller MSE reflects a superior goodness-of-fit of
LM3 of modeling the experimental data compared with
LM2. The inferior fitting when assuming a fixed zero
pressure asymptote not only provokes a larger MSE but also has an
important consequence on the estimated
values: whereas
LM3 always provides an unbiased estimation of
,
LM2 results in a biased estimation of
when analyzing
pressure decays with a nonzero pressure asymptote. The Monte Carlo
simulation showed that in the case of a negative (positive) pressure
asymptote, using LM2 leads to a significant underestimation (overestimation) of the exact coefficients. In the animal experiment, under baseline conditions, the
values obtained with LM2
were smaller compared with the values obtained using LM3.
In contrast, during either isoproterenol or esmolol infusion,
values obtained with LM3 were smaller compared with the
values obtained using LM2. According to the Monte Carlo
simulations, this suggests a negative pressure asymptote and a
significant underestimation of
with LM2 for baseline
conditions and a positive pressure asymptote and a significant
overestimation of
for isoproterenol and esmolol. This was indeed
confirmed by the values of the calculated pressure asymptotes using
LM3 (cf. Table 1). Figure 5
shows the relationship between the actual pressure asymptote and
under/overestimation of the
values using LM2 in more
detail. An excellent correlation (r2 = 0.92, P < 0.001) was observed between the difference
in
calculated with LM2 and LM3 and the
pressure asymptote as calculated using LM3 for the animal
data under baseline and during isoproterenol and esmolol infusion. This
correlation explains the differences in relative change of
during
drug infusion when using LM2 compared with LM3.

View larger version (31K):
[in this window]
[in a new window]
|
Fig. 5.
Regression between the method-dependent difference in and the pressure asymptote in the animal study. Lines: 99% confidence
interval for the mean and 99% prediction interval for an individual
difference in estimates.
|
|
In contrast to the superior MSE and the unbiased estimation of
, a
drawback of LM3 is the larger standard error on the
estimated coefficients. This is primarily the consequence of the
increased degree of freedom and error propagation in the algorithm for
determining
when using a free moving asymptote. The trade off
between the magnitude of the standard error on the estimates (smaller
when using LM2) and the closer fit that guarantees an
unbiased estimation (when using LM3) can be evaluated
quantitatively by calculation of a MSE that combines the bias and
standard error on the estimates. This calculation demonstrates for the
simulated data that in the case of a zero pressure asymptote,
LM2 always has a closer estimate of
. However, with an
increasing positive or negative pressure asymptote and decreasing
,
LM3 becomes superior. This is also observed in the animal
data. During drug infusion, the pressure asymptotes are small, and,
consequently, the bias on the estimates is small. The
values
obtained using either LM2 or LM3 did not significantly differ. In contrast, the standard error of the estimates was significantly smaller when using LM2. At baseline, a
significant negative pressure asymptote was found and, as expected, the
bias on the estimates was large.
values obtained using either
LM2 or LM3 were significantly different. The
standard error of the estimates remained higher when using
LM3 compared with LM2. With the use of the
results of Table 1, we calculated the mean relative efficiency for the
different groups. For the baseline data, the relative efficiency was
20.70, indicating that LM3 provides the most reliable
estimate. During isoproterenol and esmolol infusion, the mean relative
efficiency became 0.22 and 0.85, which indicates that LM2
provides the most reliable estimate.
The problem of choosing a two (fixed zero pressure asymptote assumed)-
or three (free moving asymptote assumed)-parameter model is still a
matter of debate because conflicting results are reported. Several
authors (1, 5, 11, 12) have demonstrated the use of a
variable asymptote to be a more rigorous and physiologically rational
method of modeling LV pressure decline during the isovolumic relaxation
period. Bernardi and associates (1) demonstrated that the
Levenberg-Marquardt algorithm with a variable asymptote is a most
accurate method for modeling LV pressure decline during the isovolumic
relaxation periods. Martin and colleagues (5) demonstrated
that a variable asymptote method of determining
was more sensitive
to
-adrenergic blockade or stimulation than to drugs that altered
cardiac loading conditions.
On the other hand, Yellin and colleagues (16) demonstrated
that
determined from an exponential model using a fixed asymptote method is comparable with
determined from an exponential model using a measured or best-fit asymptote. Yellin and colleagues (16) further concluded that as long as it is consistently
used in the same study,
resulting from any method provides useful information related to diastolic function. Also, Kettunen and colleagues (4) advocate the use of a fixed zero asymptote
method for practical clinical use to determine
. This recommendation is based on their observation that
determined using a fixed asymptote method is comparable with
determined using a variable asymptote method with the exception of conditions of
-adrenergic blockade or stimulation. The zero asymptote method was advocated on the
basis of a less complicated mathematical algorithm for most practical
clinical purposes. Yamakado et al. (15) calculated
with and without a pressure asymptote to investigate the influence of
age on ventricular relaxation. No significant differences between the
different approaches were observed. In contrast, Davis et al.
(3) obtained opposite conclusions when analyzing
ventricular relaxation rate using the zero or nonzero asymptote model.
Despite closer fitting of the pressure curves when using
LM3 and despite the bias accompanying LM2,
estimates with LM2 may show better correlations with other
physiological parameters than LM3. This is demonstrated for
the baseline animal data in Fig. 6,
showing the regression between maximum negative dP/dt and
when using LM2 (A) and LM3
(B). Although LM3 better fit the pressure decays, the better correlation was obtained using LM2
(r2 = 0.51 vs.
r2 = 0.45). We speculate that this
phenomenon is due to the biased estimates of
with LM2
for the analysis of nonzero asymptote pressure decays. This leads to
over- or underestimation of
and thus to a broader range of
values, automatically enhancing the correlation.

View larger version (20K):
[in this window]
[in a new window]
|
Fig. 6.
Regression between and the maximum negative LV
pressure change ( dP/dtmax). was determined
using LM2 (A) and LM3
(B). Open squares: dP/dtmax vs. for baseline animal data. Lines: 99% confidence intervals for the mean
and 99% prediction intervals for an individual
dP/dtmax estimates.
|
|
In previous studies, linear methods were used to determine
. We also
calculated
values under baseline conditions with linear regression,
assuming a fixed zero pressure asymptote as proposed by Weiss et al.
(13) and Nagueh et al. (7). The results were compared with the results obtained using the Levenburg-Marquardt technique with zero pressure asymptote (LM2). The
values obtained using LM2 were slightly, although
significantly, larger (63.85 ± 17.12 vs. 67.83 ± 15.47 ms,
P < 0.001) compared with the linear method.
LM2 also provided closer fits to the pressure decays, as
reflected by the smaller MSE (2.59 ± 2.06 vs. 0.90 ± 0.75, P < 0.001).
It is generally accepted that the isovolumic relaxation period of the
LV pressure curve is well approximated by the monoexponential decay
model described by Eq. 1 (1, 3-5, 9, 11-13,
15, 16) except in case of a postextrasystolic LV isovolumic
pressure decay (2) and dilated cardiomyopathy
(10). Alternative models have been proposed for fitting
the isovolumic pressure decay (6, 10). Recently, Senzaki
et al. (10) reported an improvement of quantitative
analyses when using the following more complex hybrid logistic model of
Eq. 4
|
(4)
|
The hybrid logistic model provided more consistent data fits,
especially in dilated cardiomyopathy, when a nonlinear relationship between dP/dt and P was observed. We also fitted this model
to the pressure decays of the animal study. In accordance with the results reported by Senzaki et al. (10), the
values
obtained using this model (44.29 ± 5.38 ms) were significantly
smaller compared with the values obtained with the other methods. Also, the mean pressure asymptote remained positive (1.82 ± 2.79 mmHg). Assuming an exponential model, the physical meaning of the
value is
the time needed for the pressure to decrease to 37% of its initial
value. In contrast, when assuming a hybrid logistic model, the physical
meaning of
is the time needed for the pressure to decrease to 54%
of its initial value. Therefore, the hybrid logistic function provides
values of another magnitude compared with the monoexponential
function. Thus comparing values obtained using the different models is
difficult. The MSE (0.45 ± 0.35) was significantly
(P < 0.001) larger compared with the values obtained
using LM3 (0.17 ± 0.33). Thus, in this animal
experiment, the monoexponential model provides the closer fit. The
standard error of the estimated
(1.58 ± 1.60 ms) was,
however, smaller (P < 0.001) compared with the
standard error when using LM3 (3.44 ± 1.85 ms).
Therefore, the hybrid logistic method might be a valuable alternative
for LM3, especially for pressure decays with a nonlinear relationship between dP/dt and P.
In conclusion, in this study, Monte Carlo simulations of
monoexponential pressure decays provided a reference, allowing an objective comparison of different methods for estimation of the relaxation constant
of LV pressure fall. Comparison of the
experimental data with the results of the Monte Carlo simulations
demonstrated a trade off between the nonlinear Levenburg-Marquardt
fixed zero approach on one hand and the nonlinear Levenburg-Marquardt
method with a free moving asymptote on the other hand. The latter
method closer fits the pressure curve and has the advantage of
producing an extra coefficient (P
) with potential
diagnostic information. On the other hand, this method suffers from
larger standard errors on the estimated coefficients. The nonlinear
Levenburg-Marquardt method with fixed zero asymptote produces values of
within a narrow confidence interval. However, in case of a nonzero
negative (positive) pressure asymptote, this method significantly
underestimates (overestimates) the real values. Quantitative evaluation
of the trade off between bias (when using LM2) and the
magnitude of the standard error on the estimates (larger when using
LM3) demonstrates that, in case of a zero pressure
asymptote, LM2 always has a closer estimate of
.
However, LM3 becomes superior with an increasing pressure
asymptote (both positive or negative) and decreasing
.
 |
ACKNOWLEDGEMENTS |
We thank P. Segers for critically reviewing the manuscript.
 |
FOOTNOTES |
S. De Mey was a recipient of Grant IWT-971096 from the Flemish
Institute for the Promotion of Scientific-Technological Research in the
Industry. This study was also supported in part by National Aeronautics
and Space Administration Grant NCC 9-60 (to J. D. Thomas) and by
National Heart, Lung, and Blood Institute Grant R01-HL-56688-01A1 (to
J. D. Thomas).
Address for reprint requests and other correspondence: S. De
Mey, Hydraulics Laboratory, St.-Pietersnieuwstraat 41, 9000 Gent, Belgium (E-mail: stefaan.demey{at}navier.rug.ac.be).
The costs of publication of this
article were defrayed in part by the
payment of page charges. The article
must therefore be hereby marked
"advertisement"
in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 16 August 2000; accepted in final form 23 January 2001.
 |
REFERENCES |
1.
Bernardi, L,
Uretsky B,
Reddy P,
and
Boudereau R.
Modeling the isovolumic relaxation period.
Cathet Cardiovasc Diagn
11:
255-268,
1985[ISI][Medline].
2.
Courtois, M,
Barzilai B,
Hall A,
and
Ludbrook P.
Postextrasystolic left ventricular isovolumic pressure decay is not monoexponential.
Cardiovasc Res
35:
206-216,
1997[Abstract/Free Full Text].
3.
Davis, KLMU,
Schertel ER,
Geissler HJ,
Trevas D,
Laine GA,
and
Allen SJ.
Variation in tau, the time constant for isovolumic relaxation, along the left ventricular base-to-apex axis.
Basic Res Cardiol
94:
41-48,
1999[ISI][Medline].
4.
Kettunen, R,
Timisjarvi J,
Ramo P,
Kouvalainen E,
Heikkila J,
and
Hirvonen L.
Time constant of isovolumic pressure fall in the intact canine left ventricle.
Cardiovasc Res
20:
698-704,
1986[ISI][Medline].
5.
Martin, G,
Gimeno J,
Cosin J,
and
Guillem M.
Time constant of isovolumic pressure fall: new numerical approaches and significance.
Am J Physiol Heart Circ Physiol
247:
H283-H294,
1984.
6.
Matsubara, H,
Araki J,
Takaki M,
Nakagawa S,
and
Suga H.
Logistic time constant of isovolumic relaxation pressure-time curve in the canine left ventricle. Better alternative to exponential time constant.
Circulation
92:
2318-2326,
1995[Abstract/Free Full Text].
7.
Nagueh, S,
Middleton K,
Kopelen H,
Zoghbi W,
and
Quinones M.
Doppler tissue imaging: a noninvasive technique for evaluation of left ventricular relaxation and estimation of filling pressures.
J Am Coll Cardiol
30:
1527-1533,
1997[Abstract].
8.
National Institutes of Health (NIH). Guide for the Care and
Use of Laboratory Animals. [DHHS Publication No. (NIH) 85-23, Revised 1996, Office of Science and Health Reports, Bethesda, MD
20892].
9.
Raff, G,
and
Glantz S.
Volume loading slows left ventricular isovolumic relaxation rate: evidence of load-dependent relaxation in the intact dog heart.
Circ Res
48:
813-824,
1981[Free Full Text].
10.
Senzaki, H,
Fetics B,
Chen C,
and
Kass D.
Comparison of ventricular pressure relaxation assessments in human heart failure.
J Am Coll Cardiol
34:
1529-1536,
1999[Abstract/Free Full Text].
11.
Takeuchi, M,
Fujitani K,
Kurogane K,
Bai HT,
Toda C,
Yamasaki T,
Takano S,
and
Fukuzaki H.
A comparison of two exponential models of the time constant during left ventricular isovolumic pressure decay in coronary artery disease.
Jpn Circ J
49:
1225-1234,
1985[Medline].
12.
Thompson, D,
Waldron C,
Coltart D,
Jenkins B,
and
Webb-Peploe M.
Estimation of time constant of left ventricular relaxation.
Br Heart J
49:
250-258,
1983[Free Full Text].
13.
Weiss, J,
Frederiksen J,
and
Weisfeldt M.
Hemodynamic determinants of the time-course of fall in canine left ventricular pressure.
J Clin Invest
58:
751-760,
1976.
14.
Wonnacott, TH,
and
Wonnacott RJ.
Introductory Statistics for Business and Economics (4th Ed.). New York: Wiley, 1999.
15.
Yamakado, T,
Takagi E,
Okubo S,
Imanaka Y,
Tarumi T,
Nakamura M,
and
Nakano T.
Effects of ageing on left ventricular relaxation in humans. Analysis of left ventricular isovolumic pressure decay.
Circulation
95:
917-923,
1997[Abstract/Free Full Text].
16.
Yellin, E,
Hori M,
Yoran C,
Sonnenblick E,
Gabbay S,
and
Frater R.
Left ventricular relaxation in the filling and nonfilling intact canine heart.
Am J Physiol Heart Circ Physiol
250:
H620-H629,
1986[Abstract/Free Full Text].
Am J Physiol Heart Circ Physiol 280(6):H2936-H2943
0363-6135/01 $5.00
Copyright © 2001 the American Physiological Society