|
|
||||||||
Department of Cardiology, Leiden University Medical Center, 2300 RC Leiden, The Netherlands
| |
ABSTRACT |
|---|
|
|
|---|
Conversion of conductance
catheter data to absolute ventricular volumes requires assessment of
parallel conductance (GP). We determined the
accuracy of GP obtained by the hypertonic saline
method (G

-blockade.
G




1, R2 = 0.92), and
Bland-Altman analysis yielded a nonsignificant bias and narrow limits
of agreement (bias ± 2SD = 0.002 ± 0.112 
1). Within-animal variability of
GP was very similar with both methods and
was due to changes in blood conductivity rather than geometrical
changes. Variability between animals was significant (26.3% of mean
for G

angiography; left ventricular volume; sheep; ventricular function
| |
INTRODUCTION |
|---|
|
|
|---|
THE CONDUCTANCE CATHETER METHOD provides a continuous on-line measurement of left ventricular (LV) volume by means of a multielectrode catheter positioned in the LV. In combination with simultaneous measurement of LV pressure through a sensor on the same catheter, this instrument enables quantification of ventricular function by means of pressure-volume relations. Such relations have proven to be particularly useful, because they provide indexes of systolic and diastolic ventricular function that are relatively independent of loading conditions and, as such, mainly reflect intrinsic myocardial properties.
The conductance catheter method is based on the continuous measurement of the electrical conductance of the blood in the LV. This signal is converted to a volume signal on the basis of a stacked-cylinder model and by taking into account the specific conductivity of blood and the catheter electrode spacing. However, the conductive tissues and fluids surrounding the LV cavity [e.g., myocardial wall, blood in the right ventricle (RV), lung] also contribute to the measured conductances and introduce an offset in the relation between true LV volume and conductance-derived volume. Therefore, to obtain an absolute volume signal, this parallel conductance needs to be determined and subtracted from the raw conductance signal.
To assess parallel conductance or the corresponding offset volume,
several methods have been used. These methods are based on a direct
comparison with an independent method for absolute volume measurement
(5, 6, 20) or do not require an independent volumetric
measurement, such as the dual-frequency method (12), the
suction method (3), or the volume reduction technique
(31). However, in most studies, parallel conductance is
determined by a method, previously developed in our laboratory, that
consists of injecting a small bolus of hypertonic saline through a
balloon-flotation catheter in the pulmonary artery (3).
The highly conductive saline transiently changes the conductivity of
the blood, practically without affecting parallel conductance. The
contribution of parallel conductance to the total conductance signal
can be derived from a registration of the conductance signal during the
passage of the bolus through the LV. A number of studies indicate that
the saline method provides reliable estimates of parallel conductance, but generally the evidence is indirect, since these studies were not
specifically designed to test the accuracy of the saline method (1, 7, 20, 26, 28, 29). Some studies were more
specifically related to the calibration factors of the conductance
catheter method, but those studies largely focused on whether parallel conductance remains constant during preload reduction by caval occlusion (2, 5, 27). The present study, in anesthetized sheep, was designed specifically to test the accuracy of assessment of
parallel conductance by the hypertonic saline method and used biplane
contrast cineangiography as an independent method for absolute volume
measurements. In addition, in the present study we used the dual-field
excitation mode to generate the electric field required for the
conductance measurements. Dual-field excitation, developed by our
group, is used in most studies, because it has been shown to improve
the accuracy of the conductance catheter method (10, 24, 25,
35). However, most studies regarding parallel conductance have
been performed before the introduction of the dual-field excitation
method. The more uniform electric field, which is the rationale for
using dual-field excitation, improves the linearity of the relation
between conductance and true volume (25), and thus, at
least theoretically, the possible dependence of parallel conductance on
volume could be expected to be reduced. The more uniform electric field
implies more uniform intracavitary current density, but the "wider"
field is expected to lead to more current flow through external
structures and, thus, higher parallel conductance. A possible
advantage, however, is that volume dependence and variability of
parallel conductance between hemodynamic conditions may be reduced. To
investigate the need for repeated assessments, we determined the
variability of parallel conductance between animals and between
hemodynamic conditions. Measurements were obtained over a wide range of
hemodynamic conditions, induced by dobutamine infusion, volume loading,
and
-blockade. Finally, it is generally (implicitly) assumed that parallel conductance remains constant during the cardiac cycle. This
assumption has been tested by analyzing hypertonic saline injections
for individual points in the cardiac cycle (19, 33, 34);
however, this analysis may be theoretically incorrect (see DISCUSSION). Therefore, we directly compared the
time-varying angiographic and conductance-derived volume signals. After
subtraction of a (constant) parallel conductance factor, the remaining
difference between the two signals can be interpreted as
(time-dependent) variations in parallel conductance during the cardiac cycle.
| |
METHODS |
|---|
|
|
|---|
Animals
The study was approved by the Animal Research Committee of the University of Leiden. The investigation conforms with the Guide for the Care and Use of Laboratory Animals (National Institutes of Health Publication No. 85-23, Revised 1996). Nine sheep (body mass 38.6 ± 2.9 kg, age 3-6 mo) were premedicated with ketamine (40 mg/kg im) and atropine (0.05 mg/kg im). The animals were intubated and ventilated with a Servo 900B (Siemens, Elema, Sweden). On ventilation, atracurium (0.25 mg/kg iv) was administered to achieve adequate muscle relaxation. General anesthesia was maintained with isoflurane (1%) and continuous infusion of ketamine (4-10 mg · kg
1 · h
1 iv). Arterial
PO2 and PCO2 and pH
were checked every 30 min and kept within normal ranges by adjusting
the fraction of inspiratory O2 and tidal volume as necessary.
Instrumentation
An electrocardiogram was obtained using subcutaneous needle electrodes. To facilitate catheter placement, sheaths were placed in the left and right femoral artery, the left and right femoral vein, the left carotid artery, and the left jugular vein. A balloon-flotation catheter was positioned in the pulmonary artery for injection of the hypertonic saline via the right femoral vein. A 12-electrode dual-field conductance catheter with 10-mm spacing between the sensing electrodes (Sentron, Roden, The Netherlands) was positioned along the long axis of the LV via the left femoral artery. This catheter also incorporated a solid-state pressure transducer for measurement of high-fidelity LV pressure. A 6-F pigtail angiographic catheter (Cordis) was introduced into the LV via the right femoral artery. All catheters were placed under fluoroscopic guidance. The side ports of the sheaths in femoral and jugular veins were used for fluid infusions and administration of drugs and anesthetics. Conductance and blood conductivity measurements were performed using a signal processor (Sigma-5 DF, CD Leycom, Zoetermeer, The Netherlands). LV pressure was measured using a Sentron pressure interface. Electrocardiogram, LV pressure, and the five segmental volume signals were recorded using a personal computer-based data acquisition system and digitized at 12-bit accuracy and a sample frequency of 250 Hz. All data were acquired while the respirator was disconnected at end expiration. Data were stored on hard disk for later analysis. Data acquisition was performed using Conduct-PC (CD Leycom) and data analysis by custom-made software.Conductance Catheter
The conductance catheter technique has been described in detail previously (3, 25). Briefly, a catheter with 10 or 12 electrodes is positioned along the longitudinal axis of the LV. The electrode distance is chosen such that, with electrode 1 within the apex, electrode 9 is situated just above the aortic valve. Through the two most proximal and the two most distal electrodes, two 20-kHz currents (current ratio 1:0.25) opposite in polarity are applied, creating a dual electric field in the ventricular cavity (24, 25). The interposed electrodes are used to measure the conductances of five intraventricular segments. Total time-varying LV conductance [G(t)] is calculated as the sum of these five segmental conductances. Time-varying LV volume is calculated as follows: VCath(t) = (1/
)(L2/
b)[G(t)
GP], where
is a dimensionless slope
factor (see below),
b is the specific conductivity of
the blood measured from a blood sample using a special cuvette,
L is the catheter electrode spacing, and
GP is the parallel conductance (see below).
Nomenclature.
In the literature, the term parallel conductance is often used loosely
for the "physical" parallel conductance (GP)
and for the correction volume (VC). In the present study,
the results are described mainly in terms of GP
and the relation between absolute "true" volume and conductance [G(t)] is written as follows:
V(t) = (1/
)(L2/
b)[G(t)
GP]. Originally, Baan et al.
(3) used the following equation: V(t) = (1/
)(L2/
b)G(t)
VC; other groups have used the following equation: V(t) = (1/
)[(L2/
b)G(t)
VP]. Obviously, VC and VP can
be written in terms of GP, but since
and
b may vary, these terms cannot be used interchangeably. To avoid confusion, we will use the nomenclature as follows:
G(t) for conductance, GP
for parallel conductance,
(L2/
b)G(t)
for conductance volume, VP = (L2/
b)GP
for parallel conductance volume, and VC = (1/
)(L2/
b)GP
for correction volume.
= 1/
b), and the analog output of the system equals
(L2/
b)G(t),
rather than the raw conductance [G(t)].
Slope factor.
After correction for GP, the signal obtained by
the conductance catheter is directly proportional to absolute
ventricular volume but generally underestimates true volume by a fixed
percentage. To correct the underestimation,
was introduced. In
practice,
is determined by comparing the conductance-derived volume
[or stroke volume (SV)] with an independent measurement such as
angiography or thermodilution. In animals such as dogs or sheep,
is
typically 0.8 (3, 24). A key point in the present study is
that the approaches that are used to determine
GP, i.e., the saline dilution method and the
angiographic method (see below), do not require that
be assessed
and are independent of the actual value of
. Both methods analyze
the raw conductance signals, rather than the calibrated conductance
volume signals, to determine GP. The only
implicit assumption is that
can be regarded as constant during the
cardiac cycle, but it does not necessarily need to be 1.0. This
assumption is validated by testing the linearity of the relation
between conductance and angiographic volume.
GP Obtained by Hypertonic Saline Injection
The electric field generated by the conductance catheter is not entirely restricted to the ventricular blood volume, but current also passes through the ventricular wall, other cardiac chambers, and, to some extent, through all electrically conductive structures surrounding the heart. As a consequence, the total conductance measured is the sum of the conductance of the blood in the LV and the "parallel" conductance of the surrounding structures. Baan et al. (3) devised a method to determine GP by injecting a small bolus (2-3 ml) of hypertonic (10%) saline through a balloon-flotation catheter in the pulmonary artery. This procedure can be explained as follows. If blood conductivity in the LV could be reduced to 0, the measured total conductance would represent GP only. In practice, this is not possible, but we can transiently change conductivity (by the hypertonic saline injection), plot measured total conductance vs. blood conductivity, and extrapolate these data to the point where conductivity hypothetically would be 0 and obtain GP in this way. This approach (see APPENDIX) requires a beat-to-beat estimate of blood conductivity in the LV, which can obtained as follows:
b = (1/
)(L2/SV)SG, where
SV = VED
VES, and "stroke
conductance" (SG) = GED
GES, where the subscripts ED and ES represent
end diastole and end systole, respectively. The equation for
b shows that if hemodynamics (and thus SV) are constant
during the passage of the bolus,
b is directly
proportional to the amplitude of the conductance signal (SG). Thus
GP can be obtained by plotting
GED vs. SG for each beat during the change in
blood conductivity and extrapolating this relation to SG = 0. This
point corresponds to the hypothetical situation with
b = 0 and, therefore, yields
GP.
In the present study, GP obtained by hypertonic
saline injection (G
Angiography
Angiography was performed with a Philips DCI SX biplane X-ray system with a biplane frame rate of 25 frames/s and 7-in. image intensifiers. Simultaneous biplane images from standard 30° right anterior oblique (RAO) and 60° left anterior oblique (LAO) projections were obtained after injection of 15 ml of nonionic contrast material (Iomeron, Bracco-Byk Gulden, Konstanz, Germany) at a flow rate of 6 ml/s. All data were acquired after the respirator was disconnected at end expiration. Images were stored on CD ROM at the end of the study for off-line analysis.Angiographic dimensions were calibrated on the diameter of the
angiographic catheter using QCA-CMS View software (Medis, Leiden, The
Netherlands). Angiographic LV volumes (VAngio) were
calculated using the area-length method (9, 36) as
follows: VAngio = (8/3)
(ARAOALAO)/LRAO,
where ARAO and ALAO are
the areas enclosed by the LV contours in the RAO and LAO projections,
respectively, and LRAO and
LLAO are the lengths of the LV long axis in the
RAO and LAO projections, respectively. The contours were drawn manually in all frames from two consecutive, well-opacified cardiac cycles using
custom-made software.
Protocol
Measurements were performed at baseline, after treatment with dobutamine (2.5 µg · kg
1 · min
1 iv),
after volume load (200 ml iv gelofusine over a 12-min period), and
after treatment with propranolol (1 mg/kg iv). In each condition,
b was measured, three consecutive hypertonic saline
injections (2 ml, 10% saline) were performed to determine
GP, and biplane angiography was performed.
Recording of simultaneous conductance signals was started ~5 s before
contrast injection and continued during the acquisition of angiographic
images. Because respiration may affect GP and
actual end-diastolic volume, all data were acquired during apnea at end
expiration. The respirator was disconnected 3-5 s before the
actual data acquisitions.
Comparison of Angiographic and Conductance Signals
Angiographically derived parallel conductance and
.
Contrast medium changes the conductivity of blood; therefore, we
compared the angiographically derived volume signals with the
conductance signals obtained just before contrast injection. To smooth
the conductance signal and to avoid selection bias, four consecutive
cardiac cycles were selected and temporally averaged. The conductance
and angiographic signals were synchronized by matching the peaks
(end-diastolic volume) of the two signals. The temporal resolution of
the angiographic signal was 40 ms and that of the conductance signal
was 4 ms; consequently, comparisons were made using data points at
40-ms intervals only. The data points from both signals were plotted
vs. each other, and a linear regression was performed. The
y-intercept represents the hypothetical conductance when
angiographic volume equals zero and thus represents an estimate of
GP based on direct comparison with instantaneous
angiographic volume. Statistical analysis of the comparative data
[G

. Given the definition of
,
= (L2/
b)(dG/dVAngio).
In addition, the linear correlation coefficient (R2) and the standard error of the y
estimate were used to test the linearity of the relation and, thus,
support the assumption that
remains constant during the cardiac cycle.
Time-varying GP.
The analysis described above determines GP
averaged over the full cardiac cycle. To investigate whether
GP varies during the cardiac cycle, the
conductance signal was calibrated using the coefficients (slope and
intercept) obtained from the linear regression, and subsequently the
angiographic signal was subtracted from this calibrated
conductance-volume signal. The resulting difference signal represents
the errors remaining after correction for mean
GP and can be interpreted as variations in
GP during the cardiac cycle. Mathematically,
GP can be written as follows:
GP(t) = GP + dGP(t), were
GP is the average (constant) parallel
conductance and dGP(t) the
time-varying component. VAngio is taken as the gold
standard; thus it is assumed that
VAngio(t) = (1/
)(L2/
b){G(t)
[GP + dGP(t)]}.
and
GP are obtained as the slope and intercept,
respectively, of the relation between angiographic and uncalibrated
conductance volume, as described above. Thus
dGP(t) = [VCath(t)
VAngio(t)]/[(1/
)(L2/
b)],
where the calibrated conductance volume
[VCath(t)] = (1/
)(L2/
b)[G(t)
GP]. To enable comparisons between animals
and between conditions despite changes in heart rate, the conductance
and angiographic signals were fitted with a cubic spline, resampled at
500 time points, and plotted on a normalized time scale.
Variability of GP
Variabilities of GP (G

a
a


b and parallel conductance volumes obtained by
hypertonic saline injection (V

| |
RESULTS |
|---|
|
|
|---|
Typical examples of uncalibrated conductance volume signals,
(L2/
b)G(t),
and angiographic volume signals in the four hemodynamic conditions are
shown in Fig. 1. Linear extrapolation of
all data points during the cardiac cycle, G(t)
vs. VAngio(t), yielded
G


1, indicating an essentially zero bias with a standard
deviation of 8.4% of mean GP.
|
|
|
|
Statistical analysis of the variabilities of
G



1 at baseline. This standard deviation does not
include the interanimal variability. The variability between animals
was 0.17 
1, or 26% of baseline
G

0.065 
1, or
9.8% of baseline) and
significantly larger during propranolol (+0.058 
1, or
+8.7% of baseline). Results for G


b varied significantly between animals (8.4%), although
much less than GP. The condition variabilities
of
b, however, were very similar to those of
GP, suggesting that these changes in
GP were largely due to underlying changes in
b. This suggestion is substantiated by the fact that
parallel conductance volume (VP), which is equal to
(L2/
b)GP,
did not differ significantly between conditions.
|
|
Figure 6 shows the variations of
GP during the cardiac cycle as a percentage of
its mean value. The average variation was small and not significantly
different from zero; it ranged from +1.8 ± 2.4% in the
midejection phase to
1.3 ± 2.6% during early filling.
|
| |
DISCUSSION |
|---|
|
|
|---|
Conversion of raw conductance catheter data to calibrated absolute volumes requires the assessment of GP. GP can be determined by direct comparison of the conductance signals with an independent measurement of absolute LV volume, such as by angiography (20), dimension crystals (2), or balloon volume (6). In addition, several methods have been proposed that enable assessment of GP without the need for independent volumetric measurements. Baan et al. (3) used a suction method in which cavity volume was reduced to zero by means of a multihole catheter in the LV or by manual compression. The results indicate reasonable correspondence with the saline method, but the nature of this intervention clearly limits its practical application. Recently, White et al. (31) introduced a novel approach that relies on the analysis of the transient reduction of volume induced, e.g., by balloon occlusion of the inferior vena cava. By extrapolating the relation between end-diastolic volume and end-systolic volume during the intervention to a point where these volumes are equal, an estimate of the offset volume corresponding to GP is obtained. The key assumption in this approach is that ejection fraction is constant during the volume reduction. This hemodynamic requirement may explain why the results show good correlation with the saline method in some study groups but poor agreement in several other groups. Gawne et al. (12) introduced the so-called dual-frequency method, which exploits the fact that, in the 2- to 100-kHz frequency range, blood conductivity is essentially constant whereas muscle conductivity varies. By comparing measured conductance catheter signals at two frequencies (3.3 and 33 kHz), an estimate of GP was obtained. An inherent disadvantage of this method is that changes in frequency-independent components of GP (i.e., the blood in the RV) will not be picked up. Although the results by Gawne et al. were promising, more recent studies in neonatal and adult pigs (32) have been disappointing. However, the method appears be applicable in mice, presumably because GP in these hearts resides almost exclusively in the relatively thick myocardial wall (13). However, the method most widely used to determine GP is the hypertonic saline method (3). The present study addressed the following three issues with regard to this method: the absolute accuracy of the hypertonic saline method for assessment of GP, the variability of GP between animals and between hemodynamic conditions, and the variability of GP during the cardiac cycle.
Accuracy of the Hypertonic Saline Method
Testing of the absolute accuracy of the hypertonic saline method requires an independent measurement of absolute volume. In the present study, the uncalibrated conductance data were plotted vs. absolute volumes obtained by biplane cineangiography for all time points during a full cardiac cycle. Extrapolation of this relation to a hypothetical zero angiographic volume yields an estimate of GP. This estimate was compared with the value obtained by saline dilution. Our results show an excellent agreement between these two methods. Linear regression showed a good correlation (R2 = 0.92), an essentially zero offset, and slope equal to 1.0, whereas Bland-Altman analysis yielded a nonsignificant bias (±2 SD) of 0.002 ± 0.112
1.
Previous results using the single-field conductance method in patients
vs. monoplane cineangiography (3) indicated a
nonsignificant underestimation of 6.5%. A somewhat larger and
significant underestimation of 14% was found by Boltwood et al.
(5). As in our study, biplane angiography was used to
obtain an independent estimate of GP, but
hemodynamic conditions were altered by partial balloon occlusions of
the aorta, inferior vena cava, or pulmonary artery, whereas we used
pharmacological interventions and volume infusions. We chose the
latter, because, in our experience, it is rather difficult to produce
well-defined steady-state conditions with partial occlusions, and thus
it was anticipated that it would be difficult to perform (multiple)
saline injections while maintaining steady-state hemodynamics. In the
study of Boltwood et al., G





A limitation in most studies lies in the dependence of the results on
. In principle, the error in the G
is determined, which introduces an error
source. Alternatively, analysis of conductance volumes corrected for
GP, but not for
, is possible, but only when
multiple data points are available from different animals or from
different hemodynamic conditions. In that case, the error in
GP can be derived from the offset of the
relation between volume by conductance and by the independent method.
Although the latter approach does not require measurement of
, it
does implicitly assume that the variability in
(between animals or
between conditions) is small. In contrast, in our study, the relation
with angiography was obtained by using all data points from a full
cardiac cycle. Thus the only implicit assumption is that
is
constant (but not necessarily 1.0) within the cardiac cycle, and
neither the absolute value of
nor variations of
between animals
or between hemodynamic conditions affected our results. A previous
study by Szwarc et al. (26), using single-field
excitation, yielded significant, but small, changes in
during
ejection. Our results (Table 2), with
dual-field excitation, show an excellent linear relation between
conductance and angiographic volume and indicate that, for all
practical purposes,
can be regarded as constant during the cardiac
cycle.
|
Variability of GP Between Animals and Between Hemodynamic Conditions
Quantification of the variability of GP is essential to determine the need for repeated assessments. Previous studies have clearly indicated that the variability between animals is substantial even when the age and weight of the study group are within fairly narrow ranges. The most likely causes for this variability are differences in the catheter position, the size and geometry of the heart, the position of the heart in the thorax, and the geometry of other structures in the thorax. In addition, there will be between-subject differences in the conductive properties of blood and tissues. In the present study, the interanimal variability of G

Within-subject variability of GP was
investigated in the present study by repeating the assessments after
interventions aimed at inducing different hemodynamic conditions. The
results in Table 1 indicate that, during dobutamine infusion,
G
9.8%
(G
8.6%), whereas after
propranolol infusion,
G

b was also changed: during dobutamine
infusion,
b decreased by
8.5% compared with baseline, whereas after propranolol infusion,
b was +8.3% higher
than control. These changes in
b reflect changes in
hematocrit, which, during dobutamine infusion, may be increased because
of an increased transcapillary fluid shift or red cell recruitment from
the spleen (8, 11, 15, 30), whereas, conversely,
gelofusine infusion leads to a decrease in hematocrit and, thus, an
increase in
b. Thus, because the physical structures
that contribute to GP (i.e., myocardial wall, RV
cavity, lungs) contain blood, the changes in GP
may, in fact, be largely due to changes in
b rather than
geometrical changes. Interestingly, this means that the parallel
conductance volume, which is calculated as VP = (L2/
b)GP,
would be much less affected. Indeed, statistical analysis (Table 1)
shows that changes in VP were <2.4% and not statistically significant.
The dependence of GP (or VP) on
hemodynamic conditions has been investigated in several previous
studies. Szwarc et al. (27) measured VP by the
saline method in intact dogs at baseline, after volume loading, and
after bleeding and did not find significant differences despite large
changes in hemodynamic status. Boltwood et al. (5) estimated VP by repeated saline injections under a variety
of loading conditions: compared with control, VP was
unchanged during occlusion of the pulmonary artery or the aorta but was
significantly reduced (
9%) during occlusion of the inferior vena
cava. This may reflect the influence of reduced RV volume during caval
occlusion. The influence of RV volume was also demonstrated in a study
from our group (23) where embolization of the right
coronary artery, dilating the RV, caused a 20% increase in
GP. In general, these changes in
GP do not invalidate the method but indicate the
need to reassess GP after substantial changes in
hemodynamic conditions.
Variability of GP During the Cardiac Cycle
The saline method yields a single value for GP; however, RV filling and ejection, atrial filling, changes in myocardial shape, and blood content could potentially cause changes in GP during the cardiac cycle. Our results, however, indicate that such changes are very small and can, in practice, be neglected. Previously, cyclic GP (or VP) variations have only been studied for single-field excitation. White et al. (33, 34) assessed cyclic variation of VP by plotting isochronal uncalibrated conductance volume vs. conductance stroke volume: the y-intercept for each set of isochronal points was used as an estimate of VP at the time during the cardiac cycle corresponding to the isochrone. The method, however, contains a theoretical flaw: on the one hand, it aims to determine a time-varying GP; on the other hand, it assumes that conductance stroke volume is directly proportional to blood conductivity and, thus, implicitly requires that GP is equal at end diastole and end systole (or any other pair of points during the cardiac cycle that are used to calculate "apparent" stroke volume). Thus, rather than at apparent stroke volume equals zero, the y-intercept should be determined at x = V
V
Conclusions
The main finding in this study is that the hypertonic saline method accurately determines GP for dual-field conductance catheter compared with biplane angiography. The dual-field method produced G

| |
APPENDIX |
|---|
|
|
|---|
The conventional method for determination of
GP from end-diastolic conductances
(GED) and end-systolic conductances
(GES) obtained after hypertonic saline
injections consists of plotting GED vs. GES for all beats during the saline washin
period and calculating the intercept of the relation through these
points with the line of identity. The alternative approach used in the
present study (see METHODS) is mathematically identical,
but it better illustrates that GP is determined
at the hypothetical point where
b = 0, and thus the
only remaining conductance is GP. The
b cannot be directly measured, but, as shown below,
"stroke conductance" (SG = GED
GES) is directly proportional to
b. Therefore, rather than vs.
B,
GED is plotted vs.
GED
GES and the
relation is extrapolated to the point where
GED
GES = 0. This way, similar to the conventional method, this alternative approach
determines GP at the hypothetical point where
GED = GES (Fig.
7).
|
Absolute end-diastolic volume (VED) = (1/
)(L2/
b)(GED
GP), and absolute end-systolic volume
(VES) = (1/
)(L2/
b)(GES
GP). Thus SV = VED
VES = (1/
)(L2/
b)(GED
GES), from which it can be derived that
B = (1/
)L2SG/SV. Because
, L, and SV are constant, SG is directly proportional to
B.
| |
FOOTNOTES |
|---|
Address for reprint requests and other correspondence: P. Steendijk, Dept. of Cardiology, Leiden University Medical Center, PO Box 9600, 2300 RC Leiden, The Netherlands (E-mail: p.steendijk{at}lumc.nl).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 13 November 2000; accepted in final form 13 March 2001.
| |
REFERENCES |
|---|
|
|
|---|
1.
Amirhamzeh, MM,
Dean DA,
Jia CX,
Cabreriza SE,
Yano OJ,
Burkhoff D,
and
Spotnitz HM.
Validation of right and left ventricular conductance and echocardiography for cardiac function studies.
Ann Thorac Surg
62:
1104-1109,
1996
2.
Applegate, RJ,
Cheng CP,
and
Little WC.
Simultaneous conductance catheter and dimension assessment of left ventricle volume in the intact animal.
Circulation
81:
638-648,
1990
3.
Baan, J,
Van der Velde ET,
De Bruin H,
Smeenk G,
Koops J,
Van Dijk AD,
Temmerman D,
Senden J,
and
Buis B.
Continuous measurement of left ventricular volume in animals and humans by conductance catheter.
Circulation
70:
812-823,
1984
4.
Bland, JM,
and
Altman DG.
Statistical methods for assessing agreement between two methods of clinical measurement.
Lancet
1:
307-310,
1986[ISI][Medline].
5.
Boltwood, CMJ,
Appleyard RF,
and
Glantz SA.
Left ventricular volume measurement by conductance catheter in intact dogs. Parallel conductance volume depends on left ventricular size.
Circulation
80:
1360-1377,
1989
6.
Burkhoff, D,
Van Der Velde E,
Kass D,
Baan J,
Maughan WL,
and
Sagawa K.
Accuracy of volume measurement by conductance catheter in isolated, ejecting canine hearts.
Circulation
72:
440-447,
1985
7.
Cassidy, SC,
and
Teitel DF.
The conductance volume catheter technique for measurement of left ventricular volume in young piglets.
Pediatr Res
31:
85-90,
1992[ISI][Medline].
8.
Dacar, D.
Continuous blood density measurements and volume changes during extracorporeal circulation in patients undergoing cardiac surgery.
Thorac Cardiovasc Surg
43:
13-18,
1995[ISI][Medline].
9.
Dodge, HT,
Sandler H,
Ballew DW,
and
Lord JJ.
The use of biplane angiography for the measurement of left ventricular volume in man.
Am Heart J
60:
762-776,
1960[ISI][Medline].
10.
Feldman, MD,
Subbiah K,
Finnerty P,
Haber HL,
Anne A,
and
Skalak TC.
Accuracy of single vs. dual field excitation of the conductance catheter to determine absolute left ventricular volume and contractility (Abstract).
J Am Coll Cardiol
21:
372A,
1993.
11.
Fuchs, RM,
Rutlen DL,
and
Powell WJ, Jr.
Effect of dobutamine on systemic capacity in the dog.
Circ Res
46:
133-138,
1980
12.
Gawne, TJ,
Gray KS,
and
Goldstein RE.
Estimating left ventricular offset volume using dual-frequency conductance catheters.
J Appl Physiol
62:
872-876,
1987.
13.
Georgakopoulos, D,
and
Kass DA.
Estimation of parallel conductance by dual-frequency conductance catheter in mice.
Am J Physiol Heart Circ Physiol
279:
H443-H450,
2000
14.
Glantz, SA,
and
Slinker BK.
Primer of Applied Regression and Analysis of Variance. New York: McGraw Hill, 1990.
15.
Hellyer, PW,
Wagner AE,
Mama KR,
and
Gaynor JS.
The effects of dobutamine and ephedrine on packed cell volume, total protein, heart rate, and blood pressure in anaesthetized horses.
J Vet Pharmacol Ther
21:
497-499,
1998[ISI][Medline].
16.
Kass, DA,
Midei M,
Graves W,
Brinker JA,
and
Maughan WL.
Use of a conductance (volume) catheter and transient inferior vena caval occlusion for rapid determination of pressure-volume relationships in man.
Cathet Cardiovasc Diagn
15:
192-202,
1988[ISI][Medline].
17.
Kass, DA,
Yamazaki T,
Burkhoff D,
Maughan WL,
and
Sagawa K.
Determination of left ventricular end-systolic pressure-volume relationships by the conductance catheter.
Circulation
73:
586-595,
1986
18.
Klautz, RJM,
Teitel DF,
Steendijk P,
Van Bel F,
and
Baan J.
Interaction between afterload and contractility in the newborn heart: evidence of homeometric autoregulation in the intact circulation.
J Am Coll Cardiol
25:
1428-1435,
1995[Abstract].
19.
Lankford, ED,
Kass DA,
Maughan WL,
and
Shoukas AA.
Does volume catheter parallel conductance vary during a cardiac cycle?
Am J Physiol Heart Circ Physiol
258:
H1933-H1942,
1990
20.
Odake, M,
Takeuchi M,
Takaoka H,
Hata K,
Hayashi Y,
and
Yokoyama M.
Determination of left ventricular volume using a conductance catheter in the diseased human heart.
Eur Heart J
13 SupplE:
22-27,
1992
21.
Pattynama, PM,
de Roos A,
Van der Velde ET,
Lamb HJ,
Steendijk P,
Hermans J,
and
Baan J.
Magnetic resonance imaging analysis of left ventricular pressure-volume relations: validation with the conductance method at rest and during dobutamine stress.
Magn Reson Med
34:
728-737,
1995[ISI][Medline].
22.
Slinker, BK,
and
Glantz SA.
Multiple linear regression is a useful alternative to traditional analyses of variance.
Am J Physiol Regulatory Integrative Comp Physiol
255:
R353-R367,
1988
23.
Steendijk, P,
and
Baan J.
Comparison of intravenous and pulmonary artery injections of hypertonic saline for the assessment of conductance catheter parallel conductance.
Cardiovasc Res
46:
82-89,
2000
24.
Steendijk, P,
Van Der Velde ET,
and
Baan J.
Left ventricular stroke volume by single and dual excitation of conductance catheter in dogs.
Am J Physiol Heart Circ Physiol
264:
H2198-H2207,
1993
25.
Steendijk, P,
Van Der Velde ET,
and
Baan J.
Single and dual excitation of the conductance-volume catheter analysed in a spheroidal mathematical model of the canine left ventricle.
Eur Heart J
13 SupplE:
28-34,
1992
26.
Szwarc, RS,
Laurent D,
Allegrini PR,
and
Ball HA.
Conductance catheter measurement of left ventricular volume: evidence for nonlinearity within cardiac cycle.
Am J Physiol Heart Circ Physiol
268:
H1490-H1498,
1995
27.
Szwarc, RS,
Mickleborough LL,
Mizuno S,
Wilson GJ,
Liu P,
and
Mohamed S.
Conductance catheter measurements of left ventricular volume in the intact dog: parallel conductance is independent of left ventricular size.
Cardiovasc Res
28:
252-258,
1994
28.
Tjon, A,
Meeuw L,
Hess OM,
Nonogi H,
Monrad ES,
Leskosek B,
and
Krayenbuehl HP.
Left ventricular volume determination in dogs: a comparison between conductance technique and angiocardiography.
Eur Heart J
9:
1018-1026,
1988
29.
Van Der Velde, ET,
Van Dijk AD,
Steendijk P,
Diethelm L,
Chagas T,
Lipton MJ,
Glantz SA,
and
Baan J.
Left ventricular segmental volume by conductance catheter and Cine-CT.
Eur Heart J
13 SupplE:
15-21,
1992
30.
Wagner, AE,
Dunlop CI,
and
Chapman PL.
Effects of ephedrine on cardiovascular function and oxygen delivery in isoflurane-anesthetized dogs.
Am J Vet Res
54:
1917-1922,
1993[ISI][Medline].
31.
White, PA,
Brookes CI,
Ravn H,
Hjortdal V,
Chaturvedi RR,
and
Redington AN.
Validation and utility of novel volume reduction technique for determination of parallel conductance.
Am J Physiol Heart Circ Physiol
280:
H475-H482,
2001
32.
White, PA,
Brookes CI,
Ravn HB,
Stenbog EE,
Christensen TD,
Chaturvedi RR,
Sorensen K,
Hjortdal VE,
and
Redington AN.
The effect of changing excitation frequency on parallel conductance in different sized hearts.
Cardiovasc Res
38:
668-675,
1998
33.
White, PA,
Chaturvedi RR,
Bishop AJ,
Brookes CI,
Oldershaw PJ,
and
Redington AN.
Does parallel conductance vary during systole in the human right ventricle?
Cardiovasc Res
32:
901-908,
1996[ISI][Medline].
34.
White, PA,
Chaturvedi RR,
Shore D,
Lincoln C,
Szwarc RS,
Bishop AJ,
Oldershaw PJ,
and
Redington AN.
Left ventricular parallel conductance during cardiac cycle in children with congenital heart disease.
Am J Physiol Heart Circ Physiol
273:
H295-H302,
1997
35.
Wu, CC,
Skalak TC,
Schwenk TR,
Mahler CM,
Anne A,
Finnerty PW,
Haber HL,
Weikle RM, II,
and
Feldman MD.
Accuracy of the conductance catheter for measurement of ventricular volumes seen clinically: effects of electric field homogeneity and parallel conductance.
IEEE Trans Biomed Eng
44:
266-277,
1997[ISI][Medline].
36.
Wynne, J,
Green LH,
Mann T,
Levin D,
and
Grossman W.
Estimation of left ventricular volumes in man from biplane cineangiograms filmed in oblique projections.
Am J Cardiol
41:
726-732,
1978[ISI][Medline].