|
|
||||||||
Harvard-Massachusetts Institute of Technology Division of Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, Massachusetts 02139
| |
ABSTRACT |
|---|
|
|
|---|
We present a theoretical evaluation of a cardiovascular system identification method that we previously developed for the analysis of beat-to-beat fluctuations in noninvasively measured heart rate, arterial blood pressure, and instantaneous lung volume. The method provides a dynamical characterization of the important autonomic and mechanical mechanisms responsible for coupling the fluctuations (inverse modeling). To carry out the evaluation, we developed a computational model of the cardiovascular system capable of generating realistic beat-to-beat variability (forward modeling). We applied the method to data generated from the forward model and compared the resulting estimated dynamics with the actual dynamics of the forward model, which were either precisely known or easily determined. We found that the estimated dynamics corresponded to the actual dynamics and that this correspondence was robust to forward model uncertainty. We also demonstrated the sensitivity of the method in detecting small changes in parameters characterizing autonomic function in the forward model. These results provide confidence in the performance of the cardiovascular system identification method when applied to experimental data.
beat-to-beat model; mathematical modeling; cardiovascular modeling; autonomic nervous system; hemodynamics
| |
INTRODUCTION |
|---|
|
|
|---|
WHEN ONE CONSIDERS MODELING the cardiovascular system, one usually envisions constructing a model based on physical principles that is capable of generating realistic data. This type of modeling approach, which we refer to as forward modeling, is useful for enhancing one's understanding of cardiovascular physiology. One may also consider an inverse modeling approach, in which models are built from measured data. This type of modeling approach is referred to as system identification when system dynamics or memory is being considered. System identification may potentially provide a powerful means for the intelligent patient monitoring of cardiovascular function. Rather than simply recording hemodynamic signals, the signals are mathematically analyzed so as to provide a dynamical characterization of the physiological mechanisms responsible for coupling them.
To this end, we (26, 27, 29) have previously developed a cardiovascular system identification method for the analysis of short-term, beat-to-beat fluctuations in heart rate signals derived from the surface electrocardiogram (ECG) and noninvasively measured arterial blood pressure (ABP) and instantaneous lung volume (ILV) signals. The method provides a dynamical characterization of the important autonomic and mechanical mechanisms that couple the measured fluctuations specifically in terms of impulse responses. The method also provides power spectra of perturbing noise sources, which represent the residual variability in each of the signals not attributed to the variability generated through the impulse responses.
We have evaluated the cardiovascular system identification method with experimental data collected during conditions of pharmacological autonomic blockade, postural changes, and diabetic autonomic neuropathy (27, 29). We found that these three conditions altered the impulse response and power spectra of perturbing noise source estimates in a manner consistent with known physiological mechanisms. This suggests, but does not directly demonstrate, the validity of the estimated system dynamics.
To evaluate directly the system dynamics estimated by the method, it is necessary to be able to establish the impulse responses and power spectra of perturbing noise sources in a manner independent of system identification. These impulse responses and power spectra of perturbing noise sources may then be regarded as the gold standards against which the corresponding estimates from system identification may be compared. Ideally, one would validate the method based on experimental data. However, the establishment of gold standard impulse responses and power spectra of perturbing noise sources would require extreme experimental conditions that would be virtually impossible to implement in practice. Consider, for example, the establishment of an impulse response, which would require, according to mathematical definition, applying an arbitrarily narrow, unit-area input to the appropriate point of the cardiovascular system and measuring the output response of interest while all other perturbations to the output are held constant. Moreover, even if this type of experiment could be implemented, it is unreasonable to expect that the cardiovascular state and system operating point would be precisely maintained, which renders the meaning of such a comparison to be questionable.
By contrast, the method could be readily evaluated on a theoretical basis with a forward model of the cardiovascular system, because the gold standard impulse responses and power spectra of perturbing noise sources would either be precisely known or easily determined for the relevant cardiovascular state and system operating point. That is, the gold standards could be readily established according to their mathematical definitions. A forward model would also provide a powerful means to analyze the sensitivity of the cardiovascular system identification method. That is, we would be able to determine precisely how much the dynamical properties of the forward model would have to be altered before we would see a corresponding change in the estimates. The major limitation of this type of evaluation is that the results would be only as meaningful as the extent to which the forward model coincides with the actual cardiovascular system. This limitation can be attenuated by determining the robustness of the estimates over a set of models that reflect the uncertainty in the relevant properties of the forward model. We note that the general concept of analyzing inverse modeling algorithms based on forward models of the cardiovascular system is not novel. For example, investigators (9, 16) have utilized complex forward models of the systemic circulation to evaluate the estimation of lumped parameters representing systemic arterial resistance [Ra(t)] and systemic arterial compliance.
The principal goal of this study is to present a theoretical validation of our previously developed cardiovascular system identification method. To realize this aim, this study includes the following five components: 1) a brief review of the cardiovascular system identification method, which has been previously described in detail (26, 27, 29); 2) a description of a forward model of the human cardiovascular system; 3) the procedure for evaluating the system identification method against the forward model, which includes the establishment of gold standards based on their mathematical definitions; 4) the validation of the forward model in terms of power spectra of beat-to-beat variability; and 5) the evaluation of the impulse response and power spectra of perturbing noise source estimates derived by applying system identification to beat-to-beat variability generated from the forward model against the corresponding gold standards in terms of accuracy, robustness, and sensitivity. Note that the final component amounts to assessing the equivalence of the impulse responses and power spectra of perturbing noise sources determined from executing the forward model under two conditions: 1) resting conditions by applying system identification to analyze beat-to-beat variability, and 2) extreme conditions established by the literal meaning of these quantities. Because the extreme conditions cannot be implemented in practice, this study seeks to determine whether the cardiovascular system identification method is able to estimate reliably the physiologically important impulse responses and power spectra of perturbing noise sources during experimental conditions that could be readily implementable.
| |
CARDIOVASCULAR SYSTEM IDENTIFICATION REVIEW |
|---|
|
|
|---|
Figure 1 illustrates the model upon
which the cardiovascular system identification method is based. The
model includes five physiological coupling mechanisms relating
ECG-derived heart rate signals, ABP, and ILV: circulatory mechanics,
heart rate (HR) baroreflex, ILV
HR, ILV
ABP, and sinoatrial (SA)
node.
|
Circulatory mechanics represents the mechanical feedforward effects of
a pulsatile heart rate (PHR) signal [defined to be a train of
unit-area impulses occurring at the times of ventricular contraction (R
wave)] on the pulsatile ABP waveform. HR baroreflex represents the
autonomically mediated feedback effects of fluctuations in ABP on
fluctuations in an HR tachogram, which is derived from the R-R
intervals (3). ILV
HR represents the autonomically mediated coupling between respiration and HR, which is responsible for governing respiratory sinus arrhythmia. ILV
ABP represents the
mechanical effects of respiration on ABP due to the alterations in
venous return and the filling of intrathoracic vessels and heart
chambers associated with the changes in intrathoracic pressure. SA node
maps HR (the net autonomic input to the sinoatrial node) to PHR (the
onset times of each ventricular contraction) through an "integrate
and fire" device referred to as the integral pulse frequency
modulation (IPFM) model (3). Unlike the other four physiological coupling mechanisms, SA node is predefined and not identified from the measured signals.
The model also incorporates two perturbing noise sources, NHR and NABP, which are determined from analysis of the measured signals. NHR represents the fluctuations in HR not caused by fluctuations in ABP or ILV. Such fluctuations may result, for example, from autonomically mediated perturbations driven by cerebral activity. NABP represents fluctuations in ABP not caused by PHR or fluctuations in ILV. Such ABP fluctuations may result, for example, from fluctuations in Ra(t) as tissue beds adjust local vascular resistance to match local blood flow to demand.
To obtain a complete characterization of each of the coupling mechanisms over their physiological range of frequencies, we employ a broadband excitation protocol, in which subjects breathe according to a random sequence of auditory tones (4). We have found that when the subjects are breathing according to this protocol and are otherwise at rest that the fluctuations are sufficiently small and stationary such that the coupling mechanisms may be characterized by linear time-invariant (LTI) transfer functions around a given system operating point (11). We specifically represent each of the coupling mechanisms with autoregressive moving average difference equations, a highly flexible subclass of LTI models, whose parameters may be conveniently identified with the analytic methods of linear least-squares estimation. Further details of the method are provided in (26, 29).
Figure 1 includes an example of the cardiovascular system identification results computed from a standing, healthy subject breathing randomly, in which ABP and ILV are noninvasively measured with a continuous blood pressure monitor (model 2300 Finapres; Ohmeda; Englewood, CO) and a Respitrace System two-belt chest-abdomen inductance plethysmograph (Ambulatory Monitoring; Ardsley, NY) (34). The identified transfer functions are presented in their time domain form of impulse responses. For example, the HR baroreflex impulse response estimate indicates that HR would immediately decrease and then return to baseline if a unit-area impulse of ABP were applied at time 0 while the other inputs to HR (ILV and NHR) were set to 0. The estimated perturbing noise sources in the figure are depicted in their frequency domain form of power spectra. For example, the power spectrum of NHR indicates that ABP and ILV do not fully account for HR variability within ~0.05 Hz.
| |
FORWARD MODEL DESCRIPTION |
|---|
|
|
|---|
The forward model of the human cardiovascular system to which we apply the cardiovascular system identification method includes three major components: a pulsatile heart and circulation, a short-term regulatory system, and resting physiological perturbations. The pulsatile heart and circulation is a nonlinear, lumped parameter model (R. Mukkamala, D. A. Sherman, R. G. Mark, and R. J. Cohen, unpublished observations). The short-term regulatory system consists of three subcomponents: an arterial baroreflex, a cardiopulmonary baroreflex, and a direct neural coupling mechanism between respiration and HR. The resting physiological perturbations also include three subcomponents: respiratory activity, the autoregulation of local vascular beds, and 1/f HR fluctuations. We describe all of these components and subcomponents and include parameter values below. See APPENDIX A for a description of forward model implementation.
Pulsatile Heart and Circulation
The nonlinear, lumped parameter model of the pulsatile heart and circulation is illustrated in Fig. 2 in terms of its electrical circuit analog, in which charge is analogous to blood volume (Q, ml), current, to blood flow rate (
, ml/s), and
voltage, to pressure (P, mmHg). The model consists of six compartments,
which represent the left and right ventricles (l and r), systemic
arteries and veins (a and v), and pulmonary arteries and veins (pa and
pv). Each compartment consists of a conduit for viscous blood flow, which is characterized by either a linear or nonlinear resistance (R) and a volume storage element, which is characterized by
either a linear or nonlinear compliance (C) with an associated
unstressed volume (Q0). The reference (ref) pressure is
atmospheric pressure (or ground) for the systemic compartments and
intrathoracic (th) pressure for the ventricular and pulmonary
compartments. The compliances of the ventricles vary periodically over
time (t) according to the model of Suga and colleagues
(38, 39) and are responsible for driving the flow of
blood. The four ideal diodes represent the ventricular inflow and
outflow valves and ensure unidirectional blood flow. We have
demonstrated (R. Mukkamala, D. A. Sherman, R. G. Mark, and
R. J. Cohen, unpublished observations) that this model behaves
reasonably in terms of pulsatile waveforms and limiting static terminal
pressure-flow rate properties of the systemic circulation and
heart-lung unit. However, by itself, the model cannot generate the
short-term, low-frequency hemodynamic variability that the
cardiovascular system elicits during resting conditions. To account for
this variability, it is necessary to incorporate a short-term
regulatory system and resting physiological perturbations with the
model here.
|
Short-Term Regulatory System
Arterial baroreflex.
The short-term regulatory system includes a baroreflex system, which is
based on a previously developed model (13). This model,
which is fully described here, conceptualizes the arterial (A)
baroreflex arc as the feedback system illustrated in the block diagram
of Fig. 3. The feedback system is aimed
at tracking a setpoint (sp) pressure through the following sequence of
events. The baroreceptors sense Pa(t) and then
relay this pressure via autonomic afferent fibers to the autonomic
centers in the brain. Here, the autonomic nervous system (ANS) compares
the deviation between the sensed pressure and P



P
|
(1) |
|
|
|
Cardiopulmonary baroreflex.
The baroreflex system also includes a cardiopulmonary (CP)
baroreflex arc, which is conceptualized with a feedback diagram analogous to Fig. 3. However, the sensed pressure here is
defined to be the effective right atrial transmural pressure
[P
Pth(t)] of
the pulsatile heart and circulation model (see Fig. 2), where
P
Direct neural coupling mechanism between respiration and HR. The short-term regulatory system model also includes a direct neural coupling mechanism between respiration and HR. This coupling mechanism is implemented in terms of an LTI impulse response, which maps fluctuations in ILV to fluctuations in F(t). In accordance with previous studies (29, 40) with experimental data, the impulse response is defined here by a linear combination of s(t) and p(t), each of which are advanced in time by 1.5 s. The model ILV signal [Qlu(t)] that is necessary for the implementation of this mechanism is described below.
Parameter values.
The numerical values of the parameters characterizing the baroreflex
system are taken from (13 and T. Heldt, E. B. Shim, R. D. Kamm, and R. G. Mark, unpublished observations) (see below for
exceptions) and are provided in Table 1.
Note that the cardiopulmonary baroreflex control of
F(t) and
c
- and
-sympathetic sublimbs are both characterized
by s(t), the table specifies the particular sublimb
corresponding to each s(t). This permits the modeling of,
for example, the effects of propranolol, which selectively blocks the
-sympathetic sublimb.
|
0.0002 beats/s for
s(t) and 0.00012 beats/s for p(t). These values
indicate that upon inspiration, there is a withdrawal of
parasympathetic activity, followed by a withdrawal of
-sympathetic activity.
Resting Physiological Perturbations
Respiratory activity.
The resting physiological perturbations model includes both metronome
(met) and random-interval respiratory activity represented in terms of
Qlu(t). Because ILV strongly resembles a
sinusoid during metronomic breathing, Qlu(t) is
mathematically represented as follows
|
(2) |
|
(3) |
|
|
(4) |
pa(t) (R. Mukkamala, D. A. Sherman,
R. G. Mark and R. J. Cohen, unpublished observations).
|
Autoregulation of local vascular beds.
Although respiratory-induced hemodynamic fluctuations are fairly
well understood, the mechanisms responsible for hemodynamic variability
at frequencies below ~0.1 Hz have not been adequately elucidated. It
has been previously demonstrated that ABP fluctuations are not
determined by HR fluctuations at these lower frequencies, and it has
therefore been postulated that fluctuations in
Ra(t) caused by the autoregulation of
local vascular beds are responsible for these ABP fluctuations, which,
in turn, induces HR fluctuations through the baroreflex (1,
24). However, little is known about the system dynamics
characterizing autoregulatory processes except that these processes are
relatively slow with a characteristic time constant from ~5 to
20 s (6). We therefore include into the model of
resting physiological perturbations a zeroth-order model of
autoregulatory processes. This model represents these processes as a
stochastic, exogenous disturbance to
Ra(t)
[nRa(t)] defined by a
Gaussian white noise process with zero mean and
2
variance that is bandlimited to 0.1 Hz.
1/f HR fluctuations.
Figure 1 demonstrates that ABP and ILV are not the only factors
responsible for perturbing HR at frequencies within ~0.05 Hz.
However, the other perturbing factors, which may include cerebral activity impinging on the ANS, are essentially unknown. It is well
known that HR fluctuations demonstrate fractal behavior over at least
four decades of frequency, in which the highest frequency decade is
within the frequency band of interest here (7, 37). We
therefore include in the resting physiological perturbations model an
exogenous, stochastic 1/f disturbance to
F(t) [nF(t)] The
disturbance is created by passing a zero-mean, Gaussian white noise
process with variance
2 through a cascade combination of
two LTI filters. One filter is of unit-area and is designed to have a
1/f magnitude squared frequency response over four decades
of frequency starting at 10
4 Hz (12, 20,
36). The other filter is defined by a linear combination of
s(t) (
-sympathetic) and p(t) (with arbitrary
weighting values of 12 beats/s each), because HR fluctuations are
almost exclusively mediated by the ANS (2). Importantly,
nF(t) as well as
nRa(t) are considered to
be unmeasurable quantities in the forward model.
Parameter values.
The parameter values characterizing the models of
Qlu(t) and ventilation are provided in Table
2. These values are taken from (21,
41).
|
2 and
2, which characterize
nRa(t) and
nF(t), respectively, are considered to be
free parameters whose values are chosen such that the model
low-frequency hemodynamic variability matched that determined from a
set of experimental data previously obtained from 12 healthy standing
humans breathing at a fixed rate of 0.25 Hz (10). The data
set consists of simultaneous, noninvasive measurements of ECG and ABP
(Finapres), as well as uncalibrated ILV derived from the ECG for each
subject. The specific procedure for choosing the parameter values based
on this data set is as follows. We computed the mean ± SD of the
power in three nonoverlapping frequency bands for HR and ABP over the
group of 12 subjects (see Table 3) by
using autoregressive spectral analysis (22). The free
parameters
2 and
2 were then tuned during
fixed-rate breathing excitation at 0.25 Hz, such that the power in
these frequency bands for F(t) and Pa(t) were near their respective human values.
To satisfy sufficiently this matching procedure, we also had to
increase the weighting values of the parasympathetic and
-sympathetic effector mechanisms by 33% from their original values
given in (13). However, this increase did not move these
values outside of their physiological range, as established by
experimental data (14). On the basis of this procedure, we
set
= 0.16 mmHg · s · ml
1 and
= 0.0035 beats/s.
|
| |
FORWARD MODEL-BASED EVALUATION PROCEDURE |
|---|
|
|
|---|
Gold Standard Cardiovascular System Identification Results
To evaluate the performance of the cardiovascular system identification method applied to data generated from the forward model, it is necessary to establish, in a manner independent of system identification, the impulse responses and power spectra of perturbing noise sources in Fig. 1, which characterize the forward model. These impulse responses and power spectra of perturbing noise sources may then be regarded as the gold standard cardiovascular system identification results against which the estimates may be compared.We establish the gold standards according to the literal mathematical
meaning of the impulse responses and the power spectra of the
perturbing noise sources. In particular, the impulse response is
defined to represent the output response of a system to an arbitrarily
narrow, unit-area input. The impulse response also completely
characterizes the input-output relationship of the physiological
coupling mechanisms here because of our LTI assumption. Hence, in
establishing each of the gold standard impulse responses, we apply an
impulse input to the desired point in the forward model and measure the
output response of interest while holding all other perturbations to
the output constant. For example, to establish the gold standard
ILV
ABP impulse response, we measure the ABP response to an impulse
input of ILV while holding HR constant and setting the exogenous
disturbance to Ra(t) to zero.
By definition, the perturbing noise source represents the residual variability in a measured output not due to the fluctuations in the other measured inputs. So, in establishing each of the gold standard power spectra of perturbing noise sources, we first measure the output of interest in the forward model while holding the two measured input fluctuations constant and then compute the power spectrum of the output. For example, to establish the gold standard power spectrum of NABP, we first measure ABP in the forward model while holding HR and ILV constant and then compute the power spectrum of the resulting ABP.
The specific procedures for establishing each of the gold standard impulse responses and power spectra of perturbing noise sources in Fig. 1 are described in APPENDIX B. This description includes the signal processing necessary for the estimated and gold standard impulse responses to be sampled identically.
Root-Normalized-Mean-Squared Error
To provide a compact means for evaluating the similarity of each of the estimates with respect to its corresponding gold standard, we utilize a statistic, which we refer to as the root-normalized-mean-squared error (RNMSE). Let the estimate be vector x with mean vector
and covariance
matrix
x (which reflects the uncertainty in
the estimate) and its corresponding gold standard be vector
x0. The RNMSE is then defined as follows
|
(5) |
|
x is assumed to be a matrix of zeros.
Monte Carlo Simulations
Although the impulse response estimates include a measure of uncertainty in terms of a covariance matrix, this uncertainty measure is only an estimate itself (31). To account more accurately for estimation error variance, we report the mean ± SD for the impulse response estimates as well as for the power spectral estimates and RNMSE statistics as determined from 20 different realizations of forward model generated data.| |
FORWARD MODEL VALIDATION |
|---|
|
|
|---|
The forward model-based evaluation of the cardiovascular system identification method is only meaningful provided that the low-frequency power spectral content of the signals analyzed by the method is realistic. Because the forward model is tuned to represent what one may think of as an "average" subject, we initially considered demonstrating that the model spectra resemble the group average spectra of the 12 healthy humans breathing at a fixed rate of 0.25 Hz (see Parameter Values in Resting Physiological Perturbations). However, averaging tended to smear out the spectral peaks, which were usually not centered at the same frequencies for each subject. We therefore demonstrate that the model spectra resemble what one may find experimentally through a comparison with the spectra from one of the healthy subjects. This does not seem unreasonable if we keep in mind that the power in three nonoverlapping frequency bands of the model spectra quantitatively match that computed from the group average of the 12 subjects (see Table 3).
Figure 6 illustrates that the power
spectra for ILV, HR, and ABP at frequencies below the mean HR obtained
from the forward model during fixed-rate breathing excitation at 0.25 Hz indeed resemble spectra obtained from one of the 12 subjects.
Differences at the respiratory frequency in Fig. 6 and in Table 3 may
be attributed to the fact that the subject here did not precisely follow the fixed-rate breathing protocol as well as discrepancies in
tidal volume, which was not measured. Note that although the forward
model was tuned to represent the "average" subject, it could have
been tuned to match more precisely the subject here.
|
Figure 6 also demonstrates that the model spectra exhibit a spectral
peak at ~0.07 Hz, which is near the center frequency of the
"posture" peak in humans (32). Because the exogenous disturbances within 0.1 Hz are broadband (see Autoregulation of Local Vascular Beds and 1/f HR
Fluctuations), this peak implies a system resonance in the forward
model. This result supports the simple computer simulation of de Boer
(14), which demonstrated that the "posture" peak could
be due to a system resonance. de Boer specifically implicated the
system resonance to the arterial Ra(t) baroreflex arc. However, based
on a few simple experiments, in which the weighting values of the
open-loop effector mechanisms were varied, we have found that the
resonance peak is substantially diminished in the absence of the
arterial baroreflex feedback pathways responsible for controlling
Q
-sympathetic activity in general is involved
in eliciting the resonance peak.
It is also necessary to demonstrate that the cross spectra between each of these signals are realistic. This essentially amounts to demonstrating that the transfer functions, which characterize the couplings between the signals are consistent with those determined from experimental data. However, this has been virtually taken care of by the very construction of the model, which was based largely on physiological findings published in the literature.
Although it is only necessary for the purposes of this study that the
low-frequency spectral content of ILV, HR, and ABP be reasonable, we
assume that the low-frequency content of the remaining model signals
are accounted for as well by virtue of reasonably representing these
three signals. Figure 7 gives some
credence to this assumption by illustrating that the model spectrum of left ventricular flow rate (LVFR) at frequencies below the mean HR
resembles the corresponding spectrum measured with a Doppler ultrasound
technique (17) from an individual subject breathing at a
fixed rate of 0.25 Hz and tilted upright (30° with respect to the
supine posture).
|
| |
FORWARD MODEL-BASED EVALUATION |
|---|
|
|
|---|
Accuracy Analysis
We applied the cardiovascular system identification method to the beat-to-beat variability generated from the forward model during random-interval breathing excitation. Figure 8 illustrates that there is a close agreement between the resulting cardiovascular system identification estimates and the corresponding gold standards. These results indicate the validity of the cardiovascular system identification method to the extent that the forward model coincides with the actual cardiovascular system.
|
Robustness Analysis
Perhaps the major source of uncertainty in the relevant properties of the forward model involves the system dynamics responsible for controlling HR. We therefore consider here the robustness of the HR baroreflex, ILV
HR, and NHR estimates over a set of forward models, which reflect some of our uncertainty in these system dynamics.
Cardiopulmonary HR baroreflex. We first consider how these estimates would be altered if a cardiopulmonary HR baroreflex were included in the forward model. Inclusion of this reflex may be achieved simply by adjusting its weighting values to nonzero values (see Table 1). To proceed with this analysis, the gold standard NHR power spectrum, which is no longer valid in the presence of a cardiopulmonary HR baroreflex, must be redefined (see APPENDIX C for the specific procedure).
Figure 9, A-C, illustrates the RNMSE results of the HR baroreflex, ILV
HR, and
NHR estimates as a function of the ratio of the static
gains of the cardiopulmonary to arterial HR baroreflexes. The range of
this ratio is determined from published experimental data (15,
35). Note that a negative ratio indicates a Bainbridge type of
cardiopulmonary baroreflex, whereas a positive ratio indicates a type
of cardiopulmonary baroreflex, which contributes to ABP regulation. The
results indicate that the ILV
HR impulse response estimate is most
significantly and adversely influenced by the presence of the
cardiopulmonary HR baroreflex. This result implies that low-frequency
effective right atrial transmural pressure fluctuations are determined
largely by ILV fluctuations.
|
HR impulse response estimate (consider RNMSE standard
deviation). We hypothesize that this discrepancy between positive and
negative ratios is due to the size of low-frequency ABP fluctuations.
That is, in contrast to the positive ratio, which results in tighter
ABP regulation, the negative ratio leads to sufficiently large
low-frequency ABP fluctuations such that they contribute somewhat to
the generation of low-frequency effective right atrial transmural
pressure fluctuations in the model.
Figure 9B illustrates that the ILV
HR estimate becomes
progressively worse with respect to its corresponding gold standard with increasing absolute static gain of the cardiopulmonary HR baroreflex. However, this does not necessarily imply that the estimate
is meaningless for large absolute static gain values. In fact, the
estimate is quite meaningful regardless of the static gain value of the
cardiopulmonary HR baroreflex provided that we redefine the gold
standard such that it encompasses the dynamics of both the direct
neural coupling mechanism and the cardiopulmonary HR baroreflex (see
APPENDIX C for the specific procedure). Figure
9D shows that the RNMSE results for the ILV
HR estimate with respect to the new gold standard is essentially independent of the
ratio of the static gain values of the cardiopulmonary to arterial HR
baroreflexes. The results here suggest that when considering the
ILV
HR impulse response identified from experimental data, it is
probably more accurate to interpret the estimate analogously to the
gold standard as redefined here.
Arterial baroreflex saturation.
The S parameter in Eq. 1 characterizes the degree
of arterial baroreflex saturation in the forward model. This parameter
provides an upper bound on the deviation between sensed ABP and its
setpoint pressure so that the manipulated variables cannot be
controlled to arbitrarily high or low values. The maximum deviation
permitted by this parameter in the forward model is ~28 mmHg. This
pressure is approximately equal to the range of ABP, over which the
atan mapping in Eq. 1 differs from true linearity by ~10%
(that is, the linear ABP range). However, experimental data from one
study (25) indicate that the linear ABP range may actually
be only ~10 mmHg. We therefore consider the robustness of the HR
baroreflex, ILV
HR, and NHR estimates against more narrow
linear ABP ranges.
HR, and NHR estimates as a
function of the linear ABP range of the arterial baroreflex. These
results indicate that only the HR baroreflex estimate is significantly
and adversely affected by increasing degrees of saturation.
Furthermore, the deviation of the RNMSE from its nominal value only
becomes significant when the extent of saturation is no longer
substantiated by the experimental data in (25). Hence,
provided that arterial baroreflex saturation is the only significant
nonlinearity, then the assumption that the fluctuations in
cardiovascular system identification data are small enough such that
the couplings between the fluctuations are related linearly seems quite
tenable.
|
Sensitivity Analysis
The ultimate potential of the cardiovascular system identification method is to provide a clinician with an intelligent, sensitive tool for monitoring a patient's cardiovascular state over time so as to guide therapy. An analysis of the resolving power of the method is germane to the realization of this potential. We therefore consider here an analysis of the sensitivity of the method particularly in terms of detecting changes in autonomic activity, as reflected by the weighting values for p(t) and
-sympathetic s(t) in the forward model, through the HR baroreflex,
ILV
HR, and NHR estimates.
Figure 11, A and
B, shows the sensitivity results to changes in autonomic
function in terms of the estimated versus actual percentage of change
in the static gains (sum of the weighting values) of the autonomically
mediated impulse responses, where the percentage of change here is with
respect to the nominal static gain values. These results indicate that
a change of 25% in the static gain of the gold standard HR baroreflex
impulse response is required to detect reliably a change in the static
gain of the corresponding estimate. On the other hand, a 25% change in the static gain of the gold standard ILV
HR impulse response is not
sufficient to detect reliably a change in the static gain of the
corresponding estimate.
|
On the basis of our experience with experimental human data (27,
29), we have found that the absolute peak amplitude of the
impulse response estimates seems to be quite reliable in detecting changes in autonomic function. Figure 11, C and
D, shows the sensitivity results in terms of the estimated
versus actual percentage of change in the absolute peak amplitude of
the autonomically mediated impulse responses. Because we varied the
static gains of the impulse responses simply via scaling, the actual
percentage of change in the static gain of the impulse responses is
equal to the actual percentage of change in the absolute peak
amplitude. The results indicate substantial improvement in sensitivity
with respect to the ILV
HR estimate and some improvement with respect
to the HR baroreflex estimate. That is, as small as a 10% change in
the absolute peak amplitude of the gold standard ILV
HR impulse
response is sufficient to detect reliably a change in the absolute peak amplitude of the corresponding estimate.
The substantial improvement in the sensitivity of the ILV
HR impulse
response may be explained by considering the ratio of the spectra of
ILV to the spectra of the unmeasured HR disturbance (signal-to-noise
ratio as a function of frequency) in the model. This ratio is smallest
at frequencies near 0 Hz because of the 1/f character of the
unmeasured disturbance as well as the reduced ILV spectral content near
0 Hz due to the probability density defining the random-interval
respiratory pattern (see Respiratory Activity). The
static gain of the ILV
HR impulse response estimate, which reflects
the 0 Hz estimate, is therefore not reliable. However, the ratio is
larger at higher frequencies, and the absolute peak amplitude, which
reflects the wider bandwidth parasympathetic filter, is therefore quite
reliable. On the other hand, the spectral content of ABP is
sufficiently significant with respect to that of the unmeasured
1/f HR fluctuations at frequencies near 0 Hz such that the
static gain of the HR baroreflex impulse response is fairly reliable.
Figure 12 illustrates the sensitivity
results in terms of the estimated and actual percentage of change in
total, low-frequency (0-0.15 Hz), and high-frequency
(0.15-0.4 Hz) power of the NHR spectra. These results
emphasize the superior reliability of the high-frequency components of
the estimates due to the 1/f character of the unmeasured HR
disturbance. Note that the high-frequency power of NHR is
just as sensitive a measure of parasympathetic function as the absolute
peak amplitude of ILV
HR. However, we believe that this absolute peak
amplitude may be a better measure of parasympathetic function, because
the perturbing noise source NHR, unlike ILV
HR, is not
normalized for inputs, e.g., cerebral activity.
|
| |
DISCUSSION |
|---|
|
|
|---|
Previous Forward Models
A few forward models of the cardiovascular system have been previously developed for the purposes of eliciting and analyzing low-frequency hemodynamic variability. de Boer (14) presented a beat-to-beat model, which incorporated windkessel and Starling properties with arterial baroreflex control of HR and systemic arterial resistance. The model was perturbed with sinusoidal respiration and stochastic disturbances to HR and pulse pressure to generate resting hemodynamic fluctuations. Madwed et al. (23) developed a model to study the generation of Mayer waves. Their model included windkessel properties, constant stroke volume, arterial baroreflex control of HR and systemic arterial resistance, and sinusoidal respiration. Cavalcanti (8) presented a model that incorporated windkessel properties with nonlinear arterial baroreflex control of HR and assumed stroke volume and systemic arterial resistance to be constant. Despite the absence of exogenous perturbations, the model was demonstrated to elicit HR fluctuations due to the nonlinear delayed structure of closed-loop control.None of these simple models are nearly as comprehensive as our forward model in terms of accounting for pulsatility, the short-term regulatory system, and resting physiological perturbations. Our forward model thus provides a more complete means for studying the mechanisms responsible for eliciting beat-to-beat variability compared with these oversimplified models. Consider, for example, our more thorough analysis of the physiological mechanisms responsible for the system resonance phenomenon with respect to the study of de Boer (14) (see FORWARD MODEL VALIDATION).
Forward Model Limitations
The mechanisms responsible for eliciting low-frequency hemodynamic variability, particularly at frequencies below ~0.1 Hz, have not been fully elucidated. Nevertheless, we constructed a forward model of the human cardiovascular system that is capable of generating reasonable low-frequency hemodynamic variability and is largely consistent with previous experimental findings. However, the forward model presented here is not the unique model for simultaneously realizing this behavior and being consistent with experimental findings. That is, we could alter the properties of the forward model and still generate realistic low-frequency hemodynamic variability. For example, the resonance peak at ~0.07 Hz could also be elicited with a stochastic, exogenous disturbance to Q
Robustness to Other Forward Model Uncertainties
Although we evaluated how the presence of a cardiopulmonary HR baroreflex and varying degrees of arterial baroreflex saturation would affect the autonomically mediated estimates, we did not address the effects of nonstationarities or other types of nonlinearities on these estimates. However, the stationarity property of the forward model seems quite tenable given that the data are considered over short time periods during stable experimental conditions. The stationarity assumption is further supported by preliminary studies that we have conducted, in which the impulse response estimates do not change much from adjacent time periods of data collection. Although hysteresis and nonlinear interaction between the arterial and cardiopulmonary baroreflexes have been previously reported (25, 30), such complex behaviors are usually elicited during extreme experimental conditions. Furthermore, nonlinear models have been shown to provide only a modest improvement with respect to linear models in accounting for HR fluctuations measured during the relatively stable experimental conditions of system identification data collection (11). Hence, the inclusion of only the ubiquitously reported arterial baroreflex saturation as a regulatory system nonlinearity in the forward model seems quite reasonable. However, we acknowledge that if significant system nonstationarities and/or nonlinearities excluding baroreflex saturation were present, then the autonomically mediated estimates could be substantially altered with respect to their gold standards.We also did not include an analysis of the effects of the size of the
1/f HR fluctuations on the estimates. This disturbance may
be thought of as the unmeasured noise in the identification of the
autonomically mediated impulse responses, HR baroreflex and ILV
HR.
The relative contribution of this noise term with respect to HR
fluctuations due only to ABP and ILV fluctuations reflects the
signal-to-noise ratio of this identification problem. This relative
contribution, specifically in terms of the ratio of the standard
deviation of those HR fluctuations not attributed to linear couplings
with ABP and ILV to the standard deviation of all HR fluctuations, has
been reported to be 0.31 ± 0.14 based on experimental data obtained
from humans breathing randomly (11). However, the ratio of
the standard deviation of the 1/f HR fluctuations to the
standard deviation of HR fluctuations in the forward model is 0.49 ± 0.07. In fact, we have found that none of the autonomically mediated
estimates in Fig. 8 are significantly altered for ratios up to ~0.6.
We also did not analyze the robustness of the mechanically
mediated estimates (circulatory mechanics, ILV
ABP, and
NABP) to uncertainties in the pulsatile heart and
circulation model (i.e., nonlinearities in systemic arterial
compliance). However, we believe that this is not too critical because
linear lumped parameter models are quite reasonable when small,
low-frequency hemodynamic fluctuations are being analyzed.
Sensitivity Analysis Results Supported By Experimental Data Studies
From experimental human data (27, 29), we have found the absolute peak amplitude of ILV
HR to be the most sensitive
measure of autonomic function, followed by the power in the
NHR spectra (especially the high-frequency power), the
absolute peak amplitude of HR baroreflex, and the static gain of HR
baroreflex. In contrast, we have never found the static gain of
ILV
HR to be predictive of changes in autonomic function. The fact
that the more precise sensitivity analysis here (see Sensitivity
Analysis) retrospectively predicts the resolving power of the
estimates determined from experimental human data helps confirm the
validity of the low-frequency hemodynamic fluctuations generated by the
forward model during random-interval breathing excitation.
In addition to the static gain and absolute peak amplitude of the
impulse response estimates, we have considered another parameter for
detecting autonomic changes in experimental human data, which we refer
to as the characteristic time parameter and define as follows
|
(6) |
-sympathetic function.
For example, simply scaling the impulse response, as we have done (see
Sensitivity Analysis), does not change the characteristic
time because there is no shift in balance in the activity of the two
autonomic limbs. On the basis of our sensitivity analysis results, we
expect the ILV
HR impulse response estimate to be unable to measure
true shifts in balance through the characteristic time parameter,
because only the parasympathetic limb can be estimated accurately.
However, we do expect the characteristic time parameter of the HR
baroreflex impulse response estimate to be a somewhat sensitive measure
of true shifts in balance between parasympathetic and
-sympathetic function. The validity of these expectations is supported by our analysis with experimental human data (27, 29), in which
only the characteristic time parameter of the HR baroreflex impulse response estimate was found to be sensitive to changes in autonomic function.
In conclusion, based on the results of the forward model-based
evaluation of the cardiovascular system identification method, we make
the following inferences about the performance of the method with
respect to experimental data. First, each of the cardiovascular system
identification estimates is likely to reflect the system dynamics of
actual physiological mechanisms. Second, the ILV
HR impulse response
estimate encompasses both direct neural coupling and cardiopulmonary HR
baroreflex mechanisms. Third, arterial baroreflex saturation and the
relative contributions of the HR fluctuations independent of ILV and
ABP are unlikely to affect significantly the autonomically mediated
estimates. Fourth, the absolute peak amplitude of the ILV
HR impulse
response estimate is a very sensitive measure of parasympathetic
function. Finally, the HR baroreflex impulse response estimate is a
reasonably sensitive measure of both parasympathetic function
(through absolute peak amplitude and static gain) and
-sympathetic
function (through static gain) and consequently shifts in balance
between the two autonomic limbs (through characteristic time).
The theoretical evaluation therefore provides confidence in the performance of the method with respect to experimental data and demonstrates the power of the cardiovascular system identification method. That is, the method is able to estimate reliably physiologically important impulse responses and perturbing noise sources by analyzing beat-to-beat hemodynamic variability obtained during near normal conditions rather than extreme experimental conditions, which would be required by the literal mathematical meaning of these quantities. This study supports system identification as a powerful approach for the intelligent patient monitoring of cardiovascular function. Moreover, the forward model, on which the theoretical validation is based, provides a convenient test bed of data, which may facilitate the development of new methods that could be incorporated with the cardiovascular system identification method so as to provide a more detailed picture of cardiovascular state (26).
| |
APPENDIX A |
|---|
|
|
|---|
Forward Model Implementation
The model of the pulsatile heart and circulation may be mathematically represented by a set of coupled, first-order differential equations, which are obtained with Kirchoff's Current Law and the constitutive relationship of each of the elements (R. Mukkamala, D. A. Sherman, R. G. Mark, and R. J. Cohen, unpublished observations). Given an initial set of pressures, the ensuing pressures, volumes, and flow rates are determined by numerically integrating this set of equations with a fifth-order Runge-Kutta method with adaptive step size (33). The average time step is ~0.008 s; however, the models of the short-term regulatory system and resting physiological perturbations are bandlimited to frequencies within the mean HR. For the purposes of reducing unnecessary computations, these models are implemented at a uniform sampling period of 0.0625 s. To implement simultaneously the entire forward model, it is therefore necessary to transform nonuniformly sampled signals to uniformly sampled signals and vice versa. The former transformation is achieved by averaging over the most recent 0.25 s every 0.0625 s, whereas the latter transformation is accomplished with linear interpolation.| |
APPENDIX B |
|---|
|
|
|---|
Specific Procedure For Establishing Gold Standards
The gold standard circulatory mechanics impulse response, which represents the ABP waveform that results from a single ventricular contraction, is established through the superposition principle. That is, the forward model is first executed for n, and then n
1, ventricular contractions while
nRa(t) is set to 0 and
Qlu(t) and F(t) are held
constant. The two resulting nonuniformly sampled
Pa(t) signals are resampled to 90 Hz, and their
difference is defined to be the gold standard.
To establish the gold standard HR baroreflex impulse response, the
feedback pathway of the arterial HR baroreflex arc, which is defined by
Eq. 1 and Table 1, is removed from closed-loop operation.
The input applied to this isolated feedback pathway resembles an
impulse and is mathematically expressed as follows
|
(B1) |
ABP, the area of the
"impulse," is set to the standard deviation of the ABP fluctuations
analyzed by the cardiovascular system identification method. The
resulting F(t) is then resampled to 1.5 Hz and
normalized by
ABP to arrive at the gold standard.
The gold standard ILV
HR impulse response is predefined by virtue of
its forward model implementation (see Direct Neural Coupling Mechanism Between Respiration and HR). It is only necessary to resample this impulse response to 1.5 Hz to establish the gold standard.
The gold standard ILV
ABP impulse response is determined by applying
an impulse of Qlu(t) to the forward model while
nRa(t) is set to 0 and
F(t) is held constant. The precise input applied here is analogous to Eq. B1 with the area of the
"impulse" set to the standard deviation of the ILV fluctuations
analyzed by the cardiovascular system identification method
(
ILV). The resulting Pa(t) is
resampled to 1.5 Hz with its mean removed and normalized by
ILV to arrive at the gold standard.
The gold standard NHR perturbing noise source is precisely
given by nF(t). The power spectrum of this
signal may therefore be given by the product of
2 and
the magnitude squared frequency response of the 1/f filter and the filter defined by the linear combination of s(t) and
p(t) (see 1/f HR Fluctuations).
The gold standard NHR perturbing noise source is determined by executing the forward model while F(t) and Qlu(t) are held constant and only nRa(t) is active. The resulting Pa(t) is resampled to 1.5 Hz with its mean removed, and then the power spectrum of this signal is computed by autoregressive spectral analysis. This procedure is repeated for 10 different realizations of forward model data, with the average power spectrum defined to be the gold standard.
| |
APPENDIX C |
|---|
|
|
|---|
Specific Procedure For Redefining Gold Standards With a Cardiopulmonary HR Baroreflex
The gold standard NHR perturbing noise source in the presence of a cardiopulmonary HR baroreflex is established by first executing the forward model, in which Pa(t) is fixed to P
The gold standard ILV
HR impulse response in the presence of a
cardiopulmonary HR baroreflex is established by applying an impulse of
Qlu(t) (see Eq. B1) to the forward
model with a voltage source in lieu of the systemic arterial capacitor
as described above, while nF(t) and
nRa(t) are set
to zero. The resulting F(t) is resampled to 1.5 Hz with its mean removed and normalized by
ILV to arrive
at the newly defined gold standard.
| |
ACKNOWLEDGEMENTS |
|---|
The authors thank Dr. Janice Meck, National Aeronautics and Space Administration (NASA) Johnson Space Center, for providing the experimental data utilized in this work, and Roger G. Mark for useful comments.
| |
FOOTNOTES |
|---|
This work was supported by NASA Grant NAG5-4989, a National Space Biomedical Research Institute grant, and National Institutes of Health Grant P41-RR-13622.
Address for reprint requests and other correspondence: R. Mukkamala, MIT, E25-505, 45 Carleton St., Cambridge, MA 02139 (E-mail: rmukkama{at}mit.edu).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 12 January 2001; accepted in final form 3 August 2001.
| |
REFERENCES |
|---|
|
|
|---|
1.
Akselrod, S,
Gordon D,
Madwed JB,
Snidman NC,
Shannon DC,
and
Cohen RJ.
Hemodynamic regulation: investigation by spectral analysis.
Am J Physiol Heart Circ Physiol
249:
H867-H875,
1985
2.
Akselrod, S,
Gordon D,
Ubel FA,
Shannon DC,
Barger AC,
and
Cohen RJ.
Power spectrum analysis of heart rate fluctuation: a quantitative probe of beat-to-beat cardiovascular control.
Science
213:
220-222,
1981
3.
Berger, RD,
Saul JP,
and
Cohen RJ.
An efficient algorithm for spectra analysis of heart rate variability.
IEEE Trans Biomed Eng
33:
900-904,
1986[Web of Science][Medline].
4.
Berger, RD,
Saul JP,
and
Cohen RJ.
Assessment of autonomic response by broad-band respiration.
IEEE Trans Biomed Eng
36:
1061-1065,
1989[Web of Science][Medline].
5.
Berger, RD,
Saul JP,
and
Cohen RJ.
Transfer function analysis of autonomic regulation. I. Canine atrial rate response.
Am J Physiol Heart Circ Physiol
256:
H142-H152,
1989
6.
Berne, RM,
and
Levy MN.
Cardiovascular Physiology. St. Louis, MO: Mosby, 1986.
7.
Bigger, JT, Jr,
Steinman RC,
Rolnitzky LM,
Fleiss JL,
Albrecht P,
and
Cohen RJ.
Power law behavior of RR-interval variability in healthy middle-aged persons, patients with recent acute myocardial infarction, and patients with heart transplants.
Circulation
93:
2142-2151,
1996
8.
Cavalcanti, S.
Arterial baroreflex influence on heart rate variability: a mathematical model-based analysis.
Med Biol Eng Comput
38:
189-197,
2000[Web of Science][Medline].
9.
Cevenini, G,
Barbini P,
Cappello A,
and
Avanzolini G.
Three-element model for total systemic circulation: emphasis on the accuracy of parameter estimates.
J Biomed Eng
9:
374-378,
1987[Web of Science][Medline].
10.
Charles, JB,
Fritsch-Yelle JM,
Whitson PA,
Wood ML,
Brown TE,
and
Fortner GW.
Cardiovascular deconditioning.
In: Extended Duration Orbiter Medical Project: Final Report 1989-1995, edited by Sawin CF,
Taylor GR,
and Smith WL.. Houston, TX: National Aeronautics and Space Administration, 1999, p. 1-19.
11.
Chon, KH,
Mullen TJ,
and
Cohen RJ.
A dual-input nonlinear system analysis of autonomic modulation of heart rate.
IEEE Trans Biomed Eng
43:
530-544,
1996[Web of Science][Medline].
12.
Corsini G and Saletti R. Design of a digital 1/f fl noise
simulator. In: Noise in Physical Systems and 1/f Noise.
Singapore, 1987, p. 82-86.
13.
Davis, TL.
Teaching Physiology Through Interactive Simulation of Hemodynamics (MS thesis). Cambridge, MA: Massachusetts Institute of Technology, 1991.
14.
De Boer, RW,
Karemaker JM,
and
Strackee J.
Hemodynamic fluctuations and baroreflex sensitivity in humans: a beat-to-beat model.
Am J Physiol Heart Circ Physiol
253:
H680-H689,
1987
15.
Desai, TH,
Collins JC,
Snell M,
and
Mosqueda-Garcia R.
Modeling of arterial and cardiopulmonary baroreflex control of heart rate.
Am J Physiol Heart Circ Physiol
272:
H2343-H2352,
1997
16.
Deswysen, B,
Charlier AA,
and
Gevers M.
Quantitative evaluation of the systemic arterial bed by parameter estimation of a simple model.
Med Biol Eng Comput
18:
153-166,
1980[Web of Science][Medline].
17.
Eriksen, M,
and
Walloe L.
Improved method for cardiac output determination in man using ultrasound doppler technique.
Med Biol Eng Comput
28:
555-560,
1990[Web of Science][Medline].
19.
Jacob, F,
Ertl AC,
Shannon JR,
Furlan R,
Robertson RM,
and
Robertson D.
Effect of standing on neurohumoral responses and plasma volume in healthy subjects.
J Appl Physiol
84:
914-921,
1998
20.
Keshner, MS.
1/f noise.
Proc IEEE
70:
212-218,
1982.
21.
Lambertsen, CJ.
Physical and mechanical aspects of respiration.
In: Medical Physiology, edited by Mountcastle VB.. St. Louis, MO: Mosby, 1974, p. 1361-1371.
22.
Ljung, L.
System Identification: Theory for the User. Englewood Cliffs, NJ: Prentice Hall, 1987.
23.
Madwed, JB,
Albrecht P,
Mark RG,
and
Cohen RJ.
Low-frequency oscillations in arterial pressure and heart rate: a simple computer model.
Am J Physiol Heart Circ Physiol
256:
H1573-H1579,
1989
24.
Madwed, JB,
and
Cohen RJ.
Heart rate response to hemorrhage-induced 0.05 Hz oscillation in arterial pressure in conscious dogs.
Am J Physiol Heart Circ Physiol
260:
H1248-H1253,
1991
25.
Mancia, G,
and
Mark AL.
Arterial baroreflexes in humans.
In: Handbook of Physiology. The Cardiovascular System. Peripheral Circulation and Organ Blood Flow. Bethesda, MD: Am Physiol Soc, 1983, sect. 2, vol. III, pt. 2, chapt. 20, p. 755-793.
26.
Mukkamala, R.
A Forward Model-Based Analysis of Cardiovascular System Identification Methods (PhD thesis). Cambridge, MA: Massachusetts Institute of Technology, 2000.
27.
Mukkamala, R,
Mathias JM,
Mullen TJ,
Cohen RJ,
and
Freeman R.
System identification of closed-loop cardiovascular control mechanisms: diabetic autonomic neuropathy.
Am J Physiol Regulatory Integrative Comp Physiol
276:
R905-R912,
1999
29.
Mullen, TJ,
Appel ML,
Mukkamala R,
Mathias JM,
and
Cohen RJ.
System identification of closed-loop cardiovascular control: effects of posture and autonomic blockade.
Am J Physiol Heart Circ Physiol
272:
H448-H461,
1997
30.
Pawelczyk, JA,
and
Raven PB.
Reductions in central venous pressure improve carotid baroreflex responses in conscious men.
Am J Physiol Heart Circ Physiol
257:
H1389-H1895,
1989
31.
Perrott, MH,
and
Cohen RJ.
An efficient approach to ARMA modeling of biological systems with multiple inputs and delays.
IEEE Trans Biomed Eng
43:
1-14,
1996[Web of Science][Medline].
32.
Pomeranz, B,
Macaulay RJB,
Caudill MA,
Kutz I,
Shannon DC,
Cohen RJ,
and
Benson H.
Assessment of autonomic function in humans by heart rate spectral analysis.
Am J Physiol Heart Circ Physiol
248:
H151-H153,
1985
33.
Press, WH,
Flannery BP,
Teukolsky SA,
and
Vetterling WT.
Numerical Recipes in C: The Art of Scientific Computing. New York: Cambridge University Press, 1992.
34.
Ramsdell, CD,
Sundby GH,
Sherman D,
Maa M,
Baskin JL,
Williams GH,
and
Cohen RJ.
Non-invasive assessment of alterations in cardiovascular regulation and function and susceptibility to ventricular arrhythmias resulting from microgravity exposure.
In: Space Technology and Applications International Forum 2000, edited by El-Genk MS.. New York: American Institute of Physics, 2000, p. 281-294.
35.
Raymundo, H,
Scher AM,
O'Leary DS,
and
Sampson PD.
Cardiovascular control by arterial and cardiopulmonary baroreceptors in awake dogs with atrioventricular block.
Am J Physiol Heart Circ Physiol
257:
H2048-H2058,
1989
36.
Saletti, R.
A comparison between two methods to generate 1/f noise.
Proc IEEE
74:
1595-1596,
1986.
37.
Saul, JP,
Albrecht P,
Berger RD,
and
Cohen RJ.
Analysis of long-term heart rate variability: Methods, 1/f scaling and implications.
In: Computers in Cardiology. Silver Spring, MD: IEEE Computer Society Press, 1987, p. 419-422.
38.
Suga, H,
and
Sagawa K.
Instantaneous pressure-volume relationships and their ratio in the excised, supported canine left ventricle.
Circ Res
35:
117-126,
1974
39.
Suga, H,
Sagawa K,
and
Shoukas AA.
Load independence of the instantaneous pressure-volume ratio of the canine left ventricle and effects of epinephrine and heart rate on the ratio.
Circ Res
32:
314-322,
1973
40.
Triedman, JK,
Perrott MH,
Cohen RJ,
and
Saul JP.
Respiratory sinus arrhythmia: time domain characterization using autoregressive moving average analysis.
Am J Physiol Heart Circ Physiol
268:
H2232-H2238,
1995
41.
Vander, AJ,
Sherman JH,
and
Luciano DS.
Human Physiology. New York: McGraw-Hill, 1994.
This article has been cited by other articles:
![]() |
X. Chen, J.-K. Kim, J. A. Sala-Mercado, R. L. Hammond, R. I. Elahi, T. J. Scislo, G. Swamy, D. S. O'Leary, and R. Mukkamala Estimation of the total peripheral resistance baroreflex impulse response from spontaneous hemodynamic variability Am J Physiol Heart Circ Physiol, January 1, 2008; 294(1): H293 - H301. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Mukkamala, K. Toska, and R. J. Cohen Noninvasive identification of the total peripheral resistance baroreflex Am J Physiol Heart Circ Physiol, March 1, 2003; 284(3): H947 - H959. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |