|
|
||||||||
Cardiovascular Research Institute, Departments of 1 Biophysics and 2 Physiology, Maastricht University, 6200MD Maastricht; and Departments of 3 Cardiothoracic Surgery and 4 Radiology, Academic Hospital Maastricht, 6202AZ Maastricht, The Netherlands
| |
ABSTRACT |
|---|
|
|
|---|
Aortic valve stenosis impairs subendocardial perfusion with a risk of irreversible subendocardial tissue damage. A likely precursor of damage is subendocardial contractile dysfunction, expressed by the parameter TransDif, which is defined as epicardial minus endocardial myofiber shortening, normalized to the mean value. With the use of magnetic resonance tagging in two short-axis slices of the left ventricle (LV), TransDif was derived from LV torsion and contraction during ejection. TransDif was determined in healthy volunteers (control, n = 9) and in patients with aortic valve stenosis before (AVSten, n = 9) and 3 mo after valve replacement (AVRepl, n = 7). In the control group, TransDif was 0.00 ± 0.14 (mean ± SD). In the AVSten group, TransDif increased to 0.96 ± 0.62, suggesting impairment of subendocardial myofiber shortening. In the AVRepl group, TransDif decreased to 0.37 ± 0.20 but was still elevated. In eight of nine AVSten patients, the TransDif value was elevated individually (P < 0.001), suggesting that the noninvasively determined parameter TransDif may provide important information in planning of treatment of aortic valve stenosis.
aortic valve replacement; function; torsion
| |
INTRODUCTION |
|---|
|
|
|---|
MOST CARDIAC DISORDERS affect motion of the left ventricular (LV) wall (18, 19, 25). In aortic valve stenosis, coronary flow diminishes in the subendocardial layers relative to subepicardial ones (9, 16, 26). Ultimately, this may lead to subendocardial infarction and other structural changes (13-15). The risk to run into permanent damage is the most important reason for aortic valve replacement. Currently, this risk is estimated by combined assessment of hemodynamic parameters (peak systolic LV pressure, transvalvular pressure drop, and peak flow velocity), general LV pump function parameters (ejection fraction and maximum time derivative of LV pressure), effects of increased venous pressure (edema), and sound from turbulent flow. In the search for a clear indicative parameter, measurement of subendocardial contractile function may be promising. After all, underperfusion of myocardial tissue results in an immediate decrease of contractile function (10, 22) far before irreversible tissue damage occurs. The present study was conducted to investigate whether measurement of subendocardial contractile dysfunction is possible and whether it may provide additional clinically valuable information.
Commonly, myocardial contractile dysfunction is detected from wall motion abnormalities, as measured by echocardiography. In aortic valve stenosis patients, such abnormalities cannot be found easily, probably because the related contractile dysfunction is distributed smoothly over the whole inner side of the LV wall. In the last decade, the MRI-tagging technique has been developed to assess myocardial motion noninvasively (18, 20, 27). With this technique, during a period of a few milliseconds, the magnetization of the tissue is periodically modulated with a grid pattern, thus forming tags. During the cardiac cycle, the tags are moving with the tissue and are visualized with MRI. With the use of image analysis techniques, the tags are followed in time. Thus a map of motion of the visualized cross section is obtained. In the current study, we propose to use MRI tagging in detecting transmural differences in myocardial contractile function.
In model studies on cardiac mechanics (3, 5), it was predicted that a transmural gradient in contractile function was related to changes in the ratio of torsion to shortening (TSR) of the LV. Torsion is defined here as a base-to-apex gradient of rotation of the short-axis cross section around the LV long axis. When assuming myofiber shortening to be transmurally uniform, TSR was predicted to be a fixed number. In dogs with two-dimensional echocardiography (3) and in humans with MRI tagging (1), this prediction was confirmed. In humans using MRI tagging, the transmural relative variation of myofiber shortening was reported to be <10% (17, 23). In aortic valve stenosis patients, torsion was found to be elevated (21, 24). Torsion also changed by changes in preload, afterload, or contractility (8). We conclude that under normal conditions myofiber shortening is quite evenly distributed over the LV wall, a finding that was confirmed experimentally by finding a fixed ratio of TSR. During aortic valve stenosis, torsion may be modified by changes in the transmural distribution of myofiber shortening. In the present study, we investigated whether a parameter expressing a transmural difference in myofiber shortening can be derived from myocardial motion, as measured by MRI tagging.
In the normal heart, due to the specific helical pathways of the myofibers, torsion as induced by the subepicardial myofibers is partly counteracted by the subendocardial myofibers. With aortic valve stenosis, subendocardial myofibers are likely to be less active, causing torsion to increase. Because fewer fibers are fully active, ejection is expected to decrease. Thus TSR is expected to be sensitive to a decrease of myofiber shortening in the subendocardium relative to the subepicardium. If load of the heart changes by physiological mechanisms such as exercise or contractility variation, myofiber shortening is expected to change uniformly throughout the heart. Both torsion and shortening are then expected to change proportionally, leaving TSR unaffected.
In the present study, TSR was determined from the motion of MRI tags in two parallel short-axis cross sections. With the use of a model of LV mechanics, from the TSR value, the transmural difference in myofiber shortening (TransDif), as normalized to the mean value, was estimated. Measurements were performed in healthy volunteers (control) and in aortic valve stenosis patients before (AVSten) and 3 mo after aortic valve replacement (AVRepl). It was investigated whether TSR and TransDif were affected by aortic valve stenosis. Finally, we investigated whether subendocardial myofiber shortening recovered after aortic valve replacement.
| |
METHODS |
|---|
|
|
|---|
Patient study. The study was conducted on 9 healthy volunteers (2 women and 7 men, age 31 ± 4 yr) and 10 patients (2 women and 8 men) with severe aortic valve stenosis, characterized by an aortic valve pressure gradient >50 mmHg. The AVSten group of patients (age 64 ± 9 yr) contained nine patients before aortic valve replacement. The AVRepl group of patients (age 68 ± 6 yr) contained seven patients 3 mo after aortic valve replacement. From the latter group, six individuals also belonged to the AVSten group. Written informed consent was obtained from all subjects.
Image acquisition. MRI experiments were performed at either 0.5 or 1.5 T (Philips Gyroscan T5 II or Gyroscan NT, Philips Medical Systems; Best, The Netherlands). All images were acquired using ECG triggering while patients were breathing normally. Images of two short-axis cross sections of the heart were obtained, ~2 and 4 cm below the base. Sixteen regular cine images per slice were acquired using a nontagged steady-state gradient echo sequence, starting 20 ms after the ECG trigger on the peak of the R wave and continuing with time intervals of either 20 or 62 ms (echo time 10 ms, slice thickness 8 mm, field of view 213 mm, image size 256 × 256 pixels). Thereafter, a series of tagged images from the same two slices were obtained with time intervals of 20 ms, using spatial modulation of the magnetization (6). Initial intertag distance was 5 mm with a tag width of 2.5 mm in both horizontal and vertical directions. To obtain a sufficient contrast-to-noise ratio for a longer time interval, two overlapping series of tagged images were acquired for each slice. Acquisition of the first set of MRI-tagged images was started 20 ms after the ECG trigger. The second set of MRI-tagged images started 130-160 ms later.
Image analysis.
MR images were analyzed to quantify torsion and the decline in cavity
area, as normalized to wall area. Image analysis was performed off-line
using homemade software for Matlab 5.3.1 (MathWorks; Natick, MA). For
both slices, a nontagged reference image was chosen at ~160-180
ms after the ECG trigger. In this image, the LV wall was manually
outlined for both slices (Fig. 1).
Papillary muscles were considered not to be part of the LV wall. MRI
images were recorded in the period from 30 ms until ~260 ms after the ECG trigger with regular time intervals of 20 ms. Each image frame was
split into an image with horizontal and an image with vertical tag
lines by band filtering in the frequency domain around the first
harmonic frequency of the tag pattern (spatial frequency 0.2 mm
1, ratio of bandwidth to center frequency 1.0). With
the use of a correlation method previously applied for pulsed
ultrasonic echo signals (7), from the successive images
with horizontal tag lines, vertical displacement was calculated in each
pixel to obtain vertical displacement maps for each time interval.
Similarly, the images with vertical tags were used to obtain horizontal
displacement maps. Because the boundaries of the LV wall move with the
tags, the calculated displacement maps were used to determine the
moving contours of the LV wall during the cardiac cycle. Thus the
cross-sectional areas of the cavity (Acav) and
of the LV wall (Awall) were determined as a
function of time for slice 1 and slice 2, which
were located at the basal and apical side of the equator, respectively.
Shortening was quantified as the change (
) in the logarithm of the
normalized inner diameter (
inner). It holds that
|
(1) |
multiplied
by the outer radius (Repi) using the
approximation
|
(2) |
|
inner, a
practically linear relationship was found (1) during
ejection. In determination of the ejection phase, the beginning
ejection usually started ~40 ms after the ECG trigger. The end of
ejection occurred ~220 ms later, as marked by the end of decline of
cavity area. The TSR was calculated as the negative slope of the linear
regression line of torsion as a function of
inner during
the ejection phase
|
(3) |
|
Relation between TSR and TransDif.
In the normal heart, myofiber strain is about uniformly distributed.
Myofibers run approximately circumferentially at midwall, which is
defined here by equal cross-sectional areas inside and outside the
midwall radius (4). Mean natural myofiber strain (
f, mean) is expressed as a change (
) in
|
(4) |
|
(5) |

f, epi and 
f, endo
refer to the epicardial and endocardial myofiber strain, respectively.
A model of LV wall mechanics has been used to express TransDif as a
function of TSR and the ratio of cavity to wall area
(Acav/Awall), resulting
in the following approximation (APPENDIX)
|
(6) |
Statistical analysis. Differences were analyzed for significance using a two-tailed t-test, supposing unequal variances. Within the group of patients, differences between before and after aorta valve replacement were evaluated intraindividually. Values are presented as means ± SD.
| |
RESULTS |
|---|
|
|
|---|
Figure 2 shows representative
results of analyzing the MRI-tagged images for the LV of a healthy
volunteer. The top tracings show wall and cavity area for
the two short-axis sections of the heart as a function of time during
the heart cycle. With the use of Eq. 1, from these areas the
inner was determined. After ~260 ms,
inner reaches a plateau, indicating the end of the
ejection phase by closure of the aortic valve. In Fig. 2, rigid body
rotations for both slices and calculated torsion (Eq. 2) are
plotted as a function of time. In Fig. 3,
torsion is plotted as a function of
inner. The slope of
the linear regression line through these data points represents the
negative of the TSR value (Eq. 3).
|
|
In Fig. 4, torsion is plotted as a
function of the logarithm of
inner for the control,
AVSten, and AVRepl groups. The slopes of the linear regression lines
reflect the individual TSR values. In Table
1, the mean values and SDs for these
groups are presented for age, Awall,
Acav,
inner, torsion, TSR, and
TransDif. If applicable, values are presented for the beginning and end
of ejection. Both beginning and end of ejection values of
inner in the control group were larger
(P < 0.05) and varied less than in the AVSten and
AVRepl groups. Apparently, the LV of the patients was concentrically hypertrophied with no significant difference between the AVSten and
AVRepl groups. In the control group, calculated myofiber shortening appears to be in a narrow range (0.142 ± 0.010). In both
the AVSten and AVRepl groups, myofiber shortening is less than in the
control group, P < 0.0001 and P < 0.02, respectively. Myofiber shortening in the AVSten and AVRepl groups
was not different as a group as well as intraindividually.
|
|
In the control group, TSR (0.444 ± 0.073) was lowest with the smallest variance. In the AVSten group, the TSR value (0.85 ± 0.26) was elevated (P < 0.001) and variable. In the AVRepl group, the TSR value (0.61 ± 0.09) decreased (P < 0.02) but was still elevated relative to the control group (P < 0.001).
When estimating TransDif according to Eq. 5 (Fig.
5), this variable was 0.00 ± 0.14 in the control group, 0.96 ± 0.62 in the AVSten group, and
0.37 ± 0.20 in the AVRepl group. Statistical differences
between the groups were very similar to those of the parameter TSR. In
all patients being monitored before and after aortic valve replacement,
TransDif decreased, indicating development toward a more uniform
transmural distribution of myofiber shortening.
|
| |
DISCUSSION |
|---|
|
|
|---|
In patients, the TSR of the LV during the ejection phase was determined noninvasively using MRI tagging. In the presence of an aortic stenosis, TSR was significantly higher than in a group of healthy volunteers. After aortic valve replacement, TSR partially returned to normal. With the use of a model of cardiac mechanics (APPENDIX), the value of TSR was converted to the parameter TransDif, which expresses the epicardial-endocardial difference in myofiber shortening as normalized to mean myofiber shortening. In patients with aortic valve stenosis, TransDif was elevated relative to the control group, indicating severe contractile dysfunction of the inner layers. After aortic valve replacement, TransDif decreased to an intermediate value, indicating an important recovery of contractile function of the inner layers. MRI tagging appears the first technique by which a transmural difference in myofiber shortening can be quantified noninvasively in humans.
Patients with aortic valve stenosis have elevated end-diastolic
concentricity of the LV (P < 0.01), implying that the
wall is thick relative to the cavity diameter. Concentricity is the opposite of dilatation, which is quantified by
inner
(Eq. 1; Table 1). Three months after valve replacement,
end-diastolic concentricity was practically unchanged and still
elevated (P < 0.02). Thus, after aortic valve
replacement, recovery of subendocardial contractile function appears
faster than regression of concentricity. Besides concentricity, in the
AVSten group, the wall area was found to increase, indicating
development of hypertrophy (P < 0.01), which partly
recovered in the AVRepl group.
Because the biological variance of TransDif for normals was so small, for eight of nine AVSten patients the individual TransDif value was significantly higher than the value in the control group (P < 0.001), assuming a Gaussian distribution of the TransDif values in the control group. After aortic valve replacement, in only two of seven patients was this level of significance reached, whereas in the control group all individual values obviously were normal. Thus the presented MRI tagging technique seems well suited to detect subendocardial pathology in an individual patient.
No other parameter was found to be as conclusive as TransDif to
distinguish individual AVSten patients from the control group. TSR is
close and was able to detect seven of nine patients with P < 0.001. In the control group, calculated
f,
mean as well as torsion (Table 1) appeared to be in a narrow
range, 0.142 ± 0.010 and 0.114 ± 0.034, respectively. In the AVSten group, these values were changed
highly significantly, P < 0.0001 and P < 0.002, respectively. Nevertheless, because of the large spread in
the control group, these parameters were less suited to detect
individual patients. Only four and one of nine patients were found to
be abnormal, respectively.
During ejection, measured motion of the LV wall caused the cross-sectional area of the wall to decrease slightly (Table 1). This decrease is likely due to an artifact in the measurement of radial displacement near the inner and outer boundaries of the wall. At those locations, the MRI tags are often malformed, introducing errors in displacement calculation. This displacement has been used to follow the contours of the wall. The method of analysis has been set up such that the influence of this error is moderate and does not affect any major conclusion of this study.
Control measurements were performed in young volunteers, whereas the patients were much older (Table 1). Current insights suggest that the TSR is independent of age and species. Assuming homogeneity of mechanical load in the heart, TSR was predicted in a mathematical model of LV mechanics (3) without using any species-specific information. TSR seems a mathematically determined constant that is expected to be universal to all hearts, "from mice to elephants." The mathematically predicted value was found not to be different from experimental measurements in normal hearts of dogs (3) and of young volunteers (1). It has also been reported that both torsion (12) and ejection fraction in mice are not different from that in humans, suggesting similarity of TSR. Nevertheless, it is wise to perform measurements of TSR in an age-matched group of humans. Unfortunately, these data are not available yet.
The presented MRI-tagging technique is the first clinically feasible technique to quantify transmural differences in myocardial contraction. The MRI protocol takes ~25 min, finding the position and orientation of the heart inclusive. In the current setup, time is needed for selection and transfer of files. After image acquisition, epi- and endocardial contours have to be drawn manually in one set of images, which takes ~5 min. This step is the most cumbersome one and should be automated for easy clinical application. Finally, image and data analysis take another 5 min on a regular personal computer. Various parts of the protocol and analysis may be improved, such as the use of breath-hold in MRI tagging.
In conclusion, using MRI tagging, the motion of two cross sections of the LV was measured noninvasively in healthy volunteers (control) and in AVSten and AVRepl patients. TSR in the AVSten group was significantly larger than in the control group. In the AVRepl group, TSR declined but did not return completely. With the use of a model of LV mechanics, from TSR and Acav/Awall, TransDif was calculated as normalized to mean myofiber shortening. In the control, AVSten, and AVRepl groups, TransDif was 0 ± 0.14, 0.96 ± 0.62, and 0.38 ± 0.20, respectively. Thus, in aortic valve stenosis, decreased subendocardial contractile function was found to recover 3 mo after aortic valve replacement. Over the same period, recovery of the shape of the heart (concentricity) appeared insignificant. In the AVSten group, eight of nine patients were recognized by a significant reduction of TransDif (P < 0.001) relative to the control group.
| |
APPENDIX |
|---|
|
|
|---|
Relation between TSR and TransDif.
The value of the TSR was derived directly from the MRI tagging data. In
this appendix, TSR will be related to the transmural difference in
myofiber shortening as expressed by the parameter TransDif (Eq. 4). Consider the equatorial wall as a thick-walled cylinder
section with inner radius ri and outer radius
ro. Wall coordinates [r,
c, z] represent the radial, circumferential, and base-to-apex directions. For small rotational symmetric deformations of
an incompressible cylinder, circumferential strain
cc, base-to-apex strain
zz, and shear angle
cz can be expressed as a function of radius
r as
|
(7) |
f) depends on a change in the ratio of cavity volume
(Vcav) to wall volume (Vwall) (2).
With the use of Eq. 7, and by substituting the inner and
outer radius of the cylinder section, it is found that
|
(8) |
, for myofiber strain it holds that
|
(9) |
f is assumed to be the same
everywhere (4). Thus three equations were obtained for the
inner surface (r = ri),
for the midwall surface (r = rmid), with 2r


|
(10) |
|
mid = 0. The
parameter TSR is the ratio of shear at the outer surface and relative
inner diameter decrease
|
(11) |
i, and
o. This system of
equations was solved using Mathematica 4.1 (Wolfram Research;
Champaign, IL). As a result, the fiber angles
i and
o are known.
In the control group for midejection, the ratio of cylinder cavity to
cylinder wall volume was
(Acav/Awall)me
0.64 and TSR0 = 0.444 (Table 1). With the use of
these data, it was derived that
i = 0.68 rad and
o =
0.94 rad. Given a different value of TSR, the
value of C in Eq. 7 changes, causing only shear
zc to change. With Eq. 9, myofiber
strain was calculated numerically at the inner and outer surfaces. With
Eq. 5, the parameter TransDif was calculated. For the
occurring range (Table 1), the following approximation was found
|
(12) |
|
| |
ACKNOWLEDGEMENTS |
|---|
This work was supported by The Netherlands Heart Foundation Grant 98.035.
| |
FOOTNOTES |
|---|
Address for reprint requests and other correspondence: T. Arts, Dept. of Biophysics, PO Box 616, 6200 MD Maastricht, The Netherlands (E-mail: t.arts{at}bf.unimaas.nl).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
June 20, 2002;10.1152/ajpheart.00239.2002
Received 25 February 2002; accepted in final form 12 June 2002.
| |
REFERENCES |
|---|
|
|
|---|
1.
Aelen, FWL,
Arts T,
Sanders DGM,
Thelissen GRP,
Muijtjens AMM,
Prinzen FW,
and
Reneman RS.
Relation between torsion and cross-sectional area change in the human left ventricle.
J Biomech
30:
207-212,
1997[Web of Science][Medline].
2.
Arts, T,
Bovendeerd PHM,
Prinzen FW,
and
Reneman RS.
Relation between left ventricular cavity pressure and volume and systolic fiber stress and strain in the wall.
Biophys J
59:
93-103,
1991[Web of Science][Medline].
3.
Arts, T,
Meerbaum S,
Reneman RS,
and
Corday E.
Torsion of the canine left ventricle during the ejection phase in the intact dog.
Cardiovasc Res
18:
183-193,
1984[Web of Science][Medline].
4.
Arts, T,
Prinzen FW,
Snoeckx LHEH,
Rijcken JM,
and
Reneman RS.
Adaptation of cardiac structure by mechanical feedback in the environment of the cell, a model study.
Biophys J
66:
953-961,
1994[Web of Science][Medline].
5.
Arts, T,
Veenstra PC,
and
Reneman RS.
A model of the mechanics of the left ventricle.
Ann Biomed Eng
7:
299-318,
1979[Web of Science][Medline].
6.
Axel, L,
and
Dougherty L.
MR imaging of motion with spatial modulation of magnetization.
Radiology
171:
841-845,
1989
7.
De Jong, PGM,
Arts T,
Hoeks APG,
and
Reneman RS.
Determination of tissue motion velocity by fast correlation of pulsed ultrasonic echo signals.
Ultrason Imaging
12:
84-98,
1990[Web of Science][Medline].
8.
Dong, SJ,
Hees PS,
Huang WM,
Buffer SA, Jr,
Weiss JL,
and
Shapiro EP.
Independent effects of preload, afterload, and contractility on left ventricular torsion.
Am J Physiol Heart Circ Physiol
277:
H1053-H1060,
1999
9.
Feigl, EO.
Coronary physiology.
Physiol Rev
63:
1-206,
1983
10.
Gallagher, KP,
Osakada G,
Hess OM,
Koziol JA,
Kemper WS,
and
Ross J.
Subepicardial segmental function during coronary stenosis and the role of myocardial fiber orientation.
Circ Res
50:
352-359,
1982
11.
Geerts, L,
Bovendeerd P,
Nicolay K,
and
Arts T.
Characterization of the normal cardiac myofiber field in goat measured with MR-diffusion tensor imaging.
Am J Physiol Heart Circ Physiol
283:
H139-H145,
2002
12.
Henson, RE,
Song SK,
Pastorek JS,
Ackerman JJ,
and
Lorenz CH.
Left ventricular torsion is equal in mice and humans.
Am J Physiol Heart Circ Physiol
278:
H1117-H1123,
2000
13.
Hoffman, JIE
The effect of intramyocardial forces on the distribution of intramyocardial bloodflow.
J Biomed Eng
1:
33-40,
1979[Medline].
14.
Hoffman, JIE,
and
Spaan JAE
Pressure flow relations in coronary circulation.
Physiol Rev
70:
331-390,
1990
15.
Hutchins, GM,
Kuhajda FP,
and
Moore GW.
Myocardial injury in patients with aortic stenosis.
Am J Cardiovasc Pathol
1:
31-37,
1987[Medline].
16.
Lewis, AB,
Heymann MA,
Stanger P,
Hoffman JI,
and
Rudolph AM.
Evaluation of subendocardial ischemia in valvar aortic stenosis in children.
Circulation
49:
978-984,
1974
17.
MacGowan, GA,
Shapiro EP,
Azhari H,
Siu CO,
Hees PS,
Hutchins GM,
Weiss JL,
and
Rademakers FE.
Noninvasive measurement of shortening in the fiber and cross-fiber directions in the normal human left ventricle and in idiopathic dilated cardiomyopathy.
Circulation
96:
535-541,
1997
18.
Maier, SE,
Fischer SE,
McKinnon GC,
Hess OM,
Krayenbühl HP,
and
Boesinger P.
Evaluation of left ventricular segmental wall motion in hypertrophic cardiomyopathy with myocardial tagging.
Circulation
86:
1919-1928,
1992
19.
Matheijssen, NA,
de Roos A,
Doornbos J,
Reiber JH,
Waldman GJ,
and
van der Wall EE.
Left ventricular wall motion analysis in patients with acute myocardial infarction using magnetic resonance imaging.
Magn Reson Imaging
11:
485-492,
1993[Web of Science][Medline].
20.
McVeigh, ER.
MRI of myocardial function: motion tracking techniques.
Magn Reson Imaging
14:
137-150,
1996[Web of Science][Medline].
21.
Nagel, E,
Stuber M,
Burkhard B,
Fischer SE,
Scheidegger MB,
Boesiger P,
and
Hess OM.
Cardiac rotation and relaxation in patients with aortic valve stenosis.
Eur Heart J
21:
582-589,
2000
22.
Prinzen, FW,
Arts T,
Hoeks APG,
and
Reneman RS.
Discrepancies between myocardial blood flow and fiber shortening in the ischemic border zone as assessed with video mapping of epicardial deformation.
Pflügers Arch
415:
220-229,
1989[Web of Science][Medline].
23.
Rademakers, FE,
Buchalter MB,
Rogers WJ,
Zerhouni EA,
Weisfeldt ML,
Weiss JL,
and
Shapiro EP.
Dissociation between left ventricular untwisting and filling. Accentuation by catecholamines.
Circulation
85:
1572-1581,
1992
24.
Stuber, M,
Scheidegger MB,
Fischer SE,
Nagel E,
Steinemann F,
Hess OM,
and
Boesiger P.
Alterations in the local myocardial motion pattern in patients suffering from pressure overload due to aortic stenosis.
Circulation
100:
361-368,
1999
25.
Van der Wall, EE.
Twisting, untwisting and diastolic function in aortic valve disease.
Eur Heart J
21:
512-513,
2000
26.
Vincent, WR,
Buckberg GD,
and
Hoffman JI.
Left ventricular subendocardial ischemia in severe valvar and supravalvar aortic stenosis. A common mechanism.
Circulation
49:
326-333,
1974
27.
Zerhouni, EA,
Parish MS,
Rogers WJ,
Yang A,
and
Shapiro EP.
Tagging of the human heart by magnetic resonance.
Radiology
16:
59-63,
1989.
This article has been cited by other articles:
![]() |
P. H. M. Bovendeerd, W. Kroon, and T. Delhaas Determinants of left ventricular shear strain Am J Physiol Heart Circ Physiol, September 1, 2009; 297(3): H1058 - H1068. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. T. Esch and D. E. R. Warburton Left ventricular torsion and recoil: implications for exercise performance and cardiovascular disease J Appl Physiol, February 1, 2009; 106(2): 362 - 369. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Dal-Bianco, B. K. Khandheria, F. Mookadam, F. Gentile, and P. P. Sengupta Management of Asymptomatic Severe Aortic Stenosis J. Am. Coll. Cardiol., October 14, 2008; 52(16): 1279 - 1292. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Nottin, G. Doucende, I. Schuster-Beck, M. Dauzat, and P. Obert Alteration in left ventricular normal and shear strains evaluated by 2D-strain echocardiography in the athlete's heart J. Physiol., October 1, 2008; 586(19): 4721 - 4733. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Kirn, A. Jansen, F. Bracke, B. van Gelder, T. Arts, and F. W. Prinzen Mechanical discoordination rather than dyssynchrony predicts reverse remodeling upon cardiac resynchronization Am J Physiol Heart Circ Physiol, August 1, 2008; 295(2): H640 - H646. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Delhaas, W. Kroon, W. Decaluwe, M. Rubbens, P. Bovendeerd, and T. Arts Structure and torsion of the normal and situs inversus totalis cardiac left ventricle. I. Experimental data in humans Am J Physiol Heart Circ Physiol, July 1, 2008; 295(1): H197 - H201. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. P. Sengupta, A. J. Tajik, K. Chandrasekaran, and B. K. Khandheria Twist mechanics of the left ventricle principles and application. J. Am. Coll. Cardiol. Img., May 1, 2008; 1(3): 366 - 376. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Lumens, T. Delhaas, T. Arts, B. R. Cowan, and A. A. Young Impaired subendocardial contractile myofiber function in asymptomatic aged humans, as detected using MRI Am J Physiol Heart Circ Physiol, October 1, 2006; 291(4): H1573 - H1579. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Vernooy, X. A.A.M. Verbeek, M. Peschar, H. J.G.M. Crijns, T. Arts, R. N.M. Cornelussen, and F. W. Prinzen Left bundle branch block induces ventricular remodelling and functional septal hypoperfusion Eur. Heart J., January 1, 2005; 26(1): 91 - 98. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |