|
|
||||||||
1 Institut für Experimentelle Gynäkologie and 2 Klinik und Poliklinik für Frauenheilkunde und Geburtshilfe, Universitätsklinikum Hamburg-Eppendorf, 20246 Hamburg, Germany
| |
ABSTRACT |
|---|
|
|
|---|
The pressure drop and pressure pulses in the isthmus of the ductus venosus (DV) in fetal sheep have not been measured directly and related to flow. In eight acutely anesthetized fetal sheep, a 3-Fr tip pressure transducer (TP) was inserted from the external jugular into the umbilical vein (UV). Ultrasound Doppler flow velocities, TP position, and intravenous pressures were recorded in the UV, DV, and inferior vena cava (VC) while the TP was withdrawn. Flow was steady in the UV, but small pressure fluctuations (<0.4 mmHg) could be detected. Time-averaged pressure dropped 1.9 mmHg (mean; 0.5-3.3 mmHg 95% confidence interval) across the DV isthmus. Pressure pulses increased from 1.7 mmHg (mean; 1.2-2.1 mmHg 95% confidence interval) in the DV to 3.9 mmHg (mean; 1.8-6.0 mmHg 95% confidence interval) in the inferior VC. The pressure wave from the heart arrived later [0.053 s (mean; 0.025-0.080 s 95% confidence interval)] in the isthmus of the DV than in the diaphragmatic inferior VC, indicating a wave velocity of ~1.1 m/s. At all locations, pressures and flow velocities were inversely related.
fetal sheep; pressure wave; flow velocity; ductus venosus
| |
INTRODUCTION |
|---|
|
|
|---|
IN THE FETUS, flow is pulsatile in both the venae cavae (VC) (22) and ductus venosus (DV) (10) and normally nonpulsatile in the portal sinus or umbilical vein (UV). Variations of the flow rates and/or velocity patterns predominantly in the DV, and their significance for the well-being of human fetuses, have been investigated by Doppler ultrasound and shown to be of diagnostic value (2-4, 8-10, 25). Cardiac and circulatory dysfunction as disclosed by an abnormal flow pattern may be associated with pressure variations, which at present are mostly unknown. However, only the knowledge of both flow and pressure permit sufficient understanding of physiological and pathological conditions in the central venous circulation (11, 14, 20, 22). Mathematical (1, 6, 7, 18-20) and in vitro models (5, 13) have been used to gain insight into the complex situation, but empirical in vivo data on these pressure-wave forms and their relation to flow in the fetus are remarkably rare.
Pressure pulses were derived indirectly from the wall movement of the inferior vena cava (VC) in human fetuses (16). In fetal sheep, pulsatile pressure and flow were recorded from the superior VC with indwelling catheters and flow probes (22). It was found that local pressure fluctuations were biphasic, as was the flow, but flow and pressure pulses were inverse. It was concluded that flow pulsatility in central veins is generated by pressure waves originating in the heart. The pressure waves were thought to travel from the heart in a direction opposite to the bloodstream toward the inferior VC and DV. It could be demonstrated accordingly (21) that the delay of flow pulsations in central veins increased with distance from the heart, but it has not been shown directly so far that a pulse wave actually exists, nor has its velocity been determined. Pressure pulses in combination with flow have not been measured in the DV or UV. In this introductory study, we present a technique to measure in fetal sheep pressure and flow pulses at the same time and the same location in various parts of the central venous circulation.
| |
METHODS |
|---|
|
|
|---|
The experiments with eight pregnant sheep followed the guidelines for animal research of our institution. The animals (median gestational age, 121.5 days) were sedated with xylazin (Rompun; 0.25 mg/kg im) and intubated after intravenous injection of thiopental sodium (Trapanal; 1 g). Inhalation anesthesia was maintained by artificial ventilation with 1% halothane in an O2-N2O mixture.
The animal was placed on a heating pad (38°C) in the supine position,
and the maternal abdomen was opened by a midline incision. The fetal
head was delivered after hysteromy, and the myometrium was attached
loosely to the fetal neck to avoid losses of amniotic fluid. From the
right external jugular vein, a catheter was inserted into the UV as
described previously (27). The catheter allowed introduction of a tip pressure transducer with side openings. In three
animals 3-Fr Sensodyn pressure transducers (Braun) and in five animals
3-Fr Millar pressure transducers (model SPR-524) were used (both
3
db, >5 kHz). The right carotid artery was catheterized for pressure
(Ohmeda P23XL) and heart rate measurements and for fetal blood gas
analysis (ABL 710, Radiometer Kopenhagen). A fluid-filled catheter
connected to a pressure transducer (Ohmeda P23XL) was inserted into the
amniotic cavity to monitor amniotic fluid pressure.
In the first three experiments, only the ECG could be recorded in conjunction with the Doppler signal, and ECG electrodes were attached to the fetal scalp and myometrium to allow synchronization of pressure and Doppler velocity signals. In later experiments, the signal of the tip pressure transducer was displayed directly on the screen of the ultrasound system (alternating current-coupled "physiological" input, Acuson Aspen). In one animal, after insertion of the first tip pressure transducer (Millar), a second Millar catheter was introduced as described above in an attempt to measure intravascular pressure differences.
Experimental Procedures
Central venous, arterial, and amniotic pressures and the ECG signals were continuously acquired by a MP 100 data-acquisition system (sampling rate 250 Hz, in two experiments 1,000 Hz; Biopac Systems) and stored on disk. Either the ECG or the signal from the tip pressure transducer amplifier was led into the ultrasound system and displayed on screen simultaneously with the Doppler velocity signal. All output to the screen was continuously recorded on videotape.The ultrasound probe was placed directly on the uterine wall, and the tip of the pressure transducer was localized. The Doppler gate was positioned to cover the vessel as close to the pressure transducer tip as possible (within 0.5 cm) but avoiding direct inclusion. Velocity pulses were recorded when the signal quality was judged as sufficient. The parameters of the maximum velocity envelope curve [minimum (Vmin), maximum (Vpeak), and time-averaged mean velocities (TAMX)] were determined. The tip pressure transducer was then withdrawn, and its new position was verified by ultrasound B-scan. Venous pressures and flow velocities were observed at the following five positions, which covered a distance of 6-8 cm: the UV, 1 cm upstream of the isthmus; DV1, immediately downstream of the isthmus; DV2, outlet; VC1, the inferior VC close to the outlet of the DV; and VC2, the inferior VC close to the right atrium.
The experiments were terminated by intravenous injection of 15 ml of euthanasia solution T61 (Hoechst) into the ewe. The fetus was removed, and the tip pressure transducer was calibrated.
Calibration of Tip Pressure Transducers
The tip pressure transducers were calibrated before and immediately after an experiment. Typically, there was a baseline shift ranging from 2 to 6 mmHg but a constant "gain" (slope of the calibration line). Therefore, pressure changes that occurred within 1 min or less were measured accurately.The arterial and amniotic pressure recording systems were also calibrated with zero adjustments during the experiment.
Data Processing
Pressures.
Digitized data were evaluated using Acqknowledge analysis software
(Biopac Systems). On the basis of postexperiment calibration data, all
pressure signal voltages were transformed to pressure values (in mmHg).
Depending on the signal-to-noise ratio, pulsatile venous pressure
signals mostly from the UV were digitally filtered by a 15- or 25-Hz
low-pass filter, which did not affect phases. Pressure pulses were
measured during time intervals when no ultrasonographic recording took
place. The pressure pulse (difference between minimum and maximum
pressure during one cardiac cycle) of venous pressure was determined
over five cardiac cycles to minimize the influence of ventilation (cf.
Fig. 1).
|
Doppler flow velocities.
On the basis of the maximum velocity envelope curve, the
Vmin, Vpeak, and TAMX of
one to three consecutive cardiac cycles (depending on the signal
quality) were determined (cf. Figs. 5 and 6). From these points, the
pulsatility index [PI = (Vpeak
Vmin)/TAMX] was calculated. Doppler angles
averaged 46° (mean; 42-50° 95% confidence interval) and did
not exceed 60°.
Motion artifacts. The tip of a Millar catheter was inserted 5 cm into water and oscillated perpendicular to the sensitive area (1-3 cm/s). Apparent pressure decreases and increases of maximal 0.5 mmHg were observed. The time-averaged mean pressure was accurately recorded as 5.0 cmH2O.
Statistics
Average results are described as means and 95% confidence intervals (in parentheses) unless stated otherwise. To test for differences between the five locations, ANOVA and Fisher's post hoc test were performed (Statistica, Statsoft; Tulsa, OK). P < 0.05 was accepted as significant.| |
RESULTS |
|---|
|
|
|---|
Gestational age of the fetuses ranged from 112 to 127 days (term 145 days), and mean body weight was 2.5 kg (1.6-3.3 kg). In four twin pregnancies, one of the two siblings was included. The experiments lasted between 45 and 150 min.
Fetal arterial blood gas values approximately in the middle of the
experimental procedure were pH 7.23 (7.16-7.30 pH),
PO2 25.3 mmHg (19.6-31.0 mmHg),
PCO2 66.2 mmHg (55.1-77.2 mmHg), and base
excess
1.1 mmol/l (
3.4 to 1.2 mmol/l), which indicates a moderate,
mostly respiratory acidosis.
A typical pressure recording is displayed in Fig.
1 when the tip pressure transducer
(middle) moved (as judged from the ultrasound B-scan) from
the portal sinus into the DV with a sudden small pressure drop (arrow)
at the isthmic region. At this moment and position, pressure pulses
appeared, which were linked to the arterial pressure pulses (Fig. 2,
see Figs. 5 and 6). In five experiments with these three criteria, the time-averaged pressure drop across the
isthmus of the DV was 1.9 mmHg (0.5-3.3 mmHg). This pressure drop
was significant (paired t-test). In three experiments, the above signs could not be detected in combination.
|
In the UV, small pressure pulses were visible in four experiments (Fig. 2, left); they could, however, not be detected in two animals and were questionable in another two animals. The occurrence of detectable pressure pulses was not related to the fetal arterial PO2. Flow measured by ultrasound Doppler was nonpulsatile in the UV with one exception, where pressure pulses were also observed.
Mean pressure pulses in the UV, DV1, DV2, VC1, and VC2 are summarized
in Table 1; the pressure pulse at VC2 was
significantly larger than that at DV1.
|
Corresponding points of the venous pressure curve were observed 53 ms
(25-80 ms) later at DV1 than at VC2 (cf. Fig.
3). Appearance at DV2 was 37 ms
(21-52 ms) and at VC1 was 23 ms (9-37 ms) later than that at
VC2. Nonparametric
-statistics (Statistica) showed that it was
significantly more probable that the tip location and time differences
were in the same order than not.
|
Figure 4 demonstrates a simultaneous
measurement of two venous pressures at two different locations
(top) and at the same location (bottom).
|
Typical simultaneous recordings of ultrasonographic Doppler flow
velocity and intravascular venous pressure at the same location are
illustrated in Figs. 5 (VC2) and
6 (DV2). Pressures and velocities are
inversely related, and the pulsatility of both pressure and flow (cf.
PI values on Figs. 5 and 6) is higher in the VC than in the DV.
|
|
In Table 2, mean values and confidence
intervals are listed for Vmin,
Vpeak, TAMX, and PI at the five positions. There
were no significant differences between these groups, but when DV1 was
compared with the flow profile in the VC (both VC1 and VC2), the
velocity parameters at DV1 were significantly different from the VC.
|
The mean heart rate was 146.6 beats/min (141.1-152.1 beats/min)
and was not associated with time or with positions of the pressure
transducer tip (
-statistics). The apparent mean carotid pressure was
53.7 mmHg (51.5-55.9 mmHg), and the apparent mean amniotic fluid
pressure was 11.1 mmHg (9.2-12.9 mmHg).
In three experiments, the time intervals between the R wave peak and the increase of the systolic arterial pressure (cf. Fig. 3) were 51 ms (45-56 ms), 62 ms (59-65 ms), and 54 ms (52-57 ms). There were no consistent changes related to the position of the tip pressure transducer.
| |
DISCUSSION |
|---|
|
|
|---|
Pulsatile Venous Pressure
The present study confirms and expands the results of two previous reports (21, 22) on fetal sheep (chronic preparation). A mainly inverse relationship between the biphasic time course of pressure and flow was observed in the superior VC (22). From its temporal relation to arterial pressure and the ECG R wave, the large and short-lasting peak of venous pressure (cf. Fig. 5) was thought most likely to be caused by atrial contraction and to cause the "a-depression" in the flow profile. It is apparent from the figures (22) that pressure pulses normally were in the range of 10-15 mmHg and that flow pulsatility matched pressure pulsatility.The present results confirm the basic flow and pressure profiles in the inferior VC and their inverse relationship (Fig. 5). Mean pressure pulses were distinctly lower (~4 mmHg, Table 1) than reported previously, however (22). This may reflect the differences between a chronic and an acute preparation or the heavily invasive instrumentation with flow probes, which may impede venous return.
Our data demonstrate, moreover, that in the DV, pressure is also pulsatile with a lesser pressure pulse than in the inferior VC (Table 1). Again, pressure and flow pulsations are inversely related (Fig. 6). The differences in pressure pulses between the DV and inferior VC (Table 1) are reflected as differences in flow pulsatility (Table 2), which have also been observed by others (24).
There are no other direct measurements of DV pressure pulses available. On the basis of physical principles, model calculations predict pressure pulses of 1.3 mmHg (approximately from 2.8 mmHg minimum to 4.1 mmHg peak pressure; UV pressure constant at 4.5 mmHg) (1, 19). These model estimates are close to our measurement of 1.7 mmHg (Table 1), which may improve the confidence in the results of this mathematical model.
It cannot be excluded that movement artifacts affect the pressure pulses to some extent. The predicted effect, however, is moderate (~0.5 mmHg) even with wide tip oscillations (1-3 cm). Tip movements (2-4 Hz) typically did not exceed 0.3 cm, as judged from the video recordings.
Because pressure in the UV (driving pressure) was mainly constant, flow variations in the DV and in the inferior VC were caused by changes in the opposing pressure, which explains the inverse relationship between pressure and flow. Pressure profiles distant from the heart were delayed with respect to more downstream locations, which is in agreement with the delay of the flow pulses (21). On average, the pressure wave arrived 53 ms later at the entrance of the DV (DV1) than at the inferior VC close to the heart (VC2), and the average time delay for the positions in between varied accordingly. Simultaneous pressure measurements at different locations in one experiment are in agreement. In Fig. 4, top, catheter MIL1 was located at DV1 and MIL2 was located in the inferior VC (in between VC1 and VC2). The time difference between corresponding points (arrows) was ~30 ms (dotted vertical line). About 50 min later (Fig. 4, bottom), both catheter tips were at the same location in the VC, and the corresponding points now occurred at the same time.
On the basis of ultrasound B-scans and postmortem observations, the average distance from VC2 to DV1 was taken as 6 cm. The mean velocity of the pressure wave was then calculated to be 1.1 m/s. This agrees well with the pulse wave velocity of 1-3 m/s, which was derived from the measurement of vascular stiffness of the DV and UV in an in vitro model (5) (fetal sheep). In adult animals, values of 1.2-2.5 m/s in the thoracic VC have been reported (15).
Nonparametric
-statistics (Statistica) confirmed the relation
between time delay and location, but data scatter prevented establishment of statistically significant differences between time
delays using ANOVA. There are several reasons that explain the large
variation in the measurement of time differences. Time resolution was
restricted by the sampling rate (250 Hz or 4 ms), but most influential
were variations in the pressure profile with shifts in the position of
the respective peaks or nadirs (cf. Fig. 3). It is possible that motion
artifacts of the tip pressure transducers contributed to these
displacements. As judged from the variation of the time difference
between the R peak and the end-diastolic pressure rise (10% or less of
the mean value), the time delay measurements were not biased by
systematic shifts of the time reference point (end-diastolic carotid
pressure). Because time differences were determined, the constant delay
between the true beginning of the right ventricular systolic ejection
phase and the systolic (carotid artery) pressure rise in the external pressure transducer is of no concern.
Pressure pulses existed occasionally in the UV (Fig. 2, left). Their amplitude was small (~0.2-0.4 mmHg in 4 animals), but the pulses were related to the arterial pressure pulses and were not random. These pressure fluctuations are likely transmitted from the DV, and it has been shown in model calculations (7) that the pulse amplitude depends on the compliance of the UV and on the diameter of the isthmus. Both are increased in fetal hypoxemia, as are UV flow pulsations (9), but our data did not reveal an influence of the arterial PO2 on the occurrence of UV pressure pulses. Pressure pulsations may be imposed by the umbilical arteries on the UVs and then travel from the cord toward the portal sinus. It is also possible that motion artifacts were generated, although the catheter tips appeared to move very little in the UV.
Time-Averaged Venous Pressure
When the site of venous pressure recording was shifted from the portal sinus into the DV (DV1), there was in most cases a sudden pressure drop, which amounted to ~2 mmHg (Fig. 1, arrow). This value is the first direct measurement of the isthmic pressure gradient and puts previous indirect estimates in human fetuses [0.3-1.3 mmHg (28) and 0.5-2.5 mmHg (12)] and model calculations (6, 19) with comparable results on a firm basis. In fetal sheep, a pressure drop of 3.1 ± 0.5 mmHg from the portal sinus to the diaphragmatic inferior VC was determined, which includes the isthmic region (17).In three fetuses, the three criteria to identify the pressure drop across the isthmus could not be met. In one case, the pressure transducer tip was pushed into a small branch of the UV during ultrasonographic measurements. The tip shifted directly into the DV when the catheter was withdrawn with an increase of time-averaged pressure instead of a decrease. In another case, time-averaged pressure dropped 1.3 mmHg, but pressure pulses did not appear or increase immediately. In the third fetus, the tip pressure transducer could not be visualized during its passage from the portal sinus into the DV.
Because of baseline shifts of the tip pressure transducer and the peculiar experimental situation (fetal head exteriorized, uterus incompletely closed) that prevents the accurate measurement of the reference (amniotic) pressure, we are unable to present reliable data on time-averaged continuous central venous, arterial, and amniotic fluid pressures.
Relationship Between Pressure and Flow Pulses
As demonstrated in Figs. 5 and 6, a transient increase of local pressure was associated with a transient decrease of flow velocity and vice versa, although minor differences between both profiles were sometimes visible. Typically, the pressure a wave before onset of the ventricular systole was followed by a smaller systolic pressure rise of longer duration (Figs. 5 and 6), as described in a previous report (22). However, variations of this basic pattern did occur, and, especially in the DV close to the isthmic region (DV1), both pressure peaks could be of similar size and duration.Small time differences (up to about 50 ms; cf. Fig. 5 and 6) between corresponding characteristic points of the flow and pressure wave could also be detected, but we could not find systematic phase differences. These time differences were seen, although our technique allows us to observe pressure and flow velocity simultaneously and at nearly identical locations, which is not possible with outside pressure transducers (22). Model calculations (19) also appear to predict small time differences between peaks and troughs of the flow and pressure wave. The phase shifts (cf. Fig. 6) indicate that, as in the arterial system, pulsatile flow in fetal central veins is affected by viscous, inertial, and elastic forces. Although, in principle, the simultaneous observation of pressure and velocity pulses would permit a dynamic analysis of the fetal central venous circulation (e.g., calculation of vascular impedances; cf. Ref. 7), the technical requirements to do so are not sufficiently satisfied at present.
In conclusion, contractions of the right atrium and ventricular movements (22) generate a pressure wave that travels at a speed of 1.1 m/s to the DV. In the UV upstream of the isthmus, the size of the pressure pulse suddenly decreases (from 1.7 mmHg in the DV to 0-0.4 mmHg), whereas the time-averaged pressure increases by ~2 mmHg. When pulsatile flow and pressure are observed at the same location, local pressure rises are associated with flow reductions and vice versa. Differences in pulsatility of pressure are reflected as differences in the pulsatility of flow. In some aspects, the geometry of central veins in the ovine fetus (23) is different from the human situation where, for example, the DV outlet almost directly connects with the right atrium (9). Because flow velocity and velocity pattern in the sheep fetus (26) are very similar to velocities in the DV of human fetuses, it seems likely that our results are valid in principle also for the human fetus.
The instrumentation of the sheep fetus with thin (3-Fr), moveable tip pressure transducer catheters and external ultrasonographic Doppler flow measurements is less invasive than previously reported (22). It can be used also in chronic fetal sheep experiments (21). This may help to explore, in nonanesthetized fetuses, the effects of experimental cardiac failure, for example, on flow and pressures in fetal central veins, and thus to improve for diagnostic purposes the understanding of pathological flow-velocity patterns.
| |
ACKNOWLEDGEMENTS |
|---|
We are grateful to Acuson (Nuremberg, Germany) and Advanced Technology Laboratories (Munich, Germany), who made the ultrasonographic equipment available. The editorial help of G. Power (Perinatal Biology, Loma Linda, CA) is gratefully acknowledged.
| |
FOOTNOTES |
|---|
Address for reprint requests and other correspondence: H. J. Schröder, Institut für Experimentelle Gynäkologie, Universitätsklinikum Hamburg-Eppendorf, Martinistrasse 52, 20246 Hamburg, Germany (E-mail: schroede{at}uke.uni-hamburg.de).
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
First published December 27, 2002;10.1152/ajpheart.00969.2002
Received 8 November 2002; accepted in final form 20 December 2002.
| |
REFERENCES |
|---|
|
|
|---|
1.
Bellotti, M,
Pennati G,
Pardi G,
and
Fumero R.
Dilatation of the ductus venosus in human fetuses: ultrasonic evidence and mathematical modeling.
Am J Physiol Heart Circ Physiol
275:
H1759-H1767,
1998
2.
Gembruch, U,
Krapp M,
and
Baumann P.
Changes of venous blood flow velocity waveforms in fetuses with supraventricular tachycardia.
Ultrasound Obstet Gynecol
5:
394-399,
1995[Medline].
3.
Hecher, K,
and
Campbell S.
Characteristics of fetal venous blood flow under normal circumstances and during fetal disease.
Ultrasound Obstet Gynecol
7:
68-83,
1996[Medline].
4.
Hecher, K,
Campbell S,
Doyle P,
Harrington K,
and
Nicolaides K.
Assessment of fetal compromise by Doppler ultrasound investigation of the fetal circulation.
Circulation
91:
129-138,
1995
5.
Hellevik, LR,
Kiserud T,
Irgens F,
Stergiopulos N,
and
Hanson M.
Mechanical properties of the fetal ductus venosus and umbilical vein.
Heart Vessels
13:
175-180,
1998[Medline].
6.
Hellevik, LR,
Kiserud T,
Irgens F,
Ytrehus T,
and
Eik-Nes SH.
Simulation of pressure drop and energy dissipation for blood flow in a human fetal bifurcation.
J Biomech Eng
120:
455-462,
1998[Web of Science][Medline].
7.
Hellevik, LR,
Stergiopulos N,
Kiserud T,
Rabben SI,
Eik-Nes SH,
and
Irgens F.
A mathematical model of umbilical venous pulsation.
J Biomech
33:
1123-1130,
2000[Medline].
8.
Kiserud, T.
In a different vein: the ductus venosus could yield much valuable information.
Ultrasound Obstet Gynecol
9:
369-372,
1997[Medline].
9.
Kiserud, T.
The ductus venosus.
Semin Perinatol
25:
11-20,
2001[Web of Science][Medline].
10.
Kiserud, T,
Eik-Nes SH,
Blaas HGK,
and
Hellevik LR.
Ultrasonographic velocimetry of the fetal ductus venosus.
Lancet
338:
1412-1414,
1991[Web of Science][Medline].
11.
Kiserud, T,
Eik-Nes SH,
Blaas HG,
and
Hellevik LR.
Foramen ovale: an ultrasonographic study of its relation to the inferior vena cava, ductus venosus and hepatic veins.
Ultrasound Obstet Gynecol
2:
389-396,
1992[Web of Science][Medline].
12.
Kiserud, T,
Hellevik LR,
Eik-Nes SH,
Angelsen BAJ,
and
Blaas HGK
Estimation of the pressure gradient across the fetal ductus venosus based on Doppler velocimetry.
Ultrasound Med Biol
20:
225-232,
1994[Web of Science][Medline].
13.
Kiserud, T,
Stratford L,
and
Hanson MA.
Umbilical flow distribution to the liver and the ductus venosus: an in vitro investigation of the fluid dynamic mechanisms in the fetal sheep.
Am J Obstet Gynecol
177:
86-90,
1997[Web of Science][Medline].
14.
Maulik, D.
Hemodynamic interpretation of the arterial Doppler waveform.
Ultrasound Obstet Gynecol
3:
219-227,
1993[Medline].
15.
Minten, J,
van de Werf F,
Aubert AE,
Kestelboot H,
and
de Geest H.
Apparent pulse wave velocity in the canine superior vena cava.
Cardiovasc Res
17:
627-632,
1983[Medline].
16.
Mori, A,
Trudinger B,
Mori R,
Reed V,
and
Takeda Y.
The fetal central venous pressure waveform in normal pregnancy and in umbilical placental insufficiency.
Am J Obstet Gynecol
172:
51-57,
1995[Web of Science][Medline].
17.
Paulick, RP,
Meyers RL,
Rudolph CD,
and
Rudolph AM.
Venous responses to hypoxemia in the fetal lamb.
J Dev Physiol
14:
81-88,
1990[Web of Science][Medline].
18.
Pennati, G,
Bellotti M,
Ferrazzi E,
Bozzo M,
Pardi G,
and
Fumero R.
Blood flow through the ductus venosus in human fetus: calculation using doppler velocimetry and computational findings.
Ultrasound Med Biol
24:
477-487,
1998[Web of Science][Medline].
19.
Pennati, G,
Bellotti M,
Ferrazzi E,
Rigano S,
and
Garberi A.
Hemodynamic changes across the human ductus venosus: a comparison between clinical findings and mathematical calculations.
Ultrasound Obstet Gynecol
9:
383-391,
1997[Web of Science][Medline].
20.
Pennati, G,
Bellotti M,
and
Fumero R.
Mathematical modelling of the human foetal cardiovascular system based on Doppler ultrasound data.
Med Eng Phys
19:
327-335,
1997[Web of Science][Medline].
21.
Reed, KL,
Chaffin DG,
Anderson DF,
and
Newman AT.
Umbilical venous velocity pulsations are related to atrial contraction pressure waveforms in fetal lambs.
Obstet Gynecol
89:
953-956,
1997[Medline].
22.
Reuss, ML,
Rudolph AM,
and
Dae MW.
Phasic blood flow patterns in the superior and inferior venae cavae and umbilical vein of fetal sheep.
Am J Obstet Gynecol
145:
70-78,
1983[Medline].
23.
Rudolph, CD,
and
Rudolph AM.
Fetal and postnatal hepatic vasculature and blood flow.
In: Fetal and Neonatal Physiology, edited by Polin RA,
and Fox WW.. Philadelphia, PA: Saunders, 1998.
24.
Schmidt, KG,
Silverman NH,
and
Rudolph AM.
Assessment of flow events at the ductus venosus-inferior vena cava junction and at the foramen ovale in fetal sheep by use of multimodal ultrasound.
Circulation
93:
826-833,
1996
25.
Sherer, DM,
Fromberg RA,
and
Divon MY.
Prenatal ultrasonographic assessment of the ductus venosus: a review.
Obstet Gynecol
88:
626-632,
1996[Medline].
26.
Tchirikov, M,
Hecher K,
Deprest J,
Zikulnig L,
Devlieger R,
and
Schröder HJ.
Doppler ultrasound measurements in the central circulation of anesthetized fetal sheep during obstruction of umbilical-placental blood flow.
Ultrasound Obstet Gynecol
18:
565-661,
2001.
27.
Tchirikov, M,
and
Schröder HJ.
Single case report: late gestational fetal sheep may survive blockage of the ductus venosus for 1 week.
Placenta
19:
333-336,
1998[Medline].
28.
Van Splunder, IP,
Stijnen T,
and
Wladimiroff JW.
Fetal pressure gradient estimations across the ductus venosus in early pregnancy using Doppler ultrasonography.
Ultrasound Obstet Gynecol
6:
334-339,
1995[Medline].
This article has been cited by other articles:
![]() |
C. Ebbing, S. Rasmussen, K. M. Godfrey, M. A. Hanson, and T. Kiserud Hepatic Artery Hemodynamics Suggest Operation of a Buffer Response in the Human Fetus Reproductive Sciences, February 1, 2008; 15(2): 166 - 178. [Abstract] [PDF] |
||||
![]() |
M Nii, R M Hamilton, L Fenwick, J C P Kingdom, K S Roman, and E T Jaeggi Assessment of fetal atrioventricular time intervals by tissue Doppler and pulse Doppler echocardiography: normal values and correlation with fetal electrocardiography Heart, December 1, 2006; 92(12): 1831 - 1837. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Acharya, T. Erkinaro, K. Makikallio, T. Lappalainen, and J. Rasanen Relationships among Doppler-derived umbilical artery absolute velocities, cardiac function, and placental volume blood flow and resistance in fetal sheep Am J Physiol Heart Circ Physiol, April 1, 2004; 286(4): H1266 - H1272. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |