1Carl Ludwig Institute of Physiology, University of Leipzig, D-04103 Leipzig, Germany; and 2Millar Instruments, Houston, Texas 77023-5417
Submitted 7 April 2003 ; accepted in final form 17 July 2003
| ABSTRACT |
|---|
|
|
|---|
90° or less to the shaft. The measuring catheter is advanced across the pulmonary valve into the PA. Measurements of pulmonary arterial pressure were made in five male Long Evans (364 ± 7 g body wt) and five female Sprague-Dawley (244 ± 7 g body wt) rats under control conditions. The effects of infusion of norepinephrine (0.1 mg·kg1·h1 iv for 20-min duration) were tested in Long Evans rats. Pulmonary arterial systolic pressure measurements were 34.0 ± 0.8 and 29.5 ± 0.4 mmHg, and diastolic pressure values were 23.6 ± 0.8 and 18.1 ± 0.6 mmHg in male Long Evans and female Sprague-Dawley rats, respectively. Norepinephrine induced an increase in pulmonary arterial systolic (40.8 ± 0.1 mmHg) and diastolic (28.6 ± 0.4 mmHg) pressures and an elevation in pulmonary vascular resistance from a control value of 0.093 ± 0.003 to 0.103 ± 0.004 mmHg·kg·min·ml1.
cardiac output; contractility; norepinephrine; right heart
In a one-step approach, a 3.5-Fr umbilical vessel catheter prefilled with heparinized saline had a 90° angle to the shaft at the distal end. When it was inserted into the right jugular vein of a rat, it straightened out and then resumed its original shape once it was located in the right ventricle. It was further advanced so that the tip slipped through the pulmonary valve (16). A similar one-step technique was used thereafter by other investigators (9, 11, 13, 17, 18). Several approaches were made using an introducer or a cannula. In 1972, a method was developed that made use of a cannula-catheter assembly. A fluid-filled catheter with a pigtail shape at the distal end was contained in a well-fitting cannula. This assemly was introduced into the right ventricle. From this point, the inner tubing was pushed out of the cannula whereupon it resumed its original shape and was introduced by delicate manipulation into the PA. The tubing was filled with saline and heparin solution and was connected to a pressure transducer (6). In another technique, the tip of a catheter had a "shepherd's crook" shape that allowed the catheter arch to press against the RV wall while the tip remained free. This catheter, with a cannula sleeve, was placed in the right ventricle, and the tip of the catheter was manipulated into the PA (5). A fluid-filled catheter was also used in an approach whereby measurements of PA were obtained in awake rats. An introducer, which was a blunted 7.5-cm, 19-gauge needle with the tip turned up 30°, was passed inside by a Silastic catheter that had a notched end hole. This was flushed with heparinized saline and was attached by a 25-gauge blunted needle to a pressure transducer. After the introducer had been placed in the RV cavity, the catheter was introduced into the PA, whereupon the introducer was removed, and the catheter was exteriorized in the back of the animal's neck (15). In a recent study, a similar technique was applied for recording of pulmonary arterial pressure in rats over 1 day (7). Several other studies that involved measurements of pulmonary arterial pressure also used this type of technique (4, 10, 12, 14).
In mice, a catheter with an outer diameter of 0.25 mm and a curved tip was straightened with an angioplasty guide wire to facilitate insertion in the right atrium under fluoroscopic control. The straight wire was removed, and the catheter with the curved tip was passed into the right ventricle. A soft-tip coronary artery angioplasty guide wire was then inserted, and the catheter was passed over the guide wire into the main PA for the measurement of pressure using a pressure transducer (2).
As to the previous attempt with an ultraminiature catheter pressure transducer, a 3-Fr Millar catheter was inserted into the right femoral vein and advanced via the inferior cava vein into the right atrium. By rotating the catheter tip, it was possible to place it in the right ventricle and then further advance it into the PA. This technique was applied in studies on the functional effects of hypoxia (20) and triiodothyronine (19). The procedure associated with this approach, however, required much time, was not always successful, and was not considered to be suitable for routine measurements (23).
Therefore, considerable effort during the past years has been put into development of a microtip catheter that can be applied in rats via the jugular vein for the measurement of pulmonary arterial pressure. With the use of new materials and miniaturization techniques, such an ultraminiature catheter pressure transducer was recently constructed and used in rats. In this report, we describe this catheter and the technique for pulmonary arterial catheterization. The first results obtained in two rat strains under control conditions and after infusion of norepinephrine (NE) are presented.
| MATERIALS AND METHODS |
|---|
|
|
|---|
0.023 in.; outer diameter, 0.029 in.; and length, 7 cm. There is a bifurcation near the proximal end of the sheath, and a heavy-walled polyimide tubing contains the pulling wire, which exits and terminates in a deflection controller. The deflection controller has a knurled knob that can be turned with the thumb to produce a tip deflection as required. The connection between the knurled wheel and the tip consists of a 3-mm nickel-titanium wire that was previously bonded securely to the catheter tip. This guiding wire is brought up along the outside of the sheath and fastened at the tip of the sheath with a small polyimide ring and some hard epoxy. Although it is bonded at the tip, it is free to move in the consecutive support rings down the sheath and along the handle.
The experimental procedure used conforms with the Guide for the Care and Use of Laboratory Animals published by the US National Institutes of Health [DHEW Publication No. (NIH) 85-23, Revised 1996, Office of Science and Health Reports, DRR/NIH, Bethesda, MD 20205] and was approved by the appropriate state agency of Saxony. Male Long Evans rats (n = 5; 364 ± 7 g body wt) and female Sprague-Dawley rats (n = 5; 244 ± 7 g body wt) were used. All rats were obtained from Charles River (Sulzfeld, Germany). Rats were allowed to move freely in their cages and had access to tap water and rat chow diet (Altromin C 100, Altromin; Lage, Germany).
The animals were anesthetized with thiopental sodium (Trapanal, 60 mg/kg ip, Byk Gulden; Konstanz, Germany). A cannula was inserted into the trachea to control respiration of the spontaneously breathing animals. First, the 1.4-Fr measuring catheter was inserted through the right jugular vein into the right ventricle as evidenced by the typical RV pressure curve; the introducer was then slipped over it. The tip of the introducer was deflected to an angle of
90° or less to the shaft by pulling the guiding wire. The measuring catheter was advanced across the pulmonary valve into the PA. Finally, the deflection was released, and the introducer was removed. The successive steps are illustrated in Fig. 1.
|
After collection of the RV and pulmonary arterial data, the left ventricular (LV) catheter (model SPR-249, 3-Fr, Millar) was placed in the right carotid artery and advanced upstream in the aorta into the left ventricle. Thereafter, NE dissolved in 0.9% NaCl at a dose of 0.1 mg·kg1·h1 iv was applied as a continuous infusion via a catheter (Vygon; Aachen, Germany) positioned in the left jugular vein. To prevent oxidation, 100 mg/l ascorbic acid was added. After 15 min, the pulmonary arterial catheter was withdrawn into the right ventricle to measure RV pressure under steady-state conditions of NE application. Cardiac output was measured using the thermodilution method (19) at the beginning of the experiment (basal conditions, CTRL) and at the end of the 20-min NE infusion period using a Cardiomax II thermodilution machine (Columbus Instruments; Columbus, OH).
Heart rate (HR); RV, LV, and pulmonary arterial pressures; and the rates of rise and fall of ventricular pressures (LV and RV dP/dtmax and dP/dtmin, respectively) were recorded continuously on a personal computer at a sampling rate of 2 kHz as described previously (3). Total peripheral and pulmonary vascular resistance values were calculated by dividing mean aortic and pulmonary arterial pressures, respectively, by cardiac output normalized to body weight.
All data are presented as means ± SE. For statistical comparison before and after NE treatment, a paired t-test was used (SigmaStat 2.0, SPSS; Chicago, IL). A P value <0.05 was considered significant.
| RESULTS |
|---|
|
|
|---|
|
Pulmonary arterial systolic pressure (PASP) values were 34.0 ± 0.8 and 29.5 ± 0.4 mmHg, and pulmonary arterial diastolic pressure (PADP) measurements were 23.6 ± 0.8 and 18.1 ± 0.6 mmHg in male Long Evans and female Sprague-Dawley rats, respectively (Table 1). RV systolic pressure (RVSP) and PASP were of the same magnitude, which indicates that no obstruction of the pulmonary valve had been induced when the measuring catheter was located in the PA (Fig. 2C). In addition, there was no change in RV end-diastolic pressure (RVEDP). Pulmonary vascular resistance (PVR) was comparable in Long Evans and Sprague-Dawley rats (Table 1). The functional parameters of the left ventricle (Table 2) were comparable to those obtained in earlier studies (3, 21).
|
|
Infusion of NE in male Long Evans rats accelerated HR and elevated RVSP, LVSP, RV dP/dtmax, and LV dP/dtmax values (Table 3). It induced an increase in both PASP (40.8 ± 0.1 mmHg) and PADP (28.6 ± 0.4 mmHg; Fig. 3 and Table 3) and an elevation in PVR from 0.093 ± 0.003 to 0.103 ± 0.004 mmHg·kg·min· ml1 (Table 3; P = 0.053). Similar effects were observed when NE was administered in female Sprague-Dawley rats (data not shown).
|
|
| DISCUSSION |
|---|
|
|
|---|
In previous attempts, preformed catheters (16), introducers (5, 6, 15), or guide wires (2) were used that released the preformed shape of the measuring catheter when they were removed from it. In the present approach, the tip of the introducer becomes mechanically deflected when it is placed in the right ventricle (see Fig. 1). This seems to be an appropriate and straightforward approach that is reproducible in the hands of experienced researchers and may ultimately result in a routine method. At the present stage, this catheterization procedure is successful when much experience has already been obtained with similar techniques in small laboratory animals or after extensive practice and training periods.
The shape of the PA pressure curve resembled that obtained in the aorta, although at a lower level. There was a prominent dicrotic notch during the diastolic decrease in pressure (see Fig. 2). The PASP was of the same height as the RVSP (see Fig. 2C); this is similar to the situation in LV catheterization where LVSP is the same as systolic aortic pressure (21). RVEDP did not increase when the catheter was placed in the PA as indicated by unchanged RVEDP before and after pulmonary arterial catheterization. Also in LV catheterization, LVEDP did not increase when the catheter was positioned in the left ventricle even for extended periods of time (21). Thus there is no obstruction of the pulmonary or aortic valve during these catheterization procedures. The values for PASP and PADP agree quite well with those published (5, 6, 15, 16). The rhythmic respiratory changes (see Fig. 2) were clearly seen on all records of pulmonary arterial pressure, which is in accordance with previous data published thus far for rats (5, 7, 15).
Stimulation of male Long Evans rats with NE induced the well-known positive chronotropic and inotropic effects of both heart chambers (Table 3). Also, RVSP was elevated in this rat strain, although not to the same extent as previously observed in Sprague-Dawley rats (24). The stimulation intervals with NE, however, were longer in most of the studies in which NE had been shown to elevate RVSP (1, 8). A dose-dependent increase in pulmonary arterial pressure was observed earlier (16), whereas a variable response was reported in another investigation (5). In this study, PASP, PADP, and PVR increased in parallel with the increase in RVSP (Table 3). That the values of PVR can be obtained in all experimentally induced disturbances of pulmonary circulation is important for future studies.
It may be envisaged that the pulmonary arterial catheter that is now available for rats may be miniaturized to such an extent that it will become applicable to mice. Because transgenic and knockout mice are generated in increasing numbers, pulmonary arterial catheterization is needed to examine the effects of overexpressing or deleting a protein or physiological process on pulmonary circulation.
| DISCLOSURES |
|---|
|
|
|---|
| FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
| REFERENCES |
|---|
|
|
|---|
expression. Pflügers Arch 445: 289296, 2002.[ISI][Medline]
This article has been cited by other articles:
![]() |
P. Pokreisz, G. Marsboom, and S. Janssens Pressure overload-induced right ventricular dysfunction and remodelling in experimental pulmonary hypertension: the right heart revisited Eur. Heart J. Suppl., December 1, 2007; 9(suppl_H): H75 - H84. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |