Departments of 1Cardiology and 2Cardiothoracic Surgery, Leiden University Medical Center, 2300 RC Leiden, The Netherlands; and 3Department of Cardiac Surgery, San Raffaele Hospital, 20132 Milan, Italy
Submitted 12 June 2003 ; accepted in final form 8 October 2003
| ABSTRACT |
|---|
|
|
|---|
ventricular function
| METHODS |
|---|
|
|
|---|
All patients gave informed consent to participate in the study, and the procedures were conducted in accordance with institutional guidelines. The investigation conforms to the principles outlined in the Declaration of Helsinki (34). Twelve CHF patients (New York Heart Association class III/IV) with LBBB were studied during diagnostic catheterization. Six CAD patients were studied in the operating room before undergoing coronary artery bypass grafting.
Protocol
CHF patients underwent diagnostic catheterization, including thermodilution cardiac output, left ventriculography, and coronary angiography. In addition, a conductance catheter was placed in the LV via the femoral artery, and a temporary pacing lead was positioned in the right atrium.
Before catheterization, the CHF patients were studied by echocardiography. We performed tissue-Doppler imaging as described in detail elsewhere (3) to determine myocardial velocities in basal septal and lateral segments. The time delay between peak systolic velocity in the septum and the lateral wall was determined as an index of mechanical dyssynchrony.
CAD patients received total intravenous anesthesia with target-controlled infusion of propofol and remifentanyl (1.52 µg/ml, respectively, 510 ng/ml blood concentration). A continuous cardiac output catheter was placed in the pulmonary artery via the jugular vein. After a midline sternotomy and before cardiopulmonary bypass was started, a conductance catheter was placed in the LV via a purse-string suture on the ascending aorta. External pacing leads were placed on the right atrium.
Measurements. The conductance catheter enables online measurement of five segmental volume (Vseg,i) slices perpendicular to the LV long axis. We used 7-Fr combined pressure-conductance catheters with 1-cm interelectrode spacing (CD Leycom; Zoetermeer, The Netherlands). The catheter was connected to a Cardiac Function Lab (CD Leycom) for online display and acquisition (sample frequency 250 Hz) of segmental and total LV volumes, LV pressure, and ECG. Total LV volume (VLV) is obtained as the instantaneous sum of the segmental volumes. VLV was calibrated using thermodilution and hypertonic saline dilution, as previously described (2). Periods of
10 s at a paced heart rate of 80 beats/min were selected for off-line analysis using custom-designed software.
Global cardiac function and nonuniform mechanical performance. Global LV function was measured by cardiac index (CI), end-diastolic and end-systolic volume index (EDVI and ESVI), ejection fraction (EF), end-systolic and end-diastolic pressure (ESP and EDP), maximal and minimal rate of pressure change (dP/dtmax, dP/dtmin), and the time constant of relaxation (
). LV systolic elastance was estimated by ESP/ESVI, and (dP/dtmax)/EDVI was calculated as a relatively load-independent index of systolic function.
Nonuniform LV performance was determined from the segmental LV conductance signals and characterized by the following indexes.
Mechanical dyssynchrony. At each time point, a segmental signal was defined as dyssynchronous if its change (i.e., dVseg/dt) was opposite to the simultaneous change in the total LV volume (dVLV/dt). Segmental dyssynchrony is quantified by calculating the percentage of time within the cardiac cycle that a segment is dyssynchronous. Overall LV dyssynchrony (DYS) was calculated as the mean of the segmental dyssynchronies (23). DYS may be calculated within each specified time interval: we determined DYS during systole (DYSS) and diastole (DYSD), with systole defined as the period between the moments of dP/dtmax and dP/dtmin.
Internal flow. Nonuniform contraction and filling is associated with ineffective shifting of blood volume within the LV. This internal flow (IF) is quantified by calculating the sum of the absolute volume changes of all segments and subtracting the absolute total volume change: IF(t) = [
|dVseg,i(t)/dt | | dVLV(t)/dt|]/2. Note that dVLV(t)/dt represents the effective flow into or out of the LV. Thus IF measures the segment-to-segment blood volume shifts, which do not result in effective filling or ejection. Division by two takes into account that any noneffective segmental volume change is balanced by an equal but opposite volume change in the remaining segments. IF fraction (IFF) is calculated by integrating IF(t) over the full cardiac cycle and dividing by the integrated absolute effective flow.
Mechanical dispersion. In the CHF patients, we expected a substantial dispersion in the onset of contraction between the segments. This dispersion was assessed by segmental lag times, tlag,i, which were determined by calculating the cross correlations between VLV(t) and Vseg,i(t + tlag,i) for all systolic time points (i.e., between dP/dtmax and dP/dtmin). For each segment we determined the tlag,i, which produced the highest linear correlation. Thus if tlag,i < 0, segment i precedes the global ejection and vice versa. Mechanical dispersion (DISP) was defined as 2·SD of the segmental lag times.
Statistical Analysis
All data are presented as means ± SD. Comparisons between the CAD and CHF groups were performed by unpaired t-tests. We performed receiver-operating characteristic (ROC) curve analysis to test the diagnostic performance of the various indexes to discriminate the patient groups (38). Sensitivities and specificities at the optimal cutoff point were determined. Comparison between conductance-derived and tissue-Doppler-derived dyssynchrony indexes was made by linear regression analysis.
| RESULTS |
|---|
|
|
|---|
|
|
Hemodynamic data are summarized in Table 1. EF and dP/dtmax indicate more pronounced systolic dysfunction, whereas ESP/ESVI and (dP/dtmax)/EDVI show depressed contractile state, whereas
and EDP indicate impaired diastolic function in CHF. Differences in EDVI, ESVI, and CI were present but did not reach statistical significance. Pronounced differences between CAD and CHF were found in DYS, IFF, and DISP. For both groups, dyssynchrony and IFs were highest in diastole, and the apical segments were the most affected (Fig. 3A). In both groups, mechanical dispersion in the longaxis direction was present, but it was twice as large in CHF. Figure 3B shows that contraction started in the basal segment and, on the average, subsequent segments (1 cm slices), followed by 5.9 ms for CAD and by 12.4 ms in CHF patients.
|
|
The diagnostic value of the various indexes to discriminate the two patient groups was tested using ROC analysis. Table 1 shows the results with the optimal cutoff values, and corresponding sensitivities and specificities. As expected, QRS duration accurately delineates the groups with a cut-off value of 107 ms. The dyssynchrony indexes DYS and IFF show excellent sensitivity/specificity values, which are higher than the best hemodynamic indexes EF and dP/dtmax. The other hemodynamic indexes show lower sensitivity/specificity reflecting a substantial overlap of the values between the two groups.
Tissue-Doppler measurements were performed in the CHF patients and revealed a significant difference in the timing of peak systolic velocities of the septum and the lateral wall. The average septal-to-lateral delay was 89 ± 43 ms, indicating a dyssynchronous intraventricular contraction pattern. We compared the septal-to-lateral delay times with the conductance-derived dyssynchrony indexes using linear regression analysis. The results (Fig. 4) show highly significant correlations with DYS (r2 = 0.59, P = 0.003) and IFF (r2 = 0.63, P = 0.002). The relation with DISP did not reach statistical significance (r2 = 0.26, P = 0.089).
|
| DISCUSSION |
|---|
|
|
|---|
We introduce novel indexes to quantify dyssynchrony based on volume signals acquired with the conductance catheter during cardiac catheterization. The conductance catheter was validated previously (2), and the segmental signals reflect instantaneous volume slices perpendicular to the LV long axis as obtained by cine-computerized tomography (32). Currently, the conductance catheter is used mainly to assess global systolic and diastolic function (9, 11, 12, 17). Quantification of nonuniform mechanical function and dyssynchrony may lead to a more complete diagnosis of ventricular dysfunction (23, 24). Moreover, it may guide therapy, because patients with extensive dyssynchrony are likely to benefit from resynchronization therapy (15).
We compared CHF versus CAD patients. The groups show pronounced differences for DYS, IFF, and DISP, which indicate a high sensitivity and specificity of these dyssynchrony indexes. QRS duration, dP/dtmax, and
show a similar discrimination between the groups and may also partly reflect dyssynchrony. However, whereas the conductance catheter-derived indexes directly measure regional mechanical events throughout the cardiac cycle, QRS duration reflects the underlying electrical activation, and studies indicate that mechanical and electrical synchrony may diverge (14).
and dP/dtmax have also been shown to be markers of dyssynchrony but they more indirectly reflect the integrated effects of spatially dispersed mechanical (de)activation during the isovolumic relaxation and contraction periods. Dyssynchrony is likely to be most pronounced in the isovolumic phases, which explains the sensitivity of parameters that reflect these periods. However, the consequences of dyssynchrony on the effectiveness of ejection and filling are important for cardiac pump performance, so that indexes selectively reflecting those cardiac phases may be of high value.
In the CHF patients, we compared the conductance-derived dyssynchrony indexes with the delay in timing of peak systolic velocity between the septal and lateral wall as obtained by tissue-Doppler echocardiography. Septal-to-lateral delay has recently been introduced as an index of mechanical dyssynchrony. We found a significant correlation for both DYS and IFF, but DISP did not reach a statistically significant correlation. The various indexes measure different characteristics. Whereas the tissue-Doppler method compares the timing of peak velocity between two regions that are likely to show the largest phase shift, the conductance-derived indexes are based on a comparison of the volume changes of short axis slices and global LV volume changes. Apparently, patients with a larger septal-to-lateral delay also show more segmental dyssynchrony, as reflected by DISP and IFF. It is unknown whether this correspondence is specific for LBBB-CHF patients or is more generally valid. The lack of correlation with DISP is unclear. It may be because the index is less sensitive than DISP or IFF as shown in the comparison between CAD and CHF patients, or the index may inherently be more prone to errors. Interestingly, within the group of CHF patients, neither septal-to-lateral delay nor the conductance-derived indexes showed a significant correlation with QRS duration (Fig. 5). This finding is consistent with other reports (14, 18) indicating that electrical dyssynchrony does not necessarily predict mechanical dyssynchrony, which prompts a need for methods to accurately detect mechanical dyssynchrony.
|
Our approach may offer several technical advantages. After catheter placement, the signals are obtained continuously without operator interaction. In the present study, the analysis was performed off-line, but real-time display of dyssynchrony indexes is technically feasible and should enable immediate quantification of the effects of interventions, and, e.g., the effects of changes in pacemaker settings. The method is invasive, but positioning of the catheter in the LV largely eliminates problems with through-plane motion inherent in most imaging methods. Heart failure is often associated with substantial beat-to-beat hemodynamic variations due to changes in cycle length, cardiopulmonary interaction and conduction disturbances. Thus techniques (like magnetic resonance imaging) that require hemodynamic steady-state and beat averaging to increase signal-to-noise may filter out important components of dyssynchrony. Furthermore, the temporal resolution of the conductance signals (4 ms) is relatively high.
Determination of absolute LV volume from the conductance catheter requires careful calibration (2). In the present study, calibration factor parallel conductance was obtained by the hypertonic saline method and slope factor-
by thermodilution. Slope factor-
was significantly lower in the CHF patients than in the CAD patients (0.38 ± 0.22 vs. 0.67 ± 0.08, P < 0.01), and parallel conductance was significantly higher (214 ± 60 vs. 131 ± 48 ml; P < 0.01). These findings are consistent with previous studies and reflect more electrical field inhomogeneity in the enlarged hearts in the CHF group. However, the conductance catheter has been used extensively in enlarged hearts, and validation studies show that accurate volumes estimates can be obtained, provided that the appropriate calibration is performed (19). Another advantage is that the dyssynchrony indexes can be calculated from the raw segmental conductance signals and do not require calibration. Correction for parallel conductance (offset factor) is not required because the calculations are based on volume changes, and correction for slope factor-
is not required because segmental volume changes are judged relative to the global LV volume changes. The latter, however, implicitly assumes that the segmental slope factors are all the same (and thus equal to the slope factor for global volume). This assumption may be a concern because theoretical studies indicate that volume in the segments closest to the current electrodes may be relatively underestimated due to electric field inhomogeneity, especially in enlarged hearts (26, 27, 35). To test the effects of such underestimations, if present, on our dyssynchrony indexes we recalculated DYS, IFF, and DISP after correcting segments 1 and 5 for an assumed underestimation of 20% and segments 2 and 4 for an assumed underestimation of 10%. Theoretical studies indicate that underestimation in this order of magnitude may be present (26, 35). The results were compared with the original data using Bland-Altman analysis (4) (Fig. 6). The analysis shows no significant bias and fairly narrow limits of agreement for each of the indexes, indicating that the influence of a potential underestimation of the outer segments on the dyssynchrony indexes is relatively small. Although the mean dyssynchronies were higher in the CHF patients, the differences as detected by the Bland-Altman analysis were not systematically different between the two groups.
|
The methods for quantifying dyssynchrony presented in this study show similarities with an approach previously published by Strum and Pinsky (29). They used segmental volume signals obtained from conductance catheters to quantify regional wall motion abnormalities and referenced amplitudes and phase angles of the segmental signals to the global LV volume signal. The phase angle analysis is comparable to our DISP index. However, Strum and Pinsky measured (in degrees) the relative distances between time points of regional minimal volume and global end systole, whereas we used the entire systolic wave forms and used cross correlation to determine the lag time between segmental and global volume signals. In addition, they compared regional maximal stroke volume with effective stroke volume. The latter was measured using maximums and minimums of the total LV volume as gated markers of the time when regional contraction would contribute to total LV ejection. This effective stroke volume analysis is comparable to our IF calculation (IFF index), which determines at each time point throughout the cardiac cycle whether segmental volume changes are effective (i.e., contributing to global volume changes) or lead to ineffective (segment to segment) IF. Strum and Pinsky (28, 30) applied these concepts in animal studies where reversible regional myocardial dysfunction was induced by intracoronary infusion of esmolol and global inotropy was modulated by dobutamine infusions.
Limitations
Optimally, the conductance catheter is placed in a straight position from the aortic valve to the LV apex. In the operating room, we used transesophageal echocardiography (31) and in the catheterization laboratory, we used angiography to guide positioning. However, occasionally arrhythmias necessitate pulling back the catheter slightly from the apical position. In addition, the distance from the pigtail to the first measurement electrode is
2 cm. Thus volume changes in the most apical part of the LV are not measured. If this region is highly dyssynchronous, as might be the case in patients with apical infarcts, underestimation of dyssynchrony by our methodology may be expected.
The patient groups in our study were investigated under different conditions. For practical purposes, we studied the CAD patients in the operating room during anesthesia and after sternotomy, whereas the CHF patients were awake and studied in the catheterization laboratory. These differences may have affected the comparisons between the two groups. Propofolremifentanyl anesthesia is known to have myocardial depressant and vasodilating properties (16, 22), whereas sternotomy and pericardiotomy are associated with alterations in loading conditions (7). Given the anesthesia-related cardiodepression in the CAD patients, one may expect that the differences in the hemodynamic indexes would have been more pronounced in case both groups had been studied awake. Whether these changes affect the level of dyssynchrony is not well known, but studies in dogs with regional stunning show unchanged LV wall asynchrony after systemic inotropic stimulation (21). Thus we do not expect that the differences in mechanical dyssynchrony between the groups were importantly influenced by the different experimental conditions.
Furthermore, we did not study normal subjects. Thus future studies are required to establish a normal range for the dyssynchrony indexes.
Finally, the segmental conductance catheter signals do not provide an anatomic view but represent the total volume of slices perpendicular to the LV long axis. Thus, e.g., in CAD patients, abnormal regional wall motion might be obscured by compensatory wall motions within the same circumferential segment. The proposed dyssynchrony indexes therefore reflect intersegmental differences in contraction and filling and may underestimate phase changes obtained by comparing regional lateral and septal wall motions, e.g., using tissue Doppler imaging.
In conclusion, the proposed indexes quantify various aspects of mechanical dyssynchrony using conductance catheter methodology which, at the same time, can be used for assessment of global systolic and diastolic (dys)function. Diagnostic and prognostic value of the dyssynchrony indexes requires further investigation.
| ACKNOWLEDGMENTS |
|---|
The study was supported by The Netherlands Heart Foundation Grant NHS-2002B133.
| FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
B. Kirn, A. Jansen, F. Bracke, B. van Gelder, T. Arts, and F. W. Prinzen Mechanical discoordination rather than dyssynchrony predicts reverse remodeling upon cardiac resynchronization Am J Physiol Heart Circ Physiol, August 1, 2008; 295(2): H640 - H646. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Gupta, F. Khan, M. Shapiro, S. G. Weeks, S. E. Litwin, and A. D. Michaels The associations between tricuspid annular plane systolic excursion (TAPSE), ventricular dyssynchrony, and ventricular interaction in heart failure patients Eur J Echocardiogr, April 19, 2008; (2008) jen147v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Rao, Y. Ling, R. He, A. L. Gilbert, N. G. Frangogiannis, J. Wang, S. F. Nagueh, and D. S. Khoury Integrated multimodal-catheter imaging unveils principal relationships among ventricular electrical activity, anatomy, and function Am J Physiol Heart Circ Physiol, February 1, 2008; 294(2): H1002 - H1009. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Arita, G. P. Sorescu, B. T. Schuler, L. S. Schmarkey, J. D. Merlino, J. Vinten-Johansen, A. R. Leon, R. P. Martin, and D. Sorescu Speckle-tracking strain echocardiography for detecting cardiac dyssynchrony in a canine model of dyssynchrony and heart failure Am J Physiol Heart Circ Physiol, July 1, 2007; 293(1): H735 - H742. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Lieberman, L. Padeletti, J. Schreuder, K. Jackson, A. Michelucci, A. Colella, W. Eastman, S. Valsecchi, and D. A. Hettrick Ventricular Pacing Lead Location Alters Systemic Hemodynamics and Left Ventricular Function in Patients With and Without Reduced Ejection Fraction J. Am. Coll. Cardiol., October 17, 2006; 48(8): 1634 - 1641. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. P. Strum and M. R. Pinsky Modeling Ischemia-Induced Dyssynchronous Myocardial Contraction Anesth. Analg., October 1, 2006; 103(4): 846 - 853. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A.F. Tulner, P. Steendijk, R. J.M. Klautz, J. J. Bax, M. J. Schalij, E. E. van der Wall, and R. A.E. Dion Surgical ventricular restoration in patients with ischemic dilated cardiomyopathy: Evaluation of systolic and diastolic ventricular function, wall stress, dyssynchrony, and mechanical efficiency by pressure-volume loops. J. Thorac. Cardiovasc. Surg., September 1, 2006; 132(3): 610 - 620. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Steendijk, S. A. Tulner, J. J. Bax, P. V. Oemrawsingh, G. B. Bleeker, L. van Erven, H. Putter, H. F. Verwey, E. E. van der Wall, and M. J. Schalij Hemodynamic Effects of Long-Term Cardiac Resynchronization Therapy: Analysis by Pressure-Volume Loops Circulation, March 14, 2006; 113(10): 1295 - 1304. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Edvardsen, O. Rodevand, K. Endresen, and H. Ihlen Interaction between left ventricular wall motion and intraventricular flow propagation in acute and chronic ischemia Am J Physiol Heart Circ Physiol, August 1, 2005; 289(2): H732 - H737. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A.F. Tulner, P. Steendijk, R. J.M. Klautz, J. J. Bax, M. I.M. Versteegh, E. E. van der Wall, and R. A.E. Dion Acute hemodynamic effects of restrictive mitral annuloplasty in patients with end-stage heart failure: Analysis by pressure-volume relations J. Thorac. Cardiovasc. Surg., July 1, 2005; 130(1): 33 - 40. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. J. Schreuder, F. Maisano, A. Donelli, J. R.C. Jansen, P. Hanlon, J. Bovelander, and O. Alfieri Beat-to-Beat Effects of Intraaortic Balloon Pump Timing on Left Ventricular Performance in Patients With Low Ejection Fraction Ann. Thorac. Surg., March 1, 2005; 79(3): 872 - 880. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. J. Schreuder, A. Castiglioni, F. Maisano, P. Steendijk, A. Donelli, J. Baan, and O. Alfieri Acute decrease of left ventricular mechanical dyssynchrony and improvement of contractile state and energy efficiency after left ventricular restoration J. Thorac. Cardiovasc. Surg., January 1, 2005; 129(1): 138 - 145. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Steendijk, S. A.F. Tulner, M. Wiemer, R. A. Bleasdale, J. J. Bax, E. E. van der Wall, J. Vogt, and M. J. Schalij Pressure-volume measurements by conductance catheter during cardiac resynchronization therapy Eur. Heart J. Suppl., August 1, 2004; 6(suppl_D): D35 - D42. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||