HBOC-201 (Biopure; Cambridge, MA) is a glutaraldehyde-polymerized, bovine Hb solution. HBOC-201 has a similar structure to human Hb but has been biochemically altered to enhance O2 unloading (17). Unlike human Hb, bovine Hb blood substitutes do not require 2,3-diphosphoglycerate to lower the O2 affinity. The O2 affinity of HBOC-201 (P50,
40 mmHg) is regulated by the chloride ion content of human plasma, which is adequate to lower O2 affinity of the bovine Hb to a satisfactory level for unloading in human tissue. The Bohr effect has also been shown to be more pronounced in bovine Hb, which allows an even better O2 delivery at lower pH values. It has also been shown that this blood substitute does not change hemodynamics with respect to PCO2 and pH (1, 6, 12, 17, 19, 23). HBOC-201 is modified by cross-linking of the
,
-dimers to form a tetrameric structure that does not dissociate as rapidly as free Hb (14). Similarly, glutaraldehyde polymerization in HBOC-201 results in a prolonged intravascular half-life of polymeric Hb due to reduced clearance of the solution (8).
The Hb in HBOC-201 is stroma free and is not bound to red blood cells; therefore, the Hb may be easily transported in blood plasma to places where red blood cell flow is not capable, which is an exciting characteristic in the possibility of an ischemic event. This characteristic of HBOC-201 along with its heightened ability to unload O2 into tissue and the easy acquisition and use of the product may allow bovine Hb to be more beneficial in emergency settings than whole blood. It may even prove beneficial in hypoxic settings such as acute coronary artery occlusion, where its unique ability to unload O2 while traveling in blood plasma may be exploited. The present study was designed to determine whether administration of HBOC-201 before acute coronary artery occlusion could alter the extent of myocardial injury.
| MATERIALS AND METHODS |
|---|
|
|
|---|
HBOC-201. Glutaraldehyde-polymerized bovine Hb in a buffered physiological solution of electrolytes was obtained from Biopure (Cambridge, MA).
Animal preparation. Twenty-two heartworm-free mongrel dogs of either sex that weighed 16.327.0 kg (average wt, 20.2 ± 0.6 kg) were fasted overnight. The dogs were randomized and anesthetized with pentobarbital sodium (40 mg/kg iv) and were subsequently endotracheally intubated and ventilated with room air by use of a Harvard respirator (model 613; Harvard Apparatus; Boston, MA). Respiratory rate and stroke volume were adjusted to maintain arterial blood gas and pH levels within normal physiological ranges. Anesthesia was maintained by intravenous administration of pentobarbital sodium as needed. A standard lead II ECG was used to continuously monitor heart function. A polyethylene catheter was placed in the right femoral vein for fluid and drug administration. Both femoral arteries were cannulated for collection of hemodynamic data and blood samples. In the right femoral artery, an 8-Fr pigtailed pressure-transducer catheter (MPC-500; Millar; Houston, TX) was advanced to the left ventricle. In the left femoral artery, a 5-Fr pressure-transducer catheter (MPC-350; Millar) was advanced into the abdominal aorta. The chest was opened by a median sternotomy, and the pericardium was incised and tented to cradle the heart. A proximal portion of the left anterior descending artery (LAD) was dissected away from the epicardium to allow a microaneurysm clip to be placed on the vessel to induce complete coronary artery occlusion.
Experimental protocol. Thirty minutes were allowed for stabilization after initial preparation and surgical procedures were complete. Hemodynamic data were obtained at baseline and at 30-min intervals thereafter until the end of the experiment. Blood samples were taken at baseline, 1 h of ischemia, and 1, 2, and 4 h of reperfusion.
On the day of the experiment, the animals were randomly divided into two groups to compare 1) myocardial ischemia-reperfusion (MI/R) + vehicle and 2) MI/R + HBOC-201. During the initial surgery, 10% of each animal's total blood volume was removed and replaced with either HBOC-201 or 0.9% saline at a rate of 0.2 ml·kg1·min1 during the 30-min baseline period. Ten minutes before occlusion of the LAD, a 50-mg bolus of lidocaine (10 ml) was given to minimize ventricular fibrillation; a similar bolus of lidocaine was also given to the animal 10 min before reperfusion. The LAD remained occluded for a period of 90 min and was then reperfused for 4.5 h after removal of the microanuerysm clip.
Hemodynamic data. Hemodynamic data were recorded on a 16-channel MacLab (ADInstruments) system. The hemodynamic and functional data were simultaneously logged onto a Power Macintosh 8500 computer (Apple Computer) using MacLab Chart 3.5 software. The following variables were recorded: heart rate, mean arterial blood pressure, left ventricular (LV) systolic and end-diastolic pressures, and peak positive and negative first-time derivatives of LV pressure (maximum and minimum dP/dt, respectively). Lead II of the ECG was used to simultaneously monitor heart rate, arrhythmias, and ST segment alterations.
Regional myocardial blood flow determination. In 10 additional animals, regional myocardial blood flow was measured using sterile 15-µm radiolabeled microspheres (EI du Pont de Nemours) by the reference-withdrawal method. Microspheres were injected into animals at baseline and at 75 min of ischemia. Flow determination was made using one of five available isotopes (141Ce, 113Sn, 85Sr, 95Nb, or 46Sc) in a 0.01% Tween 80 suspension in 10.0% dextran.
Each vial of microspheres was thoroughly mixed by sonication and vortex agitation before injection. Approximately one to two million microspheres were injected via the left atrium (acute MI/R) or directly into the left ventricle via the catheter (extended MI/R) for each flow measurement. The reference arterial samples were obtained from the catheter inserted into the aorta via the femoral artery; withdrawal of samples commenced before the injection of microspheres and continued at a constant rate for 4 min using a peristaltic pump (Rainin Instruments).
Two adjacent transverse slices of the left ventricle that correspond to the central ischemic region were sampled for myocardial blood flow determinations after overlay tracings were completed for quantitation of infarct size. Two transmural sections were obtained from each of the nonischemic and central ischemic regions of each slice. Sections were taken at least 1 cm away from the boundaries between regions demarcated by Evans blue staining. Each transmural wedge of tissue was subdivided into two pieces of approximately equal thickness (endocardial and epicardial); after the location of each piece of tissue was recorded, samples were weighed and placed into counting vials for radioactivity assay in a gamma counter (1282 CompuGamma; LKB Wallac). Corrections for background and overlap were performed using the CompuGamma UltroTerm II system software. Myocardial blood flow was then calculated using the equation
m = (Cm x
r)/Cr, where
m is myocardial blood flow (in ml/min), Cm is the number of counts in the tissue sample (in cpm),
r is the withdrawal rate of the reference arterial sample (in ml/min), and Cr is the number of counts in the reference arterial sample (in cpm). The amount of flow per gram of tissue was calculated by dividing the flow by the weight of the appropriate tissue sample.
Determination of area at risk and infarct size. At the conclusion of reperfusion, the LAD was reoccluded, and in vivo area-at-risk (AAR) delineation proceeded by injection of 40 ml of 10% Evans blue stain through the 8-Fr pigtailed catheter. A bolus of potassium chloride was injected with the Evans blue to aid in euthanizing the animal. The heart was then excised, the LAD was cannulated, and small polyethylene tubing was inserted. One liter of 2,3,5-triphenyltetrazolium chloride was infused at a constant 100-mmHg pressure through the cannulated LAD for ex vivo determination of necrotic tissue within the ischemic zone (the AAR). The heart was then fixed in 10% formalin for 6 h. After fixation, the hearts were sliced into 7-mm-thick serial cross-sections starting at the apex (14). The myocardial slices were then imaged using a dissecting microscope (model SZ4045; Olympus America) equipped with a charge-couple device color video camera (Iris; Sony Electronics). The LV area, AAR, and area of infarction for each slice were then determined by computer planimetry using NIH Image 1.57 software. The right ventricle was removed from each imaged section, and the LV weight was recorded. The size of the myocardial infarction was determined by the following previously described equation (16): weight of infarct = (A1 x W1) + (A2 x W2) + (A3 x W3) +... (An x Wn), where A is the percent area of infarction by planimetry, subscripted numbers 1n represent section numbers, and W is the weight of the same numbered sections. Similar calculations were used to determine weights of the noninfarcted AARs and nonischemic zones within the myocardium.
Hematology of peripheral blood. Tests for creatine kinase MB isoform (CK-MB) and troponin (Tn-I) and complete blood counts were performed at the Louisiana State University Health Sciences Center clinical laboratories. Whole blood samples were obtained at baseline, 1 h of ischemia, and 1, 2, and 4 h of reperfusion from the catheterized femoral vein (7, 18).
Myocardial histology. Routine histological staining was performed on multiple midventricular cardiac sections to determine the extent of polymorphonuclear (PMN) infiltration into the ischemic reperfused myocardium. Vehicle- and HBOC-201-treated groups were subjected to 90 min of myocardial ischemia and 4.5 h of reperfusion and were then stained as described. The tissue sections were immediately fixed and stored in a 10% neutral-buffered formalin solution. The tissue slices were paraffin embedded, cut into 1-mm sections, placed on slides, and stained with Gill's no. 3 hematoxylin and eosin. The slides were then viewed microscopically, and the number of neutrophils (PMN cells) per high-power field was determined. For each of the hearts examined, the number of neutrophils was counted in six fields on three separate tissues.
Statistical analysis. All values represent means ± SE of the indicated number of independent experiments. The infarct size, AAR, left ventricle size, leukocyte count, CK-MB, Tn-I, and hemodynamic data were analyzed with ANOVA coupled with post hoc analysis (Scheffé's test of significance). All statistics were calculated using StatVeiw 4.5 (Abacus Concepts). Statistical significance was set at P < 0.05.
| RESULTS |
|---|
|
|
|---|
Hematology. The circulating levels of leukocytes, neutrophils, and platelets for the vehicle and HBOC-201 groups at all time points are presented in Table 1. Total counts for leukocytes and neutrophils (PMN cells) did not differ significantly between the vehicle and HBOC-201 groups at anytime during the experimental protocol. Platelet counts were significantly lower at 1 h of reperfusion in HBOC-201-treated canines but were not different at other time points. Hematocrit was reduced in HBOC-201-treated compared with vehicle-treated canines; hematocrit was significantly lower at 1 h of ischemia as well as at 1 and 2 h of reperfusion and was nonsignificantly reduced at 4 h of reperfusion.
|
|
|
|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
56% in hearts treated with HBOC-201. Furthermore, treatment with HBOC-201 also resulted in significant reductions in myocardial neutrophil infiltration and myocardial release of CK-MB and Tn-I. This protective effect cannot be attributed to changes in circulating neutrophils, because there was no significant difference between the vehicle and HBOC-201 groups, although HBOC-201 did inhibit neutrophil infiltration into the ischemic reperfused myocardium. It has been previously shown (7) that neutrophil accumulation is a mechanism of tissue injury capable of extending myocardial infarction. Finally, the cardioprotection observed with HBOC-201 treatment is not attributable to changes in hemodynamic status or myocardial blood flow during myocardial ischemia. Our data strongly suggest that pretreatment with this glutaraldehyde-polymerized bovine Hb solution (i.e., HBOC-201) attenuates myocardial reperfusion injury. Reperfusion injury is the conversion of normal or injured cells to more severely injured cells during the restoration of blood flow; this resultant damage is mediated by molecular O2 (23). It has been shown that ischemia is an initiator of inflammatory mediators and O2 radicals that lead to irreversible tissue injury (15). Introduction of an O2-rich product such as bovine Hb before ischemia may reduce the severity of the ischemia by preloading the myocardium with O2 or by aiding in O2 delivery to hypoxic tissue. In either case, a reduction in the severity of ischemia would result in a reduction in ischemia-induced inflammation and a subsequent reduction in reperfusion injury.
Previous studies of cerebral, skeletal muscle, and gut ischemia (5, 11, 22) have demonstrated a reduction in tissue injury. Horn et al. (11) and Standl et al. (21) reported that intravenous administration of HBOC-201 restored the decreased skeletal muscle tissue O2 tensions after artificial arterial stenosis. Their results demonstrate that bovine Hb solution improves poststenotic tissue oxygenation. It has also been shown (22) that a Hb solution administered to pigs in severe hemorrhagic shock was able to restore systemic blood pressure and also improve microvascular oxygenation in the gut. Cole et al. (5) showed that infusion of a Hb solution was able to significantly reduce cerebral ischemia with a prolonged reperfusion of 72 h. All of these studies as well as ours indicate that infusion of Hb solution decreases the severity of ischemic end-organ tissue injury.
Nitric oxide (NO) is involved the regulation of vascular tone. The use of Hb-containing agents such as HBOC-201 are thought to cause vasoconstriction by scavenging of NO in the vascular lumen (10). This reduction in local NO concentration leads to vascular constriction (9, 13). One mechanism that appears to induce vasoconstriction is the high-affinity binding of NO by acellular Hb (8). In the present study, we did observe an increase in blood pressure after HBOC-201 infusion in our model of MI/R. However, the increase in mean arterial blood pressure did not achieve statistical significance and did not adversely affect myocardial injury after ischemia and reperfusion.
The present study demonstrates that HBOC-201 has a number of positive attributes in the setting of MI/R injury. One limitation in the present study is that HBOC-201 was administered as a pretreatment, and this does not closely mimic what occurs during the evolution of acute myocardial infarction in humans. Another limitation is the short reperfusion time of 4.5 h. Additional studies to investigate 4872 h of reperfusion will provide a better real-life analysis of the potential protective effects of HBOC-201.
In summary, the present study clearly demonstrates that pretreatment with a Hb-based blood substitute significantly ameliorates MI/R injury. In addition, the extent of myocardial inflammation was also attenuated with this Hb-based blood-substitute therapy. Additional studies are indicated to define the precise cellular and molecular mechanisms involved in this cardioprotective effect.
| GRANTS |
|---|
|
|
|---|
| DISCLOSURES |
|---|
|
|
|---|
| FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
I. George, G.-H. Yi, A. R. Schulman, B. T. Morrow, Y. Cheng, A. Gu, G. Zhang, M. C. Oz, D. Burkhoff, and J. Wang A polymerized bovine hemoglobin oxygen carrier preserves regional myocardial function and reduces infarct size after acute myocardial ischemia Am J Physiol Heart Circ Physiol, September 1, 2006; 291(3): H1126 - H1137. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Burmeister, C. Rempf, T. G. Standl, S. Rehberg, S. Bartsch-Zwemke, T. Krause, S. Tuszynski, A. Gottschalk, and J. Schulte am Esch Effects of prophylactic or therapeutic application of bovine haemoglobin HBOC-200 on ischaemia-reperfusion injury following acute coronary ligature in rats Br. J. Anaesth., December 1, 2005; 95(6): 737 - 745. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |