1Laboratory of Hemodynamics and Cardiovascular Technology, Ecole Polytechnique Fédérale de Lausanne; and 2Department of Cardiovascular Surgery, Centre Hospitalier Universitaire Vaudois, Lausanne; 3Department of Cardiovascular Surgery, Inselspital, Bern; 4Departments of Cardiology and 5Anesthesiology, Centre Hospitalier Universitaire Vaudois, Lausanne, Switzerland; and 6Hydraulics Laboratory, Institute of Biomedical Technology, Ghent University, Gent, Belgium
Submitted 1 September 2005 ; accepted in final form 11 November 2005
| ABSTRACT |
|---|
|
|
|---|
echocardiography; heart failure; hemodynamics; ventricular function
Several structural and functional parameters can be assessed by using TTE, such as myocardial wall thickness, end-systolic and -diastolic volumes, left ventricular (LV) mass, ejection fraction (EF), fractional shortening, and velocity of circumferential fiber shortening (Vcf). The myocardial performance index (MPI) is a relatively recent parameter combining both systolic and diastolic functions. MPI is the ratio of total time spent in isovolumic activity (isovolumic contraction and relaxation times) to the ejection time (ET) and is measured from the mitral inflow and LV outflow time intervals. Recently, in a clinical setting, the MPI has been described as a predictor of cardiovascular mortality, independent of other measurements of cardiac function (EF and wall motion score index) and of traditional cardiovascular risk factors (smoking, diabetes, hyperlipidemia, and hypertension) in elderly men (1). Additionally, it correlates to plasma brain natriuretic peptide in patients with hypertrophic cardiomyopathy (15). It is an attractive parameter because it appears to be independent of the LV shape. Indeed, the measurement of the LV dimensions is not mandatory. It has been reported that MPI correlates with the peak positive value of the time derivative of LV pressure (dP/dtmax) and the time constant of relaxation derived from the peak systolic pressure to the next start of diastolic pressure (
) and is affected by the pre- and afterloads (3, 13, 14, 29). However, the preload dependency of MPI reported by Moller et al. (13) in patients without MI could not be reproduced by Cheung et al. (6) in a porcine model. Nevertheless, its reliability has been proven clinically (18, 19) as well as in large (12) and in small animal models of MI (14). Furthermore, despite its increasing use, MPI has never been validated against established indexes of systolic and diastolic function as assessed from pressure-volume data in chronic rodent models of MI. Only volumetric comparisons were reported (5, 8), and only one study correlated MPI with certain CC parameters in a porcine model (12). In a clinical setting, compared with other TTE parameters such as EF and ratio of peak early diastolic filling velocity to peak filling velocity at atrial contraction, the MPI appears more representative and sensitive to functional changes provoked post-MI (18, 19). Therefore, we thought it would be instructive to analyze in a follow-up study the possible interest of this parameter to capture the LV structural and functional alterations occurring during the post-MI remodeling period.
Also, because the LV fractional shortening (LVFS) is an ejection phase index (3), we hypothesized that the ratio of these two parameters (LVFS/MPI) would be more sensitive to MI-induced alterations of cardiac function as compared with MPI alone. Therefore, the aim of this study was to assess changes in MPI and the LVFS/MPI ratio in a rodent MI model and to correlate MPI as well as the related LVFS/MPI index to other invasive indexes obtained by CC.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Experimental protocol. The rats in the MI group (n = 9) had a body weight of 254 ± 19 g (means ± SD) and were anesthetized with isoflurane (Forene; Abbott, Baar, Switzerland), intubated, and ventilated with 100% oxygen at 60 cycles/min with a tidal volume of 2 ml (model 683, Harvard Apparatus, Holliston, MA) before the surgical procedure was performed. The rats were placed on a heating pad to maintain body temperature, and disinfection was performed of the thorax. A left thoracotomy was performed at the third intercostal space to gain access to the heart. The pericardium was opened, and the left anterior descending coronary artery was located (between the left atrium and the pulmonary artery) and ligated with a 4.0 polypropylene suture (Ethicon, Somerville, NJ) to provoke the MI, which was visually confirmed by a change of color of the LV from red to a purplish/gray distal to the ligation site. The ribs were closed with two to three ligatures, a chest drain was inserted to avoid any pneumothorax, and the muscles surrounding the rib cage were sutured together before the skin was closed. Analgesia was given (Pro-Dafalgan) to attenuate any pain that ensued for the rodents because of the MI (Upsamedica, Baar, Switzerland). The rats were gradually weaned from the anesthesia in the ventilating gas, and when they began to awaken and resume spontaneous breathing, the intubation tube was removed. The rats were then returned to the animal house once complete recovery was observed. A sham operation was performed on a similar control (CTRL) group (n = 10) with the rodents weighing 238 ± 7 g (means ± SD).
Invasive cardiac measurements with CC.
Eight weeks after the initial operation, the rats were again anesthetized and intubated. The right neck region was disinfected to provide access to the carotid artery. The skin was opened, and the right jugular vein and carotid artery were isolated. A 2-Fr CC (SPR 838 Aria, Millar Instruments) was inserted into the left ventricle via the right carotid artery. Parallel conductance was measured after injection of 10% saline into the jugular vein with a 1-ml syringe (20 µl Natrium Chloratum; Sintetica, Mendrisio, Switzerland) in accordance with the method by Baan et al. (2). An occlusion analysis was performed by temporarily occluding the inferior vena cava below the diaphragm via a mini laparotomy. For each animal, the CC calibration correction factor
was assessed with the use of an ultrasonic flowmeter (Transonic, Ithaca, NY) placed around the ascending aorta.
From the simultaneous measurement of LV pressure and volume during steady-state conditions, the following parameters were derived: stroke volume (SVc); end-systolic volume (ESVc); end-diastolic volume (EDVc); EF (EFc = SVc/EDVc) (subscript c refers to data measured with the CC to differentiate between the same parameters measured with TTE); isovolumetric contraction time (ICT; duration of the isovolumic contraction); dP/dtmax; peak negative value of the time derivative of LV pressure (dP/dtmin); contractility index (dP/dtmax divided by the pressure at this point); isovolumetric relaxation time (IRT; duration of the isovolumic relaxation);
; heart rate; LV end-diastolic (LVEDP) and start of diastolic pressures, which are taken as the pressures associated with the bottom left and right corner of the pressure-volume loop, respectively; systolic and end-systolic pressures; systolic ejection period; and diastolic filling period. Cardiac output (COc) was calculated from heart rate multiplied by SVc.
From the occlusion data, we calculated the slope (Ees) and volume axis intercept (V0) of the end-systolic pressure-volume relationship (ESPVR), the slope of the end-diastolic pressure-volume relationship (EDPVR), preload-recruitable stroke work (PRSW) (with stroke work assessed from the area enclosed by the pressure-volume loop), and preload adjusted dP/dtmax (slope of the relation between dP/dtmax and EDVc).
Echocardiographic measurements. TTE was performed at baseline and every 2 wk for an 8-wk period with the use of a commercially available echocardiographic system (C256 Sequoia, Acuson, Mountain View, CA) with the animal in the left lateral decubitus position. Light anesthesia was used during the analysis with isoflurane (Abbott) ventilated inside a nose cone at 0.5 l/min with 100% oxygen. Once asleep, the rat was shaven with an electrical razor (surgical clipper 9661, 3M Health Care). Ultrasound transonic blue gel (Tyco, Mirandola, Italy) was placed on the thorax to optimize visibility of the cardiac chambers. A 15-MHz linear array transducer (15L8) was used with a frame rate of 100 Hz by using bidimensional and color Doppler imaging. The probe was placed to obtain short- and long-axis and four-chamber views. From the long-axis view, an M-mode trace of the LV was obtained, and LV end-diastolic diameter (LVDed), LV end-systolic diameter (LVDes), and posterior and septal wall diastolic wall thickness (PWth and SWth) were measured. LVFS was calculated as follows: LVFS = (LVDed LVDes)/LVDed x 100. EF was calculated from a long-axis view by using planimetry as follows: EFe = 100 x (LVDed3 LVDes3)/LVDed3. Vcf was calculated by using the following formula: (LVDed LVDes)/ET x LVDed. The subscript e refers to data assessed with echocardiography. EDVe and ESVe were calculated by using Simpson's method, and SVe was calculated as EDVe ESVe. Aortic flow or cardiac output (COe) was recorded and calculated by using pulse Doppler imaging, with the smallest possible sample volume placed at the level of the aortic annulus. Doppler yields the velocity profile (aortic velocity time integral), which is multiplied with the cross-sectional area of the outflow tract to obtain flow. Integration of the velocity profile yields the aortic velocity time integral. LV mass was calculated as {[(LVDed + SWth + PWth)3 LVDed3] x 1.04} x 0.8 + 0.14 (in g, with LV dimensions expressed in mm) (20). Relative wall thickness (RWT) was assessed as RWT = (PWth + SWth)/LVDed. LV ET was measured as the time from the beginning to the end of the aortic flow wave (Fig. 1). IRT was measured as the interval between the aortic closure click and the start of mitral flow, whereas ICT was obtained as the time delay between the cessation of mitral inflow and the onset of aortic ejection. Additionally, the mitral valve closure time (MCO) was measured. MPI is then defined as (MCO ET)/ET = (ICT + IRT)/ET (Fig. 1). A newer index of LV function, proposed by Broberg et al. (3), LVFS/MPI, was also assessed (3).
|
| RESULTS |
|---|
|
|
|---|
|
increased in the MI group. The IRT was significantly prolonged in the MI group. The dP/dtmin was augmented by 10% in the MI group (P = not significant; Table 1). Echocardiography (volumetric data with subscript e) and serial follow-up data. Body weight increased with age in both groups but remained significantly lower in the MI group (Fig. 2A). LV mass increased with age in both groups; however, after 8 wk, absolute LV mass was higher in the MI group (0.98 ± 0.11 g CTRL vs. 1.29 ± 0.10 g MI; P < 0.01). The same applied for the LV weight-to-body weight ratio (1.70 ± 0.20 g/kg CTRL vs. 2.45 ± 0.45 g/kg MI; P < 0.01). The heart rate was comparable between the CTRL and MI groups (318 ± 29 and 319 ± 29 beats/min, respectively) at week 8.
|
RWT and LVFS attenuated in the MI group but remained stable in the CTRL group (Fig. 2, E and F). Vcf rose slightly in the CTRL group but was conserved in the MI group (Fig. 2G). COe was significantly less in the MI group (P < 0.05) at week 8 (89 ± 14 ml/min MI vs. 112 ± 20 ml/min CTRL).
MPI and valve opening and closure timing parameters. The ICT was stable in the CTRL group but was prolonged in the MI group. The IRT was not different between groups but did rise in the MI group when compared with baseline at weeks 6 and 8 (P < 0.001). ET remained stable over time without any relevant difference between the groups. Consequently, the MPI increased significantly in the MI group when compared with the CTRL group (Fig. 2I). LVFS/MPI decreased significantly in the MI group as compared with the CTRL group (Fig. 2J).
Correlation of invasive parameters versus MPI and LVFS/MPI (pooled data). In the univariate correlation analysis, both MPI and LVFS/MPI were significantly correlated with body weight, LVEDP, PRSW, EDV, and ESV (Table 2).
|
Correlation of invasive parameters versus MPI and LVFS/MPI (groups separated). From the above pooled data, the two groups were separated and univariate regression analysis was applied. The corresponding graphs, with the linear regressions and correlation coefficients, are shown in Fig. 3. For the majority of the parameters, the regression lines of the pooled data and of the separated groups are almost identical or similar. Overall, the correlation coefficients based on the pooled data were comparable to those derived from the separate groups (Fig. 3 and Table 2).
|
value used for the volume correction was 2.5 ± 0.8.
|
| DISCUSSION |
|---|
|
|
|---|
Serial assessment of cardiac function. Classically, EF, RWT, LVFS, and Vcf are used to evaluate systolic function with TTE in rat models of pressure and volume overload (4, 10, 26). In a rodent study with MI by Sjaarstad et al. (24), body weight, LVFS, Vcf, and CO were all compromised because of the MI. Gao et al. (9) performed TTE in mice, demonstrating that the LVFS remains constant over time in control mice but attenuates with time in a model of MI. In our study, all of the above parameters and MPI and LVFS/MPI were significantly compromised over time because of the MI (Fig. 2, A, F, and G). However, some of the above parameters have certain limitations; LVFS and Vcf are preload and afterload dependent and are limited by spatial resolution, whereas EF is frame rate dependent and assumes symmetrical contraction, which is not the case in aneurysmal and MI hearts. RWT is only useful in studies measuring myocardial wall adaptations due to hypertensive response (10). If used in MI models with an apical infarction present, but M-mode measured at the papillary muscle level, values will be falsely positive because the MI would cause wall thinning at the apex. However, in our study, all MIs were anteroapical, precluding this phenomenon. Therefore, the MPI, and recently LVFS/MPI (3), were introduced because they have been reported to be independent of load and LV geometry (3, 29). Slama et al. (26) and Salemi et al. (23) have reported the MPI in conjunction with serial studies, but only in hypertensive models. In our study, these two novel parameters fare well compared with the classic TTE parameters reported earlier and visible in Fig. 2. Moreover, the LVFS/MPI seems to visually express alterations in cardiac function post-MI more clearly, compared with the classic parameters, despite the large standard deviations (Fig. 2J). These deviations are a result of the fact that the infarct size in the MI model is difficult to homogenize and varies between 37 ± 7%, not forgetting the variations of intra- and interobserver variability. Also, it should be acknowledged that MPI is highly dependent on timing intervals and temporal resolution and hence susceptible to interpretation errors in cases when heart rate changes over time during the experimental procedure.
Noninvasive versus invasive assessment of cardiac function. We measured cardiac function using both TTE and CC in the same animal, allowing us to correlate findings from both techniques. Hence, the relation between classic (EF, LVFS, Vcf, and RWT) and novel indexes such as MPI and LVFS/MPI and established invasive cardiac function parameters could be derived (Table 2). The MPI and LVFS/MPI express the largest correlation coefficients and most significant for PRSW and LVEDP compared with the classic indexes. Similar coefficient values are found between classic and novel indexes with respect to EDV. However, ESV shows the smallest relationship for the novel indexes; nonetheless, it is still significantly comparable. LVFS correlated well to the invasive indexes but not as well as MPI and LVFS/MPI (Table 2). LVFS has also been shown to be related to dP/dtmax (and is modified with contractility and afterload manipulations) (3, 16) and dP/dtmin (9). Therefore, LVFS could be considered as a good marker to represent serial and end point cardiac function. However, Vcf has been shown to be dependent on dP/dtmax and is modified with contractility and preload and afterload manipulations (3). In our study, Vcf could not correlate to invasive indexes except for LVEDP.
With regard to the MPI, it has been reported that the MPI correlates with dP/dtmax, dP/dtmin, LVEDP, and
and that it is dependent on contractility yet also affected by preload and afterload (3, 14, 25). The dependence of MPI on preload was also reported by Moller et al. (13) in patients without MI. However, in a recent publication by Cheung et al. (6), this phenomenon could not be reproduced in a porcine model. In our study, MPI correlated with PRSW, a measure of contractility, and with LVEDP and LV volumes. Tei et al. (29) also found correlations of MPI with dP/dtmax and
when using cardiac catheterization in larger mammals, but we could not confirm this finding using our data. In an acute MI setting, Morgan et al. (14) also showed the MPI to correlate to end-systolic and end-diastolic TTE dimensional values. We advocate this by expressing a relationship between MPI, ESV, and EDV (Table 2). Furthermore, it must be stipulated that the MPI is LV geometry independent and thus ideal for studying the course of a disease associated with chamber remodeling, which is typically elaborated during MI (29).
With regard to the LVFS/MPI, it was shown by Broberg et al. (3) that this ratio correlated strongly with dP/dtmax in mice and was dependent on contractility and afterload variations but not on preload variations. However, in our study, the above ratio correlated to the load-independent contractile parameter PRSW, to LVEDP (preload), and to volumetric parameters (Table 2) but not to dP/dtmax. Therefore, the ratio is also affected by preload because LVEDP varies with MI. In our study of chronic MI, LVFS/MPI was the most sensitive to changes in contractility (PRSW) and filling (LVEDP) compared with classic TTE parameters. In multiple linear regression analysis, LVEDP appeared as the strongest predictor of MPI and LVFS/MPI, suggesting that these indexes are potentially useful noninvasive indicators of filling pressure in the rat, at least in the setting of cardiac remodeling after MI.
It is to be emphasized that the reported data were obtained in a pooled analysis, including both CTRL and MI data. When comparing EDV, ESV, EF, MPI, and SV as assessed by CC with TTE by using the Bland-Altman analysis, the mean difference is close to zero for all parameters validating the coherence of the two methods. Slight discrepancies exist between the two methods because the TTE accuracy is frame rate dependent and the CC accuracy depends on initial calibration,
estimation, and the calculation of parallel conductance (8). With this in mind, both methods are representative and suitable for evaluating changes in cardiac function. EFc correlates well to EFe (r = 0.80 with P < 0.01), and the differences in EF that are visible between the two groups are similar between the two methods (bias 6%), as seen in Table 1 and Fig. 4, respectively. Similar correlations between CC and TTE have been demonstrated, such as those we found for ESV and EDV (5). Also, MPIe has been correlated to MPIc with a bias of only 0.007 (roughly 2%). Because of the fact that at week 8, acceptable concordance was achieved between methods, one can assume that during the time course of the study, similar relations can be established, thus validating not just the end points but also the cardiac structural (EDV, ESV, and SV) and functional (EF and MPI) variations from the baseline of the study.
Study limitations. The number of animals in each of the subgroups was comparable to other studies, justifying detailed analysis of the MPI and LVFS/MPI within each subgroup. When compared with other research groups, between 6 and 12 rats are used per group (7, 9, 26) and all correlations are performed by pooling the analyzed data (14, 25). Therefore, after a detailed literature review, no other group has demonstrated acceptable correlations by using subgroup analysis techniques. Another limitation arises from the use of anesthetics that might influence the data because of their effect on inotropy and chronotropy. However, following previous reports, we used isoflurane during the complete study period because it seems to be the most appropriate technique for repeated and prolonged studies necessitating stable hemodynamic conditions (21). In addition, the simultaneous measurement of the TTE and CC parameters is technically challenging because the rat is small and the positioning of the TTE probe is performed with the animal in the lateral decubitus position.
In conclusion, we were able to follow serial changes in cardiac function post-MI with these novel parameters (MPI or LVFS/MPI) with success and as efficiently as with classic TTE parameters. Moreover, LVFS/MPI visually expressed better the serial modifications in cardiac function. Both novel parameters were correlated to the load-independent contractile parameter PRSW and to the preload parameter LVEDP, which was pertinent in following preload changes post-MI. Finally, chamber remodeling post-MI can successfully be followed because of the fact that ESV and EDV both correlate to MPI and LVFS/MPI.
| GRANTS |
|---|
|
|
|---|
| FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
F. N. Salloum, A. Abbate, A. Das, J.-E. Houser, C. A. Mudrick, I. Z. Qureshi, N. N. Hoke, S. K. Roy, W. R. Brown, S. Prabhakar, et al. Sildenafil (Viagra) attenuates ischemic cardiomyopathy and improves left ventricular function in mice Am J Physiol Heart Circ Physiol, March 1, 2008; 294(3): H1398 - H1406. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Jegger, R. F. da Silva, I. Lartaud, V. Gaillard, X. Jeanrenaud, M. Nasratullah, L. K. von Segesser, J. Atkinson, P. Segers, H. Tevaearai, et al. Effects of an aging vascular model on healthy and diseased hearts Am J Physiol Heart Circ Physiol, September 1, 2007; 293(3): H1334 - H1343. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Jegger, A. S. Mallik, M. Nasratullah, X. Jeanrenaud, R. d. Silva, H. Tevaearai, L. K. von Segesser, and N. Stergiopulos The effect of a myocardial infarction on the normalized time-varying elastance curve J Appl Physiol, March 1, 2007; 102(3): 1123 - 1129. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Jegger, R. da Silva, X. Jeanrenaud, M. Nasratullah, H. Tevaearai, L. K. von Segesser, P. Segers, V. Gaillard, J. Atkinson, I. Lartaud, et al. Ventricular-arterial coupling in a rat model of reduced arterial compliance provoked by hypervitaminosis D and nicotine Am J Physiol Heart Circ Physiol, October 1, 2006; 291(4): H1942 - H1951. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||