INNOVATIVE METHODOLOGY
Division of Cardiology, Department of Medicine, Weill Cornell Medical College, New York, New York
Submitted 18 May 2007 ; accepted in final form 18 July 2007
| ABSTRACT |
|---|
|
|
|---|
-[2-(di-n-butylamino)-6-napathy]vinyl}pyridinium (di-4-ANEPPS). Thus real-time control of cardiac activity using optical mapping techniques is feasible. Such a system is attractive because it offers greater measurement resolution than the electrode-based systems with which real-time control has been used previously.
cardiac arrhythmias; nonlinear systems; antitachycardic pacing; charge-coupled device camera
1 ms, that is capable of acquiring electrophysiological signals and measuring APD with a high degree of accuracy. This work focuses on methods of investigating real-time control of tissue-level activity. As previously reported, we have developed a hard real-time software system [originally known as RTLab; now known as real-time experimental interface (RTXI); www.rtxi.org] that accomplishes the stringent control requirements (e.g., one that is capable of acquiring electrophysiological signals and measuring APD with a high degree of accuracy) (6). This system has been used for APD alternans control in Purkinje fibers using multiple glass microelectrodes (5). Although microelectrodes are capable of acquiring high-fidelity transmembrane voltage measurements, their use is hindered by a number of factors, including the inherent difficulty in placing them in close proximity to one another and the difficulty in using them in tissue preparations due to their fragility. Thus it is impractical to use microelectrodes to obtain high spatial resolution data in cardiac tissue preparations.
On the other hand, optical mapping techniques using fluorescent voltage-sensitive dyes allow a dense, high-resolution mapping of cardiac electrical activity and have greatly contributed to our understanding of the dynamics of arrhythmias, especially regarding the role of APD alternans (3, 4, 13, 18, 21, 22, 27) (for a recent technical review, see Ref. 10). In this article, we describe a system that uses optical mapping techniques and an electrical stimulator as the sensory and effector arms, respectively, of a closed-loop, real-time control system. Whereas optical mapping control systems using custom-made synchronization hardware have been used previously for both functional cortical mapping and feedback stimulation in heart (25, 26), to our knowledge, this is the first report of a software-driven "hard" real-time control system for electrophysiological (cardiac or neuronal) optical mapping with a deterministic delay of <1 ms that is necessary to achieve alternans control.
One example of the utility of such a system is to test various control algorithms designed to suppress APD alternans. Although the efficacy of control at a short distance from the stimulating electrode has been well established (3, 4), open questions remain regarding the feasibility of control farther from the electrode (9). Characterizing the efficacy of such distant control requires quantifying electrophysiological properties, such as conduction velocity restitution and repolarization heterogeneities, at fine spatial resolution over the entire spatial control region. Because of the spatial resolution limitations inherent to the microelectrode approach, real-time optical mapping would be invaluable for such an application.
This article is divided into three main parts. First, we discuss the technical design and characteristics of the system. Second, we test the ability of the system to control, in real time, simulated optical signals of electrical alternans. Such simulations 1) establish a lower limit on alternans detectability and controllability, 2) provide information necessary to interpret the results of future experiments, and 3) can aid the design and verification of control algorithms by enabling quantification of noise sensitivity. Third, we demonstrate the feasibility of real-time control in a real optical mapping environment by implementing an antitachycardia pacing (ATP) system and testing it in in vitro porcine experiments. The ATP system is used as a proof of principle since it requires acquisition of action potential data and real-time calculation (as a function of the acquired action potential data) of the delivery time of the next stimulus. (Note that the alternans control was not tested in vitro here because the available porcine tissue is relatively resistant to alternans occurrence.)
| DESIGN AND IMPLEMENTATION |
|---|
|
|
|---|
Second, considering the large size of the two-dimensional images, communication between the charge-coupled device (CCD) camera and computer memory can act as a bottleneck in high-speed signal acquisitions.
Third, commonly used operating systems, such as Microsoft Windows or Linux, are not capable of providing deterministic timing with a well-defined upper limit on interrupt latencies, as is required for real-time control applications. Latency in excess of 10 ms is not uncommon with either operating system.
Fourth, for control applications, we require mapping over a relatively large area (4 to 5 cm). Using a standard, single-lens optical setup over such a large area would necessitate a large distance between the front of the lens and the sample, which in turn would require a high-intensity excitation light, causing excessive photobleaching and phototoxicity and limiting the recording duration and sampling frequency (due to low-light intensity reaching the CCD).
To address these challenges, we have used a hard real-time operating system (RTAI Linux) to achieve accurate timing. To solve the latency problem while preserving the high sampling frequency, we have chosen a line-scan CCD camera (also called a push-broom camera). Line-scan cameras have a one-dimensional light sensor (2,048 x 1 pixels in our case) that greatly reduces the size of each frame (a few kilobytes instead of hundreds of kilobytes) and therefore decreases the transfer latency. In addition to low latency, our line-scan CCD camera had a long sensor (28 mm) compared with a typical area-scan CCD (a few millimeters), allowing us to use a tandem-lens assembly optical setup that greatly reduced the distance between the lens and the sample.
Imaging hardware.
The main components of the optical mapping system were a line-scan CCD camera connected to a tandem-lens assembly (10,23) and a light-emitting diode (LED) -based light source. We used an AVIIVA M2 CL 2014 line-scan camera (Atmel, San Jose, CA) with a 28.6 mm-long sensor and 2,048 x 1 pixels. Each pixel is 14 x 14 µm in size and has a well-size capacity of
300 kiloelectrons. The camera could read up to 28,700 lines/s, although it was used mainly at 500–1,500-Hz sampling rates. The light intensity was digitized at 12 bits and transmitted to the computer via a CameraLink interface and a 60-MHz framegrabber (microEnable III, Silicon Software, Mannheim, Germany). Supplement 1 (note: supplemental material may be found with the online version of this article) shows the CCD noise characteristics (1). The predominant noise in our operational range (i.e., sampling rate and dynamic range) was shot noise, with dark noise and readout noise contributing to <20% of the total noise level. Therefore, using a cooled-CCD camera would not have reduced the total noise to any significant degree (as it primarily affects the dark noise), whereas improving the light-gathering capability by using a tandem-lens assembly drastically improved the signal-to-noise ratio.
The lens assembly was composed of a 105 mm f/1.8 lens (Nikon) reverse mounted on a 50 mm f/1.4 lens (Nikon), providing x0.47 magnification at a working distance of 40 mm. With a consideration of the pixel pitch of 14 µm, this magnification is translated into a theoretical resolution of 30 µm/pixel, though in practice the optical system point-spread function (PSF) limited the resolution to
0.5 mm. A 600-nm emission filter (R60, Nikon) was placed between the two lenses (more details in supplement 2).
We used a solid-state illumination system (high-power LEDs) in place of standard Quartz Tungsten Halogen lamps (11, 17). This decision simplified the design by removing the excitation filter and the mechanical shutter. Eight 5-W high-power Luxeon V LEDs (Lumileds, San Jose, CA) with a peak output at 505 nm provided the emission light. No excitation filter was used; the light leakage through the emission filter was measured as <0.1%. The lack of an excitation filter and a mechanical shutter allowed us to mount the LEDs between the front end of the lens assembly and the sample at a distance of 15–20 mm from the tissue. Nevertheless, these LEDs suffer from initial light output instability because of high thermal load since their light output is temperature dependent, which complicates the signal processing (see Signal processing).
A National Instruments PCI-6254 analog-to-digital board provided the link between the RTXI kernel module and the ECG recorder and the stimulator.
Software. We chose the Linux operating system (kernel 2.4.24) augmented by RTAI patch (version 3.1) as our hard real-time platform. The software system was written around RTXI signal acquisition and real-time control software, which was previously used to implement the aforementioned electrode-based control system (5). The software was composed of three kernel space and two user space modules: RTXI and the optical mapping interface in the user space and the framegrabber driver, helper driver, and RTXI kernel modules (Fig. 1).
|
In our system, we used a boxcar spatial filter of width 16 to improve the signal-to-noise ratio. In effect, this filter reduced spatial resolution to
480 µm, which is comparable with the optical system PSF.
Detrending the signal proved more complicated. Not only does a LED light source with electronic shutter follow an exponential decay rather than a linear one (Fig. 2A), the restriction imposed by the real-time nature of the control application precluded the use of an off-line polynominal detrending algorithm. As an additional practical constraint, the algorithm should have been simple and robust and preferably using only fixed-point arithmetic to be implemented as a real-time Linux kernel module.
|
| VERIFICATION: SIMULATED OPTICAL REAL-TIME ALTERNANS CONTROL |
|---|
|
|
|---|
First, transmembrane potential is inverted and added to a background based on the ratio of change in fluorescent signal due to action potentials to the baseline fluorescence (
F/F) as measured from the tissue experiments (see VERIFICATION: IN VITRO OPTICAL REAL-TIME ANTITACHYCARDIA PACING).
Second, different sources of optical distortion are simulated. Multiplying the signal by a spatial Hanning (cosine) window simulates Vignetting (darkening near the edges). The blurring effect of PSF is modeled by convoluting the resulting signal with a Gaussian kernel. All the parameters (width of the Hanning window and the kernel) are derived from actual experimental data.
Third, an exponential decay function models the LED light instability.
Fourth, Gaussian random noise is added to the signal. In supplement 1, we showed that the primary source of noise in the system is shot noise, which is proportional to the square root of the signal level.
Afterward, the simulated optical signals are processed by the routines designed to remove various sources of noise and distortion as described in Signal processing. Figure 3A demonstrates an example of such a processed signal in the presence of a typical amount of noise. Vertical axis is time (top to bottom), whereas the horizontal axis corresponds to the line-scan sensor. The data are color coded, such that blue corresponds to the baseline and red to the action potential upstroke.
|
|
![]() | (1) |
![]() | (2) |
is the feedback gain (usually 0.6 to 1.0), and
Tn is the perturbation applied to the timing of the next (n + 1) stimulus.
Figure 3B shows the effect of the APD control algorithm. APD alternans is obvious in Fig. 3A (before control), but after the algorithm is activated, control is achieved. We repeated similar experiments with different BCLs and noise levels. The result is summarized in Fig. 4B. In this figure, a noise level of 1 corresponds to the amount of noise expected in a freshly stained piece of tissue. During the course of experiments, as a result of photobleaching and washout, the
F/F ratio reduces, and the noise level gradually increases to the range of 3–5 after half an hour. The y-axis shows residual alternans, which is defined as the average level of alternans (i.e., |An – An–1|) 5 s after alternans control is commenced. Figure 4C demonstrated the effects of noise on beat-to-beat alternans level.
It is not surprising that residual alternans is linearly proportional to the noise level. Nevertheless, Fig. 4B also shows that residual alternans depends on BCL, where shorter BCLs are associated with a steeper relationship between noise and residual alternans. To explain this finding, we have to explore the coupling between noise and alternans. When we consider a situation where noise is zero and control is active for a few seconds, such a system is in a metastable state with residual alternans of 0. The primary effect of noise is to introduce error in the APD measurement. This error is fed back to Eq. 1 and affects
T, which in turn acts as a perturbation and is amplified by a factor of s, where s is the slope of the APD restitution curve at this BCL. The final result is that the effect of noise on alternans is proportional to
s. In Fig. 4B, a shorter BCL operates in a steeper portion of the restitution curve, and, consequently, it results in a larger residual alternans.
| VERIFICATION: IN VITRO OPTICAL REAL-TIME ANTITACHYCARDIA PACING |
|---|
|
|
|---|
The right ventricular free wall (Fig. 5) was removed, and the right coronary artery was cannulated using a 6-Fr (2 mm) catheter after its distal end was securely closed by 4-0 silk sutures. Any transected epicardial coronary artery was closed to maintain perfusion pressure. Afterward, the isolated right ventricle was transferred to a custom-made recording chamber and perfused by warm oxygenated Tyrode solution containing (in mmol/l) 137 NaCl, 2.7 KCl, 1.0 MgCl2, 1.8 CaCl2, 0.4 NaHPO4, 16.0 NaHCO3, and 5.6 glucose at a perfusion pressure of 40–60 mmHg. A bipolar pacing electrode was attached to the endocardial surface. A continuous electrogram was recorded using two silver chloride electrodes in the bath. After spontaneous contraction was noted in the perfused right ventricle, it was paced at twice the threshold at a rate of 0.5 Hz. The pacing interval was progressively shortened, while maintaining 1:1 capture, until the pacing interval equaled 500 ms. If at any time during the experiment, except when the ATP protocol was active, reentrant tachycardia was noted, the tissue was defibrillated at 2–5 J to terminate the arrhythmia. After pacing at 500 ms for at least 15 min to achieve equilibrium, the tissue was stained by an injection of 200 µl of 5 mmol/l solution of voltage-sensitive dye (di-4-ANEPPS; Molecular Probes, Eugene, OR) in dimethyl sulfoxide into the perfusate over a 5-min period. At the same time, the perfusion solution was changed to one containing an electromechanical decoupler: either 5 mmol/l of 2,3-butanedione monoxime or 5 µmol/l of cytochalasin-D in Tyrode solution (both from Sigma, St. Louis, MO) (29). Any residual contraction was stopped by gently pressing a transparent cover over the tissue.
|
|
We induced arrhythmia in the tissue by rapid pacing. The most likely mechanism for the arrhythmia was reentry, since 1) burst pacing was the initiation mechanism and 2) the arrhythmias were fast (cycle length down to 100 ms) and usually regular and monomorphic, although we occasionally observed polymorphic arrhythmias (Fig. 7B).
|
Figure 7 shows the stack diagram of three different reentries. In this situation, the shape of the wavefront is different from simple pacing and cannot be predicted from the model presented in supplement 5, since the assumption that all waves diverge from a single point no longer holds.
To test the real-time control system, we applied a variant of commonly used ATP algorithms to reentry occurrences. ATP compromises a group of techniques employed in some implantable cardiac devices to terminate reentrant tachyarrhythmias by overdrive pacing (14). Our algorithm is based on a closed-loop control system (in contrast to open-loop ATP algorithms common in clinical settings) that monitors cardiac electrical activity through an optical channel and then, based on analyses of this activity, systematically triggers precisely timed feedback-controlled stimuli to entrain the reentry (see supplement 6).
Figure 8 shows an example. The different shape of the fourth and fifth action potential in the map (fusion complexes), combined with the timing information (e.g., 5th action potential is early, whereas the interval between the 4th and 6th is twice the baseline cycle length), strongly suggests that entrainment of the fifth stimulus into the reentry loop occurred. Hence, the real-time control system is capable of detecting the action potentials and synchronizing the firing of the stimuli accordingly.
|
| DISCUSSION |
|---|
|
|
|---|
We used the data obtained from this system to develop a realistic model for the optical signals. Applying APD control to the model, we were able to show that control is achievable over a wide range of noise levels encountered in real experiments. Nonetheless, noise sets a lower limit on the realizable level of alternans control. Having this knowledge will help with the design and interpretation of future experiments, when the system is used for APD control.
Such a hybrid system also possesses some of the problems it inherits from the base technologies. Similar to traditional optical mapping methods, photobleaching and phototoxicity remain problematic. Additionally, the need to use electromechanical decouplers potentially interferes with the shape and duration of the action potentials (15, 24). Also, the timing and processing-load restrictions currently imposed by real-time control algorithms and available hardware limited us to use a linear CCD, rather than a full two-dimensional area-scan camera.
Nevertheless, as we showed in the EXPERIMENTAL RESULTS, the signal acquired from a linear CCD can be processed to extract global information (including anisotropy ratio) about the tissue under study. We expect that technological advances in CCD cameras, processors, and data transfer hardware will allow construction of a real-time optical mapping system using area-scan cameras in the near future. Such a system should be useful in implementing and testing complex control algorithms to modulate or terminate cardiac arrhythmias.